BACKGROUND: Patellar tendon ruptures have routinely been repaired with transosseous suture tunnels. The use of knotless suture anchors for repair has been suggested as an alternative. PURPOSE: To compare the load to failure and gap formation of patellar tendon repair at the inferior pole of the patella with knotless suture anchor tape versus transosseous sutures. A secondary objective was to investigate whether either technique shows an association between bone density and load to failure. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 20 human tibias with attached patellar and quadriceps tendons were sharply incised at the bone-tendon junction at the inferior pole of the patella. A total of 10 tendons were repaired using 2 knotless suture anchors in the inferior pole of the patella and a single suture tape with 2 core sutures. The other 10 tendons were repaired using No. 2 suture passed through 3 transosseous tunnels. A distracting force was then applied through the suture in the quadriceps tendon. Gap distance through load cycling at the repair site and maximum load at repair failure were then measured. Bone density was measured using computed tomography scanning. RESULTS: No difference was found in the mean load to failure of knotless patellar tendon repair versus transosseous suture repair (367.6 ± 112.2 vs 433.9 ± 99 N, respectively; P = .12). After 250 cycles, the mean repair site gap distance was 0.85 ± 0.45 mm for the knotless patellar tendon repair versus 2.94 ± 2.03 mm for the transosseous suture repair (P = .03). A small correlation, although not statistically significant, was found between bone density and load to failure for the knotless tape repair (R 2 = 0.228; P = .66). No correlation was found between bone density and load to failure for the transosseous repair (R 2 = 0.086; P = .83). CONCLUSION: Suture tape repair with knotless anchors for repair of patellar tendon rupture has comparable load to failure with less gap formation than transosseous suture repair. There is a small correlation between bone density and failure load for knotless anchor repair, which may benefit from further investigation. CLINICAL RELEVANCE: Using knotless suture anchors for patellar tendon rupture repair would allow for a smaller incision, less dissection, and likely shorter operating time.
BACKGROUND: Patellar tendon ruptures have routinely been repaired with transosseous suture tunnels. The use of knotless suture anchors for repair has been suggested as an alternative. PURPOSE: To compare the load to failure and gap formation of patellar tendon repair at the inferior pole of the patella with knotless suture anchor tape versus transosseous sutures. A secondary objective was to investigate whether either technique shows an association between bone density and load to failure. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 20 human tibias with attached patellar and quadriceps tendons were sharply incised at the bone-tendon junction at the inferior pole of the patella. A total of 10 tendons were repaired using 2 knotless suture anchors in the inferior pole of the patella and a single suture tape with 2 core sutures. The other 10 tendons were repaired using No. 2 suture passed through 3 transosseous tunnels. A distracting force was then applied through the suture in the quadriceps tendon. Gap distance through load cycling at the repair site and maximum load at repair failure were then measured. Bone density was measured using computed tomography scanning. RESULTS: No difference was found in the mean load to failure of knotless patellar tendon repair versus transosseous suture repair (367.6 ± 112.2 vs 433.9 ± 99 N, respectively; P = .12). After 250 cycles, the mean repair site gap distance was 0.85 ± 0.45 mm for the knotless patellar tendon repair versus 2.94 ± 2.03 mm for the transosseous suture repair (P = .03). A small correlation, although not statistically significant, was found between bone density and load to failure for the knotless tape repair (R 2 = 0.228; P = .66). No correlation was found between bone density and load to failure for the transosseous repair (R 2 = 0.086; P = .83). CONCLUSION: Suture tape repair with knotless anchors for repair of patellar tendon rupture has comparable load to failure with less gap formation than transosseous suture repair. There is a small correlation between bone density and failure load for knotless anchor repair, which may benefit from further investigation. CLINICAL RELEVANCE: Using knotless suture anchors for patellar tendon rupture repair would allow for a smaller incision, less dissection, and likely shorter operating time.
