C Jiang, S R de Rijk1, G G Malliaras2, M L Bance. 1. Department of Clinical Neurosciences, University of Cambridge, Cambridge CB2 0AH, United Kingdom. 2. Division of Electrical Engineering, Department of Engineering, University of Cambridge, Cambridge CB3 0FA, United Kingdom.
Abstract
Cochlear implants (CIs) have tremendously helped people with severe to profound hearing loss to gain access to sound and oral-verbal communication. However, the electrical stimulus in the cochlea spreads easily and widely, since the perilymph and endolymph (i.e., intracochlear fluids) are essentially electrolytes, leading to an inability to focus stimulation to discrete portions of the auditory nerve, which blurs the neural signal. Here, we characterize the complex transimpedances of human cadaveric cochleas to investigate how electrical stimulus spread is distributed from 10 Hz to 100 kHz. By using electrochemical impedance spectroscopy (EIS), both the resistive and capacitive elements of human cochleas are measured and modeled with an electrical circuit model, identifying spread-induced and spread-independent impedance components. Based on this electrical circuit model, we implement a Laplace transform to simulate the theoretical shapes of the spread signals. The model is validated by experimentally applying the simulated stimulus as a real stimulus to the cochlea and measuring the shapes of the spread signals, with relative errors of <0.6% from the model. Based on this model, we show the relationship between stimulus pulse duration and electrical stimulus spread. This EIS technique to characterize the transimpedances of human cochleas provides a new way to predict the spread signal under an arbitrary electrical stimulus, thus providing preliminary guidance to the design of CI stimuli for different CI users and coding strategies.
Cochlear implants (CIs) have tremendously helped people with severe to profound hearing loss to gain access to sound and oral-verbal communication. However, the electrical stimulus in the cochlea spreads easily and widely, since the perilymph and endolymph (i.e., intracochlear fluids) are essentially electrolytes, leading to an inability to focus stimulation to discrete portions of the auditory nerve, which blurs the neural signal. Here, we characterize the complex transimpedances of human cadaveric cochleas to investigate how electrical stimulus spread is distributed from 10 Hz to 100 kHz. By using electrochemical impedance spectroscopy (EIS), both the resistive and capacitive elements of human cochleas are measured and modeled with an electrical circuit model, identifying spread-induced and spread-independent impedance components. Based on this electrical circuit model, we implement a Laplace transform to simulate the theoretical shapes of the spread signals. The model is validated by experimentally applying the simulated stimulus as a real stimulus to the cochlea and measuring the shapes of the spread signals, with relative errors of <0.6% from the model. Based on this model, we show the relationship between stimulus pulse duration and electrical stimulus spread. This EIS technique to characterize the transimpedances of human cochleas provides a new way to predict the spread signal under an arbitrary electrical stimulus, thus providing preliminary guidance to the design of CI stimuli for different CI users and coding strategies.
Cochlear implants (CIs) have helped hundreds of thousands of people who have
severe-to-profound hearing loss to improve access to sounds by directly stimulating their
cochlear nerves with electrical stimuli.
However, these current pulse stimuli cannot be focused at a narrow region of desired
cochlear nerves due to the high electrical conductivity of the fluids inside cochlear ducts
(i.e., perilymph and endolymph). This
phenomenon is known as electrical stimulus spread or current spread and results in a severe
spectral blurring of the input current stimuli at the neuronal level, as depicted in Figs. 1(a)–1(c). Due to this issue, most CI users experience difficulties in
challenging listening conditions (e.g., with background noise or multi-speakers) and have
few actual independent information channels, and their musical perception is limited. Clinically, different CI users have
different patterns and levels of spread, which result from their individual unique cochlear
anatomy and physiology. These
properties are difficult to analyze and correlate with stimulus spread, but cochlear
impedance is an electrical property that can provide some insight into cochlear stimulus
conditions and can be measured clinically.
FIG. 1.
(a) Schematic of a cochlear implantation in a human head. (b) Schematic of the
electrical stimulus spread that excites auditory nerves in an unwrapped cochlea. (c) An
example of electrical stimulus spread measured with clinical software on a cadaveric
cochlea, with the solid curve demonstrating the spread when stimulating electrode 8. (d)
The schematic of the setup for cochlear transimpedance spectroscopy and spread-induced
voltage measurements.
