Daniel Overhoff1, Meike Weis1, Philipp Riffel1, Sonja Sudarski1, Matthias F Froelich1, Peter Fries2, Stefan Schönberg1, Joshua Gawlitza3.
Abstract
BACKGROUND: One main challenge in pediatric imaging is to reduce motion artifacts by calming young patients. To that end, the Radiological Society of North America (RSNA) as early as 1997 stated the necessity of adults accompanying their child during the child's examination. Nonetheless, current research lacks data regarding radiation dose to these chaperones.Entities:
Keywords: Children; Computed tomography; Dosimetry; Imaging protocols; Radiation dose reduction; Radiation exposure
Mesh:
Substances:
Year: 2020 PMID: 32415324 PMCID: PMC7329757 DOI: 10.1007/s00247-020-04681-6
Source DB: PubMed Journal: Pediatr Radiol ISSN: 0301-0449
Computed tomography parameters
| Protocol 1 | Protocol 2 | |
|---|---|---|
| Contrast | Chest, non-enhanced | Chest, contrast-enhanced |
| Detector rows | 2×192 | 2×192 |
| Tube voltage (kVp) | 100 | 70 |
| Reference tube current (mAs) | 100 | 64 |
| Rotation time (s) | 0.25 | 0.25 |
| Pitch | 3.2 | 3.2 |
| Collimation (mm) | 192×0.6 | 192×0.6 |
| Tin filtration (mm) | 0.6 | None |
Fig. 1Diagram shows the locations of the three dosimeters next to the CT scanner. One was right next to table and gantry opening (Dosimeter 1); the second was in the diagonal prolongation, 1 m away (Dosimeter 2); and the last was beside the gantry (Dosimeter 3)
Cumulative radiation doses (μSv) measured over 10 consecutive scans for every dosimetera during the scan of the thorax phantom
| Dosimeter 1 | Dosimeter 2 | Dosimeter 3 | |
|---|---|---|---|
| Protocol 1b | 3.0 | 1.0 | 0 |
| Protocol 2b | 2.0 | 1.0 | 0 |
aDosimeter placement is shown in Fig. 1
bScan protocols are specified in Table 1
Mean values of the two pediatric patient protocols examined
| Protocola | Doseb (μSv) | Topogram DLP/CTDIvol | Examination DLP/CTDIvol | Total DLP | ||
|---|---|---|---|---|---|---|
| Dosimeter 1 | Dosimeter 2 | Dosimeter 3 | ||||
| Protocol 1 | 0.75±0.70 | 0.13±0.40 | 0 | 0.49±0.08 | 3.50±1.10 | 3.9±1.2 |
| 0.02±0.01 | 0.17±0.04 | |||||
| Protocol 2 | 0.25±0.50 | 0.50±0.60 | 0 | 0.33±0.12 | 9.40±4.30 | 9.9±4.1 |
| 0.01±0.01 | 0.42±0.13 | |||||
CTDI volumetric computed tomography dose index, DLP dose–length product
aProtocol specifications in Table 1
bDosimeter position is shown in Fig. 1