An Analysis of 2019 FDA Adverse Events for Two Insulin Pumps and Two Continuous Glucose Monitors.

Jan S Krouwer1.   

Abstract

US Food and Drug Administration adverse event data for 2019 were analyzed for two insulin pumps and two continuous glucose monitors (CGMs). The analyses were selective-they were guided by the text described in the adverse events. They included (1) percent using auto mode for the Medtronic 670G pump, (2) distributions of hyper and hypo glucose values for Medtronic and Tandem pumps, (3) a Parkes error grid for Dexcom CGM vs glucose meter when the complaint was inaccuracy, and (4) the most frequent events for Abbott Freestyle. We found that for the 670G pump, there were more hypo events when auto mode was on than when auto mode was off. With Dexcom CGMs, users complained about inaccurate result when most results were in the B zone. With the Abbott Freestyle, the most frequent adverse event was an allergic skin reaction.

Entities:  

Keywords:  MAUDE; Parkes glucose meter error grid; adverse event; continuous glucose monitor; insulin pump

Mesh:

Substances:

Year:  2020        PMID: 32880188      PMCID: PMC8875056          DOI: 10.1177/1932296820951872

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  5 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

2.  Reducing Glucose Meter Adverse Events by Using Reliability Growth With the FDA MAUDE Database.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2018-12-23

3.  Real world hybrid closed-loop discontinuation: Predictors and perceptions of youth discontinuing the 670G system in the first 6 months.

Authors:  Laurel H Messer; Cari Berget; Tim Vigers; Laura Pyle; Cristy Geno; R Paul Wadwa; Kimberly A Driscoll; Gregory P Forlenza
Journal:  Pediatr Diabetes       Date:  2020-01-03       Impact factor: 4.866

4.  Retrospective Analysis of 3-Month Real-World Glucose Data After the MiniMed 670G System Commercial Launch.

Authors:  Michael P Stone; Pratik Agrawal; Xiaoxiao Chen; Margaret Liu; John Shin; Toni L Cordero; Francine R Kaufman
Journal:  Diabetes Technol Ther       Date:  2018-08-30       Impact factor: 6.118

5.  One Year Clinical Experience of the First Commercial Hybrid Closed-Loop System.

Authors:  Rayhan A Lal; Marina Basina; David M Maahs; Korey Hood; Bruce Buckingham; Darrell M Wilson
Journal:  Diabetes Care       Date:  2019-09-23       Impact factor: 19.112

  5 in total
  1 in total

1.  Adverse Event Data for Years 2018 to 2020 for Diabetes Devices.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2021-05-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.