Biomechanical Comparison of Different Fixation Techniques for Typical Acetabular Fractures in the Elderly: The Role of Special Quadrilateral Surface Buttress Plates.

Kaifang Chen1, Fan Yang, Sheng Yao, Zekang Xiong, Tingfang Sun, Xiaodong Guo.   

Abstract

BACKGROUND: Anterior column-posterior hemitransverse fractures are prevalent in the elderly and are often associated with quadrilateral surface (QLS) comminution. Several special QLS buttress plates have been introduced, but evidence of their comparability with traditional fixation devices is lacking. This biomechanical study aimed to compare special QLS buttress plates with traditional fixation devices.
METHODS: Anterior column-posterior hemitransverse fractures with an isolated QLS fragment were created on 24 composite hemipelves and were allocated to 4 fixation groups: (1) infrapectineal QLS buttress plate, (2) suprapectineal QLS buttress plate, (3) suprapectineal reconstruction plate with 3 periarticular long screws, and (4) infrapectineal reconstruction plate with 3 periarticular long screws. Specimens were loaded to simulate partial weight-bearing (35 to 350 N) or full weight-bearing (75 to 750 N). A testing machine was synchronized with a 3-dimensional video tracking system to optically track displacement at the points of interest and to calculate construct stiffness. The fixation systems were compared using the Kruskal-Wallis and Mann-Whitney U tests.
RESULTS: The experimental results of the partial and full weight-bearing simulations were surprisingly similar. During 40 loading cycles, the maximum displacement on the 6 predetermined points did not exceed 1.1 mm. Multiple-group comparisons of relative displacements of each predetermined measurement point did not differ significantly (p > 0.05). The suprapectineal reconstruction plate with 3 periarticular long screws demonstrated the greatest construct stiffness and significantly greater stiffness than the infrapectineal plate with long screws (p < 0.017). However, no significant difference (p > 0.017) in stiffness was identified between the infrapectineal QLS buttress plate and the suprapectineal reconstruction plate with long screws.
CONCLUSIONS: The suprapectineal pelvic brim plate with 3 periarticular long screws remains the gold standard to treat anterior column-posterior hemitransverse fractures. The special infrapectineal QLS buttress plate provides stiffness and stability comparable with those of standard fixation. However, moving the pelvic brim plate from the suprapectineal border to the infrapectineal border is not recommended for anterior column-posterior hemitransverse fractures because it significantly decreases fixation stiffness. CLINICAL RELEVANCE: Special QLS buttress plates may be an alternative fixation method for anterior column-posterior hemitransverse acetabular fractures in the elderly, especially when a less invasive anterior intrapelvic approach is selected.

Entities:  

Mesh:

Year:  2020        PMID: 32675678      PMCID: PMC7508287          DOI: 10.2106/JBJS.19.01027

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   6.558


As the worldwide population ages and the elderly remain active later in life, the incidence of acetabular fractures among patients >60 years of age has increased by 2 to 3 times over the past 3 decades[1-3]. The anterior column-posterior hemitransverse fracture pattern, the classic acetabular fracture in the geriatric population, is often associated with protrusion of the femoral head and medial migration of the quadrilateral surface (QLS)[4-6]. To reduce and stabilize the medialization of the femoral head and QLS, stable fixation is indispensable, but the relative osteopenia in the elderly creates difficulty achieving adequate fracture fixation[7,8]. Early on, the most commonly reported fixation method for anterior column-posterior hemitransverse fractures was pelvic brim plates with long periarticular screws or additional infra-acetabular screws through the classic ilioinguinal approach[9]. Later, to improve direct visualization of the QLS, anterior intrapelvic approaches such as the modified Stoppa[10,11], pararectus[12,13], and supra-ilioinguinal[14] have been popularized for infrapectineal and QLS buttress plating even closer to the ischial spine, all of which were parallel to the pelvic brim and perpendicular to the suprapectineal plate[15]. Placing the infrapectineal plate in the same plane as the displacement rather than perpendicular to the displacement (suprapectineal plate) may be more helpful for preventing protrusion in older patients[16]. Based on our team’s years of clinical experience in the surgical treatment of acetabular fractures, 2 special QLS buttress plates, the infrapectineal plate and the suprapectineal plate, were introduced and achieved good clinical results in our patients. No comparative biomechanical data on these newly developed QLS buttress plates and traditional constructs are available with regard to the fixation of typical acetabular fractures in the elderly. Therefore, this biomechanical study aimed to determine which of 4 selected fixation constructs is the stiffest and best able to prevent secondary displacement of an anterior column-posterior hemitransverse fracture model with a free QLS fragment.