(a) Schematic of a cochlear implantation in a human head. (b) Schematic of the
electrical stimulus spread that excites auditory nerves in an unwrapped cochlea. (c) An
example of electrical stimulus spread measured with clinical software on a cadaveric
cochlea, with the solid curve demonstrating the spread when stimulating electrode 8. (d)
The schematic of the setup for cochlear transimpedance spectroscopy and spread-induced
voltage measurements.Cochlear impedances can be measured with the electrode array that is inserted into a
cochlea during implant surgery. This electrode array is designed to inject current stimuli
into the cochlea and is also integrated with a voltage measurement unit to characterize the
impedance of the cochlea. Cochlear impedance measurement can be categorized into two types:
contact impedance and transimpedance. The contact impedance is the measured impedance
between one intracochlear electrode and the extracochlear ground electrode in the cranial
temporal region [Fig. 1(a)], provides useful
information to check if the intracochlear electrode is functioning properly, and can reveal
open circuits or short circuits in the implant wiring. Contact impedance measurement is used
by all of the major CI companies using the stimulus-induced voltage amplitude divided by the
input current level, which however does not reveal the true complex impedance. In addition, contact impedances do not
indicate any stimulus spread information, and they are also affected by the
electrode/electrolyte interface impedance, therefore not presenting the true cochlear
electrical impedances.
Transimpedance is the inverse ratio between the delivered current at a stimulating electrode
and the measured voltage at another intracochlear electrode, measured in ohms. Since the
electrode array of a CI is placed along the cochlea, transimpedance measurement indicates
the electrical potential (resulting from the stimulus) built up at the locations away from
where a stimulating electrode should focus the stimulus, i.e., the stimulus spread. For
commercial CIs, transimpedance measurement is used by several CI companies, using these
stimulation-current-induced non-stimulating electrode voltage (SCINSEV) recordings, and is
known as the transimpedance matrix (TIM) by Cochlear Corp®, electrical field imaging (EFI)
by Advanced Bionics®, and Impedance Field Telemetry (IFT) by MED-EL®. An example of EFI
results is shown in Fig. 1(c). However, all these
measurements have limitations in providing complete information about stimulus spread.
Basically, for TIM/EFI/IFT, the transimpedance is calculated from the measured voltage at
the non-stimulating electrode in response to a current pulse at certain timing after the
pulse starts, but the measured voltage waveform does not depict a steep rising edge and a
plateau. These characteristics are different from the current pulse waveform, since the
cochlea as a biological tissue is not purely electrically resistive but contains capacitive
elements. In order to extract more
information from impedances measured from CIs, some studies have decomposed the impedance
into access resistance and polarization impedance from the measured voltage waveforms, which
partly analyzes the shape of the slow rise at the beginning of stimulus pulse and the
additional gradual increasing during the constant stimulation level (Fig. S8). However, the measured values of
access resistance and polarization impedance are dependent on the duration of pulses due to
the non-linearity in the voltage waveforms for either access resistance or polarization
impedance measurements, and therefore, this decomposition is not really applicable to short
(e.g., <5 µs) or long (e.g., >100 µs) pulses.In this study, we propose a new method for cochlear impedance characterization using
electrochemical impedance spectroscopy
(EIS) to measure cochlear transimpedance. This method can provide the actual complex
impedances of cochleas over a range of frequencies and therefore are not conditioned by
pulse duration effects. In addition, based on the EIS measurements, an equivalent circuit
model for the cochlear transimpedance can be found, which can be applied to any types of
pulses with different pulse shapes and pulse durations so that modeling of spread-induced
voltage signals under any current stimulus becomes possible. Here, we characterize the
cochlear transimpedances of three human cadaveric specimens using EIS and develop a
universal circuit model that can be parameterized to fit different specimens. We also use a
Laplace transform to simulate spread-induced voltage signals under square pulses and verify
the circuit model by comparing the simulated results with experimental measurements from the
specimens. To the best of our knowledge, this work is the first study using EIS to
characterize cochlear transimpedances and demonstrates a novel and universal method to
advance CI stimulus design, which is also applicable to other electrical implants.
