A potentially implantable enzyme electrode for amperometric measurement of glucose.

W Kerner1, H Zier, G Steinbach, J Brückel, E F Pfeiffer, T Weiss, K Cammann, H Planck.   

Abstract

A membrane limited amperometric enzyme electrode suitable for glucose measurement in biological fluids was developed. The sensor consists of a central platinum wire (0.3 mm) surrounded by a stainless steel tubing (0.8-1.0 mm outer diameter; 2-4 cm length). By successive dipcoating procedures, layers from cellulose acetate, glucose oxidase (crosslinked with glutaraldehyde) and polyurethane are placed on its surface. The platinum is polarized at +700 mV against steel. In vitro results: Electrodes are stable for at least 6 days. They exhibit a linear range extending to 500 mg/dl glucose. Response times are less than 100 sec. The sensors are not dependent on stirring and are relatively insensitive to changes of pH. Dependency of glucose measurement upon dissolved oxygen is negligible at oxygen concentrations above 0.5 mg/l. In vivo results: Preliminary studies in sheep using subcutaneously implanted needles indicate that short term glucose monitoring is feasible.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3248792

Source DB:  PubMed          Journal:  Horm Metab Res Suppl        ISSN: 0170-5903


  2 in total

1.  In vivo measurement of subcutaneous glucose concentrations with an enzymatic glucose sensor and a wick method.

Authors:  J Brückel; W Kerner; H Zier; G Steinbach; E F Pfeiffer
Journal:  Klin Wochenschr       Date:  1989-05-02

2.  On line continuous monitoring of subcutaneous tissue glucose in men by combining portable glucosensor with microdialysis.

Authors:  C Meyerhoff; F Bischof; F Sternberg; H Zier; E F Pfeiffer
Journal:  Diabetologia       Date:  1992-11       Impact factor: 10.122

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.