Materials and Methods

Specimens and Fracture Model

Twenty-four fourth-generation composite right large hemipelves (Sawbones; Pacific Research Laboratories) were used in this study. Referring to the study by May et al.[5], a model of a typical acetabular fracture in the elderly (i.e., an anterior column-posterior hemitransverse fracture with disruption of the QLS) was created for the biomechanical assessment of 4 different fixation constructs. In pursuit of a consistent fracture model, computed tomographic (CT) data of the Sawbones hemipelvis were obtained to reconstruct a 3-dimensional model. An anterior column-posterior hemitransverse fracture model with QLS disruption and a corresponding osteotomy template similar to a jig were then designed (Mimics Suite 17.0, Materialise; and Geomagic Studio 2012, Raindrop Geomagic). The computer-aided design template was subsequently printed by rapid prototyping technology using nylon material, and a thin wire saw was used to create an identical osteotomy along the template.

Instrumentation

A total of 4 fixation constructs were tested and fluoroscopy was used to verify proper implant placement and fracture fixation (Figs. 1-A through 1-D): (1) special infrapectineal QLS buttress plate (Fig. 1-A); (2) special suprapectineal QLS buttress plate with an infra-acetabular screw (3.5-mm cortical screw, 80 mm) in hole 3 and a posterior column lag screw (7.3-mm cannulated lag screw, 90 mm) in hole 7 (Fig. 1-B); (3) 12-hole suprapectineal pelvic reconstruction plate with an infra-acetabular screw (3.5-mm cortical screw, 80 mm) in hole 4 and 2 posterior column screws (3.5-mm cortical screw, 90 mm) in holes 9 and 10 (Fig. 1-C); and (4) 10-hole infrapectineal pelvic reconstruction plate with 3 long periarticular screws, the same as in the suprapectineal reconstruction plate fixation construct (Fig. 1-D). An additional 3.5-mm cortical screw was inserted at the iliac crest to fix the high anterior column fracture in all groups. Figs. 1-A through 1-D Photographs and fluoroscopic images showing the 4 different constructs for stabilization of the anterior column-posterior hemitransverse fracture. A special infrapectineal QLS buttress plate. A special suprapectineal QLS buttress plate with an infra-acetabular screw in hole 3 and a posterior column lag screw in hole 7. A 12-hole suprapectineal pelvic reconstruction plate with an infra-acetabular screw in hole 4 and 2 posterior column screws in holes 9 and 10. A 10-hole infrapectineal pelvic reconstruction plate with 3 long periarticular screws (black arrows), the same as in the suprapectineal reconstruction plate fixation construct.

Biomechanical Setup

We adopted a loading methodology similar to those mentioned by Chang et al.[17] and Kistler et al.[18]. Similarly, the composite hemipelvis and femur (Sawbones; Pacific Research Laboratories) were coupled in a 3-dimensional model and the axis of the femoral neck was used as the biomechanical loading vector (SOLIDWORKS 2017; Dassault Systèmes SolidWorks). The axis should be oriented vertically in this coordinate system because of the design of the testing machine; meanwhile, a fixture for constraining the posterior aspect of the iliac bone and a block for supporting the symphysis were designed to maintain the hemipelvis model in a horizontal position. Thus, the symphysis was allowed sliding action in the horizontal plane, and movement was prevented in the direction of the applied force but was unrestricted in all directions perpendicular to the joint reaction force. For the femoral head, a custom-designed metal sphere with a 52-mm diameter was used to freely articulate with the acetabulum, similar to the test by Vigdorchik et al.[19].