EXPERIMENTAL
Cadaveric specimen surgery
Fresh-frozen human cadaveric heads were procured from the Anatomy Gifts Registry (USA)
for surgical training and research within a longstanding surgical training facility in our
institution. Three fresh-frozen human cadaveric heads were implanted with the same
Advanced Bionics® HiFocus 1J lateral-wall electrode array. The cochlea was implanted as it
is clinically; a mastoidectomy, posterior tympanotomy, and round window approach were
performed. Prior to implantation, the cochlea was flushed with 1.0% saline (GIBCO
distilled water, sodium chloride S/3160/60, Fisher Scientific, MA) through a 1-mm opening
in the lateral semi-circular canal (LSCC) to remove any debris inside the cochlea. This has similar electrical conductivity
to the normal perilymph that is there. After flushing the cochlea and inserting the
electrode array through the round window, the 1-mm opening created in the LSCC was closed
off with a non-conductive material (Blu Tack). A ground electrode was placed underneath
the temporal muscle. The Advanced Bionics® 1J electrode array has 16 electrodes, with
electrode 1 being the most apical electrode and electrode 16 the most basal electrode. As
fabricated by its commercial provider, the electrode array used in this study only had
odd-number electrodes actually connected to pins out to a D-25 connector. In addition, due
to the damage of the electrode array after many implantations, electrode 1 (the most
apical one) was broken, and we only had seven working electrodes from the electrode array
(i.e., electrode Nos. 3, 5, 7, 9, 11, 13, and 15 [Fig.
1(d)].
After implantation, the 1J electrode array was connected to an impedance analyzer (RS PRO
LCR-6100) using a three-terminal configuration with the simulating electrode connected to
the high-force terminal, the recording electrode connected to the high-sense terminal, and
the ground electrode connected to both the low-force and low-sense terminals. The
frequency range was from 10 Hz to 100 kHz, with 10 frequencies per decade (at equivalent
logarithmical intervals). An AC signal with the amplitude of 100 mV was used. This
frequency range encompasses most of the frequency components of the usual clinical
stimulus signals in which the phase durations are typically around 20
µs–200 µs. In addition, the measurements from 10 Hz to 100 Hz can be used to
characterize the low-frequency polarization effect that is commonly seen in voltage
waveform recording. The measurement results were analyzed and fitted with an equivalent
circuit model, a parallel circuit of a resistor (R1), and a
constant phase element (CPE1) in series with another parallel
circuit of a resistor (R2) and a constant phase element
(CPE2), as shown in Fig.
2(a). All the data were fitted by ZView® (Scribner Associated Inc.). The
measurements were performed at the odd-number electrodes, since the 1J electrode array
provided by the manufacturer (Advanced Bionics®) only had these connections from these
electrodes to the connector block.
FIG. 2.
(a) Schematic of the equivalent circuit model for the EIS measurements. (b) The
Nyquist plot of EIS measured from three human cadaveric specimens when stimulating
from electrode 3 and recording at electrode 5. [(c) and (d)] The Bode plot of EIS
measured from three human cadaveric specimens when stimulating from electrode 3 and
recording at electrode 5, showing (c) the impedance magnitudes and (d) phases.
(a) Schematic of the equivalent circuit model for the EIS measurements. (b) The
Nyquist plot of EIS measured from three human cadaveric specimens when stimulating
from electrode 3 and recording at electrode 5. [(c) and (d)] The Bode plot of EIS
measured from three human cadaveric specimens when stimulating from electrode 3 and
recording at electrode 5, showing (c) the impedance magnitudes and (d) phases.
Spread-induced voltage measurements
The 1J electrode array was connected to an CI emulator in which stimuli can be generated
and controlled by Bionic Ear Data Collection System (BEDCS) research software. Both the
emulator and research software were sourced from Advanced Bionics®. The current stimuli
used were charge-balanced biphasic cathodic-leading square pulses, with a duration of 32
µs for each phase and an amplitude of 800 µA.
Spread-induced voltage signals were recorded with an oscilloscope (Teledyne LeCroy
HDO4054A-MS). The sampling rate was 1 GHz. The results were transmitted to a LabView
program and conditioned with a digital Butterworth low-pass filter with a cut-off
frequency of 6.25 MHz to remove radio-frequency noises from the CI.
