Xiaoyu Wang1, Zhenglin Zhu1, Haozuo Xiao1, Changqi Luo1, Xiaoji Luo1, Furong Lv2, Junyi Liao1, Wei Huang1. 1. Department of Orthopaedic Surgery, The First Affiliated Hospital of Chongqing Medical University, Chongqing 400016, China. 2. Department of Radiology, The First Affiliated Hospital of Chongqing Medical University, Chongqing 400016, China.
Abstract
To investigate the biocompatibility and bone ingrowth properties of a novel trabecular bone mimic porous tantalum scaffold which holds potential for bone tissue engineering, a novel three-dimensional, multiscale interconnected porous tantalum scaffold was designed and manufactured. The morphology of the novel scaffold was observed with the use of scanning electron microscopy (SEM) and industrial computerized tomography. Mesenchymal stem cells (MSCs) were cultured with novel porous tantalum powder, SEM was carried out for the observation of cell morphology and adhesion, and cytotoxicity was evaluated by the MTT assay. Canine femoral shaft bone defect models were established, and novel porous tantalum rods were used to repair the bone defect. Repair effects and bone integration were evaluated by hard tissue slice examination and push-out tests at the indicated time. We found that the novel porous tantalum scaffold is a trabecular bone mimic, having the characteristics of being three-dimensional, multiscaled, and interconnected. The MSCs adhered to the surface of tantalum and proliferated with time, the tantalum extract did not have a cytotoxic effect on MSCs. In the bone defect model, porous tantalum rods integrated tightly with the host bone, and new bone formation was found on the scaffold-host bone interface both 3 and 6 months after the implantation. Favorable bone ingrowth was observed in the center of the tantalum rod. The push-out test showed that the strength needed to push out the tantalum rod is comparable for both 3 and 6 months when compared with the normal femoral shaft bone tissue. These findings suggested that the novel trabecular bone mimic porous tantalum scaffold is biocompatible and osteoinductive, which holds potential for bone tissue engineering application.
To investigate the biocompatibility and bone ingrowth properties of a novel trabecular bone mimic porous tantalum scaffold which holds potential for bone tissue engineering, a novel three-dimensional, multiscale interconnected porous tantalum scaffold was designed and manufactured. The morphology of the novel scaffold was observed with the use of scanning electron microscopy (SEM) and industrial computerized tomography. Mesenchymal stem cells (MSCs) were cultured with novel poroustantalum powder, SEM was carried out for the observation of cell morphology and adhesion, and cytotoxicity was evaluated by the MTT assay. Canine femoral shaft bone defect models were established, and novel porous tantalum rods were used to repair the bone defect. Repair effects and bone integration were evaluated by hard tissue slice examination and push-out tests at the indicated time. We found that the novel porous tantalum scaffold is a trabecular bone mimic, having the characteristics of being three-dimensional, multiscaled, and interconnected. The MSCs adhered to the surface of tantalum and proliferated with time, the tantalum extract did not have a cytotoxic effect on MSCs. In the bone defect model, porous tantalum rods integrated tightly with the host bone, and new bone formation was found on the scaffold-host bone interface both 3 and 6 months after the implantation. Favorable bone ingrowth was observed in the center of the tantalum rod. The push-out test showed that the strength needed to push out the tantalum rod is comparable for both 3 and 6 months when compared with the normal femoral shaft bone tissue. These findings suggested that the novel trabecular bone mimic porous tantalum scaffold is biocompatible and osteoinductive, which holds potential for bone tissue engineering application.
