Direct Comparison of SARS-CoV-2 Analytical Limits of Detection across Seven Molecular Assays.

Becky Fung1, Allan Gopez1,2, Venice Servellita1,2, Shaun Arevalo1,2, Coral Ho1, Anne Deucher1, Ed Thornborrow1, Charles Chiu1,2, Steve Miller3,2.   

Abstract

Analytical sensitivity for SARS-CoV-2 detection is a key performance metric for the evaluation of viral detection assays. We determined analytical limits of detection for seven SARS-CoV-2 assays using serial dilutions of pooled patient material quantified with droplet digital PCR. Limits of detection ranged from ≤10 to 74 copies/ml for commercial high-throughput laboratory analyzers (Roche Cobas, Abbott m2000, and Hologic Panther Fusion) and 167 to 511 copies/ml for sample-to-answer (DiaSorin Simplexa, GenMark ePlex) and point-of-care instruments (Abbott ID NOW). The CDC assay yielded limits of detection ranging from 85 to 499 copies/ml, depending on the extraction method and thermocycler used. These results can help to inform the assay choice for testing approaches to manage the current COVID-19 outbreak.
Copyright © 2020 American Society for Microbiology.

Entities:  

Keywords:  SARS-CoV-2; limit of detection

Mesh:

Substances:

Year:  2020        PMID: 32651238      PMCID: PMC7448668          DOI: 10.1128/JCM.01535-20

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


INTRODUCTION

Molecular testing for SARS-CoV-2 viral shedding is a critical part of the response to the COVID-19 outbreak. These tests are applied for symptomatic patients with expected high levels of virus present in respiratory samples (1) but are increasingly being used for screening of asymptomatic individuals for whom the kinetics of viral shedding are less well documented (2). An accurate understanding of test performance is important to evaluate the utility of molecular viral detection in a variety of clinical settings, and the analytical limit of detection (LOD) is a key characteristic for method comparison. Due to variability in the reference standard material and methods of quantitation used during the evaluation of limits of detection, it has been challenging to compare performance among test methods. This study was performed to directly compare analytical limits of detection using a pool of positive patient material quantified with droplet digital PCR (ddPCR). The use of ddPCR allows for a standard independent assessment of viral load based on the number and volume of droplets measured in a droplet emulsion PCR assay, which has been shown to be a more sensitive approach for SARS-CoV-2 detection in low-viral load samples (3).

MATERIALS AND METHODS

Ten residual nasopharyngeal (NP) swab or combined NP/oropharyngeal (OP) swab raw patient samples containing high-titer SARS-CoV-2 virus in universal transport media (UTM) were combined to form patient pooled material (PPM). Nine of these were initially positive using the CDC assay with LightCycler 480 II amplification, and one was positive using Hologic Panther Fusion assay. The PPM was extracted over 3 dilutions in triplicate using EZ1 Virus 2.0 kit (Qiagen) with 400 μl sample input and 90 μl elution, and each extract was quantified using novel coronavirus (2019 nCoV) digital PCR detection kit (ApexBio Technology, Houston TX) per the manufacturer instructions. The viral concentration in stock PPM was calculated based on the average concentration over the 3 dilutions, assuming 100% extraction efficiency, and was stored at −80°C in individual aliquots to ensure only one freeze-thaw cycle. For testing, stock PPM was diluted with UTM or viral transport media (VTM) in serial 1:10 dilutions to 3.5-log copies/ml (3,160 copies/ml) and then in serial 0.5-log (1:3.16) dilutions to calculated concentrations between 1.0- and 3.5-log copies/ml (10 to 3,160 copies/ml) to span the expected range of concentrations needed to determine LOD. Sufficient volume was prepared for individual testing of at least 20 replicate samples at each concentration, and samples were held refrigerated until they were tested within 72 h. Preliminary LOD was determined with 5 replicates over multiple concentrations, and at least 15 additional replicates were performed at concentrations spanning the level with 95% detection. Experimental LOD was determined as the lowest concentration tested with 95% detection, and Probit LOD was determined using Probit analysis (4). Each commercial assay was performed according to the manufacturer’s instructions as follows: Abbott m2000, RealTime SARS-CoV-2 assay (emergency use authorization [EUA]), high-throughput dual-target assay for RdRp and N genes (Des Plaines, IL). Roche Cobas, SARS-CoV-2 test (EUA), high-throughput dual-target assay for Orf1ab and E genes (Indianapolis, IN); Hologic Panther Fusion, SARS-CoV-2 assay (EUA), high-throughput dual-target assay for ORF1ab region 1 and ORF1ab region 2 (Marlborough, MA); DiaSorin Simplexa, COVID-19 direct kit (EUA), sample-to-answer dual-target assay for S gene and ORF1ab gene (Cypress, CA); GenMark ePlex: SARS-CoV-2 test (EUA), sample-to-answer dual-target assay for two different regions of the N gene (Fremont, CA); and Abbott ID NOW, COVID-19 assay (EUA), sample-to-answer, point-of-care single-target assay for the RdRp gene (Scarborough MA). To mimic the introduction of viral material from respiratory swab samples on the ID NOW platform, quantified samples in UTM were added 1:100 to achieve the stated final concentration in ID NOW elution buffer and immediately tested. This allows for direct comparison between the viral concentration in UTM/VTM and the viral concentration in elution buffer after addition of the swab material. The CDC assay (EUA) is a dual target assay for N1 and N2 genes and requires separate extractions; these were performed using the EZ1 Virus 2.0 kit (Qiagen, Germantown MD) with 400 μl sample input and 90 μl elution volume or the MagNA Pure MP24 kit (Roche) with 200 μl sample input and 100 μl elution volume. CDC assay reagents were obtained from IDT Technologies, and the CDC assay was performed on the ABI 7500 Fast thermocycler (Applied Biosystems, Foster City CA) or the LightCycler 480 II thermocycler (Roche Diagnostics, Indianapolis IN). Results with one of two viral gene targets detected (indeterminate interpretation) were considered positive on the CDC assay.