Spread-induced voltage modeling
A Laplace transform was performed on the EIS equivalent circuit model and the biphasic
current stimulus, and their product gave the Laplace transform of the spread-induced
voltage signal. Then, an inverse Laplace transform was performed to obtain the transient
form of the spread-induced voltage signal. We performed the modeling on MATLAB (version
R2018b). The modeled results were compared to the experimental measurements.
RESULTS AND DISCUSSION
The EIS results of three human cadaveric cochleas in full heads are depicted in Fig. 2. To avoid confusion from presenting all measurements
in one graph, we show the EIS results of three different specimens measured at the same
stimulating and recording electrodes (No. 3 and No. 5, respectively), which represented a
typical stimulating/recording pair. Comparing the impedances of the three specimens, they
are in the same order of magnitude, but specimen No. 1 shows lower overall impedances, as
shown in Fig. 2(c). The variations of impedances among
different specimens can result from bone compositions, cadaver age, anatomical geometries,
CI insertion depth, etc. These
factors could not be easily separately investigated. However, impedances of tissues are in
the same order of magnitude. The lower impedances of specimen No. 1 than the other two could
be attributed to the possible high ion concentrates in bones and/or a shorter electrical
pathway from cochlea to the ground electrode. As shown in Fig.
2(b), the Nyquist plot of each specimen dataset demonstrates one pole, where the
imaginary component becomes insignificant when compared to the real component. This
indicates that the Nyquist plots are likely to contain two semicircles and that the EIS data
can be fitted by an impedance circuit model with two resistor–capacitor (RC) circuits in
series. Since objects under test were
human specimens and not electrical elements, a capacitor is likely an imperfect
representation and we used constant phase elements (CPEs) instead of capacitors, as shown in
Fig. 2(a). The mathematical model for the complex
impedance of a CPE can be expressed aswhere
Y is the admittance magnitude at 1 rad/s and p is the
parameter for the constant phase (θ = −90p°). The
equivalent circuit fitted well to the measured data of the three specimens in the Nyquist
plots in Fig. 2(b) and the Bode plots in Figs. 2(c) and 2(d).
This good agreement suggested that the equivalent circuit might be applicable to different
specimens by using parametric variations in the components of the equivalent circuit. In the
Bode plots, the cochlear transimpedances demonstrated relatively flat curves in magnitude
and close-to-zero results in phase between ∼1 kHz and ∼10 kHz, which corresponded to the
zero-imaginary results showed in the Nyquist plots. This indicates that the cochlear
transimpedances are dominated by different R-CPE circuits at low frequencies (<1 kHz) and
high frequencies (>10 kHz).We further investigated the fittings of the equivalent circuit and found the origins of
these two R-CPE circuits. We extracted the parameters for the resistors and CPEs in the
measured EIS results of one specimen from all the measurement pairs, as shown in Fig. 3. From the extracted parameters, we observed quite
different behaviors between two R-CPE circuits. The first R-CPE circuit
(R1 and CPE1) seemed constant in
all the measured EIS results, i.e., regardless of the locations of the stimulating and
recording electrodes, as shown in Figs. 3(a)–3(c). In
this case, R1 and CPE1 were
spread-independent. In contrast, the second R-CPE circuit (R2
and CPE2) was more spread-dependent. Obviously, stronger
stimulus spread should be seen at places closer to the stimulating electrode and therefore
result in higher impedances in the EIS measurements. As shown in Fig. 3(f), R2 demonstrated hill-like curves
with the virtual peaks at each stimulating electrode. Inversely, for
CPE2, the admittance magnitude at 1 rad/s
(Y2) demonstrated valley-like curves with the virtual dips at
each stimulating electrode. Interestingly, the constant phases of
CPE2 (θ2) for different
measurement pairs were also independent of the locations of stimulating and recording
electrodes. This could be explained by the properties and compositions of a cochlea for
ionic conduction being similar along a given cochlear duct. Therefore, although different
locations result in the degree of spread being different, the phase delay of spread-induced
longitudinal impedances as a result of ion movement in the 1% NaCl solution is not affected.