Trauma, tumor resection, bone infection,
and regenerative disorder-caused
bone defects are a challenge to the medical fraternity.[1−4] Autograft and allograft transplantations are the common methods
applied by surgeons to treat bone defects; however, limited bone sources
and immunological reactions are still a cause for worry.[4−6] Therefore, surgeons have long been interested in taking advantage
of the properties of various metals for the reconstruction of bone
defects.[7−11] In recent years, titanium and its alloys have been widely utilized
to produce orthopedic implants such as total hip arthroplasty (THA)
and total knee arthroplasty (TKA) prostheses, spinal fusion cage,
and bone plate and so forth.[7,12] Nevertheless, high
elastic modulus, low porosity, and low surface friction coefficient
and of titanium blocked their clinical applications for bone defect.[7,13,14] Therefore, a material with higher
porosity, higher surface friction coefficient, and lower elastic modulus
compared with titanium holds potential for bone tissue engineering.Tantalum is a bioinert metal which holds the characteristic of
anticorrosion in vivo compared with other metal materials.[7,15,16] As a bone repair material, poroustantalum was applied to arthroplasty, dental implants, spinal fusion
surgery, foot and ankle surgery, necrosis of femoral head treatment,
and so forth.[7,9,14,17−21] Its high porosity, high surface friction coefficient,
favorable compatibility, and proper elastic modulus approximate to
cancellous bone make it a promising material for bone tissue engineering,[7,22] while an ideal material for bone tissue engineering should be trabecular
bone mimic, osteoinductive, and hold the capacity for bone ingrowth,[8,13,23,24] so we developed a novel trabecula bone mimic porous tantalum, which
is three-dimensional, multiscale, and interconnected. Compared with
the porous tantalum manufactured by Zimmer Corporation (Warsaw, IN,
USA),[17,20−22,25] our novel porous tantalum holds a wider range of porosity from 60
to 80% (Zimmer, 75–85%), wider range of pore diameter from
200 to 500 μm (Zimmer, 400–600 μm), and lower elastic
modulus range from 0.5 to 4.0 GPa (Zimmer, 2.5–3.9 GPa). These
differences make the novel porous tantalum property closer to the
cancellous bone.In this study, we investigated physical properties,
biocompatibility,
cytotoxicity, and osteoinductive characteristics of the novel poroustantalum scaffold by in vitro and/or in vivo tests, the results indicated
that the novel porous tantalum has potential for bone tissue engineering
application.
Results
Structural Characterization
of Novel Porous Tantalum Scaffold
The biomechanical indexes
are shown in Table . To understand the tridimensional structure
and microstructure of the novel porous tantalum scaffold, the poroustantalum was subjected to scanning electron microscopy (SEM) analysis.
As shown in Figure , the material was three-dimensional and in a honeycomb pattern with
pore diameters ranging from 200 to 500 μm (Figure A); meanwhile, the material
was interconnected with microscopic pores with diameters ranging from
1 to 10 μm (Figure A–C). The honeycomb structure was also confirmed by
industrial computerized tomography (CT) analysis. As shown in Figure C, the longitudinal
and transverse images manifested that the material consisted of mesh
shape pores in different sizes. Macroscopic images of porous tantalum
powders and rods revealed that the material was gray and smooth, and
pinpoint-sized pores in different diameters distributed on the surfaces.
These results suggested that the novel porous tantalum scaffold is
trabecular bone mimic, having the characteristic of being three-dimensional,
multiscaled, and interconnected.
Table 1
Physical Parameters of Novel Porous
Tantalum
parameters
range
density (g/cm3)
3.5–7.0
porosity
(%)
60–80%
pore diameter (μm)
200–500
microcosmic pore
diameter
(μm)
1–10
elasticity modulus (GPa)
0.5–4.0
yield strength
(MPa)
75–120
ultimate tensile strength
(MPa)
110–210
tensile strength (MPa)
80–120
Figure 1
Physical properties and morphology of
porous tantalum (photograph
courtesy of “Chongqing Runze Pharmaceutical Co., Ltd”.
Copyright 2020). (A) SEM micrographs of novel porous tantalum scaffold.
Honeycomb pattern with pore diameters range from 200 to 500 μm
in 400 magnification (a, scale bar = 500 μm). Microscopic pores
(range from 1 to 10 μm) in 5000 magnification (b, scale bar
= 1 μm) and 10,000 magnification (c, scale bar = 1 μm).
Arrows indicate the pores in the millimeter level, circles indicate
the interconnected microscopic pore in the micron level. (B) Industrial
CT images for the tantalum rod, scale bar = 1 μm. (C) Morphology
of novel porous tantalum powder and tantalum rod, scale bar = 1 mm.
Physical properties and morphology of
porous tantalum (photograph
courtesy of “Chongqing Runze Pharmaceutical Co., Ltd”.
Copyright 2020). (A) SEM micrographs of novel porous tantalum scaffold.
Honeycomb pattern with pore diameters range from 200 to 500 μm
in 400 magnification (a, scale bar = 500 μm). Microscopic pores
(range from 1 to 10 μm) in 5000 magnification (b, scale bar
= 1 μm) and 10,000 magnification (c, scale bar = 1 μm).
