Literature DB >> 31885123

Real world hybrid closed-loop discontinuation: Predictors and perceptions of youth discontinuing the 670G system in the first 6 months.

Laurel H Messer1, Cari Berget1, Tim Vigers1,2, Laura Pyle1,2, Cristy Geno1, R Paul Wadwa1, Kimberly A Driscoll1,3, Gregory P Forlenza1.   

Abstract

OBJECTIVE: To describe predictors of hybrid closed loop (HCL) discontinuation and perceived barriers to use in youth with type 1 diabetes.
SUBJECTS: Youth with type 1 diabetes (eligible age 2-25 y; recruited age 8-25 y) who initiated the Minimed 670G HCL system were followed prospectively for 6 mo in an observational study. RESEARCH DESIGN AND METHODS: Demographic, glycemic (time-in-range, HbA1c), and psychosocial variables [Hypoglycemia Fear Survey (HFS); Problem Areas in Diabetes (PAID)] were collected for all participants. Participants who discontinued HCL (<10% HCL use at clinical visit) completed a questionnaire on perceived barriers to HCL use.
RESULTS: Ninety-two youth (15.7 ± 3.6 y, HbA1c 8.8 ± 1.3%, 50% female) initiated HCL, and 28 (30%) discontinued HCL, with the majority (64%) discontinuing between 3 and 6 mo after HCL start. Baseline HbA1c predicted discontinuation (P = .026) with the odds of discontinuing 2.7 times higher (95% CI: 1.123, 6.283) for each 1% increase in baseline HbA1c. Youth who discontinued HCL rated difficulty with calibrations, number of alarms, and too much time needed to make the system work as the most problematic aspects of HCL. Qualitatively derived themes included technological difficulties (error alerts, not working correctly), too much work (calibrations, fingersticks), alarms, disappointment in glycemic control, and expense (cited by parents).
CONCLUSIONS: Youth with higher HbA1c are at greater risk for discontinuing HCL than youth with lower HbA1c, and should be the target of new interventions to support device use. The primary reasons for discontinuing HCL relate to the workload required to use HCL.
© 2019 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  artificial pancreas; continuous glucose monitor; hybrid closed loop; insulin pump; pediatrics; type 1 diabetes

Year:  2020        PMID: 31885123      PMCID: PMC7204392          DOI: 10.1111/pedi.12971

Source DB:  PubMed          Journal:  Pediatr Diabetes        ISSN: 1399-543X            Impact factor:   4.866


  38 in total

1.  Variation and trends in incidence of childhood diabetes in Europe. EURODIAB ACE Study Group.

Authors: 
Journal:  Lancet       Date:  2000-03-11       Impact factor: 79.321

2.  Research electronic data capture (REDCap)--a metadata-driven methodology and workflow process for providing translational research informatics support.

Authors:  Paul A Harris; Robert Taylor; Robert Thielke; Jonathon Payne; Nathaniel Gonzalez; Jose G Conde
Journal:  J Biomed Inform       Date:  2008-09-30       Impact factor: 6.317

Review 3.  Adolescent psychosocial, social, and cognitive development.

Authors:  Renata Arrington Sanders
Journal:  Pediatr Rev       Date:  2013-08

Review 4.  6. Glycemic Targets: Standards of Medical Care in Diabetes-2019.

Authors: 
Journal:  Diabetes Care       Date:  2019-01       Impact factor: 19.112

5.  A Clinical Training Program for Hybrid Closed Loop Therapy in a Pediatric Diabetes Clinic.

Authors:  Cari Berget; Sarah E Thomas; Laurel H Messer; Katelin Thivener; Robert H Slover; R Paul Wadwa; G Todd Alonso
Journal:  J Diabetes Sci Technol       Date:  2019-03-12

6.  Sensor-augmented pump therapy for A1C reduction (STAR 3) study: results from the 6-month continuation phase.

Authors:  Richard M Bergenstal; William V Tamborlane; Andrew Ahmann; John B Buse; George Dailey; Stephen N Davis; Carol Joyce; Bruce A Perkins; John B Welsh; Steven M Willi; Michael A Wood
Journal:  Diabetes Care       Date:  2011-09-20       Impact factor: 17.152

7.  Hybrid Closed-Loop Insulin Delivery in Type 1 Diabetes During Supervised Outpatient Conditions.

Authors:  Benyamin Grosman; Jacob Ilany; Anirban Roy; Natalie Kurtz; Di Wu; Neha Parikh; Gayane Voskanyan; Noa Konvalina; Chrystaleni Mylonas; Rebecca Gottlieb; Francine Kaufman; Ohad Cohen
Journal:  J Diabetes Sci Technol       Date:  2016-05-03

8.  Assessing patient-reported outcomes for automated insulin delivery systems: the psychometric properties of the INSPIRE measures.

