Literature DB >> 32794005

Effects of modular ion-funnel technology onto analysis of breath VOCs by means of real-time mass spectrometry.

Giovanni Pugliese1, Felix Piel2,3,4, Phillip Trefz1, Philipp Sulzer2, Jochen K Schubert1, Wolfram Miekisch5.   

Abstract

Proton transfer reaction time-of-flight mass spectrometry (pan class="Gene">PTR-ToF-MS) is a powerful tool for real-time monitoring of trace concentrations of volatile organic compounds (VOCs). The sensitivity of PTR-ToF-MS also depends on the ability to effectively focus and transmit ions from the relatively high-pressure drift tube (DT) to the low-pressure mass analyzer. In the present study, a modular ion-funnel (IF) is placed adjacent to the DT of a PTR-ToF-MS instrument to improve the ion-focusing. IF consists of a series of electrodes with gradually decreasing orifice diameters. Radio frequency (RF) voltage and direct current (DC) electric field are then applied to the electrodes to get the ions focused. We investigated the effect of the RF voltage and DC field on the sensitivity of a pattern of VOCs including hydrocarbons, alcohols, aldehydes, ketones, and aromatic compounds. In a proof-of-concept study, the instrument operating both as normal DT (DC-mode) and at optimal IF conditions (RF-mode) was applied for the breath analysis of 21 healthy human subjects. For the range of investigated VOCs, an improvement of one order of magnitude in sensitivity was observed in RF-mode compared with DC-mode. Limits of detection could be improved by a factor of 2-4 in RF-mode compared with DC-mode. Operating the instrument in RF-mode allowed the detection of more compounds in the exhaled air compared with DC-mode. Incorporation of the IF considerably improved the performance of PTR-ToF-MS allowing the real-time monitoring of a larger number of potential breath biomarkers. Graphical abstract.

Entities:  

Keywords:  Breath analysis; Ion-funnel; PTR-ToF-MS; Real-time mass spectrometry; VOCs

Mesh:

Substances:

Year:  2020        PMID: 32794005      PMCID: PMC7497501          DOI: 10.1007/s00216-020-02846-8

Source DB:  PubMed          Journal:  Anal Bioanal Chem        ISSN: 1618-2642            Impact factor:   4.142


Introduction

Proton transfer reaction mass spectrometry (pan class="Gene">PTR-MS) is an analytical technique that allows real-time monitoring of volatile organic pan class="Chemical">compounds (VOCs) at low concentrations. It is widely used, e.g. in environmental sciences, food chemistry, homeland security, and breath analysis [1]. Since its introduction in the 1990s [2], pan class="Gene">PTR-MS has been improved in many ways. Inclusion of time-of-flight (ToF) mass analyzers has substantially overcome the limitations of the first generation of PTR quadrupole-MS (QMS) such as limited mass range and low mass resolution [3-5]. Modifications of the hollow cathode discharge ion source allowed to successfully use different chemical ionization agents such as H3O+, NO+, O2+, Kr+, Xe+, and NH4+, improving versatility and selectivity of the instrument [6, 7]. Sensitivity of pan class="Gene">PTR-MS is not solely determined by mass analyzers and detectors but it also depends on the ability to effectively focus and transmit ions from the relatively high-pressure drift tube (DT) to the low-pressure mass analyzer. As most of the ions crossing the DT do not exit through the small orifice at the MS interface, a large quantity of ion signal is lost. Ion-funnels (IF) represent a kind of ion guide that enhances sampling of ions through an orifice [8]. In an IF, a radio frequency (RF) voltage and a direct current (DC) electric field are applied to a series of electrodes with decreasing aperture sizes. The electrodes provide strong repulsive potentials at the edge of the electrode, radially focusing the ions. The first demonstration of an IF in PTR-MS was shown by Barber et al. [9]. In their instrument, the whole DT was set up as an IF with the first half used as a standard DT reactor running at a lower reduced electric field compared with the traditional DT. The RF electric field was only applied to the second section with decreasing orifice sizes. González-Méndez et al. [10] used the IF to manipulate the ion-molecule reactions and enhance the selectivity of PTR-MS. Brown et al. [11] reported that in this instrument, ion-focusing and proton transfer reaction both occurred in the IF region. This resulted in vastly different sensitivities for different compounds and in unusual fragmentation patterns. Recently, IONICON Analytik implemented a modular IF into proton transfer reaction time-of-flight mass spectrometry (PTR-ToF-MS) instruments. The aim of the present study was to characterize and optimize the IF-PTR-ToF instrument for trace VOC analysis. In a proof-of-concept setup, the instrument was then applied for real-time breath analysis in human subjects. The following questions were addressed in detail: How does modification of IF parameters affect primary and VOC product ions? Are pan class="Gene">PTR sensitivity and detection limits in VOC analysis significantly improved by the IF? Are benefits of the IF technique suited to support applications such as real-time breath gas analysis in pan class="Species">humans?

