Literature DB >> 18833019

Electromyographical recording of the electrically elicited stapedius reflex via a bipolar hook electrode.

Hans Wilhelm Pau1, Thorsten Zehlicke, Uwe Sievert, Daniel Schaudel, Detlef Behrend, Rüdiger Dahl.   

Abstract

OBJECTIVE: To prove the feasibility of recording stapedius reflexes by bipolar electromyography intra-operatively during cochlea implant via a bipolar hook needle electrode. STUDY
DESIGN: Prospective study.
SETTING: Tertiary referral center. PATIENTS: Adults receiving a MED-EL cochlear implant. INTERVENTION: Diagnostic. MAIN OUTCOME MEASURES: Stapedius reflex responses can be recorded by electromyography.
RESULTS: A double hook electrode could be inserted into the muscle tissue via the natural opening of the pyramid process along with the stapedius tendon. In 7 out of 10 patients supplied with MED-EL CIs (PULSARCI100 or SONATATI100), electrically elicited stapedius reflex potentials were recorded after eliminating stimulation artifacts caused by electromagnetic fields during electrical stimulation by low pass filtering.
CONCLUSION: Intraoperative stapedius reflex recording by bipolar electromyography via a hook electrode is feasible. Further research will show whether these signals may be used for establishing a self-adjusting speech processor.

Entities:  

Mesh:

Year:  2009        PMID: 18833019     DOI: 10.1097/MAO.0b013e31818a0898

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  3 in total

1.  Dyna-CT of the temporal bone for case-specific three-dimensional rendering of the stapedial muscle for planning of electrically evoked stapedius reflex threshold determination during cochlear implantation directly from the stapedius muscle via a retrofacial approach: a pilot study.

Authors:  Gerd Fabian Volk; René Aschenbach; Maria Gadyuchko; Thomas Bitter; Sven Koscielny; Ulf Teichgräber; Orlando Guntinas-Lichius
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-01-02       Impact factor: 2.503

2.  Accessing the stapedius muscle via novel surgical retrofacial approach during cochlear implantation surgery: Intraoperative results on feasibility and safety.

Authors:  Orlando Guntinas-Lichius; Dirk Arnold; Gerd Fabian Volk; Daniela Korth; Rene Aschenbach; Johann-Martin Hempel; Fritz Schneider; Thore Schade-Mann; Philipp Gamerdinger; Anke Tropitzsch; Hubert Löwenheim
Journal:  PLoS One       Date:  2022-08-11       Impact factor: 3.752

3.  Automated detection of electrically evoked stapedius reflexes (eSR) during cochlear implantation.

Authors:  Nora M Weiss; Attila Óvári; Tobias Oberhoffner; Laurent Demaret; Atabek Bicer; Sebastian Schraven; Karsten Ehrt; Rüdiger Dahl; Armin Schneider; Robert Mlynski
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-08-03       Impact factor: 2.503