Torsten Straßer1, Sandra Wagner1, Eberhart Zrenner1,2. 1. Centre for Ophthalmology, Institute for Ophthalmic Research, University of Tuebingen, Tuebingen, Germany. 2. Werner Reichardt Centre for Integrative Neuroscience (CIN), University of Tuebingen, Tuebingen, Germany.
Abstract
Presbyopia and myopia research shows a growing interest in ciliary muscle biometry using optical coherence tomography (OCT). Until now, segmentation of the ciliary muscle is often performed manually using either custom-developed programs or image processing software. Here we present a novel software for semi-automatic segmentation of the ciliary muscle. It provides direct import of OCT images in DICOM format, a standardized procedure for segmentation, image distortion correction, the export of anatomical ciliary muscle landmarks, like ciliary muscle apex and scleral spur, as well as a continuous thickness profile of the ciliary muscle as a novel way of analysis. All processing steps are stored as XML files, fostering documentation and reproducibility of research through the possibility of replicating the analysis. Additionally, CilOCT supports batch processing for the automated analysis of large numbers of images and the respective data export to tabulated text files based on the stored XML files. CilOCT was successfully applied in several studies and their results will be summarized in this paper.
Presbyopia and myopia research shows a growing interest in ciliary muscle biometry using optical coherence tomography (OCT). Until now, segmentation of the ciliary muscle is often performed manually using either custom-developed programs or image processing software. Here we present a novel software for semi-automatic segmentation of the ciliary muscle. It provides direct import of OCT images in DICOM format, a standardized procedure for segmentation, image distortion correction, the export of anatomical ciliary muscle landmarks, like ciliary muscle apex and scleral spur, as well as a continuous thickness profile of the ciliary muscle as a novel way of analysis. All processing steps are stored as XML files, fostering documentation and reproducibility of research through the possibility of replicating the analysis. Additionally, CilOCT supports batch processing for the automated analysis of large numbers of images and the respective data export to tabulated text files based on the stored XML files. CilOCT was successfully applied in several studies and their results will be summarized in this paper.
Accommodation, the eye’s adjustment to different viewing distances, is achieved by a contraction of the ciliary muscle, a smooth, ring-shaped muscle surrounding the crystalline lens, controlling its shape to keep the retinal image in focus. Due to a worldwide increase in myopia prevalence [1] and innovative approaches for presbyopia correction [2], research on a deeper understanding of the ciliary muscle’s morphology and mechanisms has increased during the last years. Optical coherence tomography (OCT) has been frequently used to image the ciliary muscle morphology and its changes during accommodation [3-8].To date, the shape changes of the ciliary muscle during accommodation were analyzed by measuring the ciliary muscle thickness (CMT) either in equidistant steps posterior to the scleral spur [3,5] or proportionally to the length of the muscle [9]. Alternatively, the change of the ciliary muscle’s area during accommodation is measured [8,10,11]. Both approaches require the segmentation of the ciliary muscle, which is usually performed manually, either using built-in calipers of the OCT-system [7,8], custom-developed programs [9,12-14], or image processing software [10,11]. The diversity of tools and methods for the segmentation of the ciliary muscle as well as of the analyzed parameters results in difficulties in comparing outcomes of different studies. Furthermore, custom-developed programs used for segmentation are often not made available, which impedes the reproduction of study results.Here we present the first software for semi-automatic segmentation and standardized analysis of the ciliary muscle, CilOCT, which is free to use and fully open-source. Biometric parameters of the ciliary muscle are automatically calculated after segmentation and image distortion correction, accounting for the different refractive indices of the tissue layers. Additionally, CilOCT performs the formation of continuous CMT profiles, which allow for the detailed comparison of the ciliary muscle morphology between different subjects or different accommodative states. CilOCT also provides an audit trail documenting all processing steps, fostering reproducibility of research. The application of CilOCT is demonstrated by summarizing the results of three of our previously published studies.By providing this open-source software, we aim to lay the foundation for an improved comparability of studies in the field of accommodation and ciliary muscle research using OCT, thereby fostering a better understanding of both anatomy and functionality of the ciliary muscle and its role in myopia and presbyopia development.