Methods

Ion-funnel PTR-ToF-MS instrument

All investigations were carried out using an online pan class="Gene">PTR-ToF-MS instrument equipped with a modular IF (PTR-TOF 1000 ultra, IONICON Analytik GmbH, Innsbruck, Austria; first-generation model (2017)). Figure 1 shows a schematic view of the instrument. The general working principle of PTR-ToF-MS has been described in several studies [4, 5]. Concisely, hydronium ions (H3O+) are produced in a hollow cathode glow discharge ion source from electron ionization of water vapour and are drawn by an electric field into the ion-molecule reaction region (DT). Here, the analyte sample is injected and the proton transfer reaction between the formed H3O+ and neutral analyte molecules (M) occurs: M + H3O+ → MH+ + H2O. Only molecules with proton affinities higher than water (PA (H2O) = 691 kJ mol−1) are ionized, a criterion that excludes the major constituents of air such as N2, O2, and CO2 but includes many trace gases such as most VOCs. The DT of the PTR-TOF 1000 ultra consists of a 7 cm long tube made of electrically isolated stainless steel rings. The rings are connected with resistors, and a drift voltage (Udrift) can be applied over the entire set of rings to induce an electric field (Edrift) in the DT. The modular IF is 2.2 cm long, it is placed adjacent to the DT and consists of 12 electrodes (6 with RF+ and 6 with RF−) with gradually decreasing orifice diameters, from 1 to 0.2 cm, placed at 0.1 cm distance of each other. In order to avoid the trapping of ions in axial potential wells, particularly those with low m/z, the IF electrode geometries used in this work fulfilled the following conditions:where ρ is the electrode orifice radius and δ = d/π where d is the electrode spacing [12]. A radio frequency (RF) voltage and a direct current (DC) electric field are then applied to the electrodes. The DC electric field drives the ions axially through the IF toward the exit aperture. An additional alternating current (AC) is superimposed on the electrodes, with the RF on neighbouring electrodes being phase-shifted by 180°. In this way, the RF field creates a strongly repulsive potential near the surface of each electrode. In combination with the progressively decreasing aperture size, this serves to focus the ions radially. Table S1 (see Electronic Supplementary Material, ESM) summarizes details and operating conditions of the modular IF. The protonated VOCs then enter the pulse extraction region of the orthogonal acceleration reflectron ToF analyzer via a transfer lens system. The DT is interfaced to the transfer lens region via a pinhole of ~ 0.1 cm I.D. with the cone toward the transfer lens. The operating pressure in the DT (buffer gas number density, N) and the Edrift strength are important parameters, more commonly combined and expressed in terms of the reduced electric field (E/N). Edrift accelerates the ions but at the same time collisions with the buffer gas tend to slow them down. The E/N affects the reagent ion distribution. Increasing the E/N ratio results in more energetic collisions, which reduces the proportion of the water cluster ions (H3O+(H2O)) in the DT but at the same time can increase the fragmentation of analytes [13]. Typical E/N values are in the range 90–150 Td, where 1 Td = 10−17 V cm2.
Fig. 1

Schematic view of the PTR-ToF 1000 ultra setup: (a) hollow cathode ion source, (b) drift tube, (c) ion-funnel, (d) ToF mass analyzer

Schematic view of the pan class="Gene">PTR-ToF 1000 ultra setup: (a) hollow cathode ion source, (b) drift tube, (c) ion-funnel, (d) ToF mass analyzer Within the standard pan class="Gene">PTR DT, Edrift variation is about 15% due to the ratio of the inner diameter and axial distance between the drift rings [14]. The additional RF voltage considerably increases this variation and E/N cannot be properly calculated any more. However, it is possible to define certain voltage settings in RF-mode which enable branching ratios of distinct analytes to be obtained that are comparable with the ones obtained with a classic PTR-ToF-MS instrument operated at a certain actual E/N (“TRU-E/N” method). Nevertheless, in the present paper, the E/N definition is omitted.

