Literature DB >> 32748185

Automated detection of electrically evoked stapedius reflexes (eSR) during cochlear implantation.

Nora M Weiss1, Attila Óvári1,2, Tobias Oberhoffner1, Laurent Demaret3, Atabek Bicer3, Sebastian Schraven1, Karsten Ehrt1, Rüdiger Dahl1, Armin Schneider3, Robert Mlynski4.   

Abstract

INTRODUCTION: In cochlear implantation, objective fitting methods are needed to optimize audiological results in small children or patients with poor compliance. Intraoperatively measured electrically evoked stapedius reflexes (eSR) can be used as a marker for the patient's discomfort level. The aim of this study was to develop and evaluate an automated detection method for eSR and to compare it to the detection rate of the surgeon and independent observers.
METHODS: Cochlear implantation using a fully digital surgical microscope was performed. Movements of the stapedius tendon were recorded and analyzed by means of computer vision technique. Differences in eSR elicited by stimulating electrodes at different cochlear locations (basal, middle and apical) were analyzed. The eSR detection rate of the image processing algorithm was compared to the surgeon's detection rate and to those of two less experienced observers.
RESULTS: A total of 387 electrically impulses were applied. The stimulation of middle turn electrodes showed significantly higher detection rates (50.4%) compared to the basal (40.0%; p = 0.001) and apical (43.6%; p = 0.03) turn. The software identified significantly more of the applied stimuli (58.4%) compared to the surgeon (46.3%; p = 0.0007), the intermediate observer (37.7%; p < 0.0001) and the unexperienced observer (41.3%; p < 0.0001).
CONCLUSION: The feasibility of an automated intraoperative software-based detection of eSR is demonstrated. By improving the eSR detection methods and their clinical applicability, their utility in objective cochlear implant fitting may be substantially increased.

Entities:  

Keywords:  Cochlea implantation; Electrically evoked stapedius reflex thresholds; Objective cochlear implant fitting; eSRT

Mesh:

Year:  2020        PMID: 32748185     DOI: 10.1007/s00405-020-06226-x

Source DB:  PubMed          Journal:  Eur Arch Otorhinolaryngol        ISSN: 0937-4477            Impact factor:   2.503


  15 in total

1.  Measuring the electrical stapedius reflex with stapedius muscle electromyogram recordings.

Authors:  Ryan S Clement; Paul M Carter; Daryl R Kipke
Journal:  Ann Biomed Eng       Date:  2002-02       Impact factor: 3.934

2.  How reliable is visual assessment of the electrically elicited stapedius reflex threshold during cochlear implant surgery, compared with tympanometry?

Authors:  H W Pau; K Ehrt; T Just; U Sievert; R Dahl
Journal:  J Laryngol Otol       Date:  2010-11-05       Impact factor: 1.469

3.  Threshold of reflex contractions of muscles of middle ear and recruitment of loudness.

Authors:  O METZ
Journal:  AMA Arch Otolaryngol       Date:  1952-05

4.  Effect of stylet removal on neural response telemetry and stapedial reflex thresholds during cochlear implantation.

Authors:  Mohsen Rajati; Mohammad Mehdi Ghassemi; Mehdi Bakhshaee; Mohammad Reza Tale; Hamid Tayarani
Journal:  Auris Nasus Larynx       Date:  2013-12-17       Impact factor: 1.863

5.  Intraoperative electrically evoked stapedius reflex thresholds in children undergone cochlear implantation: round window and cochleostomy approaches.

Authors:  Erkan Karatas; Murat Deniz Aud; Tekin Baglam; Cengiz Durucu; Elif Baysal; Muzaffer Kanlikama
Journal:  Int J Pediatr Otorhinolaryngol       Date:  2011-07-07       Impact factor: 1.675

6.  Would an endosteal CI-electrode make sense? Comparison of the auditory nerve excitability from different stimulation sites using ESRT measurements and mathematical models.

Authors:  Hans Wilhelm Pau; Annekathrin Grünbaum; Karsten Ehrt; Rüdiger Dahl; Tino Just; Ursula van Rienen
Journal:  Eur Arch Otorhinolaryngol       Date:  2013-05-09       Impact factor: 2.503

7.  Measurement of comfort levels by means of electrical stapedial reflex in children.

Authors:  M Bresnihan; G Norman; F Scott; L Viani
Journal:  Arch Otolaryngol Head Neck Surg       Date:  2001-08

8.  The Value of Electrically Evoked Stapedius Reflex in Determining the Maximum Comfort Level of a Cochlear Implant.

Authors:  Kelly Cristina Lira de Andrade; Lilian Ferreira Muniz; Pedro de Lemos Menezes; Silvio da Silva Caldas Neto; Aline Tenório Lins Carnaúba; Mariana de Carvalho Leal
Journal:  J Am Acad Audiol       Date:  2018-04       Impact factor: 1.664

9.  Electromyographical recording of the electrically elicited stapedius reflex via a bipolar hook electrode.

Authors:  Hans Wilhelm Pau; Thorsten Zehlicke; Uwe Sievert; Daniel Schaudel; Detlef Behrend; Rüdiger Dahl
Journal:  Otol Neurotol       Date:  2009-01       Impact factor: 2.311

10.  Influence of EEG monitoring on intraoperative stapedius reflex threshold values in cochlear implantation in children.

Authors:  B Schultz; F A Beger; B P Weber; O Niclaus; E Lüllwitz; U Grouven; A Schultz
Journal:  Paediatr Anaesth       Date:  2003-11       Impact factor: 2.556

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.