Biomechanical Testing

A testing machine (MTS Bionix Servohydraulics Test Systems Model 370.02; MTS Systems) was used to perform the mechanical loading. A 3-dimensional video tracking system (VICON VERO; Vicon Motion Systems) was used to track the 3-dimensional motion of the selected points on the hemipelvis and to measure displacement of the spatial coordinates[20]. A total of 12 markers were used, including 11 on the pelvis model and 1 on the testing machine as the calibration point. Eleven markers were attached to these selected points in pairs (Figs. 2-A and 2-B). Six pairs of markers located on opposite sides of the fracture site (1-2, 3-4, 5-6, 7-8, 7-9, and 7-10) were considered of interest, and their relative displacements were calculated in the 3-dimensional spatial coordinates. The spatial resolution of the system is in the order of 0.1 mm. Figs. 2-A and 2-B Photographs showing a total of 11 infrared light-reflecting markers attached to the predetermined points along the fracture line of the anterior column-posterior hemitransverse fracture model. Markers 1, 2, 3, 4, 5, 6, 10, and 11. Markers 7, 8, and 9.

Loading Protocol

The 4 different fixation constructs (6 Sawbones specimens per fixation group) were compared in terms of displacements along the fracture sites, especially of the QLS. Three phases of mechanical loading were applied to each sample based on previously published data using the following protocol: (1) 250 settling cycles of 17.5 to 175 N at a machine velocity of 60 mm/min, (2) 20 cycles of 35 to 350 N at a machine velocity of 10 mm/min with optical tracking simulating the postoperative phase of recommended partial weight-bearing (according to May et al.[5]), and (3) 20 cycles of 75 to 750 N at a machine velocity of 10 mm/min with optical tracking simulating the single-leg stance (according to Culemann et al.[4]).

Data Collection

For each loading cycle, force and displacement data were recorded by a personal computer connected with the MTS Bionix. We defined the axial stiffness (in N/mm) of the construct as the slope of the force-displacement curves generated from the test system. The marker data were synchronized in MATLAB (MathWorks) for characterizing the motion at each pair of markers crossing the fracture lines. The relative displacement (in millimeters) of the fracture line was defined according to the maximum change of the relative distance of each pair of markers in the spatial coordinates. The first loading phase (17.5 to 175-N loading, 250 cycles) was our preloading test, which aimed to achieve a relatively stable state between the system and the constructs, so no data collection was performed. Data were collected during the subsequent 2 loading phases (35 to 350 N and 75 to 700 N).

Statistical Analysis

The statistical analysis was performed using MATLAB software. Nonparametric tests were used because the sample values were not normally distributed according to the homogeneity test of variance. First, a multiple-group comparison of the 4 groups were performed using the Kruskal-Wallis test with significance set at p = 0.05. Next, if the multiple comparisons showed a significant difference (p < 0.05), the Mann-Whitney U test was used to compare the group with the suprapectineal reconstruction plate and long screws with each of the 3 other groups. The Bonferroni correction was utilized in these post hoc pairwise comparisons, with p < 0.017 being considered significant.

Results

The relative displacements of the 6 predetermined pairs of measurement points for all 4 groups are presented as the median and the range in Table I (35 to 350-N loading cycle) and Table II (75 to 750-N loading cycle). These tables also show the p values for the multiple-group comparisons. The maximum displacement of the measurement point pairs among all specimens did not exceed 1 mm except for points 1 and 2 in the 75 to 750-N loading test (1.08 mm). From the distribution trend of the bar graph (Figs. 3 and 4), the displacement value of each measurement point pair in the other 3 groups was greater than that in the group with the suprapectineal reconstruction plate and long screws; however, multiple-group comparisons showed no significant difference among the 4 groups (p > 0.05).
Fig. 3

Bar graph showing the values of relative displacement in millimeters for the 4 groups at the 6 predetermined measurement points under the 35 to 350-N loading test. The differences with significance are indicated with a p value. The bars indicate the mean and the error bars indicate the standard deviation.

Fig. 4

Bar graph showing the values of relative displacement in millimeters for the 4 groups at the 6 predetermined measurement points under the 75 to 750-N loading test. The bars indicate the mean and the error bars indicate the standard deviation.

