Ventricular Tachycardia Ablation in Patients with Left Ventricular Assist Devices.

Kaustubha D Patil1, Jonathan Chrispin1.   

Abstract

In this complex case study, we discuss a patient who underwent successful catheter ablation for ventricular tachycardia following left ventricular assist device placement. We discuss the technique and review existing literature in an effort to explore the feasibility and safety of this procedure in this clinical setting. Copyright:
© 2019 Innovations in Cardiac Rhythm Management.

Entities:  

Keywords:  Catheter ablation; electroanatomical mapping; left ventricular assist device; ventricular tachycardia

Year:  2019        PMID: 32477712      PMCID: PMC7252754          DOI: 10.19102/icrm.2019.101101

Source DB:  PubMed          Journal:  J Innov Card Rhythm Manag        ISSN: 2156-3977


Introduction

Because of limited organ availability, the number of heart transplantations performed per year is plateauing.[1] Thus, the use of mechanical support devices—in particular, left ventricular assist devices (LVADs)—for the treatment of end-stage heart failure has increased. However, it is well-recognized that ventricular arrhythmia (VA) burden may actually rise following the implantation of an LVAD,[2] which is associated with a decrease in patient survival.[3] Catheter ablation for recurrent, drug-refractory ventricular tachycardia (VT) in patients with structural heart disease is well-established,[4-6] resulting in fewer appropriate implantable cardioverter-defibrillator (ICD) shocks and repeat hospitalizations. In this report, we discuss the case of a patient with drug-refractory VA following LVAD placement that was successfully mapped and terminated with catheter ablation.

Case presentation

A 68-year-old male with a finding of ischemic cardiomyopathy after undergoing HeartMate™ II LVAD (Abbott Laboratories, Chicago, IL, USA) implantation in October 2013 presented to the Johns Hopkins Hospital with one week of worsening fatigue and palpitations as well as multiple ICD shocks on the morning of admission. His previous history included implantation of a dual-chamber ICD, prior ablation of scar-mediated VT, persistent atrial fibrillation, and amiodarone-induced thyrotoxicosis. The patient’s rhythm upon presentation to the emergency department was ventricular fibrillation (VF), which was hemodynamically tolerated, given his LVAD (. He underwent successful external defibrillation with one shock of 200 J and a restoration of sinus rhythm. ICD interrogation showed a prolonged episode (more than five hours) of tachycardia. Atrial lead electrograms were consistent with atrial fibrillation (atrial cycle length: 180–210 ms). Ventricular lead electrograms were consistent with a simultaneous, regular VT with a ventricular cycle length of approximately 280 ms to 290 ms. His device appropriately detected VT, and an ICD shock converted his rhythm to polymorphic VT/VF. There were then five subsequent additional shocks that failed to terminate the polymorphic VT/VF. He was admitted to the cardiac intensive care unit, where he experienced recurrent sustained monomorphic VT despite intravenous lidocaine, propanolol, and sotalol (. The electrocardiogram morphology of his VT showed right bundle branch block, a rightward axis, and an indeterminate vertical axis (negative in II, isoelectric in III, and