Literature DB >> 32804668

Evaluation of the automated cartridge-based ARIES SARS-CoV-2 Assay (RUO) against automated Cepheid Xpert Xpress SARS-CoV-2 PCR as gold standard.

Konstantin Tanida1,2, Lars Koste1, Christian Koenig1, Werner Wenzel1, Andreas Fritsch2, Hagen Frickmann1,3.   

Abstract

INTRODUCTION: To evaluate the automated cartridge-based PCR approach ARIES SARS-CoV-2 Assay targeting the ORF-sequence and the N-gene of SARS-CoV-2.
METHODS: In line with the suggestions by Rabenau and colleagues, the automated ARIES SARS-CoV-2 Assay was challenged with strongly positive samples, weakly positive samples and negative samples. Further, intra-assay and inter-assay precision as well as the limit-of-detection (lod) were defined with quantified target RNA and DNA. The Cepheid Xpert Xpress SARS-Cov-2 Assay was used as gold standard.
RESULTS: Concordance between the ARIES assay and the Cepheid assay was 100% for strongly positive samples and for negative samples, respectively. For weakly positive samples as confirmed applying the Cepheid assay, a relevant minority of 4 out of 15 samples (26.7%) went undetected by the ARIES assay. Intra- and inter-assay precision were satisfactory, while the lod was in the 103 DNA copies/reaction-range, in the 103 virus copies/reaction-range, or in the 103-104 free RNA copies/reaction-range in our hands.
CONCLUSIONS: The automated ARIES assay shows comparable test characteristics as the Cepheid assay focusing on strongly positive and negative samples but a slightly reduced sensitivity with weakly positive samples. Decisions on diagnostic use should include considerations on the lod.

Entities:  

Keywords:  COVID-19; SARS-CoV-2; automated PCR; molecular rapid testing; point-of-care-testing; testing comparison

Year:  2020        PMID: 32804668      PMCID: PMC7592518          DOI: 10.1556/1886.2020.00017

Source DB:  PubMed          Journal:  Eur J Microbiol Immunol (Bp)        ISSN: 2062-509X


INTRODUCTION

The global medical and public health efforts associated with the emergence of the COVID-(Corona Virus Disease-) 19-associated SARS-CoV-2 (Severe Acute Respiratory Syndrome-Corona Virus-2) starting from Wuhan, China, in December 2019 [1] makes the availability of rapid and reliable diagnostic approaches desirable. First polymerase chain reaction (PCR) protocols were rapidly established [2] and made available by organizations like the Center of Disease Control (CDC) [3, 4] or the China CDC [5]. However, traditional in-house real-time PCR is demanding both regarding infrastructural conditions and availability of skilled and well-trained technical assistants [6]. Accordingly, there is demand for rapid and automated diagnostic devices. With positive molecular SARS-CoV-2 test results within less than 10 minutes, the isothermal Abbott ID Now (Abbott Laboratories, Abbott Park, Illinois, USA) approach is presently the most rapid well evaluated molecular point-of-care-testing assay for the diagnosis of COVID-19 but it is outperformed regarding sensitivity by PCR-based assays [7-10]. One well-evaluated, highly sensitive automated PCR-based testing approach is the cartridge-based Xpert Xpress SARS-CoV-2 Assay (Cepheid, Sunnyvale, California, USA) with a limit of detection of less than 100 copies [11-16]. Recently, another automated cartridge-based system ARIES SARS-CoV-2 Assay (Luminex Corporation, Austin, Texas, United States) was cleared by the US Food and Drug Administration (FDA) for in-vitro-diagnostics under emergency use authorization (EUA-IVD), while the system was distributed under a research-use-only (RUO) label in Europa. A validation of the ARIES SARS-CoV-2 Assay against the well described above-mentioned Cepheid assay in line with the recommendations according to Rabenau and colleagues [17, 18] was performed to assess its suitability for in-vitro diagnostic use.