Ruptures of the patellar tendon are rare, accounting for approximately 0.6% of soft
tissue musculoskeletal injuries.[4] They are most common in males younger than 40 years of age and are usually the
result of direct trauma.[1] Zernicke et al[23] estimated that a healthy tendon can withstand up to 17.5 times a subject’s body
weight before failing; however, other factors may predispose patients to tendon rupture,
such as chronic renal failure, rheumatoid arthritis, repetitive microtrauma, lupus
erythematous, fluoroquinolone use, or history of steroid use.[16] Patellar tendon ruptures are categorized into 3 types by location of the rupture:
inferior pole of the patella, midsubstance, and tibial tuberosity. The inferior pole of
the patella is also referred to as the osteotendinous junction, and ruptures in that
location are the most common.[2,14] Therefore, we focused our study on repairs of this type of rupture. Surgery is
required to fix complete patellar tendon ruptures at the osteotendinous junction.[12,15,21] Surgical options include transosseous sutures, primary end-to-end opposition with
protective metal wiring, and suture anchors.[12]The most commonly used method for a patellar tendon rupture at the inferior pole of the
patella is repair via transosseous sutures.[10] However, this method is quite invasive and requires exposure of the entire
patella and multiple passes through the intact patellar tendon. This may damage the
patellar tendon and lengthen recovery time.[11] Other drawbacks associated with this procedure include risk of penetrating the
articular surface of the patella and the technical difficulty of attaching the patellar
tendon anteriorly over the inferior pole of the patella.[11]The suture anchor repair is a newer technique that has become commonly and successfully
used in the upper extremity for tendon and ligament attachment to bone.[2] Suture anchors provide several advantages over the traditional transosseous
approach, including the need for a smaller incision, less tissue dissection, shorter
operation time, and no involvement of the intact quadriceps tendon.[3] Suture anchors are also believed to allow earlier return of range of motion and a
faster recovery compared with other techniques.[6] In addition to these clinical benefits, suture anchors have been shown to be
biomechanically comparable with, if not superior to, the traditional transosseous technique.[2,6,14]Another promising, newer surgical technique uses suture tape. Suture tape is composed of
similar materials as high-tensile strength suture but is wider and therefore has more
surface area. The increased surface area in contact with the tissue theoretically allows
for more dispersion of force throughout the repaired tendon.[8] In a head-to-head comparison, Gnandt et al[9] demonstrated that suture tape had a greater mean load to failure than suture of
the same size in both whipstitch and Krackow-stitch models. Using a quadriceps tendon
model, Kindya et al[13] found that suture anchors with suture tape significantly outperformed
transosseous repair and standard suture anchor repair in initial tendon displacement,
stiffness, and ultimate load to failure. Although that study demonstrated the benefit of
a knotless suture tape repair for the quadriceps tendon, the investigators did not
analyze the same suture tape technique on the patellar tendon.[13]The aim of our study was to compare the load to failure and gap formation of patellar
tendon repair at the inferior pole of the patella with knotless suture anchor tape
versus transosseous sutures. A secondary objective was to investigate whether either of
these techniques would show an association between bone density and load to failure. We
hypothesized that there would be no difference in load to failure between knotless
anchor fixation using suture tape versus transosseous repair for inferior pole patellar
tendon rupture and that knotless tape repair would result in less gap formation.
Additionally, we hypothesized that there would be a correlation between load to failure
and bone density with the suture anchor repair.
Methods
The biomechanics laboratory at Louisiana State University Health provided 20
fresh-frozen cadaveric human tibias with attached patellar tendon, patella, and
quadriceps tendon harvested from fresh-frozen cadavers. Before testing, computed
tomography (CT) scanning was performed to establish bone density. After CT scanning,
the patellar bones were segmented in 3D Slicer (www.slicer.org), and mean density
was measured in Hounsfield units (HU) using the built-in quantification tool.[7]The tibial components of the specimens were potted using Bondo polyester resin (Bondo
Corporation) in metal boxes. The patellar tendons were sharply incised with a
scalpel at the bone-tendon junction at the inferior pole of the patella. Excess
tissue was then debrided from the inferior pole of the patella to allow for accurate
placement of sutures and anchors as well as to allow for more accurate testing
measurements.A total of 10 patellar tendons were repaired using a single suture tape (Fibertape,
Arthrex, Naples, FL) and 2 knotless suture anchors (4.75 mm; SwiveLock, Arthrex)
placed in the inferior pole of the patella, based on a method previously described.[13] To do this, the suture tape was started by passing the attached needle
through the proximal end of the tendon, either medially or laterally. Next, 5
Krackow stitches were made with the tape running distally down the side of the
tendon. Another 5 Krackow stitches were then made with the tape running proximally
up the opposite side of the tendon, leaving 2 strands coming out of the proximal
aspect of the tendon. All the suture in the soft tissue was pretensioned by the same
surgeon (P.A.M.) A drill and tap were then used at the junction of the medial and
middle third and junction of the lateral and middle third of the inferior pole of
the patella to prepare the bone for the suture anchor. Each end of the suture tape
was loaded within the eyelet of the suture anchor going into the respective side of
the patella. It should be noted that the tape was initially loaded with all of the
slack pulled out of the tape so that the eyelet was juxtaposed to the cut end of the
tendon. When malleting the anchor handle, it was observed that the tape cut into the
bone and caused the anchor to pull out. To prevent this, the tape was loaded with 20
mm of slack for all repairs to accommodate the 20-mm insertion stem for the 19.1--mm
length anchor (Figure
1).
