| Literature DB >> 32637377 |
Karbalaee M1, Shahbazi-Gahrouei D1, Tavakoli M B1.
Abstract
BACKGROUND: An accurate and fast radiation dose calculations method is the main part of treatment planning for successful radiation therapy.Entities:
Keywords: DOSXYZnrc; Dosimetry; Parallel Processing; Radiation Dosimeters; Radiotherapy Treatment Planning; Monte Carlo Method
Year: 2020 PMID: 32637377 PMCID: PMC7321392 DOI: 10.31661/jbpe.v0i0.716
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure 1Setup of the experimental measurement including the linac head, Mapcheck dosimeter and the designed and produced rotary phantom.
Figure 2Results of MCPDC simulation with fully tuned energy spectrum compared to DOSXYZnrc and experimental measurements in the water phantom.
Figure 3PDD of MCPDC and DOSXYZnrc in the water virtual phantoms with a layer of bone (a) and lung (b) heterogeneity.
Figure 4Dose distribution of MCPDC and DOSXYZnrc in anthropomorphic phantom with a layer of bone (a) and lung (b) heterogeneity.
The results of statistical analysis of dose discrepancies between MCPDC and DOSXYZnrc at different depths of phantom with lung (a) and bone (b) heterogeneity.
| a | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Discrepancy test in percent | Gamma evaluation test | ||||||||
| Depth(cm) | 0-1 | 1-2 | 2-3 | 3-5 | γ2mm 2% | γ3mm 3% | γ3mm 5% | γ5mm 5% | |
| Ascending part before build up region | 0 - 1.4 | 82.7 | 16.3 | 0.9 | 0.1 | 80.4 | 97.6 | 98.8 | 99.9 |
| Buildup region | 1.4 - 1.6 | 85.7 | 14.2 | 0.1 | 0.0 | 80.4 | 97.7 | 98.9 | 99.9 |
| Descending part before heterogeneity | 1.6 - 4 | 82.3 | 16.7 | 1.0 | 0.0 | 78.2 | 97.4 | 98.7 | 99.8 |
| Within the heterogeneity | 4 - 12 | 53.3 | 36.7 | 9.2 | 0.8 | 70.4 | 96.1 | 98.0 | 99.7 |
| Beneath the heterogeneity | 12 - 18 | 22.7 | 42.2 | 27.7 | 7.4 | 70.2 | 96.4 | 98.2 | 99.7 |
| Depth(cm) | 0-1 | 1-2 | 2-3 | 3-5 | γ2mm 2% | γ3mm 3% | γ3mm 5% | γ5mm 5% | |
| Ascending part before build up region | 0 - 1.4 | 40.4 | 49.8 | 8.3 | 1.5 | 80.3 | 97.7 | 98.9 | 99.9 |
| Buildup region | 1.4-1.6 | 46.1 | 49.2 | 4.7 | 0.0 | 80.1 | 97.7 | 98.9 | 99.9 |
| Descending part before heterogeneity | 1.6-4 | 50.7 | 45.6 | 3.7 | 0.0 | 79.5 | 97.6 | 98.8 | 99.8 |
| Within the heterogeneity | 4-12 | 75.6 | 23.0 | 1.4 | 0.0 | 76.4 | 97.3 | 98.6 | 99.8 |
| Beneath the heterogeneity | 12-18 | 62.4 | 30.5 | 6.4 | 0.7 | 67.9 | 96.1 | 98.0 | 99.7 |
Indicating 0 to 1% discrepancy between MCPDC and DOSXYZnrc
Demonstrating 0 to 1% discrepancy, for 82.7% of the points at depth of less than 1.4 cm
Figure 5Dose distribution of MCPDC and DOSXYZnrc on a CT slice of the anthropomorphic phantom.
Figure 6Results of MCPDC simulations and experimental measurements using Mapcheck dosimeter in different views, considering the phantom rotations at 0° (a), 90° (b) and 180° (c) of the produced rotary phantom.
Results of statistical analysis of dose discrepancies between MCPDC results and Mapcheck dose measurements in different steps of the rotation.
| Discrepancy test in percent | Gamma evaluation test | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Beam angle (Degree) | 0-1 | 1-2 | 2-3 | 3-5 | >5 | γ2mm 2% | γ3mm 3% | γ3mm 5% | γ5mm 5% |
| 0 | 29.8 | 53.2 | 12.7 | 2.9 | 1.4 | 70.3 | 97.6 | 99.5 | 99.8 |
| 90 | 43.5 | 43.7 | 7.2 | 4.1 | 1.5 | 80.3 | 98.6 | 99.7 | 99.9 |
| 180 | 30.1 | 54.1 | 5.3 | 7.3 | 1.7 | 72.5 | 98.3 | 99.1 | 99.8 |