RESULTS

Results of replicate testing of dilutions of PPM quantified using ddPCR are shown in Table 1. For commonly available SARS-CoV-2 methods, limits of detection determined by Probit analysis ranged from ≤ 10 copies/ml for the Roche Cobas assay to 511 copies/ml for Abbott ID NOW, which is a point-of-care (POC) assay yielding results in 13 min. There was a trend with more rapid sample-to-answer assays yielding slightly higher LOD (167 to 511 copies/ml) compared to high-throughput laboratory analyzers. The CDC assay showed LOD between 85 and 499 copies/ml, depending on the extraction instrument and thermocycler used. The effective sample volume tested for each assay was calculated based on the amount of sample processed, elution volume, and the amount of eluate added to the amplification reaction (Table 2). The Roche Cobas instrument processes 400 μl per sample, with 50 μl elution volume and all 50 μl of that used for PCR, for an effective tested sample volume of 400 μl. Abbott m2000 processes 500 μl sample, with 80 μl elution and 40 μl used for PCR, for an effective tested sample volume of 250 μl. Hologic Panther Fusion processes 360 μl of sample after 500 μl is diluted into 710 μl lysis buffer, with 50 μl elution and 5 μl used for PCR, for an effective tested sample volume of 25.4 μl. For the CDC assay, 5 μl of elution was added to PCR after EZ1 extraction with 400 μl sample input and 90 μl elution volume (effective tested volume of 22.2 μl) or MP24 extraction with 200 μl sample input and 100 μl elution volume (effective tested volume of 10 μl).
TABLE 1

Analytical limits of detection for SARS-CoV-2

AssayNo. of SARS-CoV-2 detected/no. tested at a viral concn (cp/ml) of:
Exp. LODa Probit LOD
3,1601,0003161003210
Abbott m2000NDND20/2019/2020/2016/203253
Roche CobasNDND20/2020/2020/2020/20≤10≤10
Panther Fusion5/55/55/520/2015/201/2010074
DiaSorin Simplexa5/55/520/2019/2014/20ND100167
GenMark ePlex5/55/55/519/208/20ND100190
Abbott ID NOW5/55/524/2514/259/200/5316511
CDC ABI 7500 (MP24)5/55/520/2015/201/5ND316237
CDC LC 480 (MP24)5/55/519/2014/201/5ND316499
CDC ABI 7500 (EZ1)5/55/520/2020/2013/20ND10085
CDC LC 480 (EZ1)5/55/520/2019/2010/20ND100185

Exp. LOD, experimental limit of detection; Probit LOD, Probit limit of detection; ND, not done; ABI 7500, Applied Biosystems 7500 Fast real-time PCR system; LC 480, Roche LightCycler 480 II real-time PCR system; MP24, Roche MagaPure 24 extractor; EZ1, Qiagen EZ1 extractor.