METHODS

Study protocol

The study was conducted as a test validation using the batch number AB0629 of the automated cartridge-based ARIES SARS-CoV-2 Assay (Luminex, referred to as “ARIES assay” in the following) targeting the ORF-sequence and the N-gene with the well-evaluated Xpert Xpress SARS-CoV-2 Assay (Cepheid, referred to as “Cepheid assay” in the following) targeting the E-gene and the N2-region of the N-gene as the gold standard according to the validation protocol as suggested for laboratory-developed real-time PCR assays by Rabenau and colleagues [17, 18]. This implies comparative testing of strongly positive samples (defined as Cepheid cycle threshold (Ct) value ≤34 for all target genes), weakly positive samples (defined as Cepheid Ct value >34 for at least one target gene), negative samples according to the gold standard, inter-assay and intra-assay precision testing as well as the definition of the limit of detection (lod). Here, the lod was determined using patient samples with defined SARS-CoV-2 copy numbers, positive control plasmids and synthesized positive control RNA. Both assays were applied as described by the manufacturer with 200 μL sample used for the ARIES system and 300 μL sample used for the Cepheid approach.

Positive control oligonucleotides

Positive control DNA and RNA was designed based on sequence fragments of the N-gene (N1, N2, N3, China CDC), the RdRp-gene, the E-gene and the ORF-region derived from the genome of the Severe acute respiratory syndrome coronavirus 2 isolate Wuhan-Hu-1 (GenBank accession number NC_045512.2) as identified applying the previously described primers [2-5], prolonged by ≥10 bases at the 5'-end and the 3'-end, and linked by EcoR1 restriction sites (sequence details in Table 1). A pEX-A128 vector backbone was used to construct the positive control DNA plasmid (eurofins Genomics, Ebersberg, Germany). Positive control RNA of 1,005 bases length was transcribed with a T7 RNA polymerase from a linear template generated by PCR (AmpTec GmbH, Hamburg, Germany). Serial dilution was performed with eSwab collection & preservation of Aerobic, Anaerobic & Fastidious Bacteria (COPAN, Brescia, Italia) transport and storage medium spiked with respiratory sample material from patients; for the positive control RNA, it was also attempted with PCR grade water to prevent preterm RNA degradation due to RNAses included in the human patient samples.
Table 1.

Sequence insert fragments used for the positive control DNA plasmid and the positive control RNA as derived from the genome of the Severe acute respiratory syndrome coronavirus 2 isolate Wuhan-Hu-1 (GenBank accession number NC_045512.2)

RdRp-gene fragment sequence
TTCTATAGATTAGCTAATGAGTGTGCTCAAGTATTGAGTG AAATGGTCATGTGTGGCGGTTCACTATATGTTAAACCAGG TGGAACCTCATCAGGAGATGCCACAACTGCTTATGCTAAT AGTGTTTTTAACATTTGTCAAGCTGTCACGGCC
E-Gene fragment sequence
ATGTACTCATTCGTTTCGGAAGAGACAGGTACGTTAATAG TTAATAGCGTACTTCTTTTTCTTGCTTTCGTGGTATTCTTG CTAGTTACACTAGCCATCCTTACTGCGCTTCGATTGTGTG CGTACTGCTGCAATATTGTTAACGTGAGTCTTGTA
N1/N2/N3 fragment sequence (CDC)
CAAAGACGGCATCATATGGGTTGCAACTGAGGGAGCCTT GAATACACCAAAAGATCACATTGGCACCCGCAATCCTGC TAACAATGCTGCAATCGTGCTACAACTTCCTCAAGGAACA ACATTGCCAAAAGGCTTCTACGCAGAAGGGAGCAGAGGC GGCAGTCAAGCCTCTTCTCGTTCCTCATCACGTAGTCGCA ACAGTTCAAG
N-gene fragment sequence (China-CDC)
TTCAACTCCAGGCAGCAGTAGGGGAACTTCTCCTGCTAG AATGGCTGGCAATGGCGGTGATGCTGCTCTTGCTTTGCT GCTGCTTGACAGATTGAACCAGCTTGAGAGCAAAATGTC TGGTAAAGGCCAACAACAACAA
ORF fragment sequence (China-CDC)
CCTACAACTTGTGCTAATGACCCTGTGGGTTTTACACTTA AAAACACAGTCTGTACCGTCTGCGGTATGTGGAAAGGTT ATGGCTGTAGTTGTGATCAACTCCGCGAACCCATGCTTCA GTCAGCTGATGCACAATCGTTTTTAAACGGGTTTGCGGTG

Origin of samples

Patient samples were partly derived from the diagnostic routine of the Bundeswehr Hospital Hamburg. Further, patient samples with quantified viral copy numbers were provided by the University Hospital Regensburg, the institution from where external laboratory control assessment schemes for molecular diagnostic tests (“Ringversuche”) are provided in Germany. Finally, well characterized material from the external laboratory control assessment scheme “INSTAND EQA scheme (340) virus genome detection of SARS-CoV-2 - April 2020” by INSTAND e.V. (Düsseldorf, Germany) was applied.

