Increased sensitivity in proton transfer reaction mass spectrometry by incorporation of a radio frequency ion funnel.

Shane Barber1, Robert S Blake, Iain R White, Paul S Monks, Fraser Reich, Steve Mullock, Andrew M Ellis.   

Abstract

A drift tube capable of simultaneously functioning as an ion funnel is demonstrated in proton transfer reaction mass spectrometry (PTR-MS) for the first time. The ion funnel enables a much higher proportion of ions to exit the drift tube and enter the mass spectrometer than would otherwise be the case. An increase in the detection sensitivity for volatile organic compounds of between 1 and 2 orders of magnitude is delivered, as demonstrated using several compounds. Other aspects of analytical performance explored in this study include the effective E/N (ratio of electric field to number density of the gas) and dynamic range over which the drift tube is operated. The dual-purpose drift tube/ion funnel can be coupled to various types of mass spectrometers to increase the detection sensitivity and may therefore offer considerable benefits in PTR-MS work.

Year:  2012        PMID: 22612412     DOI: 10.1021/ac300894t

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  4 in total

1.  PTR-ToF-MS Coupled with an Automated Sampling System and Tailored Data Analysis for Food Studies: Bioprocess Monitoring, Screening and Nose-space Analysis.

Authors:  Vittorio Capozzi; Sine Yener; Iuliia Khomenko; Brian Farneti; Luca Cappellin; Flavia Gasperi; Matteo Scampicchio; Franco Biasioli
Journal:  J Vis Exp       Date:  2017-05-11       Impact factor: 1.355

2.  High-Throughput Volatilome Fingerprint Using PTR-ToF-MS Shows Species-Specific Patterns in Mortierella and Closely Related Genera.

Authors:  Anusha Telagathoti; Maraike Probst; Iuliia Khomenko; Franco Biasioli; Ursula Peintner
Journal:  J Fungi (Basel)       Date:  2021-01-19

3.  Selective Reagent Ion Mass Spectrometric Investigations of the Nitroanilines.

Authors:  David Olivenza-León; Chris A Mayhew; Ramón González-Méndez
Journal:  J Am Soc Mass Spectrom       Date:  2019-09-09       Impact factor: 3.109

4.  Effects of modular ion-funnel technology onto analysis of breath VOCs by means of real-time mass spectrometry.

Authors:  Giovanni Pugliese; Felix Piel; Phillip Trefz; Philipp Sulzer; Jochen K Schubert; Wolfram Miekisch
Journal:  Anal Bioanal Chem       Date:  2020-08-13       Impact factor: 4.142

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.