TABLE 2

Sample volumes tested for SARS-CoV-2

AssayMinimum input vol (μl)Sample processed (μl)Elution vol (μl)Eluate added to reaction (μl)Effective sample vol(s) tested (μl)
Abbott m20007605008040250
Roche Cobas6004005050400
Panther Fusion500a 36050525.4
DiaSorin Simplexab 50
GenMark ePlex200
Abbott ID NOW100 (target), 100 (internal control)
CDC ABI 7500 (MP24)350200100510
CDC LC 480 (MP24)350200100510
CDC ABI 7500 (EZ1)40040090522.2
CDC LC 480 (EZ1)40040090522.2

A 500-μl sample + 710-μl lysis buffer = 1,210 μl input volume.

Sample-to-answer tests.

Analytical limits of detection for SARS-CoV-2 Exp. LOD, experimental limit of detection; Probit LOD, Probit limit of detection; ND, not done; ABI 7500, Applied Biosystems 7500 Fast real-time PCR system; LC 480, Roche LightCycler 480 II real-time PCR system; MP24, Roche MagaPure 24 extractor; EZ1, Qiagen EZ1 extractor. Sample volumes tested for SARS-CoV-2 A 500-μl sample + 710-μl lysis buffer = 1,210 μl input volume. Sample-to-answer tests.

DISCUSSION

Accurate assessment of analytical sensitivity for SARS-CoV-2 detection is needed to assess the comparative ability of various assays to diagnose infection and to understand the relative risk for infection in patients with negative test results. Previous comparisons have shown varied results, with laboratory-based methods generally performing better than sample-to-answer and point-of-care methods (5–8). The standard material used and methods of calibration differ for most available studies, making it difficult to assess the true limits of detection in patient samples. Here, we used the standard independent calibration method of ddPCR to quantify a pool of positive patient material and assess limits of detection across seven different assays. The inclusion of multiple positive patient samples in the pool reduces but does not eliminate the possibility of viral mutations present that might adversely affect specific assay detection. Our initial quantification method used extraction prior to ddPCR quantification, assuming 100% extraction efficiency to calculate the viral concentration in the original pooled patient sample. If the extraction efficiency were lower than 100%, the actual viral concentration would be somewhat higher than the calculated value used in this study. Overall assay sensitivity appears to be primarily related to the effective sample volume tested in the amplification reaction with some influence from assay design and detection instrumentation. The effective volume tested is dependent on the sample input volume and extraction and reaction parameters and can be directly compared for assays having separate extraction and amplification reactions. This relationship is less consistent for sample-to-answer platforms, where the nucleic acid purification, processing, and chemistry used have a substantial effect on assay LOD. The CDC SARS-CoV-2 assay is a commonly used method, but there are several variations possible, including the extraction method and real-time thermocycler used. We show that the most sensitive detection is achieved using EZ1 extracted material with the ABI 7500 Fast thermocycler, with slight decreases in sensitivity using the Roche MP24 extractor or the Roche LightCycler 480 II thermocycler. Other modifications may have somewhat varying performance. While the high-throughput laboratory-based instruments (Roche Cobas, Abbott m2000, and Hologic Panther Fusion) achieved the best analytical sensitivity, other sample-to-answer methods (DiaSorin Simplexa and GenMark ePlex) were only approximately 2- to 4-fold less sensitive, indicating good performance with more rapid results. In this study, the Abbott ID NOW POC instrument achieved a sensitivity of approximately 511 copies/ml, 1 log lower than high-throughput laboratory instruments. This can be compared to other studies showing 20,000 copies/ml using Exact Diagnostics synthetic RNA quantified control with a sample dilution protocol (9) and 125 genome equivalents/ml based on the manufacturer package insert (10). Here, we mimicked the use of a dry swab put directly into elution buffer, eliminating the dilution from using VTM in comparison studies (a method that is no longer recommended in the ID NOW package insert). Analytical sensitivity for virus detection is a significant factor in the choice of assay to use but is not the only consideration. This study did not evaluate the impact of specimen matrix on the ability to detect virus or compatibility with different media types, which may be particularly relevant for methods that bypass the extraction step and permit analysis directly from original swab samples (e.g., Abbott ID NOW system). Our findings are not generalizable to other sample types, including lower respiratory sources. Using pooled patient material quantified with ddPCR, it is reassuring that measured limits of detection were consistent in the tens of copies per milliliter for high-throughput laboratory instruments and in the hundreds of copies per milliliter for sample-to-answer and POC instruments. Less sensitive methods will fail to detect virus in some cases, primarily during the long tail of viral shedding seen in some patients. However, this may be of lower clinical significance since many of these patients would likely have developed detectable antibody responses by this time, and samples with quantitative PCR (qPCR) cycle threshold (C) values > 33 tend not to be recoverable by viral culture, indicating a potential lack of infectivity (11). Further studies are needed to clarify the relationship of viral detection to transmissibility, which will help guide the choice of testing strategies.
  10 in total