Ion-funnel characterization

Gas standard generation

The IF characterization was performed using a multicomponent gas VOC mix (IONICON Analytik GmbH, Innsbruck, Austria) including n class="Chemical">methanol, n>n class="Chemical">acetonitrile, acetaldehyde, ethanol, 2-propenal, acetone, isoprene, 2-butenal, 2-butanone, benzene, toluene, o-xylene, chlorobenzene, α-pinene, and 1,2-dichlorobenzene at concentrations of ~ 1ppmV. The VOC mix was subjected to a 100-fold dynamic dilution in pure nitrogen (purity 5.0, Linde, Vienna, Austria) by means of a liquid calibration unit (LCU, IONICON Analytik, Innsbruck, Austria) to generate a standard mixture with approximately 10 ppbV of each component. The working principle of the LCU involves the introduction of a liquid standard solution into a carrier gas stream, by forcing it through a nebuliser (X175, Burgener Research Inc., UK) and spraying the solution into an evaporation chamber at a defined temperature. This results in a rapid evaporation. The generated gaseous standard mixture can then be measured or collected directly at the output of the evaporation chamber. Two liquid ports (1–50 μl min−1), one carrier gas port (1–1000 ml min−1), and two additional gas ports (1–100 ml min−1), controlled by mass flow controllers (Bronkhorst High-Tech B.V., Ruurlo, Netherlands), enable the generation of complex standard mixtures. VOC standards can be prepared from either liquid solutions or gases, or even from both at the same time. In addition, defined amounts of humidity can be added by adding pure water via one of the two liquid ports [15]. In this study, the pan class="Chemical">LCU flow was kept constant at 1000 ml min−1 for all experiments, the LCU temperature was 75 °C and the humidity was adjusted by adding pure water (HPLC grade).

Experimental design

The standard mixture was introduced into the pan class="Chemical">DT via a 1.5 m long polyether ether ketone (PEEK) transfer line (ID: 0.75 mm, Restek, Bellafonte, PA) that was directly connected to the outlet of the LCU. The transfer line temperature was 75 °C and the sampling flow was 100 ml min−1. The signal intensity was recorded for each m/z while the settings of the IF region were varied. Operating the instrument in RF-mode (RF on), the DC electric field applied to the IF was varied in the range of 4.5–27 V/cm while the RF voltage was varied in the range of 40–200 V peak-to-peak (Vp-p) at 4.5 MHz. These testing ranges were decided upon after preliminary measurements showed that these settings approximate the best operating conditions. The entire experimental design was repeated at two different Edrift strength, 66 V/cm and 48 V/cm, and using both dry and humid samples (absolute humidity 47 g m−3) (ESM Table S2). These two sampling conditions will be referred in the text as “dry” and “humid” conditions, respectively. When the instrument was operated only with the DC field applied to the IF region (DC-mode), the RF voltage was set to zero and the DC field was set at the same Edrift value in order to have a homogeneous electric field along the pan class="Chemical">DT/IF regions. For the whole experimental design, the pan class="Chemical">DT/IF pressure was 2.3 mbar, the pan class="Chemical">DT temperature was 75 °C, and the integration time was 1 s. Three replicates were measured for each experimental setup, then the results were averaged and background signals were subtracted.

Human breath samples

All experiments were performed in accordance with the guidelines laid down in the Declaration of Helsinki and approved by the ethics committee at the University Medical Center Rostock. Informed consent was obtained from 21 healthy n class="Species">human subjects (aged between 20 and 45 years). Demographic parameters such as height, body weight, age, sex, and smoking habits were recorded for each volunteer (ESM Table S3). Volunteers were asked to breathe spontaneously and continuously over 3 min through a sterile mouth piece directly connected to the n>n class="Gene">PTR transfer line in side stream mode by means of a T-piece. During the first minute of measurement, the PTR-ToF-MS instrument was operated in RF mode: Edrift was 66 V/cm, RF voltage was 120 Vp-p, and DC field was 13.5 V/cm. During the second minute of measurement, the operating mode of the instrument was switched from RF-mode to DC-mode. During the third minute of measurement, the instrument was operated in DC-mode: RF voltage was 0 Vp-p and both Edrift and DC field were 66 V/cm. For breath measurements, the pan class="Gene">PTR transfer line temperature was 75 °C, DT temperature was 75 °C, and DT pressure was 2.3 mbar. For breath measurements, the integration time was 200 ms.

Data processing

The ion yields of all m/z were measured in counts per second (cps) and pan class="Chemical">compounds were identified by means of their protonated m/z and isotopic patterns. The normalization of the measured ion intensities to the H3O+ counts in combination with the water-cluster ion counts is standard practice in PTR-ToF-MS data treatment [16]. However, in the present paper, the normalization to reagent ions was omitted in order to reflect the actual sensitivity of the instrument, which would be masked by normalization. Both breath and standard files were processed using the software pan class="Gene">PTR-MS viewer v. 3.2.8 (IONICON Analytik GmbH, Innsbruck, Austria). For breath measurements, expiratory and inspiratory phases were recognized by means of a custom-made data processing algorithm called “breath tracker” (MATLAB version 7.12.0.635, R2011a). The function of the algorithm has been described previously [17]. Briefly, an endogenous compound that has a sufficiently abundant signal intensity in expiration is used as a tracker to differentiate between alveolar and inspired phases. n class="Chemical">Acetone, n>n class="Chemical">isoprene, or carbon dioxide is usually used for this purpose. Expiratory and inspiratory phases determined by means of the algorithm were then applied to all m/z of interest.