Displacements at Predetermined Measurement Points Under 35 to 350-N Loading Tests The values are given as the median in millimeters, with the range in parentheses. Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05. Displacements at Predetermined Measurement Points Under 75 to 750-N Loading Tests The values are given as the median in millimeters, with the range in parentheses. Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05. Bar graph showing the values of relative displacement in millimeters for the 4 groups at the 6 predetermined measurement points under the 35 to 350-N loading test. The differences with significance are indicated with a p value. The bars indicate the mean and the error bars indicate the standard deviation. Bar graph showing the values of relative displacement in millimeters for the 4 groups at the 6 predetermined measurement points under the 75 to 750-N loading test. The bars indicate the mean and the error bars indicate the standard deviation. The axial stiffnesses of all 4 fixation constructs under the 35 to 350-N and 75 to 750-N loading cycles are presented as the median and the range and as the mean and the standard deviation in Table III. Box-and-whisker plots are depicted with p values when the differences were significant in Figures 5 and 6. For the 35 to 350-N loading phase, the standard suprapectineal reconstruction plating fixation construct demonstrated the greatest stiffness (838.96 ± 385.01 N/mm), a significant difference (p = 0.006) from the infrapectineal reconstruction plating fixation construct (415.12 ± 96.08 N/mm). However, there was no difference between the standard suprapectineal fixation and the infrapectineal QLS buttress plate (551.02 ± 122.46 N/mm; p = 0.132) or the suprapectineal QLS buttress plate (414.77 ± 221.27 N/mm; p = 0.026).
Fig. 5

Box-and-whisker plots showing the values of axial stiffness in N/mm for the 4 different fixation constructs under the 35 to 350-N loading test. The horizontal red line in each box indicates the median value, the top and bottom borders of the box show the 75th and 25th percentiles, the whiskers show the 10th and 90th percentiles, and the + symbols show outliers. The differences with significance are indicated with a p value.

Fig. 6

Box-and-whisker plots showing the values of axial stiffness in N/mm for the 4 different fixation constructs under the 75 to 750-N loading test. The horizontal red line in each box indicates the median value, the top and bottom borders of the box show the 75th and 25th percentiles, and the whiskers show the 10th and 90th percentiles. The differences with significance are indicated with p values.

Axial Stiffness for All 4 Constructs The values are given as the median in N/mm, with the range in parentheses. The values are given as the mean in N/mm and the standard deviation. Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05. Significant. Pairwise comparisons were performed with use of the Mann-Whitney U test with Bonferroni correction with a significance level of p = 0.017. Box-and-whisker plots showing the values of axial stiffness in N/mm for the 4 different fixation constructs under the 35 to 350-N loading test. The horizontal red line in each box indicates the median value, the top and bottom borders of the box show the 75th and 25th percentiles, the whiskers show the 10th and 90th percentiles, and the + symbols show outliers. The differences with significance are indicated with a p value. Box-and-whisker plots showing the values of axial stiffness in N/mm for the 4 different fixation constructs under the 75 to 750-N loading test. The horizontal red line in each box indicates the median value, the top and bottom borders of the box show the 75th and 25th percentiles, and the whiskers show the 10th and 90th percentiles. The differences with significance are indicated with p values. For the 75 to 750-N loading cycle, similarly, the standard suprapectineal reconstruction plating fixation (787.71 ± 151.76 N/mm) demonstrated a stiffness greater than those of the other tested groups and showed a significant difference (p = 0.016) compared with the infrapectineal reconstruction plating construct (605.09 ± 127.01 N/mm). Again, the stiffnesses of the infrapectineal QLS buttress plate (725.01 ± 78.95 N/mm, p = 0.329) and the suprapectineal QLS buttress plate (602.41 ± 128.86 N/mm, p = 0.017) were comparable with that of the standard suprapectineal fixation. The results of the post hoc power analysis are shown in Table IV. Post Hoc Power Analysis of the 4 Outcomes A value of 0.10 indicates small, 0.25 indicates medium, and 0.40 indicates large. A value of ≥80% is desired.

