| Literature DB >> 33042886 |
Njwan F Shehab1, Nawal A Al-Sabawi1, Emad F Alkhalidi1.
Abstract
BACKGROUND AND AIMS: Rising effects of temperature due to laser use during root canal disinfection may harm periodontium and alveolar bone. Therefore, the aim of this study was to estimate the surface root temperature of lower incisors throughout the application of different power levels and times of an 810-nm diode laser.Entities:
Keywords: Diode laser; root canal surface; temperature; thermocouple
Year: 2020 PMID: 33042886 PMCID: PMC7523933 DOI: 10.4103/jispcd.JISPCD_216_20
Source DB: PubMed Journal: J Int Soc Prev Community Dent ISSN: 2231-0762
A comparison in temperature rises between middle and apical region of root surface at the selected time for each output power
| Power (W) | Time (s) | Tooth segment | Mean (°C) ± SD | |
|---|---|---|---|---|
| 1.05 | 20 | Middle | 1.7800 ± 0.5206 | NS |
| Apical | 1.9400 ± 0.7519 | |||
| 60 | Middle | 3.1500 ± 0.4424 | 0.014* | |
| Apical | 3.6600 ± 0.4626 | |||
| 1.5 | 20 | Middle | 1.9300 ± 0.6098 | NS |
| Apical | 2.5000 ± 0.4163 | |||
| 60 | Middle | 3.2200 ± 0.5078 | NS | |
| Apical | 3.7300 ± 0.6292 | |||
| 1.95 | 20 | Middle | 2.5300 ± 0.6738 | 0.003* |
| Apical | 3.9100 ± 0.5021 | |||
| 60 | Middle | 3.9600 ± 0.6299 | NS | |
| Apical | 4.2000 |
SD = standard deviation, NS = not significant
Paired t test to compare between middle and apical segment of root surface for each level of time at each output power
*Significant difference existed at P≤ 0.05
Figure 1Thermal changes in the middle and apical regions at all output powers and times on the external root surface
A comparison of temperature (°C) rises at different output powers of 810 nm diode laser at the certain time
| Time (s) | Tooth segment | Power (W) | Temperature rise Mean (°C) ± SD | ||
|---|---|---|---|---|---|
| 20 | Middle | 1.05 | 1.7800 ± 0.52026 | 4.314 | 0.024 |
| 1.5 | 1.9300 ± 0.60928 | ||||
| 1.95 | 2.5300 ± 0.67338 | ||||
| Apical | 1.05 | 1.9400 ± 0.75159 | 31.215 | 0.000 | |
| 1.5 | 2.5000 ± 0.41633 | ||||
| 1.95 | 3.9100 ± 0.50211 | ||||
| 60 | Middle | 1.05 | 3.1500 ± 0.44284 | 7.114 | 0.003 |
| 1.5 | 3.2200 ± 0.50728 | ||||
| 1.95 | 3.9600 ± 0.62929 | ||||
| Apical | 1.05 | 3.6600 ± 0.46236 | 2.038 | NS | |
| 1.5 | 3.7300 ± 0.62902 | ||||
| 1.95 | 4.2000 ± 0.81240 |
SD = standard deviation, NS = not significant
* For each level of tooth segment, different letters vertically at each time have significant difference at P ≤ 0.05 according to Duncan test
Independent t test for temperature rises of 810 nm diode laser at specific output power between different times at each the middle and apical root surface
| Tooth segment | Power (W) | Time (s) | Mean (°C) ± SD | |
|---|---|---|---|---|
| Middle | 1.05 | 20 | 1.7800 ± 0.5206 | 0.000* |
| 60 | 3.1500 ± 0.4424 | |||
| 1.5 | 20 | 1.9300 ± 0.6098 | 0.000* | |
| 60 | 3.2200 ± 0.5078 | |||
| 1.95 | 20 | 2.5300 ± 0.6738 | 0.000* | |
| 60 | 3.9600 ± 0.6299 | |||
| Apical | 1.05 | 20 | 1.9400 ± 0.7519 | 0.000* |
| 60 | 3.6600 ± 0.4626 | |||
| 1.5 | 20 | 2.5000 ± 0.4163 | 0.000* | |
| 60 | 3.7300 ± 0.6292 | |||
| 1.95 | 20 | 3.9100 ± 0.5021 | NS | |
| 60 | 4.2000 ± 0.8120 |
SD = standard deviation, NS = not significant
*Significant difference existed at P ≤ 0.05