Figure 1.
A knotless anchor patellar repair. The black arrow denotes the 20 mm of slack
preloaded into the polyethylene tape between the eyelet and the tendon. The
tape is then locked on the slots of the insertion handle before insertion of
the metal shaft into the predrilled patellar holes. This slack was preloaded
to prevent the tape from cutting into the bone.
A knotless anchor patellar repair. The black arrow denotes the 20 mm of slack
preloaded into the polyethylene tape between the eyelet and the tendon. The
tape is then locked on the slots of the insertion handle before insertion of
the metal shaft into the predrilled patellar holes. This slack was preloaded
to prevent the tape from cutting into the bone.The slack was preloaded so that before insertion of the anchor, there was 20 mm of
tape between the eyelet and the tape-tendon junction. When the driver was inserted
fully, this placed the tendon directly on the patellar bone without the tape cutting
into the inferior pole of the patella. The final portion of the anchor was then
screwed into the patella to complete its placement. The surgeon passed the
additional core suture through the proximal tendon and tied the suture using a
surgeon’s knot with 7 throws (Figures 2A and 3A). The single suture tape technique was selected, as it has been
described by previous surgeons.[13] Also, the suture tape offers the benefit of minimizing the suture burden for
the tendon, and it would be difficult to fit 2 sets of suture tape in a single
tendon.
Figure 2.
Three-dimensional illustration of patella tendon repair techniques. (A) Left
patellar tendon knotless tape repair with 1 polyethylene tape and 2 knotless
anchors inserted into the inferior pole of the patella. The repair is
supplemented with 2 core sutures from each anchor. (B) Left patellar tendon
transosseous repair. A total of 3 transosseous holes are drilled through the
patella, and 4 sutures are passed through these holes with 2 knots tied over
the superior pole of the patella.
Figure 3.
Patella tendon cadaver repair constructs for biomechanical testing. (A) A
left patellar tendon knotless tape repair with 1 polyethylene tape and 2
knotless anchors inserted into the inferior pole of the patella. The repair
is supplemented with 2 core sutures from each anchor. (B) A right patellar
tendon transosseous repair. A total of 3 transosseous holes are drilled
through the patella, and 4 sutures are passed through these holes with 2
knots tied over the superior pole of the patella.
Three-dimensional illustration of patella tendon repair techniques. (A) Left
patellar tendon knotless tape repair with 1 polyethylene tape and 2 knotless
anchors inserted into the inferior pole of the patella. The repair is
supplemented with 2 core sutures from each anchor. (B) Left patellar tendon
transosseous repair. A total of 3 transosseous holes are drilled through the
patella, and 4 sutures are passed through these holes with 2 knots tied over
the superior pole of the patella.Patella tendon cadaver repair constructs for biomechanical testing. (A) A
left patellar tendon knotless tape repair with 1 polyethylene tape and 2
knotless anchors inserted into the inferior pole of the patella. The repair
is supplemented with 2 core sutures from each anchor. (B) A right patellar
tendon transosseous repair. A total of 3 transosseous holes are drilled
through the patella, and 4 sutures are passed through these holes with 2
knots tied over the superior pole of the patella.A total of 10 tendons were repaired using two No. 2 high-strength sutures (Fiberwire,
Arthrex, Naples, Florida) passed through transosseous tunnels. Krackow suturing was
performed in the same fashion as the tendons repaired with suture anchor. However,
each suture was run proximally up the central aspect of the tendon for 5 throws,
again using 5 Krackow stitches, leaving 4 suture tails. A 2.0-mm drill was then used
to create 3 longitudinal tunnels in the medial, lateral, and central aspects of the
patella from distal to proximal. A beath pin was used to pass the suture ends
through their respective tunnels. The 2 central strands were both passed through the
middle tunnel. The suture ends were then pulled taut so that the tendon was
reapproximated to the inferior pole of the patella. The ends were then tied to their
respective ends over the superior aspect of the patella using a surgeon’s knot with
7 throws (Figures 2B and
3B). The transosseous
technique was used as a control because it is consistent with multiple described,
previously tested techniques.[6,14] All knots were tied by the same orthopaedic surgeon (P.A.M.), who also tied
the knots for the tendons repaired with tape and anchors. Two No. 2 high-strength
sutures were then passed in a similar fashion, using the Krackow stitch, through the
intact quadriceps tendon. A loop was left in the ends of both tied sutures to allow
for attachment of the testing devices.The repairs were tested using a servohydraulic testing machine (model 8874; Instron).
