Influence of feature calculating parameters on the reproducibility of CT radiomic features: a thoracic phantom study.

Ying Li1, Guanghua Tan1, Mark Vangel1, Jonathan Hall1,2, Wenli Cai1.   

Abstract

BACKGROUND: Existing studies have demonstrated that imaging parameters may affect radiomic features. However, the influence of feature calculating parameters has been overlooked. The purpose of this study is to investigate the influence of feature calculating parameters (gray-level range and bin size) on the reproducibility of CT radiomic features.
METHODS: Thirty-six CT scans from an anthropomorphic thoracic phantom were acquired with different imaging parameters including effective dose, pitch, slice thicknesses and reconstruction kernels. The influence of feature calculating parameters was investigated in terms of three gray-level ranges and eleven gray-level bin sizes. Feature reproducibility was assessed by the intraclass correlation coefficient (ICC) with the cutoff value of 0.8 and the coefficient of variation (CV) with the cutoff value of 20%. The agreements of reproducible features in different ranges and bin sizes were analyzed by Kendall's W test and Kappa test. The proportions of reproducible features, in terms of two calculating, four imaging and two segmentation parameters, were evaluated using Cochran's Q test and Dunn's test.
RESULTS: For the three gray-level ranges, 50% (44/88) of features were reproducible with a perfect agreement (Kendall's W coefficient 0.844, P<0.001). Of the 72 features that may be influenced by gray-level bin size, 33.3% (24/72) were reproducible for 11 bin sizes with a perfect agreement (Kendall's W coefficient 0.879, P<0.001). For the proportions of reproducible features, there was no statistically significant difference among three ranges (P=0.420), but there was among eleven bin sizes (P=0.013). The proportions of reproducible features in feature calculating parameters were statistically significantly lower than those in imaging parameters (adjusted P<0.05).
CONCLUSIONS: Feature calculating parameters may have a greater influence than imaging parameters on the reproducibility of CT radiomic features, which should be given special attention in clinical applications. 2020 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Radiomics; X-ray; computed tomography (CT); phantom; reproducibility

Year:  2020        PMID: 32879856      PMCID: PMC7417756          DOI: 10.21037/qims-19-921

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  22 in total

1.  Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT.

Authors:  Timothy P Szczykutowicz; Andrew DuPlissis; Perry J Pickhardt
Journal:  AJR Am J Roentgenol       Date:  2017-03-07       Impact factor: 3.959

2.  Influence of gray level discretization on radiomic feature stability for different CT scanners, tube currents and slice thicknesses: a comprehensive phantom study.

Authors:  Ruben T H M Larue; Janna E van Timmeren; Evelyn E C de Jong; Giacomo Feliciani; Ralph T H Leijenaar; Wendy M J Schreurs; Meindert N Sosef; Frank H P J Raat; Frans H R van der Zande; Marco Das; Wouter van Elmpt; Philippe Lambin
Journal:  Acta Oncol       Date:  2017-09-08       Impact factor: 4.089

3.  Can we trust the calculation of texture indices of CT images? A phantom study.

Authors:  Caroline Caramella; Adrien Allorant; Fanny Orlhac; Francois Bidault; Bernard Asselain; Samy Ammari; Patricia Jaranowski; Aurelie Moussier; Corinne Balleyguier; Nathalie Lassau; Stephanie Pitre-Champagnat
Journal:  Med Phys       Date:  2018-03-13       Impact factor: 4.071

4.  Radiomics of CT Features May Be Nonreproducible and Redundant: Influence of CT Acquisition Parameters.

Authors:  Roberto Berenguer; María Del Rosario Pastor-Juan; Jesús Canales-Vázquez; Miguel Castro-García; María Victoria Villas; Francisco Mansilla Legorburo; Sebastià Sabater
Journal:  Radiology       Date:  2018-04-24       Impact factor: 11.105