Results

H3O+·(H2O) (n = 0, 1, and 2) reagent ions

Figure 2 (a, b) shows the variation of pan class="Gene">H3O+, protonated water clusters, and O2+ and NO+ measured intensities as function of RF voltage. The signal intensities of H3O+, H2O·H3O+, and O2+ are too high to be measured directly because of detector saturation. Therefore, the signal intensities at m/z = 21 corresponding to H318O+, at m/z = 39 corresponding to H2O·H318O+ and at m/z = 34 corresponding to 18O16O+ were recorded and corrected by the natural isotope abundances.
Fig. 2

Ion intensities in counts per second (cps) of the water reagent ions (H3O+·(H2O), n = 0, 1, and 2) and parasitic ions O2+ and NO+ present in the DT under dry (a) and humid (b) conditions as a function of RF voltage. DC was 13.5 V/cm and Edrift was 66 V/cm

Ion intensities in counts per second (cps) of the pan class="Chemical">water reagent ions (H3O+·(H2O), n = 0, 1, and 2) and parasitic ions O2+ and NO+ present in the DT under dry (a) and humid (b) conditions as a function of RF voltage. DC was 13.5 V/cm and Edrift was 66 V/cm pan class="Gene">H3O+ intensity showed its maximum value at RF = 80 Vp-p under dry conditions (Fig. 2 (a)) and at RF = 120 V under humid conditions (Fig. 2 (b)). As the RF voltage decreases, the H3O+ signal intensity decreases. At the same time, an increase of the water cluster intensities was observed with decreasing RF voltage. H3O+·H2O showed its maximum value at RF = 40 Vp-p under dry conditions and at RF = 80 Vp-p under humid conditions. Nevertheless, at RF = 40 Vp-p under humid conditions, H3O+·H2O showed a higher intensity than H3O+. The protonated pan class="Chemical">water trimer was only observed under humid conditions and at RF = 40 Vp-p. Intensities of parasitic ions O2+ and NO+ only increased only at high RF voltages (> 160 Vp-p) under both dry and humid conditions. H3O+·(H2O)2 and O2+ intensities were up to six orders of magnitude smaller compared with that of the protonated water dimer. Figure 3 (a, b) shows the variation of the pan class="Gene">H3O+ and H3O+·H2O intensities as function of DC field. Due to the large difference between the intensities of the two reagent ions, H3O+·H2O intensity is displayed on a second Y-axis. The H3O+ intensity increased with increasing DC field with its maximum value at DC = 27 V/cm under both dry (Fig. 3 (a)) and humid (Fig. 3 (b)) conditions. In contrast, H3O+·H2O showed its maximum value at DC = 22.5 V/cm under both dry and humid conditions.
Fig. 3

Ion intensities in counts per second (cps) of H3O+ and H3O+·(H2O) present in the DT under dry (a) and humid (b) conditions as a function of the DC field (V/cm). RF voltage was 120 Vp-p and Edrift was 66 V/cm

Ion intensities in counts per second (cps) of pan class="Gene">H3O+ and H3O+·(H2O) present in the DT under dry (a) and humid (b) conditions as a function of the DC field (V/cm). RF voltage was 120 Vp-p and Edrift was 66 V/cm ESM Fig. S1 (a, b) shows the variation of pan class="Gene">H3O+, protonated water clusters, and O2+ and NO+ measured intensities as function of RF voltage at Edrift = 48 V/cm. ESM Fig. S2 (a, b) shows the variation of the H3O+ and H3O+·H2O intensities as function of DC voltage at Edrift = 48 V/cm. H3O+ intensity showed similar trends of those showed at Edrift = 66 V/cm. In contrast, substantial differences were found for the protonated water clusters. Under dry conditions (ESM Fig. S1 (a)) at RF = 40 Vp-p and under humid conditions (ESM Fig. S1 (b)) at RF = 80 Vp-p, H3O+·H2O became the most abundant reagent ion in the DT. Under humid conditions at RF = 40 Vp-p, H3O+·(H2O)2 showed a higher intensity than H3O+. In DC-mode at Edrift = 48 V/cm under humid conditions, pan class="Gene">H3O+·H2O represent about 65% of the total water reagent ions. In contrast, under dry conditions, they represent about the 15% of the total water reagent ions. In DC-mode at Edrift = 66 V/cm, protonated water clusters are present in low concentrations under both dry and humid conditions. Under humid conditions, H3O+·H2O represent about 8% of the total water reagent ions; under dry conditions, they represent about 1% of the total water reagent ions.