The tibial bones were anchored to the machine frame while the quadriceps tendon was
loaded in tension through a No. 2 high-strength suture, which was Krackow stitched
through the quadriceps tendon (Figure 4). The knees were tested in a simulated fully extended position.
This position was chosen to decrease the variability in the data. In addition, at
our institution, the postoperative protocol for patellar tendon repair includes
initial immobilization with the knee locked in extension, so our study mimics this
time-zero, early healing phase. Our study methods were similar to the studies
performed by Ettinger et al[6] and Sherman et al,[19] which allows our results to be compared with previously published
biomechanical data.
Figure 4.
A right transosseous patellar tendon repair specimen loaded in the Instron
testing machine. The tibia is potted, and the quadriceps is attached to a
tension device via suture.
A right transosseous patellar tendon repair specimen loaded in the Instron
testing machine. The tibia is potted, and the quadriceps is attached to a
tension device via suture.Before testing, each construct was tensioned at 80 N to tighten the sutures. The
testing was performed in 2 phases. In the first phase, we tested the repair for
fatigue applying a sinusoidal load ranging from 20 to 60 N at a frequency of 1 Hz
for a total of 250 cycles. The second phase evaluated the strength to failure by
applying a ramp load at a rate of 20 mm/s. During testing, the tendon gap was
measured using a 3-dimensional snapshot sensor (Gocator 3506) configured to capture
a field of view of 27 × 45 mm with a resolution of 20 µm and an accuracy of 12 μm.
To reduce glare during image acquisition, the tissues were sprayed with Zinc
Stearate Facsure NDT Developer (Weld-Aid Products). Tendon gapping was monitored at
0, 20, 125, and 250 cycles. To measure the tendon gap, the middle third of the
patellar bone was discretized in vertical sections with 100-µm spacing, and the
largest measured gap was used as measure for the study (Figure 5).
Figure 5.
Right transosseous patellar tendon repair. The patella is colored pink and
red while the patellar tendon is green. Construct shape and deformation were
captured by a high-resolution 3-dimensional scanner. (A) Color mapping
images of a patellar tendon repair. The distance represents the distance
along the axis of the scanner. (B) Color mapping images of patellar tendon
gapping after failure. The white rectangle is around the area with the
largest measured gap within the central third of the patella.
Right transosseous patellar tendon repair. The patella is colored pink and
red while the patellar tendon is green. Construct shape and deformation were
captured by a high-resolution 3-dimensional scanner. (A) Color mapping
images of a patellar tendon repair. The distance represents the distance
along the axis of the scanner. (B) Color mapping images of patellar tendon
gapping after failure. The white rectangle is around the area with the
largest measured gap within the central third of the patella.Additional values of stiffness and energy to failure were calculated with the
resultant force-displacement graphs.
Statistical Analysis
An a priori power analysis was performed to determine the number needed in each
group, as well as a post hoc power analysis. Descriptive data such as age and
sex of the cadavers were compared using 2-sample t test and
chi-square test, respectively. Biomechanical data such as load, stiffness, and
gapping were compared between the 2 groups using 2-sample t
test. The relationship between bone density (HU) and the load to failure was
determined using the coefficient of determination (R
2).