Ethical considerations

As clarified previously (WF-011/19) by the Ethics Committee of the Medical Association of Hamburg, Germany, in line with German National Laws, use of anonymized residual sample materials for test comparison purposes in diagnostic laboratories is neither a “research project involving human beings” in line with § 9 chapter 2 of the “Hamburgisches Kammergesetz für Heilberufe” (Hamburg’s Association of Health Care Professions Act) nor a research project requiring ethical advice according to §15 chapter 1 of the “Berufsordnung für Hamburger Ärzte und Ärztinnen” (Professional Regulations for Physicians in Hamburg). Therefore, ethical clearance was not required for this assessment.

RESULTS AND DISCUSSION

Assessment with strongly positive, weakly positive and negative samples

With the Cepheid SARS-CoV-2 assay applied as gold standard, sensitivity of the automated ARIES assay with the strongly positive samples was 100% (15/15), specificity was 100% (30/30), resulting in excellent agreement according to Landis & Koch [19]. For the weakly positive samples, however, agreement of the ARIES assay with the Cepheid assay was reduced to 26.7% (4/15). Overall agreement about all tested samples between the two assays was 81.7% and Cohen’s Kappa was 0.63 indicating substantial agreement according to Landis & Koch [19]. Thereby, negative reactions of the ARIES-PCR targeting the ORF-sequence were more frequently observed, even in case of the strongly positive samples, compared with the ARIES N-gene-PCR. Details are provided in Tables 2–4.

Inter- and intra-assay variation

For the assessment of inter- and intra-assay variation, excellent concordance was observed for the strongly positive samples, while slightly more variation was seen for the weakly positive samples in the high ct-value range (Tables 5 and 6).
Table 5.

Intra-assay precision testing with a strongly positive sample (Cepheid E-Gene Ct-value 23.7; Cepheid N-Gene Ct-value 26.3), a weakly positive sample (Cepheid E-Gene Ct-value 31.0; Cepheid N-Gene Ct-value 34.4), and a negative sample

Running sample numberARIES assay
Overall ratingCt-value of the ORF-sequenceCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)
Strongly positive sample
#1positive25.527.50.0
#2positive25.527.80.0
#3positive25.527.10.0
Weakly positive sample
#1positive0.034.70.0
#2positive0.036.70.0
#3positive0.033.40.0
Negative sample
#1negative0.00.023.6
#2negative0.00.023.6
#3negative0.00.024.3

Ct = cycle threshold.

Table 6.

Inter-assay precision testing with a strongly positive sample (Cepheid E-Gene Ct-value 23.4; Cepheid N-Gene Ct-value 25.8), a weakly positive sample (Cepheid E-Gene Ct-value 35.3; Cepheid N-Gene Ct-value 38.4), and a negative sample

Running sample numberARIES assay
Overall ratingCt-value of the ORF-sequenceCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)
Strongly positive sample
#1positive23.325.40.0
#2positive23.626.00.0
#3positive22.925.90.0
Weakly positive sample
#1negative0.00.031.6
#2negative0.00.031.3
#3negative0.00.027.5
Negative sample
#1negative0.00.025.5
#2negative0.00.025.8
#3negative0.00.025.3

Ct = cycle threshold.

Limit of detection assessment with well characterized samples, DNA plasmids and positive control RNA

With plasmid DNA, the lod of the ARIES system with a positive result indicated by at least one positive PCR target as defined by the manufacturer was in the 103 copies/reaction-range. Within the 102 copies/reaction-range, target DNA in 50% of the repeated runs would have gone undetected. Focusing on the ORF-sequence target, 25% of the runs were positive in the 103 copies/reaction-range, 75% in the 104 copies/reaction-range, and 100% in the 105 copies/reaction-range. Focusing on the N-gene target, 50% of the runs were positive in the 102 copies/reaction range and 100% in the 103 copies/reaction range (Table 7).
Table 7.