Discussion

The anterior column-posterior hemitransverse fracture pattern with protrusion of the femoral head and medial migration of the QLS is prevalent in the growing elderly population. Anatomical reduction resulting in a congruent hip joint is the ultimate goal, but adequate internal fixation can maintain fracture reduction and allow painless early mobility until fracture-healing is achieved[5,18,21]. The choice of internal fixation for such acetabular fractures involving the QLS (suprapectineal or infrapectineal plating fixation and periarticular long screws or QLS buttress plating fixation) is controversial. However, to our knowledge, no direct biomechanical comparison between suprapectineal and infrapectineal fixation has been performed to date. Although biomechanical studies showed that plate fixation with long periarticular screws was comparable with or superior to QLS buttress plating in biomechanical stability, the insertion of these screws always bears the risk of joint penetration and offers inadequate stability with a high risk of failure and poor outcomes; thus, a secondary hip arthroplasty is required afterward[16,22]. Therefore, we have developed 2 special QLS buttress plates, suprapectineal and infrapectineal, that have demonstrated good clinical results in our patients. Similarly, no comparative biomechanical data on these newly developed QLS buttress plates and classic fixation constructs are available. Based on the above 2 points, a typical acetabular fracture model in the elderly (anterior column-posterior hemitransverse) with QLS disruption was created for the biomechanical assessment of 4 different fixation constructs. Because suprapectineal plating fixation with long periarticular screws represents the gold standard for stabilizing anterior column-posterior hemitransverse fractures, it was designated as the control group in this experiment[4-6]. Therefore, a total of 3 pairwise comparisons were used to explore the biomechanical stiffness and stability of each fixation construct between standard suprapectineal reconstruction plating fixation and the other 3 fixations (infrapectineal QLS buttress plate, suprapectineal QLS buttress plate, and infrapectineal reconstruction plate with 3 periarticular long screws). To simulate the postoperative phase of recommended partial weight-bearing, 20 cycles of 35 to 350-N loading were performed after 250 settling preload cycles of 17.5 to 175 N. Because the maximum displacement of the predetermined point pairs among all specimens did not exceed 1.1 mm, the medial displacement of the quadrilateral plate was sufficiently prevented by the 4 fixation constructs under the modeled biomechanical conditions[5]. The multiple-group comparison of relative displacements on each predetermined measurement point pair among the 4 groups showed no significant difference (p > 0.05). However, the comparison of construct stiffness showed that the standard fixation method performed better than the other methods. Importantly, the infrapectineal QLS buttress plate that spans both columns through the QLS was statistically comparable with the standard fixation construct, with only a trend toward a difference (p = 0.132). This finding is similar to those of previous studies demonstrating that the final stiffness of the infrapectineal QLS buttress plate was comparable with traditional fixation constructs in a transtectal transverse model[18]. In addition, the pairwise comparison showed that suprapectineal reconstruction plating fixation was significantly stiffer than infrapectineal fixation (p = 0.006). This might be explained by the mechanics of the normal hip joint during stance (i.e., the weight-bearing area of the hip joint is concentrated in the superior region). Therefore, the standard fixation construct comprised a suprapectineal plate supporting superiorly along the pelvic brim, which biomechanically constitutes a bridge between the weight-bearing acetabular roof and the pelvis, obtaining the best stability. For various reasons, partial weight-bearing is often not practical in the elderly. Therefore, another 20 cycles of 75 to 750-N loading were performed after the 35 to 350-N loading cycle to simulate the full weight-bearing phase. Interestingly, the experimental results were strikingly similar to those shown above. Overall, for partial or full weight-bearing loading simulations, the suprapectineal pelvic brim plate with 3 periarticular long screws represents the gold standard for stabilizing the anterior column-posterior hemitransverse fracture pattern. The special infrapectineal QLS buttress plate provides stiffness and stability comparable with those of the standard fixation construct. A successful acetabular fracture surgical procedure consists of more than just an appropriate fixation construct; an improved surgical approach is also necessary, especially for elderly patients. Traditionally, the standard suprapectineal plate combined with periarticular long screws is usually inserted using an ilioinguinal approach, which requires laborious exposure and is associated with a high complication rate. Recent advances in surgical approaches, including the emergence of less invasive intrapelvic approaches such as the modified Stoppa, pararectus, or supra-ilioinguinal approaches[10-14], provide improved access to the QLS and the medial aspect of the posterior column with lower patient morbidity. Afterward, buttress plating of the medial wall and spanning fixation from the posterior to anterior columns via a single anterior incision became possible[15,18]. Thus, in the present study, we compared the strength and construct stiffness of 2 such QLS buttress plates with the standard bicolumnar fixation methods using a column plate and periarticular long screws and revealed that these newer QLS buttress plates are biomechanically viable options when the less invasive anterior intrapelvic approach is desired for anterior column-posterior hemitransverse fracture stabilization. This study had several limitations. First, artificial bone models could not represent the biological changes that occur in vivo because of the absence of soft-tissue and muscular forces, which may have affected test results. Second, the Sawbones model was produced in accordance with normal bone density, which did not account for the osteopenia in the typical elderly patient with an anterior column-posterior hemitransverse fracture. Third, the artificial bones showed higher stiffness than human bones and the uniform property distribution within the foam core did not account for the natural variations in density and force distribution seen throughout the human bones[23]. Fourth, a sample size of 6 per study group may be considered relatively small, and, as a result, some biomechanically meaningful differences may have been missed. A future biomechanical study design with a larger number of samples is needed. Other limitations were that 1 marker per fragment was not sufficient to determine the movements between the fragments. Three markers per fragment would be feasible, but it is admittedly challenging. In conclusion, this biomechanical study assessed 4 fixation constructs for stabilizing anterior column-posterior hemitransverse fractures with QLS disruption. All 4 constructs could maintain stable fixation under simulated partial or full weight-bearing activity. Meanwhile, the suprapectineal pelvic brim plate with 3 periarticular long screws remains the gold standard. Importantly, the special infrapectineal QLS buttress plate provides stiffness and stability comparable with the standard fixation construct. Therefore, in situations in which a less invasive anterior approach is desired, the use of newer QLS buttress plates presents an alternative fixation method that is at least comparable with standard forms of fixation at resisting fracture motion and medial subluxation. However, it is not recommended to move the pelvic brim plate from the suprapectineal border to the infrapectineal border for anterior column-posterior hemitransverse fractures because it will significantly decrease the fixation strength.
TABLE I

