Ivana Lapić1, Désirée Coen Herak1, Snježana Prpić2, Andrea Prce3, Vanja Raščanec4, Renata Zadro5, Dunja Rogić1.
Abstract
INTRODUCTION: The aim of the study was the analytical verification of automated latex-enhanced particle immunoturbidimetric (LPIA) D-Dimer assay INNOVANCE D-dimer on Sysmex CS-5100 and Atellica COAG 360 analysers, and HemosIL D-dimer HS500 on ACL TOP 550, as well as the comparison with the enzyme-linked immunofluorescent assay (ELFA) on the miniVidas analyser.Entities:
Keywords: D-Dimer; blood coagulation tests; haemostasis; immunoturbidimetry; validation study
Mesh:
Substances:
Year: 2020 PMID: 33071556 PMCID: PMC7528643 DOI: 10.11613/BM.2020.030705
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Precision and bias of INNOVANCE D-Dimer assay on Sysmex CS-5100 and Atellica COAG 360, and HemosIL D-Dimer HS 500 assay on ACL TOP 550
| Level | Control 1 | Control 2 | Control 1 | Control 2 | Control 1 | Control 2 |
| Target value (mg/L FEU) | 0.30 | 2.69 | 0.31 | 2.81 | 0.60 | 1.97 |
| Mean ± SD (mg/L FEU) | 0.31 ± 0.01 | 2.77 ± 0.09 | 0.31 ± 0.01 | 2.76 ± 0.05 | 0.62 ± 0.05 | 2.01 ± 0.04 |
| Within-run CV (%) | 3.6 | 3.2 | 3.9 | 1.6 | 7.9 | 1.8 |
| Between-run CV (%) | 1.7 | 5.6 | 6.9 | 4.3 | 4.0 | 1.9 |
| Bias (%) | 3.3 | 3.0 | 0 | - 1.8 | 3.3 | 2.0 |
| SD - standard deviation. FEU - fibrinogen equivalent unit. CV - coefficient of variation. | ||||||
Figure 1Passing Bablok regression analysis for comparison of D-Dimer results between: (A) Sysmex CS-5100 vs. Atellica COAG 360, (C) Sysmex CS-5100 vs. ACL TOP 550 and (E) Atellica COAG 360 vs. ACL TOP 550. Corresponding scatter diagrams obtained by Bland-Altman analysis with mean biases are shown as B, D and F, respectively. CI - confidence interval. FEU - fibrinogen equivalent unit.
The results of Passing-Bablok regression analysis, Bland-Altman analysis and Spearman’s rank correlation coefficient for D-Dimer comparison of Sysmex CS-5100, Atellica COAG 360 and ACL TOP 550 with the miniVidas analyser
| Sysmex CS-5100 | 0.93 | - 0.33 | 1.60 | - 0.9 |
| Atellica COAG 360 | 0.95 | - 0.27 | 1.57 | 1.0 |
| ACL TOP 550 | 0.94 | - 0.16 | 1.37 | - 0.7 |
| ρ - Spearman’s rank correlation coefficient. CI - confidence interval. FEU - fibrinogen equivalent unit. | ||||
Agreement of D-Dimer results relative to the cut-off value of 0.5 mg/L FEU
| ACL TOP 550 | 63 | 96.8 | 0.93 | ACL TOP 550 | Sysmex CS-5100 / Atellica COAG 360 | |
| Sample 1 | 0.57 | 0.41 / 0.48 | ||||
| Sample 2 | 0.81 | 0.31 / 0.31 | ||||
| miniVidas | 43 | 97.7 | 0.85 | miniVidas | Sysmex CS-5100 / Atellica COAG 360 / ACL TOP 550 | |
| Sample 1 | 0.74 | 0.33 / 0.31 / 0.38 | ||||
| N - total number of compared samples. CI - confidence interval. FEU - fibrinogen equivalent unit. *Percentage of concordant D-Dimer results between compared analysers relative to the cut-off of 0.5 mg/L FEU. †The same two samples were discordant when comparing ACL TOP 550 with both Sysmex CS-5100 and Atellica COAG 360. ‡The same sample was discordant when comparing miniVidas with Sysmex CS-5100, Atellica COAG 360 and ACL TOP 550. | ||||||