Akram M Asbeutah1, Abdullah A AlMajran2, Ajit Brindhaban1, Saad A Asbeutah3.
Abstract
INTRODUCTION: There are increasing concerns about radiation exposure among women who undergo full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT). The main aim of this study was to compare the entrance surface dose (ESD) and average glandular dose (AGD) from FFDM and DBT for different breast thicknesses.Entities:
Keywords: Breast cancer; average glandular dose; digital breast tomosynthesis; entrance skin dose; full-field digital mammography
Mesh:
Year: 2020 PMID: 32495513 PMCID: PMC7476200 DOI: 10.1002/jmrs.405
Source DB: PubMed Journal: J Med Radiat Sci ISSN: 2051-3895
Patient characteristics and mammographic parameters for FFDM and DBT techniques for different compressed breast thicknesses.
|
Breast Thickness Parameter | Technique | 2.1–3 cm | 3.1–4 cm | 4.1–5 cm | 5.1–6 cm | 6.1–7 cm | 7.1–8 cm |
|---|---|---|---|---|---|---|---|
| Number | 5 | 14 | 42 | 80 | 46 | 13 | |
|
Age, Median (IQR), years (Range) |
45 (13) (33–52) |
46.5 (17.3) (37–71) |
49.5 (16.3) (39–81) |
48.5 (11) (36–73) |
46 (11.3) (38–70) |
50 (11) (38–61) | |
|
kV, Median (IQR) (Range) | FFDM |
27 (0) (27–27) |
28 (0.5) (27–29) |
29 (0) (27–31) |
29 (1) (29–31) |
29 (1) (29–31) |
30 (1) (29–30) |
| DBT |
26 (2) (26–29) |
29 (0) (29–29) |
29 (0) (29–29) |
29 (0) (29–31) |
29 (2) (29–31) |
31 (0) (29–31) | |
|
mAs, Median (IQR) (Range) | FFDM |
37.7 (10.6) (26.6–43.2) |
50.5 (8.3) (39.7–56.3) |
48.7 (12.6) (35.3–62.7) |
54.1 (7.9) (42.2–96.3) |
61.2 (10.4) (49.8–83.0) |
71.1 (13.1) (60.8–94.3) |
| DBT |
4 (1.5) (2.9–4.9) |
4.5 (0.3) (3–5) |
5.3 (1.2) (3.5–8.6) |
6.3 (1.7) (4.2–8.7) |
7.8 (2) (5.6–14.1) |
7.8 (10.6) (6.5–11.9) | |
| Target/Filter | FFDM | Mo/Rh | Mo/Rh | Rh/Rh | Rh/Rh | Rh/Rh | Rh/Rh |
| DBT | Rh/Rh | Rh/Rh | Rh/Rh | Rh/Rh | Rh/Rh | Rh/Rh |
DBT = digital breast tomosynthesis; FFDM = full‐field digital mammography; IQR = interquartile range; Mo = Molybdenum; Rh = Rhodium.
Figure 1Exposure parameters in individual examinations for full‐field digital mammography (FFDM) and digital breast tomosynthesis (DBT) with different target/filter material (Mo = Molybdenum, Rh = Rhodium). (a) Tube voltage (kV) versus breast thickness. (b) Tube load (mAs) versus breast thickness. (c) Tube voltage (kV) versus tube load (mAs).
The median (IQR) (min, max) for highest ESD and total AGD in mGy for FFDM and DBT techniques for different compressed breast thicknesses. There was a statistically significant difference in the median ESD and total AGD values between FFDM and DBT techniques (P <0.001).
|
Technique Parameter Breast Thickness | FFDM | DBT | Difference (%) | |||
|---|---|---|---|---|---|---|
| Highest ESD‐mGy | Total AGD‐mGy | Highest ESD‐mGy | Total AGD‐mGy | ESD | AGD^ | |
| 2.1–3 cm |
3.3 (1.3) (2.3–3.9) |
3.3 (1.5) (2.8––4.2) |
3.1 (0.5) (2.6–3.3) |
1.8 (1.0) (1.5–3.1) | 0.2 (6.1) | 1.5 (45.5) |
| 3.1–4 cm |
4.7 (0.5) (3.4–5.3) |
4.5 (1.3) (3.5–4.8) |
4.0 (0.7) (2.8‐5.2) |
2.5 (0.3) (1.9–3.0) | 0.7 (14.9) | 2.0 (44.4) |
| 4.1–5 cm |
5.5 (1.5) (3.6–8.5) |
4.6 (1.6) (4.1–5.7) |
4.7 (1.2) (2.8–9.3) |
2.6 (0.5) (1.8–3.9) | 0.8 (14.5) | 2.0 (43.5) |
| 5.1–6 cm |
6.7 (1.4) (3.9–10.5) |
5.0 (0.4) (4.1–8.3) |
5.9 (1.8) (0.2–9.9) |
3.0 (1.2) (2.1–4.8) | 0.8 (12) | 2.0 (40) |
| 6.1–7 cm |
8.1 (1.5) (5.5–11.1) |
5.6 (0.5) (4.8–6.6) |
7.8 (3.3) (5.1–16.2) |
3.6 (1.1) (2.5–7.0) | 0.3 (3.7) | 2.0 (35.7) |
| 7.1–8 cm |
9.1 (1.5) (5.9–10.0) |
6.0 (0.4) (4.7–6.3) |
8.9 (2.9) (5.9–15.5) |
4.0 (1.5) (3.0–6.2) | 0.2 (2.2) | 2.0 (33.3) |
AGD = average glandular dose; DBT = digital breast tomosynthesis; ESD = entrance skin dose; FFDM = full‐field digital mammography.
(ESD FFDM‐ ESD DBT/ESD FFDM) × 100.
(AGD FFDM‐AGD DBT/AGD FFDM) × 100.
Figure 2The highest entrance skin dose (ESD) and total average glandular dose (AGD) versus breast thickness for full‐field digital mammography (FFDM) (a and c) and digital breast tomosynthesis (DBT) (b and d).