Effect of RF voltage and DC field on VOC signal intensities

Figure 4 shows effects of RF voltage (40–200 pan class="Chemical">Vp-p) and DC field (4.5–27 V/cm) onto intensities of all investigated VOCs.
Fig. 4

Effect of RF voltage (40–200 Vp-p) and DC field (4.5–27 V/cm) on VOC intensities. The whole experiment was conducted at Edrift of 48 V/cm and 66 V/cm, with dry and humid samples (humidity of 47 g m−3). Data were normalized to maximum values to emphasize relative changes

Effect of RF voltage (40–200 pan class="Chemical">Vp-p) and DC field (4.5–27 V/cm) on VOC intensities. The whole experiment was conducted at Edrift of 48 V/cm and 66 V/cm, with dry and humid samples (humidity of 47 g m−3). Data were normalized to maximum values to emphasize relative changes Figure 5 shows signal intensities of pan class="Chemical">acetaldehyde, acetone, benzene, and dichlorobenzene as a function of RF voltage.
Fig. 5

Intensities of acetaldehyde (blue triangle), acetone (grey square), benzene (yellow cross), and dichlorobenzene (red round) as function of RF voltage. DC field was 13.5 V/cm and Edrift was 66 V/cm. Data were normalized onto respective maximum values to emphasize the relative changes

Intensities of pan class="Chemical">acetaldehyde (blue triangle), acetone (grey square), benzene (yellow cross), and dichlorobenzene (red round) as function of RF voltage. DC field was 13.5 V/cm and Edrift was 66 V/cm. Data were normalized onto respective maximum values to emphasize the relative changes Intensities of pan class="Chemical">acetaldehyde, methanol, ethanol, 2-propenal, and isoprene showed their maximum values at RF = 80 Vp-p. In contrast, intensities of acetone, acetonitrile, 2-butenal, and butanone showed their maximum values at RF = 120 Vp-p. Aromatic compounds, such as benzene, toluene, o-xylene, chlorobenzene, and α-pinene, showed their maximum intensities at RF = 160 Vp-p. Intensity of dichlorobenzene showed its maximum at RF = 200 Vp-p. Figure 6 shows the measured intensities of the pan class="Chemical">acetaldehyde, acetone, benzene, and dichlorobenzene as function of DC field. Intensities of most of the investigated compounds showed a substantial increase when the DC field was increased from 4.5 to 13.5 V/cm; when the DC voltage was further increased up to 27 V/cm, they showed variations < 10%. This was with the exception of dichlorobenzene which showed steadily increasing intensity with increasing DC voltage, with its maximum at DC = 27 V/cm.
Fig. 6

Intensities of acetaldehyde (blue triangle), acetone (grey square), benzene (yellow cross), and dichlorobenzene (red round) as function of DC voltage. RF voltage was 120 Vp-p and Edrift was 66 V/cm. Data were normalized onto respective maximum values to emphasize the relative changes

Intensities of pan class="Chemical">acetaldehyde (blue triangle), acetone (grey square), benzene (yellow cross), and dichlorobenzene (red round) as function of DC voltage. RF voltage was 120 Vp-p and Edrift was 66 V/cm. Data were normalized onto respective maximum values to emphasize the relative changes Table 1 compares sensitivities and limits of detection (LODs) capan class="Chemical">lculated for DC-mode and RF-mode for all the investigated VOCs. Sensitivity is expressed as ion count rate per second per part-per-billion volume mixing ratio of supplied analyte (cps/ppbV). LODs were calculated for 1 s of integration time using the “3σ method” with σ being the standard deviation of the background noise level [18]. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm under humid conditions. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm under humid conditions. At these conditions, switching from DC-mode to RF-mode led to an improvement in sensitivity of about 1 order of magnitude for most of the investigated compounds with the exception of methanol, ethanol, and dichlorobenzene. In contrast, only an improvement by a factor of 2–4 was observed for the LODs in RF-mode compared with DC-mode.
Table 1

Comparison of sensitivities and LODs for DC-mode and RF-mode of all investigated VOCs. LODs were calculated for 1 s of integration time. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm under humid conditions. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm under humid conditions

Sensitivity (cps ppbV−1)LOD (ppbV)
Compound (m/z)DC-modeRF-modeRF-mode/DC-modeDC-modeRF-modeRF-mode/DC-mode
Methanol (33)27.5116.54.20.7350.4321.7
Acetonitrile (42)46.7418.28.60.3370.113.1
Acetaldehyde (45)56.5458.18.10.660.2452.7
Ethanol (47)3.418.25.312.3198.311.5
2-Propenal (57)43406.39.40.3010.1462.1
Acetone (59)103.61008.39.70.260.092.9
Isoprene (69)10120.812.11.2260.4632.6
2-Butenal (71)65.870110.70.1630.0692.4
2-Butanone (73)66.1750.911.40.3250.1043.1
Benzene (79)39.2423.610.80.220.0563.9
Toluene (93)50.2620.312.40.1440.062.4
o-Xylene (107)65.4747.411.40.1150.033.8
Chlorobenzene (113)38.2413.710.80.1550.0612.5
α-Pinene (137)25.7252.19.80.2650.0873
1,2-Dichlorobenzene (147)43.9264.76.00.0870.0442
Comparison of sensitivities and LODs for DC-mode and RF-mode of all investigated VOCs. LODs were capan class="Chemical">lculated for 1 s of integration time. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm under humid conditions. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm under humid conditions

