Analytical sensitivity for SARS-CoV-2 detection is a key performance metric for the evaluation of viral detection assays. We determined analytical limits of detection for seven SARS-CoV-2 assays using serial dilutions of pooled patient material quantified with droplet digital PCR. Limits of detection ranged from ≤10 to 74 copies/ml for commercial high-throughput laboratory analyzers (Roche Cobas, Abbott m2000, and Hologic Panther Fusion) and 167 to 511 copies/ml for sample-to-answer (DiaSorin Simplexa, GenMark ePlex) and point-of-care instruments (Abbott ID NOW). The CDC assay yielded limits of detection ranging from 85 to 499 copies/ml, depending on the extraction method and thermocycler used. These results can help to inform the assay choice for testing approaches to manage the current COVID-19 outbreak.Entities:
Keywords: SARS-CoV-2; limit of detection
Mesh:
Substances:
Year: 2020 PMID: 32651238 PMCID: PMC7448668 DOI: 10.1128/JCM.01535-20
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948
Analytical limits of detection for SARS-CoV-2
| Assay | No. of SARS-CoV-2 detected/no. tested at a viral concn (cp/ml) of: | Exp. LOD | Probit LOD | |||||
|---|---|---|---|---|---|---|---|---|
| 3,160 | 1,000 | 316 | 100 | 32 | 10 | |||
| Abbott m2000 | ND | ND | 20/20 | 19/20 | 20/20 | 16/20 | 32 | 53 |
| Roche Cobas | ND | ND | 20/20 | 20/20 | 20/20 | 20/20 | ≤10 | ≤10 |
| Panther Fusion | 5/5 | 5/5 | 5/5 | 20/20 | 15/20 | 1/20 | 100 | 74 |
| DiaSorin Simplexa | 5/5 | 5/5 | 20/20 | 19/20 | 14/20 | ND | 100 | 167 |
| GenMark ePlex | 5/5 | 5/5 | 5/5 | 19/20 | 8/20 | ND | 100 | 190 |
| Abbott ID NOW | 5/5 | 5/5 | 24/25 | 14/25 | 9/20 | 0/5 | 316 | 511 |
| CDC ABI 7500 (MP24) | 5/5 | 5/5 | 20/20 | 15/20 | 1/5 | ND | 316 | 237 |
| CDC LC 480 (MP24) | 5/5 | 5/5 | 19/20 | 14/20 | 1/5 | ND | 316 | 499 |
| CDC ABI 7500 (EZ1) | 5/5 | 5/5 | 20/20 | 20/20 | 13/20 | ND | 100 | 85 |
| CDC LC 480 (EZ1) | 5/5 | 5/5 | 20/20 | 19/20 | 10/20 | ND | 100 | 185 |
Exp. LOD, experimental limit of detection; Probit LOD, Probit limit of detection; ND, not done; ABI 7500, Applied Biosystems 7500 Fast real-time PCR system; LC 480, Roche LightCycler 480 II real-time PCR system; MP24, Roche MagaPure 24 extractor; EZ1, Qiagen EZ1 extractor.
Sample volumes tested for SARS-CoV-2
| Assay | Minimum input vol (μl) | Sample processed (μl) | Elution vol (μl) | Eluate added to reaction (μl) | Effective sample vol(s) tested (μl) |
|---|---|---|---|---|---|
| Abbott m2000 | 760 | 500 | 80 | 40 | 250 |
| Roche Cobas | 600 | 400 | 50 | 50 | 400 |
| Panther Fusion | 500 | 360 | 50 | 5 | 25.4 |
| DiaSorin Simplexa | — | — | — | — | 50 |
| GenMark ePlex | — | — | — | — | 200 |
| Abbott ID NOW | — | — | — | — | 100 (target), 100 (internal control) |
| CDC ABI 7500 (MP24) | 350 | 200 | 100 | 5 | 10 |
| CDC LC 480 (MP24) | 350 | 200 | 100 | 5 | 10 |
| CDC ABI 7500 (EZ1) | 400 | 400 | 90 | 5 | 22.2 |
| CDC LC 480 (EZ1) | 400 | 400 | 90 | 5 | 22.2 |
A 500-μl sample + 710-μl lysis buffer = 1,210 μl input volume.
Sample-to-answer tests.