| Literature DB >> 32429881 |
Petr Fulin1, David Pokorny2, Jan Hert2, Antonin Sosna2.
Abstract
BACKGROUND: The lifetime implants is a key parameter that the surgeon should take into account at the time of the primary total hip arthroplasty (THA). The aim of this study was a clinical and radiographical evaluation of the Delta PF-FIT (LimaCorporate, Italy) THA system with ceramic-on-ceramic articulations. We have not found a clinical or radiographical assessment of this implant in available published literature.Entities:
Keywords: Delta PF cup; FIT stem; HHS; LimaCorporate; Total hip arthroplasty; WOMAC
Mesh:
Year: 2020 PMID: 32429881 PMCID: PMC7236923 DOI: 10.1186/s12891-020-03253-x
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Lima Implants (LimaCorporate). Delta PF Cup (left), FIT stem with ceramic insert and a BIOLOX® Delta head (right)
Fig. 2Distribution of operated patients
Indications for THA
| Indications | ||
|---|---|---|
| Diagnosis | Number | Percentage |
| Arthritis | 153 | 77.7% |
| Postdysplastic arthritis | 23 | 11.7% |
| Femoral head necrosis | 10 | 5.1% |
| Rheumatic arthritis | 4 | 2.0% |
| Femoral neck fracture | 4 | 2.0% |
| Femoral nail failure | 2 | 1.0% |
| State after coxitis | 1 | 0.5% |
Statistical values (FU-follow up, FL- flexion, ABD – abduction, ER – external rotation, IR- internal rotation, IQR interquartile range)
| Age | FU | HHS | WOMAC | FL | ABD | ER | IR | |
|---|---|---|---|---|---|---|---|---|
| mean | 57.9 | 7.7 | 97.6 | 97.4 | 104.2 | 40.7 | 28.9 | 19.1 |
| median | 59.0 | 7.7 | 100.0 | 100.0 | 100.0 | 40.0 | 30.0 | 20.0 |
| IQR | 11.0 | 1.9 | 3.0 | 4.0 | 10.0 | 5.0 | 0.0 | 2.0 |
Results of pre- and post-operatively clinical evaluation (SD – standard deviation)
| Results ( | ||||||||
|---|---|---|---|---|---|---|---|---|
| Preoperative | Postoperative | |||||||
| Mean | Min | Max | SD | Mean | Min | Max | SD | |
| HSS | 51.21 | 28 | 73 | 4.92 | 97.59 | 61 | 100 | 5.13 |
| WOMAC | 50.12 | 27 | 69 | 4.94 | 97.38 | 65 | 100 | 5.18 |
| Flexion | 89.51 | 45 | 110 | 8.2 | 104.16 | 70 | 130 | 10.61 |
| Abduction | 27.32 | 5 | 40 | 4.71 | 40.72 | 30 | 50 | 4.52 |
| External rotation | 19.81 | 0 | 30 | 5.12 | 28.95 | 0 | 40 | 5.6 |
| Internal rotation | 11.26 | 0 | 20 | 4.84 | 19.13 | 0 | 35 | 6.17 |
Fig. 3Bone trabeculae remodelling with the integration into the hydroxyapatite surface of the implant
Fig. 4Matrix of scatter plots. The highest correlations show the values closest to the linear ascending line. The lowest correlation followed by values with random scattering of values
Fig. 5The density of HHS difference (left side) and the density of WOMAC difference (right side)