Arrows indicate the pores in the millimeter level, circles indicate
the interconnected microscopic pore in the micron level. (B) Industrial
CT images for the tantalum rod, scale bar = 1 μm. (C) Morphology
of novel poroustantalum powder and tantalum rod, scale bar = 1 mm.
Cell Growth on the Novel
Porous Tantalum Scaffold
Next,
biocompatibility of the novel porous tantalum scaffold was examined
with the use of mesenchymal stem cells (MSCs). UC-MSC cells were seeded
on the porous tantalum scaffold and incubated in complete Dulbecco
modified Eagle medium (DMEM) and subjected to the SEM test at the
indicated time points. On day 3, we found the cells survived and were
attached stably on the porous tantalum scaffold (Figure Aa). On the higher magnification,
we found that the cell was stretched out and cytoplasm extended outside
to form protrusions (Figure Ab). On day 5, the cell number increased compared with on
day 3, which indicated the proliferation of the cells. The cells attached
and connected with each other to form a layer on the surface of the
porous tantalum scaffold (Figure Ac). On the higher magnification (Figure Ad), we detected that the cells
had grown into the inner side of the material and formed protrusions
connected with each other. The results demonstrated the porous tantalum
scaffold was biocompatible and suitable for cell proliferation.
Figure 2
Cell growth
on a novel porous tantalum scaffold and cytotoxicity
evaluation of porous tantalum scaffold. (A) SEM micrographs of UC-MSCs
on the porous tantalum scaffold at different time points. UC-MSCs
on the porous tantalum scaffold on day 3 (a, scale bar = 50 μm,
b, scale bar = 30 μm). UC-MSCs on the porous tantalum scaffold
on day 5 (a, scale bar = 50 μm, b, scale bar = 30 μm).
(B) Cytotoxicity evaluation of the porous tantalum extract. MTT assay
were utilized to evaluate the cell proliferation of UC-MSCs cultured
in the tantalum extract and UC-MSCs cultured in complete DMEM were
used as the control. Cell proliferation did not statistically differ
between the groups with extended cultivation time.
Cell growth
on a novel porous tantalum scaffold and cytotoxicity
evaluation of porous tantalum scaffold. (A) SEM micrographs of UC-MSCs
on the porous tantalum scaffold at different time points. UC-MSCs
on the porous tantalum scaffold on day 3 (a, scale bar = 50 μm,
b, scale bar = 30 μm). UC-MSCs on the porous tantalum scaffold
on day 5 (a, scale bar = 50 μm, b, scale bar = 30 μm).
(B) Cytotoxicity evaluation of the porous tantalum extract. MTT assay
were utilized to evaluate the cell proliferation of UC-MSCs cultured
in the tantalum extract and UC-MSCs cultured in complete DMEM were
used as the control. Cell proliferation did not statistically differ
between the groups with extended cultivation time.
Cytotoxicity of Porous Tantalum Extract
To monitor
the influence of the porous tantalum extract on the proliferation
of UC-MSCs, MTT assay was used. Absorbance values showed that from
day 1 to day 6, the growth rate of the UC-MSCs in the control group
accelerated gradually, ultimately entering a stable phase. Compared
with the control group, the experiment group showed the same trend,
and no statistically significance difference (p >
0.05) of the absorbance value was found on each day (Figure B). These results indicated
that the porous tantalum extract was not cytotoxic to UC-MSCs, achieving
the basic requirement of implant materials.
Construction of Canine
Femoral Shaft Bone Defect Models and
Implantation of the Porous Tantalum Rod
To mimic the defects
of long bone shaft, canine femoral shaft bone defect models were constructed.
Briefly, a diameter of ∼4 mm of the whole cortex bone defect
in the midpiece of the femur shaft was constructed. Then, a poroustantalum rod with a diameter of 4 mm and 8 mm in height was embedded
in the bone along with the defect, which was as close as possible
to the tunnel host bone, the porous tantalum rod was extended to the
middle of the bone marrow cavity (Figure Aa). Each animal was implanted two poroustantalum rods on the same side, X-ray was taken 1 month after the
operation to confirm that the tantalum rods were in the femur shaft
and without a peri-implant fracture (Figure Ab). 3 and 6 months after the implantation,
the tantalum rods were taken out, the specimens are listed in Figure B. The surgical sites
exhibited no swelling, bleeding, or oozing, and no implant loss was
observed. The morphology of the specimens showed that the tantalum
rods were surrounded by the host bone; what was interesting, new bone
formation in the bone marrow cavity side was seen in 3 months (Figure Bb), and the tantalum
rod was connected with the opposite bone cortex in 6 months (Figure Bd). Taking these
results together, we successfully constructed the canine femoral bone
defect model and implanted porous tantalum rods; favorable bone ingrowth
was observed by morphology analysis of the specimens.