Assessment of the limits-of-detection (lod) applying DNA plasmids as positive control in quadruplicate

Positive control DNA copy numberCt-values of the ARIES ORF-sequenceCt-values of the ARIES N-geneCt-values of the ARIES the internal control (RNAse-P-gene)
Run 1Run 2Run 3Run 4Run 1Run 2Run 3Run 4Run 1Run 2Run 3Run 4
1.4 × 10625.321.721.721.927.223.723.924.1negativenegativenegativenegative
1.4 × 10528.624.125.524.230.726.627.727.1negativenegativenegativenegative
1.4 × 10429.625.027.3negative31.530.130.432.5negativenegativenegativenegative
1.4 × 10332.3negativenegativenegative34.436.236.537.2negativenegativenegativenegative
1.4 × 102negativenegativenegativenegativenegativenegative38.338.839.125.3negativenegative
1.4 × 101negativenegativenegativenegativenegativenegativenegativenegativenegative25.726.426.8
1.4negativenegativenegativenegativenegativenegativenegativenegativenegative27.027.026.6
<1negativenegativenegativenegativenegativenegativenegativenegativenegative27.127.126.9
<1negativenegativenegativenegativenegativenegativenegativenegativenegative27.126.127.1

Ct = cycle threshold.

With free single-stranded RNA, the limit of detection of the ARIES system was in the 104 copies/reaction-range. Within the 103 copies/reaction-range, only 50% of the runs would have led to positive results. Focusing on the ORF-sequence target, 25% of the runs were positive in the 103 copies/reaction-range, and 100% in the 104 copies/reaction-range. The reason for the single failed amplification of the ORF-sequence in the 105 copies/reaction range remains unclear. Focusing on the N-gene target, 50% of the runs were positive in the 103 copies/reaction range and 100% in the 104 copies/reaction range (Table 8). Due to poor stability of free RNA molecules, their integration in the evaluation has been a particular challenge to the diagnostic assay assessed.
Table 8.

Assessment of the limits-of-detection (lod) applying free RNA as positive control in quadruplicate

Positive control RNA copy numberCt-values of the ARIES ORF-sequenceCt-values of the ARIES N-geneCt-values of the ARIES the internal control (RNAse-P-gene)
Run 1Run 2Run 3Run 4Run 1Run 2Run 3Run 4Run 1Run 2Run 3Run 4
3.8 × 10720.921.220.421.722.322.722.223.1negativenegativenegativenegative
3.8 × 10624.524.725.225.726.126.426.827.3negativenegativenegativenegative
3.8 × 10528.027.3negative27.829.729.130.829.5negativenegativenegativenegative
3.8 × 10431.230.728.932.632.732.730.534.2negativenegativenegativenegative
3.8 × 10334.1negativenegativenegative35.737.4negativenegativenegativenegative36.4negative
3.8 × 102negativenegativenegativenegativenegativenegativenegativenegative38.8negativenegativenegative
3.8 × 101negativenegativenegativenegativenegativenegativenegativenegativenegativenegativenegativenegative
3.8negativenegativenegativenegativenegativenegativenegativenegative36.5negativenegativenegative
<1negativenegativenegativenegativenegativenegativenegativenegativenegative37.0negativenegative

Ct = cycle threshold.

With quantified virus particles in patient samples, the lod of the ARIES assay was in the 103 copies/reaction-range. While all samples containing 103 and more virus copies were detected positive by the ARIES assay, virus RNA in the 102 copies/reaction-range and less went undetected.

DISCUSSION

In summary, the ARIES SARS-CoV-2 Assay did not show any non-specific reactions in our hands in the course of the performed validation procedure. Focusing on diagnostic sensitivity, however, target sequences in a considerable proportion of weakly positive samples which were still detected positive with the automated assay by Cepheid went undetected. With a limit of detection >103 virus copies, stages of the infection with lower viral copy number may be missed. Based on previous estimates [9, 20], respective high copy numbers are only detectable for few days of the infection. In contrast, highly sensitive PCR facilitates the detection of prolonged shedding of viral RNA sequences irrespective of their infectious potential [21-24]. Anyway, clinical or hygiene-related decisions based on positive results of the ARIES SARS-CoV-2 Assay should consider its limits of detection which are considerably higher than the ones of the automated Cepheid approach [11-16]. This is of particular importance in situations when maximum sensitivity is desired, both for clinical decision making or surveillance purposes. The assessment has a number of limitations. First of all, there was only a limited number of positive samples available for evaluation purposes, so samples had to be taken from different sources. Secondly, the definition of the Cepheid assay as the gold standard is a methodical limitation by itself, because this decision does not consider potential failure of the gold standard itself. For example, it cannot be excluded that there might have been individual SARS-CoV-2-containing samples being positively detected by the ARIES system but missed by the Cepheid approach in broader evaluations including higher sample numbers. However, in spite of such theoretical assumptions, the data provide a preliminary estimation of the diagnostic performance of the validated diagnostic test in line with validation standards as suggested previously [17, 18].