Displacements at Predetermined Measurement Points Under 35 to 350-N Loading Tests

Paired MarkersGroupP Value
Infrapectineal QLS Buttress Plate*Suprapectineal QLS Buttress Plate*Suprapectineal Reconstruction Plate with Long Screws*Infrapectineal Reconstruction Plate with Long Screws*
1-20.06 (0.01 to 0.14)0.05 (0.01 to 0.46)0.07 (0.04 to 0.08)0.10 (0.03 to 0.18)0.738
3-40.16 (0.02 to 0.37)0.17 (0.03 to 0.74)0.08 (0.02 to 0.23)0.09 (0.04 to 0.43)0.579
5-60.07 (0 to 0.28)0.03 (0.01 to 0.20)0.07 (0 to 0.32)0.07 (0.02 to 0.14)0.717
7-80.12 (0.04 to 0.55)0.05 (0.01 to 0.20)0.06 (0.01 to 0.18)0.03 (0 to 0.06)0.056
7-90.08 (0.05 to 0.63)0.16 (0.03 to 0.29)0.05 (0 to 0.25)0.11 (0.01 to 0.22)0.394
7-100.08 (0.03 to 0.54)0.03 (0.02 to 0.28)0.04 (0.01 to 0.17)0.05 (0.01 to 0.18)0.535

The values are given as the median in millimeters, with the range in parentheses.

Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05.

TABLE II

Displacements at Predetermined Measurement Points Under 75 to 750-N Loading Tests

Paired MarkersGroupP Value
Infrapectineal QLS Buttress Plate*Suprapectineal QLS Buttress Plate*Suprapectineal Reconstruction Plate with Long Screws*Infrapectineal Reconstruction Plate with Long Screws*
1-20.12 (0.05 to 0.26)0.21 (0 to 1.08)0.11 (0.06 to 0.18)0.16 (0.09 to 0.30)0.295
3-40.10 (0.02 to 0.14)0.45 (0.05 to 0.62)0.23 (0 to 0.29)0.10 (0.01 to 0.65)0.297
5-60.27 (0.11 to 0.49)0.03 (0 to 0.36)0.15 (0.01 to 0.55)0.17 (0.03 to 0.24)0.120
7-80.07 (0.03 to 0.66)0.15 (0.01 to 0.43)0.04 (0.01 to 0.14)0.09 (0.01 to 0.25)0.657
7-90.14 (0.02 to 0.38)0.17 (0.07 to 0.44)0.17 (0.02 to 0.19)0.14 (0.01 to 0.23)0.846
7-100.15 (0.01 to 0.55)0.09 (0.02 to 0.89)0.07 (0.04 to 0.24)0.06 (0.02 to 0.15)0.912

The values are given as the median in millimeters, with the range in parentheses.

Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05.