Material and methods
Software architecture
CilOCT was developed as a Java (Version 8) application with a graphical user interface based on JavaFX and uses several open-source libraries. The user interface leverages the ribbon-paradigm, a graphical control element consisting of a set of toolbars placed on several tabs, first introduced in Microsoft Office 2007, using the open-source library FXRibbon (version 1.0.1, BSD license) to simplify and accelerate user interaction [15]. The image import makes use of the standard Java ImageIO classes for common image file formats (BMP, JPEG, PNG) and dcm4che-imageio (version 3.3.8, MPL 1.1 license) for raw DICOM files of the Zeiss Visante™ anterior segment-OCT (AS-OCT). Mathematical calculations, especially matrix algebra and polynomial spline fitting, are implemented using the Apache commons-math3 library (version 3.6.1, Apache license 2.0). Im- and export of guiding landmark coordinates to XML leverages jackson-databind (version 2.9.7, Apache license 2.0). CilOCT uses Apache Maven 3 as project management and build-automation system [16].
Software functionalities
CilOCT enables the semi-automatic segmentation of the ciliary muscle in images acquired using OCT. It allows OCT images to be imported from raw DICOM images of the Zeiss Visante™ AS-OCT [17] or common image formats (i.e. BMP, JPEG, PNG). DICOM images of the Zeiss Visante™ AS-OCT are automatically resized and rotated for the correct display [14]. The software can easily be extended for other OCT instruments. After importing the image, its contrast can either be manually increased or decreased or adjusted by adaptive histogram equalization [18] to enhance the edges of the ciliary muscle for easier placement of guiding landmarks along the muscle’s boundaries. Changing the contrast does not affect the following automatic segmentation. The semi-automatic segmentation using the manually placed guiding landmarks results in polynomial spline fits of the borders of the ciliary muscle, the anterior chamber, the scleral-air boundary, as well as the surface of a possibly used contact lens. The coordinates of the guiding landmarks, along with any processing performed to enhance image quality, can be stored as an XML file, which allows for reliable reproduction of the segmentation results. Before the determination of ciliary muscle biometric parameters, the image distortion resulting from the refraction of the infrared light beam on boundaries with different refractive indices is corrected to enable the calculation of the true layer thicknesses. Image distortion correction is performed and applied to the segmentation using ray-tracing based on Snell’s Law and the refractive indices of the respective layers (sclera: 1.41, ciliary muscle: 1.38, aqueous fluid: 1.34 [19]). The values of the refractive indices of the different layers are configurable. Finally, based on the distortion corrected polynomial splines, the following biometric parameters are calculated: ciliary muscle apex, scleral spur, perpendicular axis (a selective anterior CMT reading taken at the apex), ciliary muscle apex shift, ciliary muscle area, and CMT profile [20]. The determined parameters can be copied to the system clipboard or saved as character separated text files. The batch-processing feature allows for the analysis of large numbers of images based on previously created and stored XML segmentation files and to subsequently export the anatomical ciliary muscle landmarks, determined parameters, and CMT profiles as tabulated text files.
Imaging of the ciliary muscle using OCT
In our previous studies, a Zeiss Visante™ AS-OCT (Carl Zeiss Meditec AG, Jena, Germany; enhanced high-resolution corneal mode, 512 A-scans, scanning size: 10 mm x 3 mm, wavelength 1310 nm) was used. To image the ciliary muscle, subjects were positioned in front of the OCT device with their head placed in the combined head and chin rest. After adjusting the imaged area using pupil center alignment, the subjects were instructed to shift their gaze about 40° to an external target without moving their head but keeping it parallel to the instrument. This allowed imaging the entire ciliary muscle as shown in Fig 1. A detailed description of the process including a sketch of the experimental setup can be found in [20].