CONCLUSIONS

The automated cartridge-based ARIES SARS-CoV-2 Assay specifically detected SARS-CoV-2 sequences in our validation approach. Compared with the automated Cepheid approach, however, the assay was considerably less sensitive for cases with low target RNA loads, a fact that should be considered in case of decisions for its diagnostic use.
Table 2.

Sensitivity assessment with strongly positive samples defined by cycle threshold values <34 in the gold standard assay

Running sample numberPrevious diagnostic results from the same patientSARS CoV-2 target gene copy-number within the sample (if available)Cepheid assay (gold standard)ARIES assay (assessed test platform)
Overall ratingCt-value of the E-geneCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)Overall ratingCt-value of the ORF sequenceCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)
#1positiven.a.positive30.433.127.9positivenegative36.3negative
#2positiven.a.positive30.132.327.9positivenegative37.5negative
#3n.a.n.a.positive23.325.527.7positive24.025.8negative
#4n.a.n.a.positive26.328.927.7positive26.229.2negative
#5n.a.n.a.positive30.232.728.1positivenegative31.3negative
#6n.a.n.a.positive23.926.328.0positivenegative28.8negative
#7n.a.n.a.positive27.229.328.1positive28.430.4negative
#8n.a.n.a.positive30.733.128.3positivenegative34.6negative
#9positive= 5.8 × 104positive28.932.728.2positive28.336.9negative
#10positive=2.6 × 103positive27.729.328.4positivenegative33.3negative
#11positive= 1.0 × 104positive30.233.528.1positive29.437.2negative
#12positive= 1.9 × 103positive30.232.928.1positivenegative38.3negative
#13positive=9.2 × 104positive27.028.828.1positivenegative29.7negative
#14positive= 7.2 × 104positive29.031.427.6positive30.533.3negative
#15positiven.a.positive26.428.828.7positivenegative30.3negative

Ct = cycle threshold. n.a. = not applicable.

Table 3.

Sensitivity assessment with weakly positive samples defined by cycle threshold values >34 for at least one target gene in the gold standard assay

Running sample numberPrevious diagnostic results from the same patientSARS-CoV-2 target gene copy-number within the sample (if available)Cepheid assay (gold standard)ARIES assay (assessed test platform)
Overall ratingCt-value of the E-geneCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)Overall ratingCt-value of the ORF sequenceCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)
#1negativen.a.positive35.938.827.5negativenegativenegative28.4
#2positiven.a.positive33.235.828.1negativenegativenegative27.7
#3positiven.a.positive35.738.730.5negativenegativenegative23.5
#4positiven.a.positive35.637.227.7negativenegativenegative25.9
#5n.a.n.a.positive32.835.627.4negativenegativenegative24.2
#6positiven.a.positive33.636.427.7negativenegativenegative26.0
#7n.a.n.a.positive33.536.928.4positivenegative36.0negative
#8positiven.a.positive34.937.428.1negativenegativenegative26.5
#9positive=9.4 × 101positive37.939.828.3negativenegativenegative30.1
#10positive= 3.6 × 102positive34.436.228.1negativenegativenegative24.8
#11positive= 8.0 × 102positive37.040.528.0negativenegativenegative35.2
#12positive= 3.6 × 103positive31.334.228.1positive29.2negativenegative
#13positiven.a.positive36.036.628.0positivenegative37.4negative
#14positiven.a.positive34.836.127.9positivenegative37.7negative
#15positiven.a.positivenegative39.528.3negativenegativenegative27.6

Ct = cycle threshold, n.a. = not applicable.

Table 4.