The current pathway for cochlear transimpedance measurements is from cochleas via temporal
bones to the remote ground electrode under the temporal muscle. This results in effectively
two types of impedances, a common and similar pathway for current returning from each
electrode (i.e., from the ground electrode to temporal bones), which is recording electrode
independent, and a stimulating/recording-location-dependent pathway (i.e., within cochleas).
Therefore, we could suggest that the common pathway corresponds to
R1 and CPE1, and what varies the
transimpedances is the inner cochlear structures that correspond to
R2 and CPE2. We simulated the
impedances of the inner cochlear structures from 10 Hz to 100 kHz and found that the
impedances were nearly resistive from 10 Hz to 10 kHz (Fig. S7), which was consistent with
the previous report by Spelman et al.
FIG. 3.
The extracted values for all the components in the equivalent circuit model fitted to
the measured EIS in specimen No. 2: (a) the admittance magnitude at 1 rad/s of
CPE1 (Y1), (b) the constant
phase of CPE1 (θ1), (c)
R1, (d) the admittance magnitude at 1 rad/s of
CPE2 (Y2), (e) the constant
phase of CPE2 (θ2), and (f)
R2.
The extracted values for all the components in the equivalent circuit model fitted to
the measured EIS in specimen No. 2: (a) the admittance magnitude at 1 rad/s of
CPE1 (Y1), (b) the constant
phase of CPE1 (θ1), (c)
R1, (d) the admittance magnitude at 1 rad/s of
CPE2 (Y2), (e) the constant
phase of CPE2 (θ2), and (f)
R2.We further validated this equivalent circuit and the origins of two R-CPE circuits by
extracting the parameters of the equivalent circuit for the EIS measurements in the other
two specimens. As shown in Figs. S1 and S2, the parameters of the first R-CPE circuit were
also independent of the stimulating and recording electrode locations, as well as the phase
of CPE2, whereas R2 and
Y2 extracted from specimen Nos. 2 and 3 demonstrated
spread-related profiles that were similar to specimen No. 1. More importantly, the extracted
parameters of the equivalent circuit were in the same range among all these three specimens.
The parameters of R2 and CPE2 were
different among different specimens due to the differences of intracochlear geometries and
slight variations in implantation depth in the specimens. We did observe some differences in
the parameters of R1 and CPE1 among
different specimens (Fig. S3), which could potentially be attributed to the differences of
specimens, including the contacts between the ground electrode and cranium, conductivities
of skulls and temporal bones, sizes of the head, differing soft tissue pathways, such as
muscle bulk, and variations in
the day-to-day temperatures during testing of specimens. The extracted parameters in the
equivalent circuit can reveal some differences not easily noticeable in Bode plots. As shown
in Fig. 3(c), specimen Nos. 2 and 3 demonstrated
similar impedances at low frequencies (10 Hz–100 Hz), which were different from that of
specimen No. 1. However, when comparing the extracted circuit parameters as shown in Fig.
S3, we only found similarity in Y1, but differences in
θ1 and R1. In particular, for
R1, specimen No. 3 demonstrated a closer value to specimen No.
1 than specimen No. 2. As previously mentioned, the independent effects of the factors of
impedance variations on these circuit parameters are difficult to understand. However, it
could be the reason that these factors, though causing differences in
Y1, θ1, and
R1, could work together and show a combined result, which has
similarity in the overall impedance results of specimens Nos. 2 and 3 from 10 Hz to 100 Hz.