Figure 3
Construction of canine
femoral shaft bone defect models and implantation
of porous tantalum rod. (A) Graphical diagram for the construction
and implantation of porous tantalum rod at canine’s midpiece
of the femur shaft (a). X-ray image of 1 month after the implantation
(b). (B) Retrieved femoral shaft samples at 3 month (a, scale bar
= 1 cm) and 6 month (c, scale bar = 1 cm), and retrieved tantalum
rod-bone samples at 3 month (b, scale bar = 1 mm) and 6 month (d,
scale bar = 1 mm).
Construction of canine
femoral shaft bone defect models and implantation
of porous tantalum rod. (A) Graphical diagram for the construction
and implantation of porous tantalum rod at canine’s midpiece
of the femur shaft (a). X-ray image of 1 month after the implantation
(b). (B) Retrieved femoral shaft samples at 3 month (a, scale bar
= 1 cm) and 6 month (c, scale bar = 1 cm), and retrieved tantalum
rod-bone samples at 3 month (b, scale bar = 1 mm) and 6 month (d,
scale bar = 1 mm).
Bone Ingrowth and New Bone
Formation by Histological Assessment
To further analyze the
bone ingrowth and new bone formation of
the porous tantalum and host bone interfaces, we took advantage to
utilize hard tissue slicing and staining. A merged panoramic version
and distinct-separated images by methylene blue and Van Gieson (VG)
staining are listed in Figure and 5. For 3 month specimens, no gap
between the tantalum scaffold and the host bone was observed in the
cortical bone side of the porous tantalum rod, and the porous tantalum
and the host bone interfaces were in close contact with each other
(Figures A and 5A). New trabecular bone was found in the bone marrow
cavity side of the porous tantalum (Figures Aa–c and 5Aa–c),
and bone ingrowth was found in the cortical bone side of the poroustantalum rod (Figures Ad and 5Ad). As for 6 month specimens, no
gap between the tantalum scaffold and the host bone was observed in
the cortical bone side of the porous tantalum rod, and the poroustantalum and the host bone interfaces were in close contact with each
other (Figures B and 5B). More new trabecular bone was formed and almost
filled the porous tantalum pore in the bone marrow cavity side of
the porous tantalum (Figures Ba–c and 5Ba-c) compared with
the 3 month specimens. More bone ingrowth was found in the cortical
bone side of the porous tantalum rod compared with the 3 month specimens
(Figures Bd and 5Bd). Quantitative analysis of new bone formation
in 3 and 6 months is shown in Figure C. Taking these results together, we identified that
the tantalum scaffold and the host bone interfaces integrated with
each other in the cortical side of the porous tantalum rod, and a
new trabecular bone filled the pore of the material in the bone marrow
cavity side of the porous tantalum rod. In other words, the poroustantalum rod possessed the property of bone induction and worked as
a good scaffold for bone ingrowth.
Figure 4
Methylene blue staining for retrieved
tantalum rod-bone samples.
(A) Methylene blue staining for retrieved tantalum rod-bone samples
at 3 months after the implantation. A merged panoramic version in
the left panel and distinct-separated images (a–d) are listed
in the right panel, arrows indicated the new bone and tantalum rod
interfaces. (B) Methylene blue staining for retrieved tantalum rod-bone
samples at 6 months after the implantation. A merged panoramic version
in the left panel and distinct-separated images (a–d) are listed
in the right panel, arrows indicated the new bone and tantalum rod
interfaces.
Figure 5
VG staining for retrieved tantalum rod-bone
samples. (A) VG staining
for retrieved tantalum rod-bone samples at 3 months after the implantation.