5.  Validation of A Method to Compensate Multicenter Effects Affecting CT Radiomics.

Authors:  Fanny Orlhac; Frédérique Frouin; Christophe Nioche; Nicholas Ayache; Irène Buvat
Journal:  Radiology       Date:  2019-01-29       Impact factor: 11.105

6.  Accuracy of the CT numbers of simulated lung nodules imaged with multi-detector CT scanners.

Authors:  Mitchell M Goodsitt; Heang-Ping Chan; Ted W Way; Sandra C Larson; Emmanuel G Christodoulou; Jeomsoon Kim
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

7.  A radiomics approach to assess tumour-infiltrating CD8 cells and response to anti-PD-1 or anti-PD-L1 immunotherapy: an imaging biomarker, retrospective multicohort study.

Authors:  Roger Sun; Elaine Johanna Limkin; Maria Vakalopoulou; Laurent Dercle; Stéphane Champiat; Shan Rong Han; Loïc Verlingue; David Brandao; Andrea Lancia; Samy Ammari; Antoine Hollebecque; Jean-Yves Scoazec; Aurélien Marabelle; Christophe Massard; Jean-Charles Soria; Charlotte Robert; Nikos Paragios; Eric Deutsch; Charles Ferté
Journal:  Lancet Oncol       Date:  2018-08-14       Impact factor: 41.316

8.  Assessing Agreement between Radiomic Features Computed for Multiple CT Imaging Settings.

Authors:  Lin Lu; Ross C Ehmke; Lawrence H Schwartz; Binsheng Zhao
Journal:  PLoS One       Date:  2016-12-29       Impact factor: 3.240

9.  Harmonizing the pixel size in retrospective computed tomography radiomics studies.

Authors:  Dennis Mackin; Xenia Fave; Lifei Zhang; Jinzhong Yang; A Kyle Jones; Chaan S Ng; Laurence Court
Journal:  PLoS One       Date:  2017-09-21       Impact factor: 3.240

10.  Decoding tumour phenotype by noninvasive imaging using a quantitative radiomics approach.

Authors:  Hugo J W L Aerts; Emmanuel Rios Velazquez; Ralph T H Leijenaar; Chintan Parmar; Patrick Grossmann; Sara Carvalho; Sara Cavalho; Johan Bussink; René Monshouwer; Benjamin Haibe-Kains; Derek Rietveld; Frank Hoebers; Michelle M Rietbergen; C René Leemans; Andre Dekker; John Quackenbush; Robert J Gillies; Philippe Lambin
Journal:  Nat Commun       Date:  2014-06-03       Impact factor: 14.919

View more
  3 in total

1.  Quantitative assessment of acquisition imaging parameters on MRI radiomics features: a prospective anthropomorphic phantom study using a 3D-T2W-TSE sequence for MR-guided-radiotherapy.

Authors:  Jing Yuan; Cindy Xue; Gladys Lo; Oi Lei Wong; Yihang Zhou; Siu Ki Yu; Kin Yin Cheung
Journal:  Quant Imaging Med Surg       Date:  2021-05

2.  Prediction of neoadjuvant chemotherapy response in high-grade osteosarcoma: added value of non-tumorous bone radiomics using CT images.

Authors:  Lei Xu; Pengfei Yang; Kun Hu; Yan Wu; Meng Xu-Welliver; Yidong Wan; Chen Luo; Jing Wang; Jinhua Wang; Jiale Qin; Yi Rong; Tianye Niu
Journal:  Quant Imaging Med Surg       Date:  2021-04

Review 3.  Deep Learning With Radiomics for Disease Diagnosis and Treatment: Challenges and Potential.

Authors:  Xingping Zhang; Yanchun Zhang; Guijuan Zhang; Xingting Qiu; Wenjun Tan; Xiaoxia Yin; Liefa Liao
Journal:  Front Oncol       Date:  2022-02-17       Impact factor: 6.244

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.