Results
Drawing on a previous study[13] that compared knotless suture tape tendon repair versus transosseous suture
for the quadriceps tendon, we performed an a priori power analysis.[12,17,23] The number needed for a 1 – B of 0.8 was determined to be 5 in each group. A
post hoc power analysis on gapping after 250 cycles was performed. With a sample
size of 10 in each group, it was determined that 1 – B was 0.94.The mean age of the cadaveric specimens in the transosseous suture repair group was
62.1 ± 4.1 years, which was not significantly different from the age of 65.5 ± 12
years in the knotless repair group (P = .40). There were 6 female
and 4 male cadaveric knees in the transosseous suture repair group versus 7 female
and 3 male knees in the knotless repair group, with no significant difference in sex
distribution between groups (P = .64).
Load to Failure
No difference was found in the mean load to failure of knotless patellar tendon
repair versus transosseous suture repair: 367.6 ± 112.2 versus 433.9 ± 99 N,
respectively (P = .12) (Figure 6). All 10 of the knotless repairs
failed due to screw pullout, whereas all 10 of the transosseous suture repairs
failed due to suture breakage. A significant difference was found in the mode of
failure between both groups (P < .001). Stiffness in the 2
groups was 36.7 ± 10 versus 33.7 ± 8.5 N/mm, respectively (P =
.7). Energy to maximum load in the 2 groups was 10.6 ± 6.5 versus 10.8 ± 4.4 J,
respectively (P = .9).
Figure 6.
Load to failure for knotless anchor repair with polyethylene tape versus
transosseous suture repair. No difference was found in the mean load to
failure of knotless patellar tendon repair versus transosseous suture
repair (P = .12). SD is represented by error bars.
Load to failure for knotless anchor repair with polyethylene tape versus
transosseous suture repair. No difference was found in the mean load to
failure of knotless patellar tendon repair versus transosseous suture
repair (P = .12). SD is represented by error bars.
Gapping
After completion of the repair, the mean repair gap was 0.30 ± 0.18 mm for the
knotless patellar tendon repair versus 0.89 ± 1.17 mm for the transosseous
suture repair. No statistical difference was found in the tendon repair gap
distance immediately after the repair between both groups (P =
.23). The knotless patellar tendon repair had a significantly smaller gap
distance as soon as 20 cycles (P = .04) as well as at 125
cycles (P = .015). After 250 cycles, the increase in gap
distance was larger for the transosseous suture repair versus the knotless
patellar tendon repair (increase of 2.16 ± 1.25 vs 0.55 ± 0.44 mm;
P = .01). After 250 cycles, the mean repair site gap
distance was 0.85 ± 0.45 mm for the knotless patellar tendon repair versus 2.94
± 2.03 mm for the transosseous suture repair (P = .03) (Figure 7). Although the
knotless group had less gapping, both the knotless group and transosseous group
had increased gapping after 250 cycles (P = .01 and
P = .001, respectively).
Figure 7.
Patellar tendon gapping after transosseous versus knotless repair.
Gapping was measured after 0, 20, 125, and 250 cycles. After 250 cycles,
the increase in gap distance was larger for the transosseous suture
repair versus the knotless patellar tendon repair (increase of 2.16 ±
1.25 vs 0.55 ± 0.44 mm, respectively; P = .01). SD is
represented by error bars.
Patellar tendon gapping after transosseous versus knotless repair.
Gapping was measured after 0, 20, 125, and 250 cycles. After 250 cycles,
the increase in gap distance was larger for the transosseous suture
repair versus the knotless patellar tendon repair (increase of 2.16 ±
1.25 vs 0.55 ± 0.44 mm, respectively; P = .01). SD is
represented by error bars.
Bone Density
No difference was found in the bone density of the patellae in the knotless
repair versus transosseous repair groups (P = .39) (see Table 1). We noted a
small correlation, although not statistically significant, between bone density
and load to failure for the knotless tape repair (R
2 = 0.228; P = .66). No correlation was seen between
bone density and load to failure for the transosseous repair (R
2 = 0.086; P = .83).
TABLE 1
Patellar Bone Density in Hounsfield Units (HU) for Knotless Anchor Versus
Transosseous Suture Patellar Tendon Repair
Repair Type
Bone Density, HU, mean ± SD
R2
P Value
Knotless anchors
270 ± 53
0.228a
.66
Transosseous suture
234 ± 53
0.086a
.83
P value
.39
Correlation coefficient of each group is reported for the
relationship of bone density and load to failure.