A merged panoramic version in the left panel and distinct-separated
images (a–d) are listed in the right panel, arrows indicated
the new bone and tantalum rod interfaces. (B) VG staining for retrieved
tantalum rod-bone samples at 6 months after the implantation. A merged
panoramic version in the left panel and distinct-separated images
(a–d) are listed in the right panel, arrows indicated the new
bone and tantalum rod interfaces. (C) Quantitative analysis of new
bone formation in 3 month and 6 month’s specimens in VG staining.
**p < 0.05.
Methylene blue staining for retrieved
tantalum rod-bone samples.
(A) Methylene blue staining for retrieved tantalum rod-bone samples
at 3 months after the implantation. A merged panoramic version in
the left panel and distinct-separated images (a–d) are listed
in the right panel, arrows indicated the new bone and tantalum rod
interfaces. (B) Methylene blue staining for retrieved tantalum rod-bone
samples at 6 months after the implantation. A merged panoramic version
in the left panel and distinct-separated images (a–d) are listed
in the right panel, arrows indicated the new bone and tantalum rod
interfaces.VG staining for retrieved tantalum rod-bone
samples. (A) VG staining
for retrieved tantalum rod-bone samples at 3 months after the implantation.
A merged panoramic version in the left panel and distinct-separated
images (a–d) are listed in the right panel, arrows indicated
the new bone and tantalum rod interfaces. (B) VG staining for retrieved
tantalum rod-bone samples at 6 months after the implantation. A merged
panoramic version in the left panel and distinct-separated images
(a–d) are listed in the right panel, arrows indicated the new
bone and tantalum rod interfaces. (C) Quantitative analysis of new
bone formation in 3 month and 6 month’s specimens in VG staining.
**p < 0.05.
Bone Integration of the Porous Tantalum and Host Bone
To
confirm the bone integration and determination of the interfacial
bond strength of the tantalum rod and the host bone, the mechanical
push-out test was used. The machine is shown in Figure A. As shown in Figure B, compared with the normal bone, there was
no statistically significant difference in strength for pushing the
porous tantalum rod out either in the 3 month or 6 month specimens.
What was interesting, the strength for pushing porous tantalum rod
out in 3 month specimens was not statistically smaller than that in
the 6 month specimens. The load-deformation curves are shown in Figure C. According to the
mode of failure, normal bone was used for the reference of maximum
strength, and the two of the 6 month specimens were pushed out by
breakage of the host bone. As we found even in the 3 month’s
specimens, the power for pushing the porous tantalum out from the
host bone was comparative with the normal femoral bone, so there is
no significant difference among the normal bone, the 3, and 6 month
specimens for the push-out test. These results indicated the bone
integration of the porous tantalum material and the host bone.
Figure 6
Push-out test
for the evaluation of bone integration of the porous
tantalum and host bone. (A) Universal testing machine for the push-out
test. (B) Push-out strength in different samples, no statistical significance
was found among groups. (C) Representative load-deformation curve
for each group (a: normal bone, b: 3 month’s specimen, c: 6
month’s specimen).
Push-out test
for the evaluation of bone integration of the poroustantalum and host bone. (A) Universal testing machine for the push-out
test. (B) Push-out strength in different samples, no statistical significance
was found among groups. (C) Representative load-deformation curve
for each group (a: normal bone, b: 3 month’s specimen, c: 6
month’s specimen).
