Literature DB >> 32688048

Over-overlapped loop arrays: A numerical study.

Ming Lu1, John C Gore2, Xinqiang Yan3.   

Abstract

Arrays of coils are commonly used in MRI both for reception and in parallel transmission to alleviate radiofrequency field inhomogeneities at high fields. Most designs typically overlap loop elements by a critical area (approximately 10%) to minimize mutual inductive couplings. With this geometrical constraint, loop sizes have to be reduced to accommodate large numbers of coils for a given coverage. However, the contribution of coil noise to total noise increases as each coil size decreases, which reduces overall signal-to-noise ratio (SNR), especially in deeper regions of the sample volume. Here we propose arrays designs using elements that overlap much more (over-overlapped), and using numerical calculations we investigate their performance compared to two kinds of conventionally overlapped arrays (one with the same coil size but smaller coil number, and one with the same coil number but smaller coil size). Our simulation results show that the over-overlapped array can considerably increase the central SNR when coil noise dominates.
Copyright © 2020 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Array; Loop; Over-overlap; Parallel imaging; RF coil; SNR

Mesh:

Year:  2020        PMID: 32688048      PMCID: PMC7443860          DOI: 10.1016/j.mri.2020.07.006

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  24 in total

1.  SENSE: sensitivity encoding for fast MRI.

Authors:  K P Pruessmann; M Weiger; M B Scheidegger; P Boesiger
Journal:  Magn Reson Med       Date:  1999-11       Impact factor: 4.668

2.  A conformal transceive array for 7 T neuroimaging.

Authors:  Kyle M Gilbert; Jean-Guy Belliveau; Andrew T Curtis; Joseph S Gati; L Martyn Klassen; Ravi S Menon
Journal:  Magn Reson Med       Date:  2011-08-16       Impact factor: 4.668

3.  Bloch-based MRI system simulator considering realistic electromagnetic fields for calculation of signal, noise, and specific absorption rate.

Authors:  Zhipeng Cao; Sukhoon Oh; Christopher T Sica; John M McGarrity; Timothy Horan; Wei Luo; Christopher M Collins
Journal:  Magn Reson Med       Date:  2013-09-04       Impact factor: 4.668

4.  Ultimate intrinsic signal-to-noise ratio in MRI.

Authors:  O Ocali; E Atalar
Journal:  Magn Reson Med       Date:  1998-03       Impact factor: 4.668

5.  On the Contribution of Curl-Free Current Patterns to the Ultimate Intrinsic Signal-to-Noise Ratio at Ultra-High Field Strength.

Authors:  Andreas Pfrommer; Anke Henning
Journal:  NMR Biomed       Date:  2017-02-10       Impact factor: 4.044

6.  Radiofrequency energy deposition and radiofrequency power requirements in parallel transmission with increasing distance from the coil to the sample.

Authors:  Cem M Deniz; Manushka V Vaidya; Daniel K Sodickson; Riccardo Lattanzi
Journal:  Magn Reson Med       Date:  2015-03-05       Impact factor: 4.668

7.  An EM Simulation-Based Design Flow for Custom-Built MR Coils Incorporating Signal and Noise.

Authors:  Andreas Horneff; Michael Eder; Erich Hell; Johannes Ulrici; Jorg Felder; Volker Rasche; Jens Anders
Journal:  IEEE Trans Med Imaging       Date:  2017-10-18       Impact factor: 10.048

8.  Approaching ultimate intrinsic signal-to-noise ratio with loop and dipole antennas.

Authors:  Riccardo Lattanzi; Graham C Wiggins; Bei Zhang; Qi Duan; Ryan Brown; Daniel K Sodickson
Journal:  Magn Reson Med       Date:  2017-07-04       Impact factor: 4.668

9.  Transceiver-Phased Arrays for Human Brain Studies at 7 T.

Authors:  Nikolai I Avdievich
Journal:  Appl Magn Reson       Date:  2011-12       Impact factor: 0.831

10.  Potential acceleration performance of a 256-channel whole-brain receive array at 7 T.

Authors:  Arjan D Hendriks; Peter R Luijten; Dennis W J Klomp; Natalia Petridou
Journal:  Magn Reson Med       Date:  2018-09-26       Impact factor: 4.668

View more
  1 in total

1.  Design and construction of an interchangeable RF coil system for rodent spinal cord MR imaging at 9.4 T.

Authors:  Ming Lu; Gary Drake; Feng Wang; Chaoqi Mu; Li Min Chen; John C Gore; Xinqiang Yan
Journal:  Magn Reson Imaging       Date:  2021-10-05       Impact factor: 2.546