Patellar Bone Density in Hounsfield Units (HU) for Knotless Anchor Versus
Transosseous Suture Patellar Tendon RepairCorrelation coefficient of each group is reported for the
relationship of bone density and load to failure.
Discussion
The current study showed the knotless anchor tape repair had less repair site gap
formation compared with the transosseous suture repair; however, there was no
significant difference in load to failure. Additionally, we found no significant
difference in stiffness or energy until failure.Previous biomechanical comparisons between knotted suture anchors and transosseous
sutures have measured the gap formation and the mean load to failure for patellar
tendon repair. Overall, the results have been mixed. For the mean load to failure,
Bushnell et al[2] found no significant difference among their suture anchor group (779 N),
transosseous Fiberwire group (730 N), and transosseous Ethibond group (763 N).
Ettinger et al[6] tested 2 suture anchor groups, titanium (597 N) and hydroxyapatite (689 N),
with both having a significantly higher average load to failure than their
transosseous group (301 N). Lanzi et al[14] also found higher average load to failure in their suture anchor group (669
N) than in their transosseous group (582 N).Sherman et al[19] compared a 3-corkscrew anchor group (258.8 N) versus a transosseous tunnel
group (287 N) with no significant differences in gapping or load to failure. Similar
to Bushnell et al[2] and Sherman et al,[19] we found no significant difference in mean load to failure between our suture
anchor group (367 N) and our transosseous group (433 N). Our load-to-failure values
for knotless anchor were lower than those reported by Bushnell et al; however, 4
Fiberwire sutures were used in that study, whereas we used a single suture tape for
the anchor repairs in the current study. Interestingly, our values were higher than
those reported by Sherman et al,[19] even though their study used 3 anchors versus 2 in our study. The main
differences in this study are the use of suture tape as well as the knotless anchor
fixation type of the tape versus a threaded corkscrew anchor. Additionally,
different methods of testing the repair may account for the difference in reported
values. Another aspect to consider for both studies is increased suture burden to
the soft tissue when more suture strands are used.Bushnell et al[2] found that suture anchors (4.1 mm) had significantly less gap formation than
either the Fiberwire (6.7 mm) or Ethibond (8.5 mm) transosseous groups at 250
cycles. Ettinger et al[6] also found significantly less gap formation in both of their suture anchor
groups, titanium (2.9 mm) and hydroxyapatite (3.7 mm), compared with the
transosseous group (17.2 mm) over the first 20 cycles. Lanzi et al[14] found that suture anchors (2.16 mm) had significantly less gap formation than
transosseous sutures (5.71 mm) after 1000 cycles. Sherman et al[19] tested patellar tendon suture anchor repair versus transosseous repair,
showing no difference in gapping at the end of their study (2.92 and 3.33 mm,
respectively; P = .120).The transosseous repair group in our study had a large amount of gapping variability,
whereas the knotless tape repair group had much less variability. This may be due to
asymmetric tensioning of suture limbs in the transosseous group. This increased
variability may also be due to the increased working length of the suture compared
with the knotless tape repair; perhaps the much longer suture material that is run
through the entire length of the patella allows more length for variability.Although our study showed less gap formation with the anchor tape repair, our gap
values were also smaller after 250 cycles compared with the studies that tested to
similar cycle numbers.[19] These differences may be accounted for by individual repair technique, method
of measuring gap, or the technique used to stress the repair. Gelberman et al[8] found that gapping of >3 mm had a significant effect on accrual of
strength and stiffness with time in tendon repairs.One study evaluated knotless suture tape repair of the quadriceps tendon but did not
evaluate patellar tendon repairs. Kindya et al[13] compared transosseous tunnels, threaded anchors, and knotless anchors with
suture tape. They demonstrated results similar to those of our study, with the
knotless tape construct showing less gapping than the other groups. However, Kindya
et al showed a higher load to failure for the knotless tape group, whereas our study
showed no difference with respect to load to failure. Sherman et al[20] showed that a threaded suture anchor repair had less gap formation than a
transosseous suture (7.7 vs 10.7 mm, respectively), although they reported similar
ultimate loads to failure (286.0 vs 250.5 N, respectively). However, Sherman et al[20] used 3 suture anchors with 3 Fiberwire sutures for the suture anchor group
and used 3 Fiberwire sutures for the transosseous suture group. Our study used only