Application in human breath samples

In a proof-of-concept study, the instrument operating both in DC-mode and RF-mode was applied for breath analysis of 21 pan class="Species">human healthy subjects. Table 2 contains the list of VOCs that could be detected in exhaled breath. pan class="Chemical">Compounds that could only be detected in RF-mode are labelled using bold italic text. Concentrations of these pan class="Chemical">compounds were below the LODs in DC-mode and above the LODs in RF-mode.
Table 2

List of VOCs that could be detected from exhaled breath in real-time. Compounds that could only be detected in RF-mode compared with DC-mode are labelled using bold italic text. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm

Peak numberMeasured mass (m/z)Exact mass (m/z)Mass accuracy (ppm)Empirical formulaConcentration range (ppbV)LOD DC-mode (ppbV)LOD RF-mode (ppbV)LOQ DC-mode (ppbV)LOQ RF-mode (ppbV)Potential compound
133.03133.033− 60.55CH4O+37.401–364.0330.7350.4322.4261.426Methanol
242.03142.034− 71.37C2H4N+2.443–76.180.3370.111.1120.363Acetonitrile
347.04647.049− 63.76C2H6O+12.332–74.09412.3198.3140.65327.423Ethanol
459.05359.04967.74C3H7O+133.39–1043.8640.260.090.8580.297Acetone
561.03161.02849.16C2H5O2+4.328–40.2980.8650.7632.8552.518Acetic acid
663.02863.02631.73C2H7S+1.435–10.8940.2030.120.670.396Dimethylsulfide
769.07269.07028.96C5H9+31.166–193.7661.2260.4634.0461.528Isoprene
871.05271.04942.22C4H7O+0.25–6.9570.1630.0690.5380.2282-Butenal
973.06973.06554.75C4H9O+1.26–3.7690.3250.1041.0730.343Butanone
1079.05279.054− 25.30C6H7+0.309–8.3620.220.0560.7260.185Benzene
1187.07387.080− 80.39C5H11O+0.62–1.6430.3490.2721.1520.898Pentanal
1289.05289.060− 89.83C4H9O2+0.469–5.1460.2030.1260.670.416Ethylacetate
1393.07393.07032.23C7H9+0.413–7.610.1440.060.4750.198Toluene
14137.129137.132− 21.88C10H17+0.911–12.4130.2650.0870.8750.287Limonene
1549.00849.010− 40.81CH5S+0.194–2.4520.3540.0731.1680.241Methyl mercaptan
1668.05368.04958.78C4H6N+0.067–0.5580.1410.0320.4650.106Pyrrole
1780.04780.049− 24.98C5H6N+0.068–0.7130.2150.0390.710.129Pyridine
1885.06385.065− 23.51C5H9O+0.331–1.2180.3230.2661.0660.8783-Penten-2-one
1991.05591.057− 21.96C4H11S+0.168–1.0910.2630.0920.8680.304Methyl propyl sulfide
2097.06197.065− 41.21C6H9O+0.248–3.970.3340.131.1020.4292,5-Dimethylfuran
21101.055101.060− 49.48C5H8O2+0.218–0.7370.2610.1950.8610.644Coffee furanone
22105.056105.05419.04C4H9O3+0.142–0.7850.2420.120.7990.396β-Hydroxybutyric acid
23106.069106.06537.71C7H8N+0.017–0.0940.0880.050.290.165Vinylpyridine
24108.074108.081− 64.77C7H10N+0.063–0.2420.2020.0770.6670.254o-Toluidine
25118.065118.058 59.29C8H8N+0.051–0.5690.0840.0140.2770.046Indole
26125.088125.096− 63.95C8H13O+0.042–0.3430.1680.0890.5540.294Acetylcyclohexene
27129.077129.07054.23C10H9+0.07–0.1530.1610.0590.5310.195Naphthalene
28133.058133.065− 52.61C9H9O+0.042–0.1420.120.040.3960.132Cinnamaldehyde
29135.088135.08059.22C9H11O+0.074–0.3170.1670.0750.5510.248Cinnamyl alcohol
30149.055149.060− 33.54C9H9O2+0.095–0.5610.2230.1260.7360.416Cinnamic acid
31151.110151.112− 13.24C10H15O+0.054–0.2330.1180.0490.3890.162Carvone
List of VOCs that could be detected from exhaled breath in real-time. pan class="Chemical">Compounds that could only be detected in RF-mode compan>red with DC-mode are labelled using bold italic text. The DC-mode data were collected at Edrift = 66 V/cm, RF = 0 Vp-p, and DC = 66 V/cm. The RF-mode data were collected at Edrift = 66 V/cm, RF = 120 Vp-p, and DC = 13.5 V/cm Concentrations and LODs and pan class="Gene">LOQs were capan class="Chemical">lculated applying the kinetic theory [19, 20].

