Literature DB >> 32509918

Virtual clinical trial for quantifying the effects of beam collimation and pitch on image quality in computed tomography.

Ehsan Abadi1, William P Segars1,2,3, Brian Harrawood1, Shobhit Sharma1,4, Anuj Kapadia1,2,4, Ehsan Samei1,2,3,4,5.   

Abstract

Purpose: To utilize a virtual clinical trial (VCT) construct to investigate the effects of beam collimation and pitch on image quality (IQ) in computed tomography (CT) under different respiratory and cardiac motion rates. Approach: A computational human model [extended cardiac-torso (XCAT) phantom] with added lung lesions was used to simulate seven different rates of cardiac and respiratory motions. A validated CT simulator (DukeSim) was used in this study. A supplemental validation was done to ensure the accuracy of DukeSim across different pitches and beam collimations. Each XCAT phantom was imaged using the CT simulator at multiple pitches (0.5 to 1.5) and beam collimations (19.2 to 57.6 mm) at a constant dose level. The images were compared against the ground truth using three task-generic IQ metrics in the lungs. Additionally, the bias and variability in radiomics (morphological) feature measurements were quantified for task-specific lung lesion quantification across the studied imaging conditions.
Results: All task-generic metrics degraded by 1.6% to 13.3% with increasing pitch. When imaged with motion, increasing pitch reduced motion artifacts. The IQ slightly degraded (1.3%) with changes in the studied beam collimations. Patient motion exhibited negative effects (within 7%) on the IQ. Among all features across all imaging conditions studies, compactness2 and elongation showed the largest ( - 26.5 % , 7.8%) and smallest ( - 0.8 % , 2.7%) relative bias and variability. The radiomics results were robust across the motion profiles studied. Conclusions: While high pitch and large beam collimations can negatively affect the quality of CT images, they are desirable for fast imaging. Further, our results showed no major adverse effects in morphology quantification of lung lesions with the increase in pitch or beam collimation. VCTs, such as the one demonstrated in this study, represent a viable methodology for experiments in CT.
© 2020 Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  CT simulator; computational human phantoms; computed tomography; in silico modeling; simulation; virtual clinical trial

Year:  2020        PMID: 32509918      PMCID: PMC7262564          DOI: 10.1117/1.JMI.7.4.042806

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  27 in total

1.  Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed.

Authors:  Bernd J Wintersperger; Konstantin Nikolaou; Franz von Ziegler; Thorsten Johnson; Carsten Rist; Alexander Leber; Thomas Flohr; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Invest Radiol       Date:  2006-05       Impact factor: 6.016

2.  Accelerating Monte Carlo simulations of photon transport in a voxelized geometry using a massively parallel graphics processing unit.

Authors:  Andreu Badal; Aldo Badano
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

Review 3.  Application of the 4-D XCAT Phantoms in Biomedical Imaging and Beyond.

Authors:  W Paul Segars; B M W Tsui; George S K Fung; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2017-08-10       Impact factor: 10.048

4.  A real-time Monte Carlo tool for individualized dose estimations in clinical CT.

Authors:  Shobhit Sharma; Anuj Kapadia; Wanyi Fu; Ehsan Abadi; W Paul Segars; Ehsan Samei
Journal:  Phys Med Biol       Date:  2019-11-04       Impact factor: 3.609

5.  Thin-section three-dimensional spiral CT: is isotropic imaging possible?

Authors:  W A Kalender
Journal:  Radiology       Date:  1995-12       Impact factor: 11.105

Review 6.  CT coronary angiography: 256-slice and 320-detector row scanners.

Authors:  Edward M Hsiao; Frank J Rybicki; Michael Steigner
Journal:  Curr Cardiol Rep       Date:  2010-01       Impact factor: 2.931

7.  High-pitch spiral computed tomography: effect on image quality and radiation dose in pediatric chest computed tomography.

Authors:  Michael M Lell; Matthias May; Paul Deak; Sedat Alibek; Michael Kuefner; Axel Kuettner; Henrik Köhler; Stephan Achenbach; Michael Uder; Tanja Radkow
Journal:  Invest Radiol       Date:  2011-02       Impact factor: 6.016

8.  CT scan parameters and radiation dose: practical advice for radiologists.

Authors:  Siva P Raman; Mahadevappa Mahesh; Robert V Blasko; Elliot K Fishman
Journal:  J Am Coll Radiol       Date:  2013-11       Impact factor: 5.532

9.  The Image Biomarker Standardization Initiative: Standardized Quantitative Radiomics for High-Throughput Image-based Phenotyping.

Authors:  Alex Zwanenburg; Martin Vallières; Mahmoud A Abdalah; Hugo J W L Aerts; Vincent Andrearczyk; Aditya Apte; Saeed Ashrafinia; Spyridon Bakas; Roelof J Beukinga; Ronald Boellaard; Marta Bogowicz; Luca Boldrini; Irène Buvat; Gary J R Cook; Christos Davatzikos; Adrien Depeursinge; Marie-Charlotte Desseroit; Nicola Dinapoli; Cuong Viet Dinh; Sebastian Echegaray; Issam El Naqa; Andriy Y Fedorov; Roberto Gatta; Robert J Gillies; Vicky Goh; Michael Götz; Matthias Guckenberger; Sung Min Ha; Mathieu Hatt; Fabian Isensee; Philippe Lambin; Stefan Leger; Ralph T H Leijenaar; Jacopo Lenkowicz; Fiona Lippert; Are Losnegård; Klaus H Maier-Hein; Olivier Morin; Henning Müller; Sandy Napel; Christophe Nioche; Fanny Orlhac; Sarthak Pati; Elisabeth A G Pfaehler; Arman Rahmim; Arvind U K Rao; Jonas Scherer; Muhammad Musib Siddique; Nanna M Sijtsema; Jairo Socarras Fernandez; Emiliano Spezi; Roel J H M Steenbakkers; Stephanie Tanadini-Lang; Daniela Thorwarth; Esther G C Troost; Taman Upadhaya; Vincenzo Valentini; Lisanne V van Dijk; Joost van Griethuysen; Floris H P van Velden; Philip Whybra; Christian Richter; Steffen Löck
Journal:  Radiology       Date:  2020-03-10       Impact factor: 29.146

10.  Lung tumor segmentation methods: Impact on the uncertainty of radiomics features for non-small cell lung cancer.

Authors:  Constance A Owens; Christine B Peterson; Chad Tang; Eugene J Koay; Wen Yu; Dennis S Mackin; Jing Li; Mohammad R Salehpour; David T Fuentes; Laurence E Court; Jinzhong Yang
Journal:  PLoS One       Date:  2018-10-04       Impact factor: 3.240

View more
  2 in total

1.  The ultrafast, high-pitch turbo FLASH mode of third-generation dual-source CT: Effect of different pitch and corresponding SFOV on image quality in a phantom study.

Authors:  Yang Zhou; Lei Hu; Silin Du; Rui Jin; Wangjia Li; Fajin Lv; Zhiwei Zhang
Journal:  J Appl Clin Med Phys       Date:  2021-11-09       Impact factor: 2.102

2.  VIRTUAL CLINICAL TRIALS IN MEDICAL IMAGING SYSTEM EVALUATION AND OPTIMISATION.

Authors:  Bruno Barufaldi; Andrew D A Maidment; Magnus Dustler; Rebecca Axelsson; Hanna Tomic; Sophia Zackrisson; Anders Tingberg; Predrag R Bakic
Journal:  Radiat Prot Dosimetry       Date:  2021-10-12       Impact factor: 0.972

  2 in total