Discussion
Bone defects or nonunion fractures induced by
trauma, regenerative
disorders, bone tumor, and infection are challenging in treatment
for surgeons.[1−4] Although abundant efforts have been devoted to develop bone mimic
materials, ideal scaffolds which are osteoinductive, biocompatible,
and induce effective bone in-growth at the repair sites are still
lacking.[3,10,11,26,27]In the present
study, we manufactured a novel trabecular bone mimic
porous tantalum scaffold, which is three-dimensional, multiscaled
and interconnected porous material. The structural features were identified
by SEM and industrial CT assays. To the best of our knowledge, the
porosity, pore diameter, and elastic modulus are closer to cancellous
bone compared with porous tantalum produced by Zimmer Corporation.[15−18] Our further canine femoral shaft bone defect models verified that
the porous tantalum scaffold is biocompatible and osteoinductive,
the integration of the tantalum scaffold and the host bone was also
confirmed by mechanical tests. Thus, the novel porous tantalum scaffold
holds great potential for further application in bone tissue engineering.A higher tridimensional surface area was proved to promote cell
adhesion, cell growth, and oxygen and nutrition transport.[11,28] In the present study, the tridimensional geometric scaffold structure
and interconnected honeycomb-like porous tantalum scaffold material
could provide a large internal surface area, and facilitated cell–scaffold
interaction in a three-dimensional environment. Except trabecular
bone are interconnected physiologically, it was reported that interconnectivity
in favor of neovascularization, which maintains sustained bone development
and ingrowth, promotes higher volume of host bone and material interaction
and offers load transfer along the bone-implant interface.[25,29,30] The novel porous tantalum scaffold
featured with interconnected micron pores (Figure A), which contributed to the new bone ingrowth
and material host bone integration.The elastic modulus of tantalum
is comparative with other metal
materials; however, elastic modulus of porous tantalum is much lower,
which makes tantalum holds potential to be manufactured as cancellous
bone mimic scaffolds.[11,25] As the elastic modulus of cancellous
bone ranges from 0.76 to 4.0 GPa, porosity of cancellous bone ranges
from 75 to 90%, and pore size ranges from 50 to 300 μm.[25,31,32] Therefore, we designed our poroustantalum scaffold with the elastic modulus ranges from 0.5 to 4.0
GPa, porosity ranges from 60 to 80% and pore size ranges from 200
to 500 μm. All these parameters are approximate to cancellous
bone. Different pore sizes influence different processes of bone ingrowth,[11,25,29−31] it was reported
that larger than 300 μm pore size enhances vascularization,
smaller than 150 μm pore size facilitates cell spanning across
the pores, and larger than 200 μm pore size facilitates occupancy
of the cells inside the pores. Besides pore size, the random and disorder
distributed pattern is preferred for bone ingrowth. Hence, we designed
the porous tantalum scaffold as a random distributed pore size range
from 200 to 500 μm, which was proved to be conducive for bone
ingrowth and host-implant integration. In addition, we found favorable
bone ingrowth in the center of the scaffold, which proved that the
random distributed pore size holds potential for continuous bone ingrowth.Although porous tantalum has been clinically applied for interbody
cervical fusions, reconstruction of large bony defects in primary
and revision TKAs or THAs and so forth,[7,17−19] it is still conflicting as high porosity lowers the mechanical strength
of the materials.[31] Thus, the balance between
the microstructure and mechanical strength need to be achieved by
regulating the porosity and pore size in order to achieve better in
vivo performance. Through our in vivo assays, we found no gap formation
between porous tantalum rod and host bone, porous tantalum and the
host bone interfaces integrated with each other and new trabecular
bone filled the pore of the porous tantalum material. The integration
of porous tantalum and host bone was further confirmed by mechanical
tests. We infer that the enough bone ingrowth may improve the mechanical
strength of materials. These results indicated that the cancellous
bone mimic scaffold may be suitable for the treatment of bone defect.To determine the osteoinductivity of the porous tantalum scaffold,
we inserted the porous tantalum rod in the center of bone marrow cavity.
As the bone marrow cavity is a good source of MSCs, this model should
also determine the osteoinductivity of the porous tantalum scaffold.
As shown in Figures and 5, new bone formation and ingrowth were
found in the bone marrow cavity side of the porous tantalum rod, these
results indicated that the porous tantalum scaffold is osteoinductive.
On the other hand, we found that more new bone formation in the bone
marrow cavity side when compared with the cortical side we inferred
that this was because instead of MSCs in the bone marrow side, osteocytes
in the cortical side were limited in the formation of cancellous bone.
In addition, porosity, pore diameter, and elastic modulus of cortical
are different from cancellous bone,[11,25,32] although bone growth was also found in the cortical
side of the tantalum rod (Figures and 5), the porous tantalum
scaffold is cancellous bone mimic, hence, density gradient and site
specific scaffolds may be more helpful for the treatment of the clinical
bone defect.In the past decades, surgeons are interested in
metal material-mediated
bone defect repair; however, metal materials cause an electrochemical
process known as corrosion which results in pain and makes it uncomfortable
for patients.[7] Recently, tantalum, a bioinert
metal was characterized to be extremely resistant to corrosion has
been applied for clinical use.[7,14] In the present study,
we designed a novel porous tantalum material, the cytotoxicity still
needs to be evaluated. Through in vitro MTT assay, we found that the
porous tantalum material is not toxic to MSCs. The in vivo pathological
scoring also supported the safety of the porous tantalum (data not
shown). That is to say, the novel porous tantalum scaffold is safe
for implantation in vivo. Further research studies focus on long-term
observation (longer than 1year) of bone ingrowth and compare with
other porous materials should be carried out.