Fig 1
OCT image of the temporal ciliary muscle of the right eye.
The ciliary muscle appears as a dark shaded bent triangular structure below the sclera. The image was acquired using a Zeiss VisanteTM anterior segment-OCT, covering a size of 10 mm length and 3 mm tissue depth.
OCT image of the temporal ciliary muscle of the right eye.
The ciliary muscle appears as a dark shaded bent triangular structure below the sclera. The image was acquired using a Zeiss VisanteTM anterior segment-OCT, covering a size of 10 mm length and 3 mm tissue depth.
Segmentation of the ciliary muscle using CilOCT
The ciliary muscle segmentation consists of a two-stage process:Guiding landmarks (about 10–15 per boundary) are placed manually roughly along the borders between (1) sclera and muscle, (2) muscle and pigmented epithelium, as well as (3) along the muscle boundary towards the iridocorneal angle (Fig 2b). For each of the borders, a polynomial spline is fitted through the coordinates of these landmarks.
Fig 2
An illustrative example of the segmentation of the ciliary muscle: (a) The imported raw DICOM OCT image. (b) Manually placed guiding landmarks for the upper (green circles), lower (blue circles), and the nasal or temporal border (yellow circles) of the ciliary muscle, and a landmark in the iridocorneal angle (purple circle) as starting point of the 8-neighbor flood-fill algorithm (guiding landmarks size is enlarged for better visibility). (c) Maximum brightness gradients determined from the first rough polynomial spline fit of the landmarks. (d) Polynomial splines of the final segmentation. (e) Distortion correction based on Snell’s Law using the refractive indices of the different tissue layers. The incoming light rays of the infrared light beam and how they are refracted at the boundary between air and scleral-conjunctival tissue are exemplified as yellow lines perpendicular to the top of the images. Red and orange lines depict the refracted light rays. (f) Distortion corrected segmentation and determined biometric parameters scleral spur, ciliary muscle apex, perpendicular axis, and ciliary muscle area (all values in pixels).
For each pixel of the fitted polynomial splines, the coordinates of the maximum brightness gradient within a distance of ± 5 pixels perpendicular to the spline are determined (Fig 2c) and used for a second, exact polynomial spline fit (Fig 2d).An illustrative example of the segmentation of the ciliary muscle: (a) The imported raw DICOM OCT image. (b) Manually placed guiding landmarks for the upper (green circles), lower (blue circles), and the nasal or temporal border (yellow circles) of the ciliary muscle, and a landmark in the iridocorneal angle (purple circle) as starting point of the 8-neighbor flood-fill algorithm (guiding landmarks size is enlarged for better visibility). (c) Maximum brightness gradients determined from the first rough polynomial spline fit of the landmarks. (d) Polynomial splines of the final segmentation. (e) Distortion correction based on Snell’s Law using the refractive indices of the different tissue layers. The incoming light rays of the infrared light beam and how they are refracted at the boundary between air and scleral-conjunctival tissue are exemplified as yellow lines perpendicular to the top of the images. Red and orange lines depict the refracted light rays. (f) Distortion corrected segmentation and determined biometric parameters scleral spur, ciliary muscle apex, perpendicular axis, and ciliary muscle area (all values in pixels).In addition to the ciliary muscle borders, the scleral-air boundary, as well as the borders of the anterior chamber are segmented by a simple 8-neighbors flood-fill algorithm using a manually placed starting point (purple landmark in Fig 2b, turquoise area in Fig 2c, orange and yellow borders in Fig 2d). These borders are required for the following distortion correction using Snell’s Law and the refractive indices of the different tissues (infrared light rays depicted as yellow lines perpendicular to the upper border of the image and as red/orange lines after refraction at the different tissue borders in Fig 2e). Similarly, a possibly worn contact lens can be segmented (see S1 Movie for an example). The thresholds of the flood-fill algorithm can be manually adjusted for the different regions. Fig 2f depicts the determined biometric parameters scleral spur, ciliary muscle apex, perpendicular axis, and ciliary muscle area after distortion correction.In addition to the biometric parameters ciliary muscle apex, scleral spur, perpendicular axis, ciliary muscle apex shift, and ciliary muscle area, the CMT profile is calculated as the distance between the intersections of a secant perpendicular to the scleral-conjunctival surface with the scleral-muscle border and the muscle-pigmented epithelium border, respectively (Fig 3). This approach is used to normalize different shapes of the scleral-conjunctival surface [20].
