Rajesh Sinha1, Pranita Sahay1, Rohit Saxena1, Nidhi Kalra1, Vinay Gupta1, Jeewan S Titiyal1.
Abstract
PURPOSE: To evaluate the visual outcomes of bilateral implantation of a new hydrophobic foldable extended depth of focus (EDOF) IOL.Entities:
Keywords: Cataract surgery; extended depth of focus; extended depth of focus intraocular lenses; phacoemulsification; presbyopia; supraphob infocus IOL
Mesh:
Year: 2020 PMID: 32971619 PMCID: PMC7727980 DOI: 10.4103/ijo.IJO_2139_19
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1Design of SupraPhob Infocus IOL
Post-Operative Visual Outcome of all cases undergoing bilateral implantation of Supraphob Infocus IOL
| 1 month | 3 months | |
|---|---|---|
| Visual Acuity(logMAR) | ||
| UDVA (Uniocular) | 0.12±0.08 | 0.11±0.08 |
| UDVA (Binocular) | - | 0.03±0.07 |
| UIVA (Uniocular) | 0.17±0.05 | 0.15±0.07 |
| UIVA (Binocular) | - | 0.14±0.06 |
| UNVA (Uniocular) | 0.41±0.09 | 0.41±0.08 |
| UNVA (Binocular) | - | 0.36±0.05 |
| Contrast Sensitivity (logCS) | 1.46±0.05 | 1.47±0.06 |
| Distance Stereopsis (3m) seconds of arc | 95.38±23.64 | 90.2±24.78 |
| Near Stereopsis (33cm) seconds of arc | 61.15±19.16 | 62.5±19.49 |
| MRSE (D) | NA | 0.04±0.37 |
| Spectacle Independance Distance (%) | NA | 96.15% |
| Spectacle Independance Near (%) | NA | 92.31% |
Footnotes. UDVA- Uncorrected distance visual acuity; UIVA- Uncorrected intermediate visual acuity; UNVA- Uncorrected near visual acuity; MRSEManifest refractive spherical equivalent; D-Diopters; NA- Not assessed
Figure 2Binocular distance corrected Defocus Curve for Supraphob Infocus IOL
Aberrometry profile of cases undergoing bilateral implantation of Supraphob Infocus IOL at 3-months
| Aberrometry Parameters | Mean±SD |
|---|---|
| HOAs (µm) | |
| HOAs | |
| Total | 0.30±0.13 |
| Internal | 0.27±0.13 |
| Coma | |
| Total | 0.17±0.09 |
| Internal | 0.16±0.08 |
| Trefoil | |
| Total | 0.14±0.08 |
| Internal | 0.09±0.07 |
| Spherical Aberration | |
| Total | 0.08±0.06 |
| Internal | 0.06±0.07 |
| Secondary Astigmatism | |
| Total | 0.05±0.04 |
| Internal | 0.05±0.05 |
| PSF | |
| Total | 0.07±0.08 |
| Internal | 0.07±0.04 |
| MTF | |
| Total (5 Cpd) | 0.48±0.16 |
| Internal (5 Cpd) | 0.54±0.17 |
| Total (10 Cpd) | 0.22±0.11 |
| Internal (10 Cpd) | 0.26±0.13 |
| Total (15 Cpd) | 0.15±0.11 |
| Internal (15 Cpd) | 0.16±0.11 |
| Average Total | 0.26±0.07 |
| Average Internal | 0.29±0.08 |
HOA-higher order aberrations; PSF- point spread function; MTF- modulation transfer function; cpd- cycles per degree
Figure 3Total and internal modulation transfer function (MTF) following implantation of Supraphob Infocus IOL
Review of literature of post-operative visual outcome with EDOF IOL
| CONCERTO study | Pedrotti | Titiyal | Current study | |
|---|---|---|---|---|
| Type of study | Ambispective, multicentric | Prospective | Prospective | Prospective |
| Sample size | *299 | **55 | 50 | 52 |
| IOL design | Technis Symfony | Technis Symfony | Technis Symfony | Supraphob Infocus |
| Binocular Visual acuity (logMAR) | ||||
| UDVA | 0.03±0.09 | 0.04±0.09 | 0.01±0.04 | 0.03±0.07 |
| UIVA | 0.13±0.16 | 0.05±0.09 | 0.09±0.06 | 0.14±0.06 |
| UNVA | 0.21±0.16 | 0.18±0.10 | 0.19±0.05 | 0.36±0.05 |
| Contrast Sensitivity (logCS) | - | CSV1000 1.7 at 6 cycles/degree | - | 1.47±0.06 |
| Distance stereopsis (3m) | - | - | 103.6±49.1 (Randot) | 90.2±24.78 (FD2) |
| Near Stereopsis (33cm) | - | - | 21.1±2.3 (Randot) | 62.5±19.49 (Randot) |
| Aberrations | - | |||
| HOA | ||||
| Total | - | 0.62±0.41 | 0.30±0.13 | |
| Internal | - | 0.64±0.43 | 0.27±0.13 | |
| PSF | ||||
| Total | 0.18±0.06 | 0.03±0.02 | 0.07±0.08 | |
| Internal | - | 0.03±0.02 | 0.07±0.04 | |
| MTF | ||||
| Total | - | 0.24±0.08 | 0.26±0.07 | |
| Internal | - | 0.24±0.07 | 0.29±0.08 | |
| MRSE (D) | -0.3±1.13 | -0.08±0.28 | within 0.5D | 0.04±0.37 |
| Spectacle Independence Distance (%) | 92.1% | 77.94±25.72 | 96% | 96.2% |
| Spectacle Independence Near (%) | 72.1% | 62% | 92.3% |
*Total - 411 cases, Emmetropia -299 cases, Intended micromonovision- 112 cases. ** Total cases - 185, Symfony IOL- 55 cases