negative in aVF) suggestive of an apical lateral left ventricular origin. The VT was subsequently successfully terminated with antitachycardia pacing via his ICD. With the LVAD introduced, all VT episodes were hemodynamically stable, but the patient complained of palpitations. Given the patient’s recurrent VT requiring ICD therapy despite antiarrhythmic drug therapy, he was referred for an electrophysiology study and catheter ablation. An endocardial shell of the left ventricle, left ventricular outflow tract, and LVAD inflow cannula insertion site was created using a CARTO-Sound® (Biosense Webster, Diamond Bar, CA, USA) intracardiac echocardiogram catheter. Endocardial left ventricular mapping was performed using CARTO® (Biosense Webster, Diamond Bar, CA, USA) by way of a single transseptal approach and using a DECA-NAV® (Biosense Webster, Diamond Bar, CA, USA) mapping catheter. Endocardial left ventricular mapping was notable for a large anterior wall scar from the base to the apex, spanning from the septum to the lateral wall (. The apical segments of the scar just lateral to the LVAD inflow cannula insertion site had multiple sites of slow conduction with isolated late potentials and low-amplitude, long-fractionated signals. On isoproterenol, the patient’s clinical VT was induced (. During VT, the mentioned sites of slow conduction showed middiastolic potentials (. Entrainment mapping from these sites was concealed with a postpacing interval that closely approximated the tachycardia cycle length (TCL) (. The stimulus-to-QRS interval length was equal to the electrogram-to-QRS interval length (. The stimulus-to-QRS interval length (159 ms) as compared with the TCL (375 ms) was consistent with a critical isthmus site being at this location (ie, stimulus to-QRS length/TCL ratio of 42%). Using an irrigated, force-sensing ablation catheter, radiofrequency ablation lesions were delivered to the sites consistent with a critical isthmus, showing slowed ventricular conduction and subsequent termination of VT during ablation (. The successful ablation site was visualized with intracardiac echocardiography (. Additional substrate-based ablation was performed to connect the critical isthmus sites to the apical scar (. Postablation, there was no inducible VT with ventricular programmed stimulation, either on or off isoproterenol. Given the documented rapid atrial rates on initial device interrogation, we proceeded with atrioventricular node ablation for the definitive rate control of rapid atrial arrhythmias. The atrioventricular node was successfully ablated, with resultant complete heart block and pacing dispensed via the patient’s previously implanted ICD. Given the multiple failed ICD shocks administered for polymorphic VT/VF after the prolonged VT prior to presentation, defibrillation threshold testing was performed. A 50-Hz pulse was delivered by the patient’s internal defibrillator, resulting in polymorphic VT/VF. The patient’s dual-chamber ICD appropriately detected VF and delivered an appropriate shock of one shock of 35 J, which successfully restored normal sinus rhythm. There were no intraoperative complications, and the patient did well postprocedure without the recurrence of the presenting clinical VT. He was discharged on the fifth day after ablation on a regimen of propranolol, sotalol, and mexiletine.