Fig 3
Ciliary muscle thickness profile.
The profile is calculated as the distance between the intersection of a secant perpendicular to the scleral-conjunctival surface with the scleral-muscle border and the muscle-pigmented epithelium border (red lines).
Ciliary muscle thickness profile.
The profile is calculated as the distance between the intersection of a secant perpendicular to the scleral-conjunctival surface with the scleral-muscle border and the muscle-pigmented epithelium border (red lines).A demonstration of the segmentation process using the open-source software CilOCT is available as a movie in the supporting information (S1 Movie). To test the segmentation of the ciliary muscle using the CilOCT software, the OCT image file in DICOM format (S1 File) along with the corresponding guiding landmarks file (S2 File) can be used.
Application of CilOCT in previous studies
Images and results reported here were collected in previously published studies [20-22] in which CilOCT was applied. The studies followed the tenets of the declaration of Helsinki and were approved by the Institutional Review Board of the medical faculty of the University of Tuebingen (376/2017BO2).
Results
In a first study, the reliability of the ciliary muscle segmentation performed with CilOCT was evaluated by analyzing the OCT images of fifteen near-emmetropic volunteers. Intra- and inter-examiner as well as intra-session repeatability using the two parameters ciliary muscle area (CMA) and perpendicular axis (PA) (Fig 3) were assessed [20]. Six OCT images of the right temporal ciliary muscle were taken in each of the volunteers during both, near and far accommodation. A short break followed by subject re-alignment after three images in both accommodation states, respectively, allowed for the intra-session repeatability analysis. The ciliary muscle segmentation was performed twice by the same examiner (SW) within one week to assess intra-examiner repeatability and additionally by a second examiner (TS) once to evaluate inter-examiner repeatability. The intra-examiner analysis of the CMA revealed a statistically significant mean difference of 0.06 mm² between the two successively performed segmentations, probably caused by a training effect of the examiner [20]. However, the analysis of the PA showed that the resulting mean difference between the two segmentations was 22.97 μm (≙ 2.94 px) and therefore within the range of the tomographer’s resolution [20]. The intra-session repeatability, analyzed using the mean value of the CMA of the first and the last three images segmented by one examiner, was found to be good with an ICC of 0.88 and a paired t-test revealed no statistically significant difference [20]. A comparison between the CMA derived from the segmentation of the two examiners showed no statistically significant difference and inter-examiner repeatability was found to be good with an ICC of 0.87 [20].The analysis of the CMT profiles as a novel parameter resulted in the identification of regions of a posterior thickening and an anterior thinning of the ciliary muscle when accommodating by 3 D. This could not be observed in such detail using selective thickness measurements based on equidistant or proportional steps as done before [20].Referring to the accumulating evidence of an association between near vision and myopic development, the second study using CilOCT aimed at comparing the accommodation process of emmetropic (n = 18) and myopic (n = 20) young adults regarding both refractive power changes of the crystalline lens and CMT changes for different accommodation demands. The OCT image analysis revealed that accommodation mechanics of the two refractive groups differ significantly with respect to the ciliary muscle’s shape, movement, and thickness changes. The myopic ciliary muscle was found to be thinner than the emmetropic one in the anterior region, but thicker posteriorly. Moreover, myopes exhibited smaller anterior thickness changes of the ciliary muscle during accommodation, but they showed a larger muscle movement compared to their emmetropic counterparts. On the contrary, refractive power changes of the crystalline lens during a dynamic target presentation of the pattern far-near-far were not found to be statistically significantly different between emmetropes and myopes [21].In this context, a third study was performed to investigate the effect of prolonged nearwork on the ciliary muscle and the refractive power changes of the lens. OCT imaging of the ciliary muscle as well as photorefraction recordings of accommodation for a demand of 4 D were performed in 17 emmetropes and 18 myopes before and after a continuous 30-min reading task at 25 cm distance. The analysis of the changes of the CMT profiles revealed an unexpected thinning of the muscle after the nearwork task in both refractive groups. However, only myopes showed a sustained relative myopic shift of their lens power after the prolonged close work, while it remained rather stable in emmetropes [22].These previous studies using CilOCT gave new insights into the myopic ciliary muscle’s anatomical structure and reactions during regular and sustained accommodation and the respective differences to the emmetropic muscle. The findings could therefore be instrumental in the further assessment of a possible role of the ciliary muscle in myopigenesis.
