A simple manual method to estimate water-equivalent diameter for calculating size-specific dose estimate in chest computed tomography.

Dimitris Mihailidis1, Virginia Tsapaki2, Pelagia Tomara3.   

Abstract

OBJECTIVES: The American Association of Physicists in Medicine (AAPM) Task Groups (TG) 204 and 220 introduced a method to estimate patient dose by introducing the Size-Specific Dose Estimate (SSDE). They provided patient size-specific conversion factors that could be applied to volumetric CT Dose Index CTDIvol to estimate patient dose in terms of SSDE based on either effective diameter (Deff) or water equivalent diameter (Dw). Our study presented an alternative method to manually estimate SSDE for the everyday clinical routine chest CT that can be readily used and does not require sophisticated computer programming.
METHODS: For 16 adult patients undergoing chest CT, the method employed an average relative electron density (ρelung = 0.3) for the lung tissue and a ρetissue of 1.0 for the other tissues to scale the lateral thickness and compute the effective lateral thickness on the patient's axial image. The proposed method estimated a "corrected" Deff (Deffcorr) to replace Dw and compared results with TG220 and a second method proposed by Huda et al, for the same set of CT studies.
RESULTS: The results showed comparable behavior for all methods. There is overall agreement especially between this study and TG220. Largest differences were +13.3% and+15.9% from TG220 and Huda values, respectively. Patient size correlation showed strong correlation with the TG220 and Huda et al methods.
CONCLUSIONS: A simple, quick manual method to estimate CT patient radiation dose in terms of SSDE was proposed as an alternative where sophisticated computer programming is not available. It can be readily used during any clinical chest CT scanning. ADVANCES IN KNOWLEDGE: The paper is novel as it presents simple, quick manual method to estimate CT patient radiation dose in chest imaging. The process can be used as alternative in cases no sophisticated computer programming is available.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 33058736      PMCID: PMC7774705          DOI: 10.1259/bjr.20200473

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  17 in total

1.  Relative electron density calibration of CT scanners for radiotherapy treatment planning.

Authors:  S J Thomas
Journal:  Br J Radiol       Date:  1999-08       Impact factor: 3.039

2.  NCRP Report No. 160, Ionizing Radiation Exposure of the Population of the United States, medical exposure--are we doing less with more, and is there a role for health physicists?

Authors:  D A Schauer; O W Linton
Journal:  Health Phys       Date:  2009-07       Impact factor: 1.316

3.  Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images.

Authors:  Jia Wang; Jodie A Christner; Xinhui Duan; Shuai Leng; Lifeng Yu; Cynthia H McCollough
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

4.  The direct use of CT numbers in radiotherapy dosage calculations for inhomogeneous media.

Authors:  R P Parker; P A Hobday; K J Cassell
Journal:  Phys Med Biol       Date:  1979-07       Impact factor: 3.609

5.  Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter.

Authors:  Shuai Leng; Maria Shiung; Xinhui Duan; Lifeng Yu; Yi Zhang; Cynthia H McCollough
Journal:  Radiology       Date:  2015-02-25       Impact factor: 11.105

6.  Lung density as measured by computerized tomography: implications for radiotherapy.

Authors:  J Van Dyk; T J Keane; W D Rider
Journal:  Int J Radiat Oncol Biol Phys       Date:  1982-08       Impact factor: 7.038

7.  Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part I. Development and validation of methods using the CT image.

Authors:  Jia Wang; Xinhui Duan; Jodie A Christner; Shuai Leng; Lifeng Yu; Cynthia H McCollough
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

8.  Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.

Authors:  Christiane S Burton; Timothy P Szczykutowicz
Journal:  J Appl Clin Med Phys       Date:  2017-11-27       Impact factor: 2.102

9.  Comparison of methods to estimate water-equivalent diameter for calculation of patient dose.

Authors:  Andrew Daudelin; David Medich; Syed Yasir Andrabi; Chris Martel
Journal:  J Appl Clin Med Phys       Date:  2018-07-07       Impact factor: 2.102

10.  Automated Calculation of Water-equivalent Diameter (DW) Based on AAPM Task Group 220.

Authors:  Choirul Anam; Freddy Haryanto; Rena Widita; Idam Arif; Geoff Dougherty
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

View more
  1 in total

1.  Real-time fully automated dosimetric computation for CT images in the clinical workflow: A feasibility study.

Authors:  Massimiliano Porzio; Choirul Anam
Journal:  Front Oncol       Date:  2022-08-11       Impact factor: 5.738

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.