Discussion

VAs are common among patients with structural heart disease, and ICDs have been shown to improve survival in both primary and secondary prevention populations.[7,8] However, despite various advances that have been made in ICD technology, ICD shocks are associated with a decrease in long-term survival and with a decline in quality of life.[9] Antiarrhythmic therapy is effective in decreasing VAs, but these medications can be associated with significant side effects. Elsewhere, while LVAD therapy has emerged as a promising option to improve survival in patients with end-stage heart failure, it has been repeatedly reported that LVADs may actually increase the VA burden.[4] In a single-center retrospective analysis of 43 patients who underwent implantation of a HeartMate™ II LVAD (Abbott Laboratories, Chicago, IL, USA), 27.9% experienced early electrical storm, defined as three or more ICD therapy applications in 24 hours, with a median time from LVAD implant of 9.1 days ± 7.8 days. Those who had early VT also had a significantly higher all-cause mortality rate at 30 days (33.3% versus 6.5%).[2] These findings were replicated in a larger study of 98 patients with ICDs who underwent placement of an LVAD. In total, 48 (49%) of the 98 patients experienced VA and experienced an average of 30 ± 98 appropriate ICD therapy applications.[10] One of the main predictors of VA post-LVAD implantation is a history of pre-LVAD VA.[2,10,11] Our patient had a history of VA prior to his LVAD placement. Unfortunately, successful elimination of VA post-LVAD placement remains difficult despite antiarrhythmic medications and, thus, catheter ablation has emerged as an adjunct to decrease the VA burden. Ideally, invasive activation and entrainment mapping during VT is the most specific way to identify the critical isthmus that contributes to the maintenance of monomorphic VT. However, in most circumstances, mapping in VT is not feasible, given that VT is not hemodynamically tolerated, particularly in a substrate with poor systolic function and advanced heart failure. Decreased end-organ perfusion results in significant metabolic acidosis, leading to an increase in periprocedural complications and mortality. A substrate-based approach of identifying areas of slow conduction in sinus rhythm has emerged as an alternative strategy. One of the benefits of mechanical hemodynamic support during catheter ablation of VA is that one can maintain organ perfusion during VA, allowing for more detailed and precise electroanatomical mapping of the VA circuits. Patients with LVADs often have hemodynamically tolerable VAs, as seen in our patient who presented with VF. Baratto et al. reported on 64 patients who underwent catheter ablation of unstable VA with extracorporeal membrane oxygenation (ECMO) for hemodynamic support. At least one VT was terminated in 81% of the procedures performed and, more importantly, the study demonstrated the safety of ECMO-supported VA catheter ablation.[12] Herweg et al. presented their center’s experience of six patients regarding the feasibility of catheter ablation of VA in patients with LVAD. In two of the six patients, the patients’ inflow cannula served as the substrate for the VA.[5] Sacher et al. reported a multicenter experience of patients who underwent catheter ablation of VA in the presence of a HeartMate™ II LVAD (Abbott Laboratories, Chicago, IL, USA). In total, 34 patients with a mix of nonischemic and ischemic cardiomyopathies, respectively, underwent 39 ablation procedures.[13] In this series, nine patients required VA ablation at less than 30 days after their LVAD placement secondary to refractory VT. Ultimately, only 9% (10/110) of the targeted VT episodes within the study population were related to the LVAD cannula. In a meta-analysis of 18 studies on the catheter ablation of VA post-LVAD implantation, 90.3% of the VAs were found to be of scar-related reentry origin, while 19.3% were related to the LVAD cannula.[6] Following successful catheter ablation, the VT storm terminated in 90% of patients and there was a significant reduction in ICD therapy applications as compared with prior to surgery (23.8% versus 57.1%). Activation/entrainment mapping was performed in 61.2% of the patients. Further, despite the presence of an aortic outflow tract cannula, which limits access into the ventricule, 36.4% of the procedures were performed using a retrograde aortic approach. In conclusion, our case demonstrated the feasibility and safety of activation/entrainment mapping in a patient with scar-related VT. In the setting of a large area of anterior scar, ablation was limited to a small area, allowing for a precision-oriented ablation strategy, further decreasing the potential for complications associated with prolonged procedures and ablation times. Of note, our patient had a VT circuit involving the LVAD inflow cannula, which is seen in only a minority of patients with VT post-LVAD. The use of CARTO-Sound® (Biosense Webster, Diamond Bar, CA, USA) allowed for real-time geometry creation and visualization of the inflow cannula throughout the procedure. The present case supports that catheter ablation for drug-refractory VT in patients with LVADs is feasible and should be included in the armamentarium of treatment strategies in specialized cardiac centers.
  13 in total

1.  Heart disease and stroke statistics--2012 update: a report from the American Heart Association.

Authors:  Véronique L Roger; Alan S Go; Donald M Lloyd-Jones; Emelia J Benjamin; Jarett D Berry; William B Borden; Dawn M Bravata; Shifan Dai; Earl S Ford; Caroline S Fox; Heather J Fullerton; Cathleen Gillespie; Susan M Hailpern; John A Heit; Virginia J Howard; Brett M Kissela; Steven J Kittner; Daniel T Lackland; Judith H Lichtman; Lynda D Lisabeth; Diane M Makuc; Gregory M Marcus; Ariane Marelli; David B Matchar; Claudia S Moy; Dariush Mozaffarian; Michael E Mussolino; Graham Nichol; Nina P Paynter; Elsayed Z Soliman; Paul D Sorlie; Nona Sotoodehnia; Tanya N Turan; Salim S Virani; Nathan D Wong; Daniel Woo; Melanie B Turner
Journal:  Circulation       Date:  2011-12-15       Impact factor: 29.690

2.  Characteristics of ventricular tachycardia ablation in patients with continuous flow left ventricular assist devices.

