Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital.

C Ting1, A A Nanji.   

Abstract

We evaluated the precision and accuracy of monitoring of the blood glucose level in the laboratory and at the bedside with one of four glucose meters by an experienced operator and by 39 nurses in a teaching hospital. For precision studies aqueous quality control materials were used. A total of 85 blood samples were tested. The precision of the glucose meters (expressed as the coefficient of variation [CV]) in the hands of the experienced operator ranged from 6.7% to 11.1%. The correlation between the values obtained by the experienced operator and the reference values obtained in the laboratory was high (0.95 to 0.98). The precision of the values obtained by the nurses using the meters ranged from 13.7% on medical wards to 45.7% in the intensive care unit (ICU). The correlation between these values and those obtained in the central laboratory ranged from 0.72 to 0.82. Twenty-four percent of the glucose values determined on medical wards and 62% of those determined in the ICU deviated from the reference value by at least 20%. Of the 85 patients 12 (14%) would have received different insulin dosages had the reference value been available at the same time as the glucose meter reading: in 3 of the patients the discrepancy was 6 units of insulin or greater. Continuous quality control of bedside monitoring of the blood glucose level is needed. In addition, personnel who use glucose meters should receive adequate training.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3275478      PMCID: PMC1267502     

Source DB:  PubMed          Journal:  CMAJ        ISSN: 0820-3946            Impact factor:   8.262


  13 in total

1.  Potential danger of extending SMBG techniques to hospital wards.

Authors:  F E Cohen; B Sater; K R Feingold
Journal:  Diabetes Care       Date:  1986 May-Jun       Impact factor: 19.112

2.  Community-based diabetes education: an outreach program.

Authors:  B Balik; P M Moynihan; B Haig
Journal:  Diabetes Educ       Date:  1985       Impact factor: 2.140

3.  Biochemical self-monitoring: promise, practice and problems.

Authors:  S Walford; K G Alberti
Journal:  Contemp Issues Clin Biochem       Date:  1985

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Performance of blood-glucose meters.

Authors:  L J Borthwick; I S Ross
Journal:  Lancet       Date:  1979-04-28       Impact factor: 79.321

6.  BM strips: how accurate are they in general wards?

Authors:  A S Hutchison; A Shenkin
Journal:  Diabet Med       Date:  1984-09       Impact factor: 4.359

7.  Comparison of various methods for rapid glucose estimation.

Authors:  R S Clements; N A Keane; K A Kirk; B R Boshell
Journal:  Diabetes Care       Date:  1981 May-Jun       Impact factor: 19.112

Review 8.  Home monitoring of diabetes mellitus--a quiet revolution.

Authors:  A L McCall; C J Mullin
Journal:  Clin Lab Med       Date:  1986-06       Impact factor: 1.935

9.  Behavioral skills in self-monitoring of blood glucose: relationship to accuracy.

Authors:  R R Wing; R Koeske; A New; D Lamparski; D Becker
Journal:  Diabetes Care       Date:  1986 Jul-Aug       Impact factor: 19.112

10.  Self glucose monitoring: a comparison of the Glucometer, Glucoscan, and Hypocount B.

Authors:  J D Nelson; M A Woelk; S Sheps
Journal:  Diabetes Care       Date:  1983 May-Jun       Impact factor: 19.112

View more
  4 in total

1.  Self monitoring of blood glucose concentrations by non-insulin dependent diabetic patients.

Authors:  R Tattersall
Journal:  BMJ       Date:  1992-11-14

2.  The need for continuous blood glucose monitoring in the intensive care unit.

Authors:  Ram Weiss; Isaac Lazar
Journal:  J Diabetes Sci Technol       Date:  2007-05

3.  Bedside monitoring of the blood glucose level: the high cost of reliable results.

Authors:  V T Innanen; N Korogyi; L Mellor; T Ambus; A Kenshole
Journal:  CMAJ       Date:  1989-04-15       Impact factor: 8.262

4.  Evaluation of bedside blood glucose monitoring by the phlebotomy team in an acute general care hospital.

Authors:  J M Johnston; P Grondziowski; R M Windisch; R A Eder; L Lytle
Journal:  Acta Diabetol       Date:  1995-12       Impact factor: 4.280

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.