Agreement between the STANDARD F COVID-19 Ag FIA and the RT-PCR assay results for 359 nasopharynx samples tested
| Detected | Not detected | Positive per cent agreement (95% CI) | Negative per cent agreement (95% CI) | |
|---|---|---|---|---|
| Positive samples ( | ||||
| STANDARD F COVID-19 Ag FIA | 49 | 55 | 47.1 (37.1–57.1) | — |
| RT-PCR assay | 104 | 0 | NA | — |
| Negative samples ( | ||||
| STANDARD F COVID-19 Ag FIA | 4 | 251 | — | 98.4 (96.0–99.6) |
| RT-PCR assay | 0 | 255 | — | NA |
CI, confidence interval; NA, not applicable.
This value corresponds to the assay's sensitivity assuming all 104 samples being from individuals diagnosed with COVID-19.
This value corresponds to the assay's specificity assuming all 255 samples being from individuals not diagnosed with COVID-19.