CilOCT availability
The source code of CilOCT is available in a GitHub repository at https://github.com/strator1/ciloct and is licensed under the open-source GNU General Public License 3 (GNU GPLv3) to ensure that any derivative work is kept as open-source.A pre-built version of the software can be downloaded from https://github.com/strator1/CilOCT/releases/tag/2.0.6. The software runs on any operating system with compatible Java Runtime Environment (JRE, version ≥ 8) installed (i.a. Windows, macOS, Linux).
Discussion
The open-source software CilOCT allows for a semi-automatic segmentation and a standardized analysis of the ciliary muscle with an audit trail documenting the processing steps for reproducible results. Reproducible research requires, as stated in a recent article, “Making computer codes available to others together with a detailed log of every action” in order to “provide[s] a level of detail greater than experimental descriptions using natural language” [23]. A minimum requirement is, in addition to the study protocol and the description of the measurement procedures, the data with descriptive metadata as well as the analysis software and, in case the original software is not supported anymore, the source code [24]. By providing a detailed log of the image processing and segmentation steps, CilOCT opens up the possibility of a reliable replication of the ciliary muscle’s segmentation and subsequent analysis. CilOCT’s batch feature enables time- and resource-saving automatic and reproducible determination of biometric parameters of the ciliary muscle in large numbers of OCT images resulting from longitudinal studies or those involving large cohorts. Such studies will help to answer questions regarding the relationship between the ciliary muscle morphology and the accommodation dynamics, and the possible involvement of the ciliary muscle in the development of myopia. In case a relationship between myopia progression and changes in the ciliary muscle morphology was found, the here presented anatomical measures might serve as additional endpoints in interventional studies aiming at myopia control. Similarly, biometric parameters of the ciliary muscle derived from the application of OCT imaging may help to answer the still open question of the ciliary muscle’s contribution to the development of presbyopia. While it is well accepted that a progressive hardening of the crystalline lens finally results in presbyopia [25], it is unclear if and to which amount the ciliary muscle is active at that stage [26], though several studies found indications for a preserved functionality well beyond the onset of presbyopia [27-30]. The analysis of the CMT profiles may add further evidence to a purely lenticular-based theory of presbyopia.Furthermore, CilOCT’s unique use of a coordinate system with the scleral spur as the point of origin and the intersections between upper ciliary muscle border and perpendiculars to the sclera as basis for the abscissa renders the analysis independent of the ciliary muscle’s position and orientation within the OCT image [20]. CMT profiles of subjects with different refractive errors, of different ages, or in different accommodative states can thereby be accurately overlaid and compared. In addition to the commonly used ciliary muscle area measurements and discrete CMT readings, CMT profiles offer a novel analysis method for visualization and measurement of fine details of the thickness changes, which might remain undetected using current analysis techniques.By providing CilOCT as free and open-source software, we aim to establish a better comparability of studies assessing the ciliary muscle’s structure, features, and changes during accommodation, thereby bringing forward both, myopia and presbyopia research.