Despite the differences, the parameters of R1 and
CPE1 among different specimens were in the same ranges of
magnitudes (Fig. S3).With this well-fitted equivalent circuit model to human cochlear transimpedances, we
simulated spread-induced voltage signals. This was the primary purpose of the
characterizations of human cochlear transimpedances, i.e., to understand the stimulus
spread, including the amplitude and shape of signals. According to its definition in
electronics, transimpedance is the ratio of the resultant voltage at the output terminal to
the current at the input terminal. Since
we have information for cochlear transimpedances, we have the relation between
spread-induced voltage and stimulating current at all stimulating/recording pairs as
, where
and
are the
stimulating current and spread-induced voltage at the jth and
kth electrode of a CI, respectively, and
is the
transimpedance when stimulating at the jth electrode and recording at the
kth electrode (j ≠ k),
s is the complex variable. Note that all these terms are in the
s domain. For the stimulus, we considered a symmetric cathodic-leading
biphasic square pulse, which is commonly used clinically. The Laplace transform of the
biphasic current stimulus from the ith electrode can be expressed as
,
where ij,amp is the current amplitude and
T0 is the duration of each phase. As for the transimpedances,
Zj,k(s) can be expressed as
,
where the subscripts j, k denote locations of the
stimulating (jth) and recording (kth) electrodes for the
parameters (j ≠ k). The spread-induced voltage in the
s domain is simply the production of current stimulus and transimpedance
in the s domain. To obtain the spread-induced voltage in the time domain,
we performed an inverse Laplace transform, ,
and foundwhere
is the Gamma function and
is the Mittag–Leffler function. All the detailed mathematical derivations can be found in
the supplementary
material [Eqs. (S1.1)–(S1.16)].A typical simulated spread-induced voltage signal is shown in Fig. 4(c) and was compared with experimental data measured from the same specimen.
The simulated data and the experimental data were almost completely overlap as plotted in
Fig. 4(c), with a root-mean-square (RMS) error of 0.3
mV, i.e., a RMS relative error of 0.08%. We also calculated the RMS errors for all the
measurement pairs and found that the errors were below 0.25%. For all the specimens, the
average RMS errors were 0.13 ± 0.06%, 0.33 ± 0.14%, and 0.36 ± 0.14%, with a maximum of
0.60% (Figs. S4–S6). The agreement between the simulated and experimental data suggested
that the equivalent circuit with two R-CPE circuits might be a good presentation of human
cochlear transimpedances, not only for EIS characterizations but also for transient
spread-induced voltage measurements. More importantly, since this equivalent circuit can
represent cochlear transimpedances at all frequencies, it can be used to simulate the
spread-induced voltage signals from stimuli of any pulse shape, duration, and amplitude.
FIG. 4.
(a) The relation of current stimulus (),
voltage response (),
and transimpedance ()
in the frequency domain. (b) The Laplace transforms of current stimulus, voltage
response, and the equivalent circuit for cochlear transimpedance. (c) The simulated (red
curve) and measured (blue dot) spread-induced voltage response
(VSI) as a function of time, showing the good agreement
between the simulated and experimental results. (d) The relative errors between the
simulated and experimental results for all the simulating and recording pairs.
(a) The relation of current stimulus (),
voltage response (),
and transimpedance ()
in the frequency domain. (b) The Laplace transforms of current stimulus, voltage
response, and the equivalent circuit for cochlear transimpedance. (c) The simulated (red
curve) and measured (blue dot) spread-induced voltage response
(VSI) as a function of time, showing the good agreement
between the simulated and experimental results. (d) The relative errors between the
simulated and experimental results for all the simulating and recording pairs.Here, we show an example of simulation and how spread is affected by different pulse
durations. Basically, the narrower the pulse durations are, the higher the frequencies of
the signals that lie in the spectrum. The cochlear transimpedance spectra under stimulation
at electrode 15 are depicted in Fig. 5(a). The cochlear
transimpedances at the low frequencies (<100 Hz) were quite similar among the
measurements at different recording locations, which indicated a similar level of voltage
response around the entire cochlea and a significant electrical spread. At the middle-range
frequencies (1 kHz–32 kHz), the cochlear transimpedances were lower at the recording
locations farther away from the stimulating electrode, which suggested that spread could be
partly reduced in this frequency range, but the spectrum at each stimulating/recording pair
was relatively constant across these frequencies. At the high frequencies (>32 kHz), it
is noteworthy that the curves were fanned out toward higher frequencies, i.e., the cochlear
transimpedances dropped faster at the recording locations farther away from the stimulating
electrode. This was probably because at the farther-away locations, the AC signal delays
could be longer, and the transimpedances contained more capacitive elements that caused the
transimpedance magnitudes to drop faster. From this, the electrical stimulus can be more
focused if the stimulus is a higher-frequency signal, which was confirmed by the plots to
visualize this spread in the normalized spread-induced voltage response distributions using
pulses with different frequencies in Fig. 5(b). As
shown in Fig. 5(c), the spatial spread was similar at
frequencies below 32 kHz and started to decrease over 32 kHz. We simulated the normalized
transient spread-induced voltage responses under symmetric biphasic stimuli with different
pulse durations (i.e., 100 µs, 10 µs, and 1
µs), as shown in Figs. 5(d)–5(f). It
can be seen that the amplitudes of voltage responses at the recording locations farther away
from the stimulating electrode were reduced in the case of 1-µs pulses
compared to 100-µs pulses. It is also noteworthy that the voltage response
became more asymmetric under pulses with shorter pulse durations. This may be useful for
auditory nerve stimulation, where the central nerves and peripheral nerves respond to
different polarities of stimuli, while the feasibility of delivering high-frequency pulses and the
actual nerve responses need to be further tested.