Authors:  J Weissberg-Benchell; J B Shapiro; K Hood; L M Laffel; D Naranjo; K Miller; K Barnard
Journal:  Diabet Med       Date:  2019-03-20       Impact factor: 4.359

9.  Hypoglycemia Reduction and Changes in Hemoglobin A1c in the ASPIRE In-Home Study.

Authors:  Ram Weiss; Satish K Garg; Bruce W Bode; Timothy S Bailey; Andrew J Ahmann; Kenneth A Schultz; John B Welsh; John J Shin
Journal:  Diabetes Technol Ther       Date:  2015-06-02       Impact factor: 6.118

10.  Type 1 diabetes through the life span: a position statement of the American Diabetes Association.

Authors:  Jane L Chiang; M Sue Kirkman; Lori M B Laffel; Anne L Peters
Journal:  Diabetes Care       Date:  2014-07       Impact factor: 19.112

View more
  29 in total

1.  Therapeutic Contract in Adolescents Using Hybrid Closed Loop System.

Authors:  Gianluca Tornese; Claudia Carletti; Chiara Barbiero; Elena Faleschini; Egidio Barbi
Journal:  J Diabetes Sci Technol       Date:  2020-08-24

2.  Six months of hybrid closed loop in the real-world: An evaluation of children and young adults using the 670G system.

Authors:  Cari Berget; Laurel H Messer; Tim Vigers; Brigitte I Frohnert; Laura Pyle; R Paul Wadwa; Kimberly A Driscoll; Gregory P Forlenza
Journal:  Pediatr Diabetes       Date:  2020-01-07       Impact factor: 4.866

3.  A Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes.

Authors:  Marc D Breton; Lauren G Kanapka; Roy W Beck; Laya Ekhlaspour; Gregory P Forlenza; Eda Cengiz; Melissa Schoelwer; Katrina J Ruedy; Emily Jost; Lori Carria; Emma Emory; Liana J Hsu; Mary Oliveri; Craig C Kollman; Betsy B Dokken; Stuart A Weinzimer; Mark D DeBoer; Bruce A Buckingham; Daniel Cherñavvsky; R Paul Wadwa
Journal:  N Engl J Med       Date:  2020-08-27       Impact factor: 91.245

4.  Impact of the Hybrid Closed-Loop System on Sleep and Quality of Life in Youth with Type 1 Diabetes and Their Parents.

Authors:  Erin C Cobry; Emily Hamburger; Sarah S Jaser
Journal:  Diabetes Technol Ther       Date:  2020-10-13       Impact factor: 6.118

5.  An Analysis of 2019 FDA Adverse Events for Two Insulin Pumps and Two Continuous Glucose Monitors.

Authors:  Jan S Krouwer
Journal:  J Diabetes Sci Technol       Date:  2020-09-03

Review 6.  Felix dies natalis, insulin… ceterum autem censeo "beta is better".

Authors:  Lorenzo Piemonti
Journal:  Acta Diabetol       Date:  2021-05-23       Impact factor: 4.280

7.  Extended Use of the Control-IQ Closed-Loop Control System in Children With Type 1 Diabetes.

Authors:  Lauren G Kanapka; R Paul Wadwa; Marc D Breton; Katrina J Ruedy; Laya Ekhlaspour; Gregory P Forlenza; Eda Cengiz; Melissa J Schoelwer; Emily Jost; Lori Carria; Emma Emory; Liana J Hsu; Stuart A Weinzimer; Mark D DeBoer; Bruce A Buckingham; Mary Oliveri; Craig Kollman; Betsy B Dokken; Daniel Cherñavvsky; Roy W Beck
Journal:  Diabetes Care       Date:  2020-12-21       Impact factor: 19.112

8.  Clinical Management and Pump Parameter Adjustment of the Control-IQ Closed-Loop Control System: Results from a 6-Month, Multicenter, Randomized Clinical Trial.

Authors:  Grenye O'Malley; Laurel H Messer; Carol J Levy; Jordan E Pinsker; Gregory P Forlenza; Elvira Isganaitis; Yogish C Kudva; Laya Ekhlaspour; Dan Raghinaru; John Lum; Sue A Brown
Journal:  Diabetes Technol Ther       Date:  2021-04       Impact factor: 6.118

9.  Automated Insulin Delivery in Real Life (AID-IRL): Real-World User Perspectives on Commercial AID.

Authors:  Dana Lewis
Journal:  J Diabetes Sci Technol       Date:  2020-09-16

10.  Real-World Use of a New Hybrid Closed Loop Improves Glycemic Control in Youth with Type 1 Diabetes.

Authors:  Laurel H Messer; Cari Berget; Laura Pyle; Timothy Vigers; Erin Cobry; Kimberly A Driscoll; Gregory P Forlenza
Journal:  Diabetes Technol Ther       Date:  2021-06-21       Impact factor: 6.118

View more