Literature DB >> 19548843

Evaluation of the Innovance D-DIMER analytical performance.

Desiree Coen Herak1, Marija Milos, Renata Zadro.   

Abstract

BACKGROUND: Widespread use of D-dimer in recent years has led to the development of a number of new fully automated quantitative D-dimer assays.
METHODS: We evaluated the analytical performance of the particle-enhanced immunoturbidimetric assay Innovance D-DIMER (Siemens Medical Solutions) on the Behring Coagulation System (BCS) analyzer.
RESULTS: Within-run coefficients of variation (CVs) for samples with low, borderline, slightly, and extremely increased D-dimer concentrations were 2.1%-5.5%, whereas between-run CVs for control samples with low and extremely increased D-dimer were 5.5%-8.4%. The assay exhibited good linearity in the working range between 0.17 mg/L and 5.45 mg/L fibrinogen equivalent units (FEU), with the lower limit of detection of 0.099 mg/L FEU. The upper reference value determined in 40 plasma samples from healthy volunteers was 0.495 mg/L FEU. The results obtained in 457 fresh plasma samples were compared with results obtained with VIDAS D-Dimer Exclusion. Passing and Bablok regression analysis demonstrated highly significant correlation (y=1.370x-0.108, r=0.952, p<0.001). Bland and Altman difference plots demonstrated slightly higher results obtained with Innovance D-DIMER that was more pronounced with increasing values. Very good agreement between both assays was observed (kappa=0.860; 95% confidence interval (CI), 0.811-0.908).
CONCLUSIONS: This study demonstrates that Innovance D-DIMER fulfills all analytical requirements for daily routine use.

Mesh:

Substances:

Year:  2009        PMID: 19548843     DOI: 10.1515/CCLM.2009.219

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  7 in total

1.  [Age-adjusted D-dimer cut-offs to diagnose thromboembolic events: validation in an emergency department].

Authors:  N Verma; P Willeke; P Bicsán; P Lebiedz; H Pavenstädt; P Kümpers
Journal:  Med Klin Intensivmed Notfmed       Date:  2013-07-13       Impact factor: 0.840

2.  Prognostic value of D-dimer in elderly patients with Pulmonary Embolism.

Authors:  Hernan Polo Friz; Chiara Buzzini; Annalisa Orenti; Veronica Punzi; Mauro Molteni; Laura Primitz; Luca Cavalieri d'Oro; Guido Arpaia; Patrizia Boracchi; Claudio Cimminiello
Journal:  J Thromb Thrombolysis       Date:  2016-10       Impact factor: 2.300

3.  Comments on point of care D-dimer testing in the emergency department: a bioequivalence study.

Authors:  Suzanne Ekelund; Eric Heilmann
Journal:  Ann Lab Med       Date:  2013-12-06       Impact factor: 3.464

4.  Association of sleep problems with neuroendocrine hormones and coagulation factors in patients with acute myocardial infarction.

Authors:  Roland von Känel; Mary Princip; Jean-Paul Schmid; Jürgen Barth; Hansjörg Znoj; Ulrich Schnyder; Rebecca E Meister-Langraf
Journal:  BMC Cardiovasc Disord       Date:  2018-11-21       Impact factor: 2.298

5.  Verification of automated latex-enhanced particle immunoturbidimetric D-Dimer assays on different analytical platforms and comparability of test results.

Authors:  Ivana Lapić; Désirée Coen Herak; Snježana Prpić; Andrea Prce; Vanja Raščanec; Renata Zadro; Dunja Rogić
Journal:  Biochem Med (Zagreb)       Date:  2020-10-15       Impact factor: 2.313

6.  Women in sports: the applicability of common national reference intervals for inflammatory and coagulation biomarkers (HemSter Study).

Authors:  Vanja Radišić Biljak; Valentina Vidranski; Lana Ružić; Ana-Maria Simundic; Tihomir Vidranski
Journal:  Biochem Med (Zagreb)       Date:  2020-12-15       Impact factor: 2.313

7.  Evaluating the Use of a Negative D-Dimer and Modified Low Wells Score in Excluding above Knee Deep Venous Thrombosis in an Outpatient Population, Assessing Need for Diagnostic Ultrasound.

Authors:  Maryam Rahiminejad; Anshul Rastogi; Shirish Prabhudesai; David Mcclinton; Peter MacCallum; Sean Platton; Emma Friedman
Journal:  ISRN Radiol       Date:  2014-03-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.