Conclusions
In
summary, we manufactured a novel three-dimensional, multiscaled,
and interconnected porous tantalum scaffold with porosity ranges from
60 to 80%, pore diameter ranges from 200 to 500 μm, elastic
modulus ranges from 0.5 to 4.0 GPa, all the parameters are closer
to trabecular bone. Through our in vitro and in vivo canine femoral
shaft bone defect model, we found the novel porous tantalum scaffold
is trabecular bone mimic, biocompatible, and osteoinductive, which
holds potential for bone tissue engineering application.
Materials and
Methods
Manufacture and Structural Characterization of the Porous Tantalum
Materials
The novel porous tantalum scaffold was manufactured
by Chongqing Runze Pharmaceutical Co.,Ltd. (Chongqing, China). Briefly,
pure tantalum powders were utilized for the manufacturing of poroustantalum materials. Tantalum powders with a specific amount of sponge
carrier and additives, which were used to regulate the pore diameter
of the porous scaffold, pore distribution, and porosity were subjected
to slip-casting forming, then underwent 1500–2100 °C high
temperature sintering, and finally approved to after treatment technology
and necessary preparation (preparation procedures has been filed patents).The porous tantalum was subjected to SEM (Nova NanoSEM, NE, USA
accelerating voltage of 15 kV and JEOL SEM, Tokyo, Japan, accelerating
voltage of 5 kV) analysis. The porous tantalum rod was also characterized
using industrial CT (TOSCANER, Tokyo, Japan) at 500 Kv and 1.4 mA.
Cell Culture and Related Chemicals
The MSC cell line
UC-MSCs (human umbilical cord-derived MSCs) was previously characterized.[33] The UC-MSC cell line was obtained from Chongqing
Engineering Research Center of Stem Cell Therapy (Chongqing, China).
The UC-MSCs were maintained in complete DMEM (Hyclone, China) containing
10% fetal bovine serum (Gibco, Australia), 100 units/mL penicillin,
and 100 μg/mL streptomycin at 37 °C in 5% CO2 incubator.
Culturing UC-MSCs on Tantalum Powder and
SEM Analysis
The UC-MSCs were digested with trypsin and resuspended
in culture
medium at a concentration of 2 × 104/μL. Around
50 μL of cell suspension (106 cells) was seeded onto
the poroustantalum powder (diameter 2–3 mm). The tantalum
powders with cells were then incubated in an incubator for 2 h, and
transferred into 6-well plates containing complete DMEM medium.On day 3 and day 5, the UC-MSCs/tantalum powders were fixed in 2.5%
glutaraldehyde buffered with 0.1 M sodiumcacodylate (NaC) solution
at 4 °C overnight. Then, the samples were washed with 0.1 M NaC,
and then subjected to secondary fixation with 1% osmium tetroxide
in 0.1 M NaC for 2 h, followed by dehydrated in gradual ethanol boxes
for 15 min, respectively, and finally subjected to SEM observation
(HITACHI SU8010, Tokyo, Japan) with an accelerating voltage of 20
kV.
Cytotoxicity Examination
Novel porous tantalum scaffold
extract preparation followed the ISO 10993-5, 2009 standard.[22] The extract was prepared according the principle
of ‘material weight/extraction transmitter = 0.2 g/mL’,
which was identified as 100% concentration. Briefly, sterile porous
tantalum powders were maintained in complete DMEM and kept in a CO2 incubator for 72 h. The extract was sterilized with the use
of 0.22 μm filter.As for the MTT assay, 200 μL
of porous tantalum extract (experimental group) or complete DMEM (control
group) were added in a 96-well plate, then UC-MSC cells were resuspended,
and seeded at a concentration of 2 × 104 per well.
At designed time points, 20 μL of MTT (5 mg/mL) medium was added
to each well (pH 7.4) and the plate was incubated in a CO2 incubator at 37 °C for 4 h, optical density value of each well
was determined at a wavelength of 490 nm.