Discussion

Incorporation of a modular IF adjacent to the pan class="Chemical">DT led to a substantial improvement in sensitivity and LODs of the PTR-ToF-MS instrument. Improved sensitivities allowed the detection of a broader range of VOCs from human breath samples in real-time. Intensities determined for pan class="Chemical">water reagent ions (H3O+·(H2O), n = 0, 1, and 2) and for protonated VOCs showed a considerable dependence on RF voltage and DC field applied to the IF region. Highest intensities for H3O+ were observed in the RF range 80–120 Vp-p and at DC = 27 V/cm. At lower and higher RF voltages, the focusing effect of the funnel was lost and ion transfer was less efficient. High RF voltages increase the kinetic energy of molecules. As binding forces in the water clusters are weak when compared with normal chemical bonding, this will lead to collisional decomposition of water clusters long before fragmentation of chemical compounds occurs. Higher H3O+ intensities in humid samples were most probably due to back diffusion of sample gas from the DT into the ion source generating additional H3O+ [15, 21, 22]. O2+ and NO+ were present in low intensities and were observed in RF-mode only at high RF voltage (> 160 Vp-p) as a result of improved ion transmission [13, 23]. VOCs showed maximum intensities at substance-specific DC field and RF voltage. In agreement with IF theory, cut-off values occurred at low (< 50 pan class="Chemical">Vp-p) RF voltages for all VOCs and at high (> 160 Vp-p) RF voltages for low-mass compounds (m/z < 90). At low RF voltages, the focusing effect of IF is lost for low and high masses. In a substance-specific way, higher masses show maximum transmission at high (> 120 Vp-p) RF voltages due to the dependency of effective potential onto m/z. Decreasing transmission of low masses at high RF voltages is attributed to diffusional loss of molecules due to the relatively high kinetic energy of low-mass molecules under these conditions [12, 24]. In addition, fragmentation may contribute to this effect, as we observed a 10% increase in acetaldehyde fragmentation with increasing RF voltage. Up to 50% fragmentation was reported by Barber et al. under similar conditions. In contrast to pan class="Chemical">oxygen-containing aliphatic substances, aromatic compounds showed increasing ion yields of the protonated monomers even at high RF voltages where non-aromatic substances already exhibited decreasing intensities. Enhanced ability of aromatic systems to stabilize ionic states may explain efficient generation of molecule ions without losses through fragmentation even at high RF voltages. This hypothesis is further confirmed by dichlorobenzene showing an almost linear increase with increasing RF voltages, most probably due to the additional charge-stabilizing effects of the chlorine atoms. In contrast to previous setups, with the IF used in this study, sensitivity increases were rather uniform, i.e. approximately one order of magnitude for all investigated pan class="Chemical">compounds. This is a strong indicator that the IF does not have major effects onto the ion chemistry within the DT itself. Therefore, the advantages of PTR-MS, e.g. quantification without calibration, are preserved. Although Brown and Barber et al. reported relative increases in sensitivity of 1–2 orders of magnitude for single compounds (acetaldehyde 45×, acetone 200×), absolute sensitivities for a broad range of compounds reported in our study were in general higher, e.g. 10 times higher for methanol and 2 times higher for acetaldehyde. Higher improvements in relative sensitivity as well as higher fragmentation rates reported by Barber and Brown et al. can thus be attributed to different geometries of IF and DT in their instruments. Characterization and optimization of pan class="Chemical">DT conditions, RF voltage, and DC field and effects of humidity are of general importance for any IF-PTR-ToF instrument and can therefore be beneficial for the whole community [20]. In addition, the modular IF described in this study can be implemented into several PTR-ToF-MS instruments. The impact of the IF onto quantification can be seen when LODs and pan class="Gene">LOQs are looked upon. As the applied IF will improve transmission of the ions, in parallel to the desired effects, increased ion yields will also induce growing background noise. Thus, the “raw” gain in ion counts will not directly translate into identical improvements of LODs and LOQs. LODs and LOQs substantially depend on noise inherent in a PTR-MS signal. This noise can be described by a Poisson distribution: the 1σ error in a measurement that is derived from counting a total of N ions is , with τ being the integration time [19, 25, 26]. Taking this into account, LODs and LOQs could effectively be improved by a factor of 2–4 when the instrument was switched from DC-mode to RF-mode. Hence, just determining increases in ion yields may lead to overestimation of the instrument performances for quantitative analysis. For real-life applications, e.g. trace gas analysis in breath, LODs and LOQs have to be determined to take into account all effects of DC field and RF voltage applied within the IF. Especially in diseased states, pan class="Disease">breath VOC concentrations may change quickly and abruptly. Hence, only real-time monitoring can provide complete and comprehensive information from breath VOC analysis [27-29]. PTR-ToF-MS with integration time of ≥ 200 ms enables breath-resolved continuous monitoring of breath volatiles. In this pilot study, the range of detectable volatile substances was significantly enlarged through application of IF technology.