Authors:  Frederic Sacher; Tobias Reichlin; Erica S Zado; Michael E Field; Juan F Viles-Gonzalez; Petr Peichl; Kenneth A Ellenbogen; Philippe Maury; Srinivas R Dukkipati; Francois Picard; Josef Kautzner; Laurent Barandon; Jayanthi N Koneru; Philippe Ritter; Saagar Mahida; Joachim Calderon; Nicolas Derval; Arnaud Denis; Hubert Cochet; Richard K Shepard; Jerome Corre; James O Coffey; Fermin Garcia; Meleze Hocini; Usha Tedrow; Michel Haissaguerre; Andre d'Avila; William G Stevenson; Francis E Marchlinski; Pierre Jais
Journal:  Circ Arrhythm Electrophysiol       Date:  2015-04-13

Review 3.  Ventricular arrhythmias after left ventricular assist device.

Authors:  Shashima Nakahara; Christopher Chien; Jill Gelow; Khidir Dalouk; Charles A Henrikson; James Mudd; Eric C Stecker
Journal:  Circ Arrhythm Electrophysiol       Date:  2013-06

4.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.

Authors:  Arthur J Moss; Wojciech Zareba; W Jackson Hall; Helmut Klein; David J Wilber; David S Cannom; James P Daubert; Steven L Higgins; Mary W Brown; Mark L Andrews
Journal:  N Engl J Med       Date:  2002-03-19       Impact factor: 91.245

5.  Catheter Ablation of Ventricular Tachycardia in Patients With a Ventricular Assist Device: A Systematic Review of Procedural Characteristics and Outcomes.

Authors:  Robert D Anderson; Geoffrey Lee; Sohaib Virk; Richard G Bennett; Christopher S Hayward; Kavitha Muthiah; Jonathan Kalman; Saurabh Kumar
Journal:  JACC Clin Electrophysiol       Date:  2018-09-26

6.  Predictors of ventricular arrhythmia after left ventricular assist device implantation: A large single-center observational study.

Authors:  Elena Efimova; Julia Fischer; Livio Bertagnolli; Borislav Dinov; Simon Kircher; Sascha Rolf; Philipp Sommer; Andreas Bollmann; Sergio Richter; Anna Meyer; Jens Garbade; Gerhard Hindricks; Arash Arya
Journal:  Heart Rhythm       Date:  2017-07-27       Impact factor: 6.343

7.  A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias.

Authors: 
Journal:  N Engl J Med       Date:  1997-11-27       Impact factor: 91.245

8.  Effect of Ventricular Arrhythmia Ablation in Patients With Heart Mate II Left Ventricular Assist Devices: An Evaluation of Ablation Therapy.

Authors:  David Snipelisky; Yogesh N V Reddy; Kevin Manocha; Aalok Patel; Shannon M Dunlay; Paul A Friedman; Thomas M Munger; Samuel J Asirvatham; Douglas L Packer; Yong-Mei Cha; Suraj Kapa; Peter A Brady; Peter A Noseworthy; Joseph J Maleszewski; Siva K Mulpuru
Journal:  J Cardiovasc Electrophysiol       Date:  2016-11-21

9.  Electrical storm in the early phase of HeartMate® II device implantation: Incidence, risk factors and prognosis.

Authors:  Jerome Corre; François Picard; Rodrigue Garcia; Adlane Zemmoura; Nicolas Derval; Arnaud Denis; Antoine Romen; Karine Nubret; Pierre Jais; Michel Haissaguerre; Pierre Dos Santos; Laurent Barandon; Frederic Sacher
Journal:  Arch Cardiovasc Dis       Date:  2017-12-06       Impact factor: 2.340

10.  Ventricular arrhythmias and implantable cardioverter-defibrillator therapy in patients with continuous-flow left ventricular assist devices: need for primary prevention?

Authors:  Arthur R Garan; Melana Yuzefpolskaya; Paolo C Colombo; John P Morrow; Rosie Te-Frey; Drew Dano; Hiroo Takayama; Yoshifumi Naka; Hasan Garan; Ulrich P Jorde; Nir Uriel
Journal:  J Am Coll Cardiol       Date:  2013-05-02       Impact factor: 24.094

View more
  1 in total

1.  Catheter ablation for atrial fibrillation in left ventricular assist device: A case report.

Authors:  Yu Jin Chung; Jin-Oh Choi; Kyoung-Min Park
Journal:  Medicine (Baltimore)       Date:  2021-06-25       Impact factor: 1.889

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.