Review 1.  Validation of laboratory-developed molecular assays for infectious diseases.

Authors:  Eileen M Burd
Journal:  Clin Microbiol Rev       Date:  2010-07       Impact factor: 26.132

2.  Comparison of Abbott ID Now and Abbott m2000 Methods for the Detection of SARS-CoV-2 from Nasopharyngeal and Nasal Swabs from Symptomatic Patients.

Authors:  Amanda Harrington; Brian Cox; Jennifer Snowdon; Jonathan Bakst; Erin Ley; Patricia Grajales; Jack Maggiore; Stephen Kahn
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

3.  Evidence Supporting Transmission of Severe Acute Respiratory Syndrome Coronavirus 2 While Presymptomatic or Asymptomatic.

Authors:  Nathan W Furukawa; John T Brooks; Jeremy Sobel
Journal:  Emerg Infect Dis       Date:  2020-06-21       Impact factor: 6.883

4.  Comparison of Two High-Throughput Reverse Transcription-PCR Systems for the Detection of Severe Acute Respiratory Syndrome Coronavirus 2.

Authors:  Arryn R Craney; Priya D Velu; Michael J Satlin; Kathy A Fauntleroy; Katrina Callan; Amy Robertson; Marisa La Spina; Beryl Lei; Anqi Chen; Tricia Alston; Anna Rozman; Massimo Loda; Hanna Rennert; Melissa Cushing; Lars F Westblade
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

5.  Comparison of Four Molecular In Vitro Diagnostic Assays for the Detection of SARS-CoV-2 in Nasopharyngeal Specimens.

Authors:  Wei Zhen; Ryhana Manji; Elizabeth Smith; Gregory J Berry
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

6.  Comparison of Commercially Available and Laboratory-Developed Assays for In Vitro Detection of SARS-CoV-2 in Clinical Laboratories.

Authors:  Joshua A Lieberman; Gregory Pepper; Samia N Naccache; Meei-Li Huang; Keith R Jerome; Alexander L Greninger
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

7.  Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards.

Authors:  Bernard La Scola; Marion Le Bideau; Julien Andreani; Van Thuan Hoang; Clio Grimaldier; Philippe Colson; Philippe Gautret; Didier Raoult
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2020-04-27       Impact factor: 3.267

8.  SARS-CoV-2 Viral Load in Upper Respiratory Specimens of Infected Patients.

Authors:  Lirong Zou; Feng Ruan; Mingxing Huang; Lijun Liang; Huitao Huang; Zhongsi Hong; Jianxiang Yu; Min Kang; Yingchao Song; Jinyu Xia; Qianfang Guo; Tie Song; Jianfeng He; Hui-Ling Yen; Malik Peiris; Jie Wu
Journal:  N Engl J Med       Date:  2020-02-19       Impact factor: 91.245

9.  ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens.