Conclusion

The spectrum of detectable VOCs in real-time breath analysis was considerably enhanced through the application of IF technology. The IF can be tuned in order either to obtain the best operating conditions for a specific compound of interest or to realize operating conditions which represent the best compromise for the acquisition of a large number of pan class="Chemical">compounds. In contrast to previous setups, the IF used in this study did not have major effects onto ion chemistry within the pan class="Chemical">DT itself and therefore offers optimal conditions for VOC screening in biomedical applications. (PDF 926 KB)
  19 in total

1.  High resolution PTR-TOF: quantification and formula confirmation of VOC in real time.

Authors:  Martin Graus; Markus Müller; Armin Hansel
Journal:  J Am Soc Mass Spectrom       Date:  2010-02-10       Impact factor: 3.109

2.  Increased sensitivity in proton transfer reaction mass spectrometry by incorporation of a radio frequency ion funnel.

Authors:  Shane Barber; Robert S Blake; Iain R White; Paul S Monks; Fraser Reich; Steve Mullock; Andrew M Ellis
Journal:  Anal Chem       Date:  2012-05-30       Impact factor: 6.986

3.  Enhancement of Compound Selectivity Using a Radio Frequency Ion-Funnel Proton Transfer Reaction Mass Spectrometer: Improved Specificity for Explosive Compounds.

Authors:  Ramón González-Méndez; Peter Watts; David Olivenza-León; D Fraser Reich; Stephen J Mullock; Clive A Corlett; Stuart Cairns; Peter Hickey; Matthew Brookes; Chris A Mayhew
Journal:  Anal Chem       Date:  2016-10-21       Impact factor: 6.986

4.  Proton-Transfer-Reaction Mass Spectrometry: Applications in Atmospheric Sciences.

Authors:  Bin Yuan; Abigail R Koss; Carsten Warneke; Matthew Coggon; Kanako Sekimoto; Joost A de Gouw
Journal:  Chem Rev       Date:  2017-10-04       Impact factor: 60.622

5.  Performance characteristics and applications of a proton transfer reaction-mass spectrometer for measuring volatile organic compounds in ambient air.

Authors:  S Hayward; C N Hewitt; J H Sartin; S M Owen
Journal:  Environ Sci Technol       Date:  2002-04-01       Impact factor: 9.028

6.  Continuous real time breath gas monitoring in the clinical environment by proton-transfer-reaction-time-of-flight-mass spectrometry.

Authors:  Phillip Trefz; Markus Schmidt; Peter Oertel; Juliane Obermeier; Beate Brock; Svend Kamysek; Jürgen Dunkl; Ralf Zimmermann; Jochen K Schubert; Wolfram Miekisch
Journal:  Anal Chem       Date:  2013-10-07       Impact factor: 6.986

7.  Validation of atmospheric VOC measurements by proton-transfer-reaction mass spectrometry using a gas-chromatographic preseparation method.

Authors:  Carsten Warneke; Joost A De Gouw; William C Kuster; Paul D Goldan; Ray Fall
Journal:  Environ Sci Technol       Date:  2003-06-01       Impact factor: 9.028

8.  The ion funnel: theory, implementations, and applications.

Authors:  Ryan T Kelly; Aleksey V Tolmachev; Jason S Page; Keqi Tang; Richard D Smith
Journal:  Mass Spectrom Rev       Date:  2010 Mar-Apr       Impact factor: 10.946

9.  Effects of humidity, CO2 and O2 on real-time quantitation of breath biomarkers by means of PTR-ToF-MS.

Authors:  Phillip Trefz; Jochen K Schubert; Wolfram Miekisch
Journal:  J Breath Res       Date:  2018-03-01       Impact factor: 3.262

10.  Extending PTR based breath analysis to real-time monitoring of reactive volatile organic compounds.

Authors:  Giovanni Pugliese; Phillip Trefz; Beate Brock; Jochen K Schubert; Wolfram Miekisch
Journal:  Analyst       Date:  2019-12-02       Impact factor: 4.616

View more
  1 in total

1.  Ultra-Sensitive Isopropanol Biochemical Gas Sensor (Bio-Sniffer) for Monitoring of Human Volatiles.

Authors:  Po-Jen Chien; Takuma Suzuki; Ming Ye; Koji Toma; Takahiro Arakawa; Yasuhiko Iwasaki; Kohji Mitsubayashi
Journal:  Sensors (Basel)       Date:  2020-11-29       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.