Conclusions
In this paper, we presented CilOCT, an open-source software for the semi-automatic segmentation of the ciliary muscle in OCT images. It provides automatic calculation of several biometric parameters for analysis of the ciliary muscle’s morphology as well as a novel approach for the creation of a continuous ciliary muscle thickness profile. The software is written in Java and therefore runs on Windows, macOS, and Linux/Unix. CilOCT can import OCT images from either the raw DICOM format of the Zeiss Visante™ AS-OCT or standard image files. Further sources can easily be added. CilOCT performs an image distortion correction accounting for the difference in the refractive index of the anatomical layers. All image processing steps and the guiding landmarks can be stored as an XML file, allowing for a reproducible segmentation and an automated batch processing. CilOCT was successfully employed in several studies and we believe that it can foster further research on the structure and the function of the ciliary muscle in myopia and presbyopia.
Demonstration of the segmentation of the ciliary muscle in an OCT image using the open-source software CilOCT.
(MP4)Click here for additional data file.
Example OCT image of the ciliary muscle of one of the authors (TS).
(DCM)Click here for additional data file.
Saved guiding landmarks file for the segmentation of the ciliary muscle in the OCT image (S1 File).
(XML)Click here for additional data file.18 Mar 2020PONE-D-20-01371CilOCT: An open-source software for semi-automatic segmentation and analysis of the ciliary muscle in OCT imagesPLOS ONEDear Dr. Straßer,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.We would appreciate receiving your revised manuscript by Apr 30 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocolsPlease include the following items when submitting your revised manuscript:A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.We look forward to receiving your revised manuscript.Kind regards,Asaf AchironAcademic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional requirements:1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.plosone.org/attachments/PLOSOne_formatting_sample_main_body.pdf and http://www.plosone.org/attachments/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. We strongly encourage authors to avoid using figures or tables that have already been published, because this can cause serious copyright concerns. We recommend that you replace figure 3 with a reference to the previous publication.If you feel that it must be included in the current submission, please confirm that you have express written permission from the copyright holder of figure 3 to publish it under the CCAL license.Anything published under this license is available online to read, download, copy, distribute, and use in any way (with attribution) without further permission. Please see http://www.PLOSone.org/static/guidelines.action#preparation with reference to the paragraph titled Figures for more information. You must ensure that the permission from the copyright holder specifically allows you to publish the content under the CCAL license.To seek permission, we recommend that you contact the original copyright holder with the following text: “I request permission for the open-access journal PLOS ONE to publish XXX under the Creative Commons Attribution License (CCAL) CC BY 4.0 (http://creativecommons.org/licenses/by/4.0/us/). Please be aware that this license allows unrestricted use and distribution, even commercially, by third parties. Please reply and provide explicit written permission to publish XXX under a CC BY license.”Please upload the granted permission to the manuscript as a supporting information file. In the figure caption of the copyrighted figure, please include the following text: “Reprinted from [ref] under a CC BY license, with permission from [name of publisher], original copyright [original copyright year].”If you are unable to obtain permission from the original copyright holder, please either i) remove the figure or ii) supply a replacement figure, which can be for illustrative purposes only. Please check copyright information on all replacement figures. If applicable, please specify in the figure caption text when a figure is similar but not identical to the original image, and is therefore for illustrative purposes only.3. Your ethics statement must appear in the Methods section of your manuscript. If your ethics statement is written in any section besides the Methods, please move it to the Methods section and delete it from any other section. Please also ensure that your ethics statement is included in your manuscript, as the ethics section of your online submission will not be published alongside your manuscript.