Lucia Calisto Farracho1, Berenice Moutinot2, Angeliki Neroladaki1, Marion Hamard1, Karel Gorican1, Pierre Alexandre Poletti1, Jean Yves Beaulieu2, Cindy Bouvet2, Sana Boudabbous1.
Abstract
PURPOSE: To assess the accuracy and reliability of using cone beam computed tomography (CBCT) compared to X-ray six weeks after injury for predicting scaphoid union.Entities:
Keywords: CBCT; Immobilisation; Scaphoid healing; X-rays
Year: 2020 PMID: 32944592 PMCID: PMC7481520 DOI: 10.1016/j.ejro.2020.100251
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Scanning parameters of the CBCT (OnSight, Carestream Health, Rochester, New York) and XR (Siemens, ISIO and Philips, DigitalDiagnost).
| Parameters | CBCT | XR |
|---|---|---|
| Current | 5mA | 2.8mAs |
| Energy | 80kVp | 44kVp |
| FOV | 216 × 216 mm | |
| Isotropic voxel size (m) | 884 × 884 | |
| Rotation time | 25.18 s | |
| Exposure time | 21 s | |
| Scanning slice thickness | 0.26mm | |
| Scan rotation angle | 216.5° | |
| Focus-detector distance | 120 cm |
CBCT: cone-beam computed tomography; XR: X-ray; s: second; dose index.
Fig. 1Manual worker aged 34 years with an unstable fracture in Schoenberg’s zone III of the right scaphoid according to Mayo criteria [21] treated with surgery to stabilize the fracture. XR (a) showed no healing, according to all readers. To the contrary, CBCT (b) showed total agreement of scaphoid consolidation.
Fig. 2CBCT (a) of a 36-year-old, office worker with a right scaphoid fracture in Schoenberg’s zone III with complete healing according to all readers. XR (b) of the same patient shows no evidence of scaphoid healing.
Comparison of statistical analysis in terms of CBCT and XR for scaphoid consolidation.
| XR | CBCT | |
|---|---|---|
| Agreement among four readers | 0.543 | 0.641 |
| Agreement among hand surgeons | 0.95 | 0.803 |
| Agreement among radiologists | 0.35 | 0.803 |
| Sensitivity | 0.75 | 0.78 |
| Specificity | 0.4 | 0.4 |
| VPN | 0.5 | 0.53 |
| VPP | 0.66 | 0.67 |
| Accuracy | 0.53- 059 | 0.61−0.63 |
CBCT: cone-beam computed tomography; XR: X-ray; VPN: negative predictive value; VPP: positive predictive value.
Comparison of qualitative analyses between CBCT and XR.
| P (Wilcoxon test) | Correlation (Spearmann r) | |
|---|---|---|
| Schenberg zone | 0.5722 | 0.5698 |
| Bridges | 0.6072 | 0.3397 |
| Overall impression | 0.7744 | 0.4783 |
| Density | 0.125 | 0.2141 |
| Translation | 0.0002 | 0.2906 |
| Humpback deformity | 0.2101 | −0.05495 |
| Communition | 0.7539 | 0.1797 |
| Cyst formation | 0.3438 | −0.09759 |
CBCT: cone-beam computed tomography; XR: X-ray.