Construction of Canine
Femoral Shaft Bone Defect Models and
Implantation of Porous Tantalum Rod
All animal experiments
were performed according to institutional guidelines under the approved
protocols by the Chongqing Administration Rule of Laboratory Animals
and the National Institutes of Health Guide for the Care and Use of
Laboratory Animals. A total of 7 adult beagle dogs (weight 6–8
Kg) were selected. After intraperitoneal anesthesia with 3% pentobarbital
sodium (1 mL/kg) and standard aseptic surgical procedures, the midpiece
of the femur shaft was exposed by a lateral incision. A bone defect
model was constructed with the use of a surgery drill bit at the middle
part of the femur shaft, which resulted in a bone defect with a width
of ∼4 mm and a depth of 8 mm (reach to the middle part of bone
marrow cavity). A porous tantalum rod (4 mm in diameter and 8 mm in
height) was implanted in the femur along with the defect, which was
as close as possible in contact with the host bone. Each Canine Femur
was implanted with two porous tantalum rods, and left thigh was selected
as the surgical side. After the surgical procedures, the wound was
sutured after adequate hemostasis. X-ray was taken after the surgery
to confirm that the tantalum rods were in the femur shaft and without
a peri-implant fracture. The animals were kept in independent cages
with standard conditions.At the indicated time points, animals
were euthanized by overdose intraperitoneal pentobarbital sodium injection
(Sigma-Aldrich, United States). All efforts were made to minimize
the suffering of the animals, the animal was kept in standard conditions
until it was confirmed that the animal was sacrificed. Then, the femurs
were extracted to observe the bone growth around the porous tantalum
rods. A hollow orthopaedic drill with a diameter of ∼5 mm was
used to completely remove the porous tantalum rods.
Hard Tissue
Slicing and Staining
All harvested specimens
were fixed immediately with 4% phosphate-buffered paraformaldehyde.
After gradient dehydration in ethanol solutions (60, 75, 88, 95, and
100%), infiltration, embedding in epoxy resin, and polymerization
were performed. Then, a metal slicer was utilized to slice the specimens
along the direction parallel to the longitudinal axis of the novel
porous tantalum rods, a porous tantalum rod plane was fully exposed.
After that the slices were polished to 90 μm in thickness. Finally,
sections were subjected to methylene blue and VG staining, osteogenesis
and bone ingrowth of the porous tantalum–bone interface were
recorded by microscopy. ImageJ software was used for the quantitative
analysis the new bone formation in 3 and 6 month’s VG staining,
the ratio of the trabecular bone and tantalum scaffold was calculated
in 5 different fields, respectively.
Push-Out Test
To confirm the interfacial bond strength
between the bone and the implant, the specimens were tested by a mechanical
push-out test at the indicated time points using universal testing
machine (Shimadzu AG-IS, Kyoto, Japan). In brief, samples were fixed
on a support fixture with a central hole (4 mm in diameter). Then,
the fixture was firmly fixed onto the machine platform. The load was
carried out collinearly along the long axis of the implant with a
3.0 mm cylindrical plunger. The test began with a consistent displacement
speed of 0.5 mm/min, the test stopped when the fracture or failure
(max load: 900 N) of the interface was observed. The push-out test
strength was recorded from the peak on the load-deformation curve.
Normal femoral shaft specimens were used as the control, max load
was calculated according to the normal specimens.
Statistical
Analysis
Quantitative data were expressed
as mean ± SD. SPSS 20.0 statistical software (IBM, Armonk, NY,
USA) was applied for data analysis. One-way analysis of variance (ANOVA)
or t-test was used to determine the statistical significance
with a cut off of p < 0.05.
Authors: S Van Bael; Y C Chai; S Truscello; M Moesen; G Kerckhofs; H Van Oosterwyck; J-P Kruth; J Schrooten Journal: Acta Biomater Date: 2012-04-07 Impact factor: 8.947
Authors: Y Li; J Zhou; P Pavanram; M A Leeflang; L I Fockaert; B Pouran; N Tümer; K-U Schröder; J M C Mol; H Weinans; H Jahr; A A Zadpoor Journal: Acta Biomater Date: 2017-12-12 Impact factor: 8.947
Authors: S M Ahmadi; R Hedayati; Y Li; K Lietaert; N Tümer; A Fatemi; C D Rans; B Pouran; H Weinans; A A Zadpoor Journal: Acta Biomater Date: 2017-11-08 Impact factor: 8.947