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #1: YesReviewer #2: NoReviewer #3: No**********2. Has the statistical analysis been performed appropriately and rigorously?Reviewer #1: YesReviewer #2: NoReviewer #3: N/A**********3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #1: YesReviewer #2: NoReviewer #3: Yes**********4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #1: YesReviewer #2: NoReviewer #3: Yes**********5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #1: Thank you for allowing me to review this interesting paper.Research article-is the very suitable for this article.Open-Source is the most talked issue in programing in the last years It’s very good to see it in our field.Remarksabstract1. I think you need to elaborate more about the “batch processing”- Most of the readers don’t know enough about it.Intro2. You are talking about different refractive Indices please elaborate more including the numbers you are talking about.3. You’re explaining how you took the images- Its not very clear- please add a photo how you’ve taken it. Moreover- you’re siting [16] your own article.Method4. When you are talking about figure 2e and 2f is not very clear. It needs to be more elaborated.5. The colors in 2b are not sharp enough- very difficult to see it and understand.6. In line 126 you are talking about “distortion correction”- it is not clear enough since you’ve already talked about the ciliary muscle in 2d.Discussion7. You are talking about 6 OCT images in 15 Volunteers – why is the discrepancy?I think it is a very interesting research article with a very updated topic. Though with a small number of subjects.Reviewer #2: editor commentsGeneral1 was intrigue by the title of this study – to evaluate the ciliary muscles – indeed a great topic as accommodation presbyopia and myopia our daily life.2 The ability to standardized reports of imaging is of benefit to all researchers as it create a common language for definition and classificationcover letter1 I understand you offer the code for the program. Can you offer the software itself for those who do not have coding experience? An online tool will be greatIntro1 some font and formatting could be improved- please go over and make sure all is correct= for example reference 15 is not in the same font2 FIG 1- please refer the reader to the figure from the text.3 “CilOCT supports the import of images of the ciliary muscle either from standard image format 54 (BMP, JPEG, PNG) or from raw Zeiss DICOM files [15]. In the case of an import from raw Zeiss 55 DICOM format, the images are automatically rotated and resized [14].”and“To image the ciliary muscle subjects were positioned in front of an anterior segment-OCT (VisanteTM 62 AS-OCT, Carl Zeiss Meditec AG, Jena, Germany; enhanced high-resolution corneal 63 mode, 512 A-scans, scanning size: 10 mm x 3 mm, wavelength 1310 nm) with their head placed 64 in a combined head and chin rest [16]. Since the infrared beam can only penetrate the sclera but 4 65 not the highly pigmented iris [17], the subjects were instructed to shift their gaze about 40° to an 66 external target while keeping their head parallel to the instrument. This allowed imaging the entire 67 ciliary muscle as shown in Fig 1 [16]. Images and results reported were collected in previously 68 published studies [16,18,19]. The studies followed the tenets of the declaration of Helsinki and 69 were approved by the Institutional Review Board of the medical faculty of the University of 70 Tuebingen (376/2017BO2)”should move to the methods4 The intro lacking the purpose of this study: please state clearly that this is a review study aiming to introduce the software- no new data is given her- this is an overview reportyou could add that to the title that it is a reviewmethods-1 CilOCT is developed---> “was”2 The user interface leverages the ribbon-paradigm to simplify and accelerate user interaction- please explain in words that average human can explain that is this ribbon? give example or photo?3 Table 1 – please refer the reader from the text- I think this could be incorporated in the text actually as the table does not add much4 ccontrast can either be manually increased or decreased or adjusted by adaptive histogram equalization--->Is different contrast affect the measurements?5 Illustrative Examples – could you provide capture screen video as a tutorial? ah S1 movie Ok6 is a windows version exist that a regulat user can download and install and work?Discussion1 no results sections?for example, from 100 scans how many were accurate? what is the false positive negative? timing of the procedure? etc.2 The intra-examiner analysis of the CMA revealed---where is this reported? how many examiners? how many test? what was the statistical testplease follow a intro- method-results-conclusion