Authors:  Tao Suo; Xinjin Liu; Jiangpeng Feng; Ming Guo; Wenjia Hu; Dong Guo; Hafiz Ullah; Yang Yang; Qiuhan Zhang; Xin Wang; Muhanmmad Sajid; Zhixiang Huang; Liping Deng; Tielong Chen; Fang Liu; Ke Xu; Yuan Liu; Qi Zhang; Yingle Liu; Yong Xiong; Guozhong Chen; Ke Lan; Yu Chen
Journal:  Emerg Microbes Infect       Date:  2020-12       Impact factor: 7.163

10.  Clinical Evaluation of Three Sample-to-Answer Platforms for Detection of SARS-CoV-2.

Authors:  Wei Zhen; Elizabeth Smith; Ryhana Manji; Deborah Schron; Gregory J Berry
Journal:  J Clin Microbiol       Date:  2020-07-23       Impact factor: 5.948

  10 in total
  35 in total

1.  Rapid and Sensitive Detection of Antigen from SARS-CoV-2 Variants of Concern by a Multivalent Minibinder-Functionalized Nanomechanical Sensor.

Authors:  Dilip Kumar Agarwal; Andrew C Hunt; Gajendra S Shekhawat; Lauren Carter; Sidney Chan; Kejia Wu; Longxing Cao; David Baker; Ramon Lorenzo-Redondo; Egon A Ozer; Lacy M Simons; Judd F Hultquist; Michael C Jewett; Vinayak P Dravid
Journal:  Anal Chem       Date:  2022-06-02       Impact factor: 8.008

2.  Ultrasensitive detection of SARS-CoV-2 RNA and antigen using single-molecule optofluidic chip.

Authors:  G G Meena; A M Stambaugh; V Ganjalizadeh; M A Stott; A R Hawkins; H Schmidt
Journal:  APL Photonics       Date:  2021-06-01

3.  Evaluating the Ability to ID (COVID-19) NOW: a Large Real-World Prospective Evaluation of the Abbott ID NOW COVID-19 Assay.

Authors:  K R Barker; L N Small; D V Thai; K Y Sohn; L C Rosella
Journal:  Microbiol Spectr       Date:  2022-05-17

4.  Revisiting the guidelines for ending isolation for COVID-19 patients.

Authors:  Yong Dam Jeong; Keisuke Ejima; Kwang Su Kim; Shoya Iwanami; Ana I Bento; Yasuhisa Fujita; Il Hyo Jung; Kazuyuki Aihara; Koichi Watashi; Taiga Miyazaki; Takaji Wakita; Shingo Iwami; Marco Ajelli
Journal:  Elife       Date:  2021-07-27       Impact factor: 8.140

Review 5.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

6.  A Multiplex One-Step RT-qPCR Protocol to Detect SARS-CoV-2 in NP/OP Swabs and Saliva.

Authors:  Joshua Bland; Ashley Kavanaugh; Lenny K Hong; Omar Perez; Shrihari S Kadkol
Journal:  Curr Protoc       Date:  2021-05

Review 7.  Applications of digital PCR in COVID-19 pandemic.

Authors:  Chianru Tan; Dongdong Fan; Nan Wang; Fang Wang; Bo Wang; Lingxiang Zhu; Yong Guo
Journal:  View (Beijing)       Date:  2021-01-01

8.  ddPCR increases detection of SARS-CoV-2 RNA in patients with low viral loads.

Authors:  Agnès Marchio; Christophe Batejat; Jessica Vanhomwegen; Maxence Feher; Quentin Grassin; Maxime Chazal; Olivia Raulin; Anne Farges-Berth; Florence Reibel; Vincent Estève; Anne Dejean; Nolwenn Jouvenet; Jean-Claude Manuguerra; Pascal Pineau
Journal:  Arch Virol       Date:  2021-07-12       Impact factor: 2.574

9.  Detection of SARS-CoV-2 at the point of care.

Authors:  Michael J Loeffelholz; Yi-Wei Tang
Journal:  Bioanalysis       Date:  2021-07-22       Impact factor: 2.681

10.  Performance evaluation of Panther Fusion SARS-CoV-2 assay for detection of SARS-CoV-2 from deep throat saliva, nasopharyngeal, and lower-respiratory-tract specimens.

Authors:  River Chun-Wai Wong; Ann Han Wong; Yolanda Iok-Ieng Ho; Eddie Chi-Man Leung; Raymond Wai-Man Lai
Journal:  J Med Virol       Date:  2020-10-08       Impact factor: 20.693

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.