TABLE III

Axial Stiffness for All 4 Constructs

Stiffness Under 35 to 350-N LoadingStiffness Under 75 to 750-N Loading
Infrapectineal QLS buttress plate
 Median*547.87 (425.68 to 768.29)686.19 (651.15 to 830.37)
 Mean551.02 ± 122.46725.01 ± 78.95
Suprapectineal QLS buttress plate
 Median*339.34 (264.71 to 851.35)607 (397.66 to 737.51)
 Mean414.77 ± 221.27602.41 ± 128.86
Suprapectineal reconstruction plate with long screws
 Median*696.43 (492.19 to 1,500)818.61 (658.15 to 839.21)
 Mean838.96 ± 385.01787.71 ± 151.76
Infrapectineal reconstruction plate with long screws
 Median*412.04 (294.39 to 492.19)595.12 (521.69 to 697.9)
 Mean415.12 ± 96.08605.09 ± 127.01
P value for multiple-group comparison0.011§0.020§
P value for pairwise comparisons#
 Infrapectineal QLS buttress plate vs. suprapectineal reconstruction plate with long screws0.1320.329
 Suprapectineal QLS buttress plate vs. suprapectineal reconstruction plate with long screws0.0260.017
 Infrapectineal reconstruction plate with long screws vs. suprapectineal reconstruction plate with long screws0.006§0.016§

The values are given as the median in N/mm, with the range in parentheses.

The values are given as the mean in N/mm and the standard deviation.

Multiple-group comparisons were performed with use of the Kruskal-Wallis test with a significance level of p = 0.05.

Significant.

Pairwise comparisons were performed with use of the Mann-Whitney U test with Bonferroni correction with a significance level of p = 0.017.

TABLE IV

Post Hoc Power Analysis of the 4 Outcomes

Parameter and Loading CycleEffect Size f*Power
Displacement
 35 to 350 N0.168.2%
 75 to 750 N0.4841.0%
Stiffness
 35 to 350 N1.4199.9%
 75 to 750 N1.0097.4%

A value of 0.10 indicates small, 0.25 indicates medium, and 0.40 indicates large.

A value of ≥80% is desired.

  23 in total

1.  Biomechanical comparison of the human cadaveric pelvis with a fourth generation composite model.

Authors:  Brandon L Girardi; Tarik Attia; David Backstein; Oleg Safir; Thomas L Willett; Paul R T Kuzyk
Journal:  J Biomech       Date:  2016-01-22       Impact factor: 2.712

2.  Treatment of protrusio fractures of the acetabulum in patients 70 years and older.

Authors:  Michael T Archdeacon; Namdar Kazemi; Cory Collinge; Bradley Budde; Scott Schnell
Journal:  J Orthop Trauma       Date:  2013-05       Impact factor: 2.512

3.  Single Modified Ilioinguinal Approach for the Treatment of Acetabular Fractures Involving Both Columns.

Authors:  Kaifang Chen; Yanhui Ji; Zhenfei Huang; Ramphul Navinduth; Fan Yang; Tingfang Sun; Zekang Xiong; Sheng Yao; Jaimo Ahn; Xiaodong Guo
Journal:  J Orthop Trauma       Date:  2018-11       Impact factor: 2.512

4.  Acute total hip arthroplasty for selected displaced acetabular fractures: two to twelve-year results.

Authors:  Dana C Mears; John H Velyvis
Journal:  J Bone Joint Surg Am       Date:  2002-01       Impact factor: 5.284

5.  Internal fixation of osteopenic acetabular fractures involving the quadrilateral plate.

Authors:  G Y Laflamme; J Hebert-Davies; D Rouleau; B Benoit; S Leduc
Journal:  Injury       Date:  2010-12-14       Impact factor: 2.586

6.  A biomechanical study of standard posterior pelvic ring fixation versus a posterior pedicle screw construct.

Authors:  Jonathan M Vigdorchik; Xin Jin; Anil Sethi; Darren T Herzog; Bryant W Oliphant; King H Yang; Rahul Vaidya
Journal:  Injury       Date:  2015-05-07       Impact factor: 2.586

7.  Biomechanical comparison of sacroiliac screw techniques for unstable pelvic ring fractures.

Authors:  C M A van Zwienen; E W van den Bosch; C J Snijders; G J Kleinrensink; A B van Vugt
Journal:  J Orthop Trauma       Date:  2004-10       Impact factor: 2.512

8.  Are quadrilateral surface buttress plates comparable to traditional forms of transverse acetabular fracture fixation?

Authors:  Brian J Kistler; Ian R Smithson; Seth A Cooper; Jacob L Cox; Aniruddh N Nayak; Brandon G Santoni; H Claude Sagi
Journal:  Clin Orthop Relat Res       Date:  2014-11       Impact factor: 4.176

9.  Acetabular fracture fixation via a modified Stoppa limited intrapelvic approach. Description of operative technique and preliminary treatment results.

