Yun Jeong Chae1, Heirim Lee2, Bokyeong Jun1, In Kyong Yi3.
Abstract
BACKGROUND: After prehospital insertion of i-gel, a popular supraglottic airway (SGA), fiberoptic-guided intubation through i-gel is often required to switch the i-gel to a definitive airway for anticipated difficult airway. The Aintree intubation catheter (AIC) was developed for this purpose yet it requires many procedural steps during which maintenance of adequate ventilation is difficult. We custom-made a long endotracheal tube (LET) which may facilitate this procedure and compared the efficacy of the AIC and LET in a cervical immobilized manikin.Entities:
Keywords: Airway management; Fiberoptic; Intratracheal intubation; Manikin; Supraglottic airway device
Mesh:
Year: 2020 PMID: 32552828 PMCID: PMC7302158 DOI: 10.1186/s12871-020-01069-9
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1a Fiberoptic-guided LET insertion, (b) Removal of i-gel, (c) Position of cuff pilot balloon during removal of i-gel, (d) Adjustment of length of the LET. Abbreviation: LET, long endotracheal tube. Abbreviations: LET, long endotracheal tube
Fig. 2Consort diagram. Abbreviations: AIC, Aintree intubation catheter; LET, long endotracheal tube
Demographics
| No. | Position | Sex | Age | Experiences | |||
|---|---|---|---|---|---|---|---|
| Anesthesiology | Fiberoptic intubation (times) | AIC | LET | ||||
| 1 | Resident | M | 30 | 3 | 20 | 0 | 0 |
| 2 | Resident | F | 35 | 4 | 20 | 0 | 0 |
| 3 | Fellow | M | 37 | 5 | 16 | 0 | 0 |
| 4 | Resident | M | 28 | 1 | 0 | 0 | 0 |
| 5 | Resident | F | 31 | 1 | 0 | 0 | 0 |
| 6 | Fellow | M | 34 | 5 | 20 | 0 | 0 |
| 7 | Resident | M | 32 | 1 | 0 | 0 | 0 |
| 8 | Resident | F | 31 | 2 | 10 | 0 | 0 |
| 9 | Resident | F | 30 | 3 | 15 | 0 | 0 |
| 10 | Resident | F | 33 | 4 | 20 | 0 | 0 |
| 11 | Resident | F | 33 | 4 | 10 | 0 | 0 |
| 12 | Resident | F | 28 | 3 | 15 | 0 | 0 |
| 13 | Fellow | F | 32 | 5 | 20 | 0 | 0 |
| 14 | Resident | F | 32 | 2 | 15 | 0 | 0 |
| 15 | Resident | F | 32 | 2 | 23 | 0 | 0 |
| 16 | Attending | M | 36 | 10 | 50 | 0 | 0 |
| 17 | Attending | F | 35 | 9 | 20 | 0 | 0 |
| 18 | Fellow | F | 34 | 5 | 15 | 0 | 0 |
| 19 | Fellow | M | 36 | 5 | 3 | 0 | 0 |
| 20 | Fellow | M | 35 | 6 | 25 | 0 | 0 |
Abbreviations: AIC Aintree intubation catheter, LET long endotracheal tube
Fig. 3Intubation time. AIC and LET resulted in similar durations for fiberoptic time, however, LET resulted in shorter duration for procedure time. Total intubation time was shorter in LET. Abbreviations: AIC, Aintree intubation catheter; LET, long endotracheal tube.
Intubation profile
| Parameter | AIC ( | LET ( | |
|---|---|---|---|
| Total intubation time, sec | 94.0 ± 28.4 | 70.8 ± 16.4 | 0.001 |
| Fiberoptic time, sec | 17.5 ± 12.3 | 19.0 ± 6.9 | 0.612 |
| Procedure time, sec | 76.5 ± 25.4 | 51.9 ± 13.8 | < 0.001 |
| Success, | 19 (95) | 20 (100) | 1.000 |
| Ease of insertion | 1.0 [1.0–1.0] | 2.0 [1.0–2.75] | < 0.001 |
| Difficulty | 4.0 [3.0–5.75] | 4.0 [3.0–6.0] | 1.000 |
Values are mean ± standard deviation or median [interquartile range] or number (%)
Abbreviations: AIC Aintree intubation catheter, LET long endotracheal tube