Specificity assessment with samples negative in the Cepheid assay

Running sample numberPrevious diagnostic results from the same patientSARS-CoV-2 target gene copy-number within the sample (if available)Cepheid assay (gold standard)ARIES assay (assessed test platform)
Overall ratingCt-value of the E-geneCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)Overall ratingCt-value of the ORF sequenceCt-value of the N-geneCt-value of the internal control (RNAse-P-gene)
#1negativen.a.negativenegativenegative27.5negativenegativenegative27.3
#2negativen.a.negativenegativenegative27.3negativenegativenegative27.1
#3negativen.a.negativenegativenegative27.7negativenegativenegative25.6
#4negativen.a.negativenegativenegative27.8negativenegativenegative25.0
#5negativen.a.negativenegativenegative27.8negativenegativenegative25.2
#6negativen.a.negativenegativenegative28.9negativenegativenegative27.2
#7negativen.a.negativenegativenegative28.2negativenegativenegative28.4
#8positiven.a.negativenegativenegative28.3negativenegativenegative26.6
#9negativen.a.negativenegativenegative28.3negativenegativenegative28.6
#10negativen.a.negativenegativenegative27.3negativenegativenegative25.5
#11positiven.a.negativenegativenegative28.4negativenegativenegative29.1
#12negativen.a.negativenegativenegative27.6negativenegativenegative28.7
#13positiven.a.negativenegativenegative27.5negativenegativenegative25.1
#14positiven.a.negativenegativenegative28.1negativenegativenegative26.6
#15negativen.a.negativenegativenegative27.7negativenegativenegative27.6
#16negativen.a.negativenegativenegative28.1negativenegativenegative23.1
#17positiven.a.negativenegativenegative28.3negativenegativenegative24.8
#18positiven.a.negativenegativenegative28.2negativenegativenegative27.7
#19positiven.a.negativenegativenegative28.0negativenegativenegative25.9
#20positiven.a.negativenegativenegative28.2negativenegativenegative31.6
#21positiven.a.negativenegativenegative27.9negativenegativenegative32.4
#22negativen.a.negativenegativenegative27.5negativenegativenegative25.2
#23negativen.a.negativenegativenegative28.1negativenegativenegative29.1
#24negativen.a.negativenegativenegative28.3negativenegativenegative25.7
#25negativen.a.negativenegativenegative28.1negativenegativenegative27.8
#26negativen.a.negativenegativenegative28.1negativenegativenegative27.8
#27negativen.a.negativenegativenegative28.1negativenegativenegative24.6
#28positiven.a.negativenegativenegative30.5negativenegativenegative31.0
#29positiven.a.negativenegativenegative29.9negativenegativenegative26.1
#30negativen.a.negativenegativenegative25.0negativenegativenegative28.7

Ct = cycle threshold, n.a. = not applicable.

  7 in total

Review 1.  Analytic and Clinical Performance of Major Commercial Severe Acute Respiratory Syndrome Coronavirus 2 Molecular Assays in the United States.

Authors:  Michelle R Campbell; Matthew J Binnicker
Journal:  Clin Lab Med       Date:  2022-02-21       Impact factor: 2.172

Review 2.  FDA authorized molecular point-of-care SARS-CoV-2 tests: A critical review on principles, systems and clinical performances.

Authors:  Brigitte Bruijns; Laura Folkertsma; Roald Tiggelaar
Journal:  Biosens Bioelectron X       Date:  2022-05-21

3.  SARS-CoV-2 E Gene Variant Alters Analytical Sensitivity Characteristics of Viral Detection Using a Commercial Reverse Transcription-PCR Assay.

Authors:  Stephen Tahan; Bijal A Parikh; Lindsay Droit; Meghan A Wallace; Carey-Ann D Burnham; David Wang
Journal:  J Clin Microbiol       Date:  2021-06-18       Impact factor: 5.948

Review 4.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

5.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Sarah Berhane; Melissa Taylor; Ada Adriano; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Julie Domen; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2021-03-24

6.  Comparison of Five Serological Assays for the Detection of SARS-CoV-2 Antibodies.

Authors:  Anja Dörschug; Julian Schwanbeck; Andreas Hahn; Anke Hillebrecht; Sabine Blaschke; Kemal Mese; Uwe Groß; Sascha Dierks; Hagen Frickmann; Andreas E Zautner
Journal:  Diagnostics (Basel)       Date:  2021-01-06

7.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.