Authors:  J D Cole; B R Bolhofner
Journal:  Clin Orthop Relat Res       Date:  1994-08       Impact factor: 4.176

Review 10.  The treatment of acetabular fractures through the ilioinguinal approach.

Authors:  E Letournel
Journal:  Clin Orthop Relat Res       Date:  1993-07       Impact factor: 4.176

View more
  12 in total

Review 1.  What's New in Geriatric Acetabular Fractures.

Authors:  Ian Hasegawa; Morgan Hasegawa; John P Livingstone; Thomas J K Kane; Lorrin Lee
Journal:  Hawaii J Health Soc Welf       Date:  2022-03

2.  Treatment of anterior column posterior hemitransverse fracture with supra-ilioinguinal approach.

Authors:  Yizhou Wan; Sheng Yao; Kaifang Chen; Lian Zeng; FengZhao Zhu; TingFang Sun; XiaoDong Guo
Journal:  J Int Med Res       Date:  2021-01       Impact factor: 1.671

3.  Internal fixation of anterior acetabular fractures with a limited pararectus approach and the anatomical plates: preliminary results.

Authors:  Sheng Yao; Kaifang Chen; Fengzhao Zhu; Jia Liu; Yulong Wang; Lian Zeng; Yizhou Wan; Yanzhen Qu; Liang Yang; Xiaodong Guo; Xu Yang
Journal:  BMC Musculoskelet Disord       Date:  2021-02-18       Impact factor: 2.362

4.  Anterior intrapelvic approach and suprapectineal quadrilateral surface plate for acetabular fractures with anterior involvement: a retrospective study of 34 patients.

Authors:  Gianluca Ciolli; Domenico De Mauro; Giuseppe Rovere; Amarildo Smakaj; Silvia Marino; Lorenzo Are; Omar El Ezzo; Francesco Liuzza
Journal:  BMC Musculoskelet Disord       Date:  2021-12-30       Impact factor: 2.362

5.  Finite element analysis of the Union Plate in treating elderly acetabular fracture patients.

Authors:  Guixiong Huang; Kaifang Chen; Yulong Wang; Xiaodong Guo
Journal:  J Orthop Surg Res       Date:  2022-01-29       Impact factor: 2.359

6.  Differences in fixation to young and elderly quadrilateral surfaces with anatomic quadrilateral surface plate (AQSP) based on cortical thickness morphological results.

Authors:  Jialiang Guo; Weichong Dong; Yali Zhou; Jinglue Hu; Pengyu Ye; Wei Chen; Yingze Zhang; Zhiyong Hou
Journal:  J Orthop Surg Res       Date:  2022-03-05       Impact factor: 2.359

7.  Primary stability of multi-hole cups compared to plate osteosynthesis in osteoporotic anterior column and posterior hemi-transverse acetabular fractures-A biomechanical comparison.

Authors:  Andreas Höch; Rebekka Reise; Philipp Pieroh; Christoph-Eckhard Heyde; Johannes Karl Maria Fakler; Stefan Schleifenbaum
Journal:  PLoS One       Date:  2022-07-27       Impact factor: 3.752

8.  The novel infra-pectineal buttress plates used for internal fixation of elderly quadrilateral surface involved acetabular fractures.

Authors:  Yizhou Wan; Sheng Yao; Yan Ma; Lian Zeng; Yulong Wang; Yanzhen Qu; Guixiong Huang; Xiaodong Guo; Kaifang Chen
Journal:  Orthop Surg       Date:  2022-06-15       Impact factor: 2.279

9.  Does use of a quadrilateral surface plate improve outcome in elderly acetabular fractures?

Authors:  Graeme Nicol; Ethan Sanders; Allan Liew; Geoffrey Wilkin; Wade T Gofton; Steven Papp; George Grammatopoulos
Journal:  J Clin Orthop Trauma       Date:  2020-10-07

10.  Recommendations of protective measures for orthopedic surgeons during COVID-19 pandemic.

Authors:  Yulong Wang; Lian Zeng; Sheng Yao; Fengzhao Zhu; Chaozong Liu; Anna Di Laura; Johann Henckel; Zengwu Shao; Michael T Hirschmann; Alister Hart; Xiaodong Guo
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-06-10       Impact factor: 4.342

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.