design- we are all used to itciliary muscle thickness profiles- you should show that this is indeed an helpful parameterconclusion- after reading the study- I must admit that the format or order of the study made me confused- what was the propose of this study? if 3 previous studies were already puclished what is new in the current one- if it a review please descrivvve in depth the previous studies (n, methods etc) summeried the results etc.I think if the aim is to present a new software the the format should be technical and present data that show that the software is more accurate them manul segmention that it reduce cost and time etc-Reviewer #3: This manuscript summarizes the authors’ experience with a software for the analysis of OCT images of the ciliary muscle. This experience is based on 3 of the authors’ previous studies (ref 16, 18 and 19). The paper is well written, and demonstration of the technique is clear and appears to be easy to reproduce. However, in my opinion, the authors did not provide adequate justification to publish this manuscript, since there is no evidence of any new original research compared to the 3 previous reports. This manuscript may benefit from being presented as a review.At list one of the figures was published in one of the authors’ previous studies, though this was adapted with permission. This seems inappropriate to me.**********6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #1: Yes: Evgeny GelmanReviewer #2: Yes: Asaf AchironReviewer #3: No[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.Submitted filename: review CilOCT.docxClick here for additional data file.30 Apr 2020We ensured that the manuscript meets PLOS ONE's style requirements, including those for file naming.We supplied a replacement figure for figure 3We added the ethics statement to the Methods and material section.Please find all other responses to reviewer questions and comments in the submitted document "Response to Reviewers.docx"Submitted filename: Response to Reviewers.docxClick here for additional data file.26 May 2020Review of the application of the open-source software CilOCT for semi-automatic segmentation and analysis of the ciliary muscle in OCT imagesPONE-D-20-01371R1Dear Dr. Straßer,We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.With kind regards,Asaf AchironAcademic EditorPLOS ONEAdditional Editor Comments (optional):Reviewers' comments:Reviewer's Responses to QuestionsComments to the Author1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.Reviewer #2: All comments have been addressedReviewer #3: All comments have been addressed**********2. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.Reviewer #2: YesReviewer #3: Yes**********3. Has the statistical analysis been performed appropriately and rigorously?Reviewer #2: N/AReviewer #3: N/A**********4. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.Reviewer #2: YesReviewer #3: Yes**********5. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.Reviewer #2: YesReviewer #3: Yes**********6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)Reviewer #2: I think this will be a good tool.the author did correct all my remarksI have no more commentsgood luckReviewer #3: Thank you for the request to re-review this article.The authors addressed and amended all issues in a satisfactory manner, most importantly changing the format of the paper to a review. No further comments. My recommendation is to accept this valuable manuscript.**********7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.Reviewer #2: NoReviewer #3: Yes: Tal Koval28 May 2020PONE-D-20-01371R1Review of the application of the open-source software CilOCT for semi-automatic segmentation and analysis of the ciliary muscle in OCT imagesDear Dr. Straßer:I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.For any other questions or concerns, please email plosone@plos.org.Thank you for submitting your work to PLOS ONE.With kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Asaf AchironAcademic EditorPLOS ONE
Authors: S A Strenk; J L Semmlow; L M Strenk; P Munoz; J Gronlund-Jacob; J K DeMarco Journal: Invest Ophthalmol Vis Sci Date: 1999-05 Impact factor: 4.799
Authors: José J Esteve-Taboada; Alberto Domínguez-Vicent; Daniel Monsálvez-Romín; Antonio J Del Águila-Carrasco; Robert Montés-Micó Journal: Graefes Arch Clin Exp Ophthalmol Date: 2017-04-20 Impact factor: 3.117
Authors: Kathryn Richdale; Loraine T Sinnott; Mark A Bullimore; Peter A Wassenaar; Petra Schmalbrock; Chiu-Yen Kao; Samuel Patz; Donald O Mutti; Adrian Glasser; Karla Zadnik Journal: Invest Ophthalmol Vis Sci Date: 2013-02-07 Impact factor: 4.799
Authors: Dinesh Kaphle; Katrina L Schmid; Leon N Davies; Marwan Suheimat; David A Atchison Journal: Invest Ophthalmol Vis Sci Date: 2022-06-01 Impact factor: 4.925