A study to determine the differences between the displayed dose values for two full-field digitalmammography units and values calculated using a range of Monte-Carlo-based techniques:a phantom study.

M Borg1, I Badr, G J Royle.   

Abstract

Modern full-field digital mammography (FFDM) units display the mean glandular dose (MGD) and the entrance or incident air kerma (K) to the breast following each exposure. Information on how these values are calculated is limited and knowing how displayed MGD values compare and correlate to conventional Monte-Carlo-based methods is useful. From measurements done on polymethyl methacrylate (PMMA) phantoms, it has been shown that displayed and calculated MGD values are similar for thin to medium thicknesses and appear to differ with larger PMMA thicknesses. As a result, a multiple linear regression analysis on the data was performed to generate models by which displayed MGD values on the two FFDM units included in the study may be converted to the Monte-Carlo values calculated by conventional methods. These models should be a useful tool for medical physicists requiring MGD data from FFDM units included in this paper and should reduce the survey time spent on dose calculations.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22927653     DOI: 10.1093/rpd/ncs167

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Mean glandular dose in digital mammography: a dose calculation method comparison.

Authors:  Moayyad E Suleiman; Patrick C Brennan; Mark F McEntee
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-24

2.  Comparison of radiation doses between diagnostic full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT): a clinical study.

Authors:  Akram M Asbeutah; Abdullah A AlMajran; Ajit Brindhaban; Saad A Asbeutah
Journal:  J Med Radiat Sci       Date:  2020-06-03

3.  Comparison between Recorded and Measured Radiation Doses in Diagnostic Full-field Digital Mammography: A Phantom Study.

Authors:  Akram Mahmoud Asbeutah; Ajit Brindhaban
Journal:  J Clin Imaging Sci       Date:  2020-11-13

4.  Comparison of Mean Glandular Dose between Full-Field Digital Mammography and Digital Breast Tomosynthesis.

Authors:  Kar Choon Teoh; Hanani Abdul Manan; Norhashimah Mohd Norsuddin; Iqbal Hussain Rizuana
Journal:  Healthcare (Basel)       Date:  2021-12-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.