Effects of humidity, CO2 and O2 on real-time quantitation of breath biomarkers by means of PTR-ToF-MS.

Phillip Trefz1, Jochen K Schubert, Wolfram Miekisch.   

Abstract

Proton transfer reaction time-of-flight mass spectrometry (PTR-ToF-MS) represents an attractive tool for the real-time analysis of VOC profiles in human breath. Quantification of breath VOCs by means of direct MS may be affected by the matrix, as human breath not only contains several hundred VOCs at the ppbV-pptV level, but is water saturated and contains percentage levels of CO2. Investigation of breath biomarkers in clinical studies requires quantitative and comparable results. We therefore systematically assessed the effect of humidity, CO2 and O2 on the results of PTR-MS analysis. We investigated more than 20 VOCs, including aldehydes, ketones, aromatic compounds and hydrocarbons with different sample humidity, CO2 and O2 content. The influence of data processing (e.g. normalization to the H3O+ ion count) was also addressed. An increase of the H3O+ count of about 20% was observed when the humidity in the sample was increased to breath levels. Large differences regarding the measured VOC intensities were found between the dry and humid samples. Data normalization to the H3O+ or water-clusters could not fully compensate for the humidity-dependent effects. However, as the determination of most VOCs linearly depends on the humidity over the whole investigated range, factor-based correction seems possible. The effects of CO2 were more pronounced in the dry samples than in the humid samples but only had a minor influence on the results. The same was true for the influence of O2. For the reliable quantification of VOCs in clinical studies and for the standardization of VOC research, well-adapted calibration standards are required for PTR-MS analysis.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29199640     DOI: 10.1088/1752-7163/aa9eea

Source DB:  PubMed          Journal:  J Breath Res        ISSN: 1752-7155            Impact factor:   3.262


  11 in total

1.  Selected ion flow tube mass spectrometry for targeted analysis of volatile organic compounds in human breath.

Authors:  Ilaria Belluomo; Piers R Boshier; Antonis Myridakis; Bhamini Vadhwana; Sheraz R Markar; Patrik Spanel; George B Hanna
Journal:  Nat Protoc       Date:  2021-06-04       Impact factor: 13.491

2.  Release Kinetics Studies of Early-Stage Volatile Secondary Oxidation Products of Rapeseed Oil Emitted during the Deep-Frying Process.

Authors:  Tomasz Majchrzak; Andrzej Wasik
Journal:  Molecules       Date:  2021-02-14       Impact factor: 4.411

3.  Physiological and metabolic effects of healthy female aging on exhaled breath biomarkers.

Authors:  Pritam Sukul; Simon Grzegorzewski; Celine Broderius; Phillip Trefz; Thomas Mittlmeier; Dagmar-Christiane Fischer; Wolfram Miekisch; Jochen K Schubert
Journal:  iScience       Date:  2022-01-06

4.  Real-Time Non-Invasive Monitoring of Short-Chain Fatty Acids in Exhaled Breath.

Authors:  Joris Meurs; Evangelia Sakkoula; Simona M Cristescu
Journal:  Front Chem       Date:  2022-04-26       Impact factor: 5.545

5.  Real-time metabolic monitoring under exhaustive exercise and evaluation of ventilatory threshold by breathomics: Independent validation of evidence and advances.

Authors:  Giovanni Pugliese; Phillip Trefz; Matthias Weippert; Johannes Pollex; Sven Bruhn; Jochen K Schubert; Wolfram Miekisch; Pritam Sukul
Journal:  Front Physiol       Date:  2022-08-12       Impact factor: 4.755

6.  Exhaled breath compositions under varying respiratory rhythms reflects ventilatory variations: translating breathomics towards respiratory medicine.

Authors:  Pritam Sukul; Jochen K Schubert; Karim Zanaty; Phillip Trefz; Anupam Sinha; Svend Kamysek; Wolfram Miekisch
Journal:  Sci Rep       Date:  2020-08-24       Impact factor: 4.379

7.  Selective Reagent Ion Mass Spectrometric Investigations of the Nitroanilines.

Authors:  David Olivenza-León; Chris A Mayhew; Ramón González-Méndez
Journal:  J Am Soc Mass Spectrom       Date:  2019-09-09       Impact factor: 3.109

8.  Non-Invasive Assessment of Metabolic Adaptation in Paediatric Patients Suffering from Type 1 Diabetes Mellitus.

Authors:  Phillip Trefz; Sibylle C Schmidt; Pritam Sukul; Jochen K Schubert; Wolfram Miekisch; Dagmar-Christiane Fischer
Journal:  J Clin Med       Date:  2019-10-26       Impact factor: 4.241

9.  Effects of modular ion-funnel technology onto analysis of breath VOCs by means of real-time mass spectrometry.

Authors:  Giovanni Pugliese; Felix Piel; Phillip Trefz; Philipp Sulzer; Jochen K Schubert; Wolfram Miekisch
Journal:  Anal Bioanal Chem       Date:  2020-08-13       Impact factor: 4.142

10.  ptairMS: real-time processing and analysis of PTR-TOF-MS data for biomarker discovery in exhaled breath.

Authors:  Camille Roquencourt; Stanislas Grassin-Delyle; Etienne A Thévenot
Journal:  Bioinformatics       Date:  2022-01-19       Impact factor: 6.937

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.