BM strips: how accurate are they in general wards?

A S Hutchison, A Shenkin.   

Abstract

In order to test the accuracy of BM Test Glycemie 20-800 strips when used in general wards, simultaneous BM strip results were obtained for 183 emergency out-of-hours plasma glucose requests. The results were considered accurate if the value obtained by the Emergency Biochemistry Laboratory (using a Beckman II glucose analyser) was within one strip-range on either side of the stated BM strip value. Fifty-nine (32.2%) of the BM strip results were considered inaccurate by this definition, and in 22 of these inaccurate results the differences were large enough to have led to inappropriate management if this was based on the BM strip result alone. The reasons for these inaccuracies are not clear, but it is suggested that improved training of ward staff and formal quality control schemes may help to improve results.

Entities:  

Mesh:

Substances:

Year:  1984        PMID: 6242803     DOI: 10.1111/j.1464-5491.1984.tb01958.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  3 in total

1.  A critical look at nursing observations.

Authors:  J Burroughs; B I Hoffbrand
Journal:  Postgrad Med J       Date:  1990-05       Impact factor: 2.401

2.  Accuracy of home blood glucose monitoring by children.

Authors:  C R Kirk; H Burke; D C Savage; A O Hughes
Journal:  Br Med J (Clin Res Ed)       Date:  1986-07-05

3.  Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital.

Authors:  C Ting; A A Nanji
Journal:  CMAJ       Date:  1988-01-01       Impact factor: 8.262

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.