FIG. 5.
(a) Simulated transimpedance magnitude spectra when stimulating electrode 15. (b)
Normalized spread-induced voltage amplitude under stimulations with different
frequencies. (c) Stimulus focusing ability under sine-wave stimulations with different
frequencies. [(d)–(f)] The transient normalized spread-induced voltage responses under
biphasic pulse stimuli with different pulse durations, (d) 100 µs, (e)
10 µs, and (f) 1 µs.
(a) Simulated transimpedance magnitude spectra when stimulating electrode 15. (b)
Normalized spread-induced voltage amplitude under stimulations with different
frequencies. (c) Stimulus focusing ability under sine-wave stimulations with different
frequencies. [(d)–(f)] The transient normalized spread-induced voltage responses under
biphasic pulse stimuli with different pulse durations, (d) 100 µs, (e)
10 µs, and (f) 1 µs.
CONCLUSION
In conclusion, we have characterized complex impedance spectra of human cadaveric cochleas
using a cochlear implant and found an equivalent circuit model for the measured impedances.
The circuit model indicated two separate parts of impedances of which one is spread-induced
and the other is spread-independent. The circuit model could be used for modeling electrical
stimulus spread signals, which were verified with stimulations and recordings from a
cochlear implant. Therefore, with the cochlear impedance spectroscopy measurements and the
equivalent circuit, one could design new cochlear implant stimuli and predict the voltage
response and spread distribution. Based on this, stimuli can be customized for different
cochlear implant users, and more types of stimuli can be tested before applying to them
in vivo. As shown in this study, using an ultra-short phase duration of
0.5 µs can significantly suppress the stimulus spread, which provides an
example for cochlear implant clinics to reduce spectral blurring and improve cochlear
implant performance. In addition, the spread-independent component could be clinically
useful to diagnose problems with the ground electrodes. Moreover, this method is also
applicable to other medical implants using electrical stimuli.When coupled to the existing electrical characterization tools integrated with cochlear
implants, this technology could reveal more information about cochleas for cochlear implant
users, with potential for cochlear implant troubleshooting, such as distinguishing problems
with intracochlear electrodes or extracochlear ground electrode and identifying the
conditions (such as fibrosis or soft tissue ingrowth) surrounding the electrodes. For future
work, we will test the universality of this method and model for cochlear implants with
different designs from different manufacturers, and further explore the stimulus design to
suppress electrical spread and enhance stimulation efficiency, and correlate the results
with psychoacoustic measurements in patients with CIs.See the supplementary
material for the complete mathematical derivations and impedance results of
other specimens that are not shown.
DATA AVAILABILITY
The data that support the findings of this study are available within the article and its
supplementary
material.
Authors: Jill B Firszt; Laura K Holden; Margaret W Skinner; Emily A Tobey; Ann Peterson; Wolfgang Gaggl; Christina L Runge-Samuelson; P Ashley Wackym Journal: Ear Hear Date: 2004-08 Impact factor: 3.570
Authors: Christopher Kenneth Giardina; Elliot Samuel Krause; Kanthaiah Koka; Douglas Carl Fitzpatrick Journal: IEEE Trans Biomed Eng Date: 2018-02 Impact factor: 4.538
Authors: Susan T Eitutis; Robert P Carlyon; Yu Chuen Tam; Marina Salorio-Corbetto; Zebunnisa Vanat; Karen Tebbutt; Rhian Bardsley; Harry R F Powell; Shibasis Chowdhury; James R Tysome; Manohar L Bance Journal: Otol Neurotol Date: 2022-04-01 Impact factor: 2.311