2 suture anchors and a suture tape, instead of wire, to balance the theoretical
trade-offs of distributing stress within the tendon while minimizing further tissue
damage with more needle passes.Although several previous biomechanical comparisons have found superiority in suture
anchors compared with transosseous suture[2,6,14] as well as suture tape compared with traditional suture,[9] we found that they were comparable, with the exception of repair site
gapping. Although evidence indicates that suture anchors can outperform transosseous
sutures biomechanically, Roudet et al[18] suggested that the clinical evidence thus far does not demonstrate
superiority of either technique. Regardless of whether suture anchors are
biomechanically superior or noninferior to transosseous sutures, suture anchors are
less invasive, involve less dissection, and entail a shorter operating time, making
them an enticing option for surgeons. More specifically, the transosseous tunnels
require access to the proximal and distal aspects of the patella, requiring longer
and larger dissection and subsequent closure.[5] Suture anchor fixation has also shown favorable clinical results. Colombelli
et al[5] reported good to excellent results at 12-month follow-up in patients with
quadriceps or patellar tendon rupture treated with suture anchor fixation, with no
major complications. However, further clinical trials should be performed comparing
transosseous sutures versus suture anchors with suture tape to determine whether the
surgical advantages are worth the increased cost and whether suture tape is more
effective in pathologic tendons.Our secondary research question analyzed the bone density of our specimens to see
whether there was a correlation with load to failure. In the upper extremity, it has
been demonstrated that increased bone density correlates with increased pull-out
strength of suture anchors.[22] Our results did show a small correlation between load to failure and bone
density in repairs performed with anchors and suture tape. This relationship was not
present between bone density and transosseous repairs. This may indicate that suture
anchor repair would have a higher load to failure for younger patients with higher
bone density; the cadaveric specimens we used were from people in their 60s, whereas
injured patients are typically younger. Further studies could be performed to see
whether there is an age–bone density correlation with anchor fixation or whether
there is a density threshold for the use of suture anchors.
Limitations
Limitations to our study include the use of cadaveric tissue, which is a
limitation shared by many biomechanical studies. First, cadaveric tissue does
not exhibit any healing processes that might affect outcomes as clinical trials
progress; however, it does simulate a time zero repair. Second, although being
able to strip away all other connective tissue except the tendon being tested is
useful in isolating the strength of the repair, it does not necessarily directly
correlate with how a repair will perform when acted upon by the surrounding
tissue, such as the extensor retinaculum, in a real patient. Third, we
manufactured these tendon ruptures in donor tendons. Therefore, it was not known
whether predisposing conditions that can result in a patellar tendon rupture or
attenuation were present, which could affect the suture-tendon interface.
Although suture tape with suture anchors was comparable with transosseous
sutures in our study, perhaps in clinical trials where some of the affected
tendons have preexisting pathology, the extra width and surface area of the
suture tape will allow it to have a more secure suture-tendon interface.Another set of limitations entailed the 2 techniques used. The knotless suture
tape repair involved a single suture tape with 2 core sutures, whereas the
transosseous repair used 2 sets of suture. As mentioned previously, the single
suture tape technique was selected because it has been described by previous surgeons.[13] Also, the suture tape offers the benefit of minimizing the suture burden
for the tendon, and it would be difficult to fit 2 sets of suture tape in a
single tendon. The transosseous technique was used as a control because it is
consistent with multiple described, previously tested techniques.[6,14]
Conclusion
Suture tape repair with knotless anchors for patellar tendon rupture repair has
comparable load to failure as transosseous suture repair, with less gap formation.
We found a small correlation between bone density and failure load for knotless
anchor repair that may benefit from further investigation.
Authors: Max Ettinger; Antonios Dratzidis; Christof Hurschler; Stephan Brand; Tilman Calliess; Christian Krettek; Michael Jagodzinski; Maximilian Petri Journal: Am J Sports Med Date: 2013-08-27 Impact factor: 6.202
Authors: Joseph T Lanzi; Justin Felix; Christopher J Tucker; Kenneth L Cameron; John Rogers; Brett D Owens; Steven J Svoboda Journal: Am J Sports Med Date: 2016-05-13 Impact factor: 6.202