Clinical evaluation of a SARS-CoV-2 RT-PCR assay on a fully automated system for rapid on-demand testing in the hospital setting.

Dominik Nörz1, Nicole Fischer2, Alexander Schultze3, Stefan Kluge4, Ulrich Mayer-Runge3, Martin Aepfelbacher2, Susanne Pfefferle2, Marc Lütgehetmann2.   

Abstract

BACKGROUND: The ongoing SARS-CoV-2 pandemic presents a unique challenge for diagnostic laboratories around the world. Automation of workflows in molecular diagnostics is instrumental for coping with the large number of tests ordered by clinicians, as well as providing fast-tracked rapid testing for highly urgent cases. In this study we evaluated a SARS-CoV-2 LDT for the NeuMoDx 96 system, a fully automated device performing extraction and real-time PCR.
METHODS: A publicly available SARS-CoV-2 RT-PCR assay was adapted for the automated system. Analytical performance was evaluated using in-vitro transcribed RNA and clinical performance was compared to the cobas 6800-based reference assay within the lab.
RESULTS: The Envelope (E) Gene-LDT displayed good analytical performance with an LoD of 95.55 cp/mL and no false positives during evaluation of cross-reactivity. A total of 176 patient samples were tested with both the E-Gene-LDT and the reference assay. Positive and negative agreement were 100 % and 99.2 % respectively. Invalid-rate was 6.3 %.
CONCLUSION: The E-Gene-LDT showed analytical and clinical performance comparable to the cobas6800-based reference assay. Due to its random-access workflow concept and rapid time-to-result of about 80 min, the system is very well suited for providing fast-tracked SARS-CoV-2 diagnostics for urgent clinical samples in the hospital setting.
Copyright © 2020 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  COVID-19; NeuMoDx; PCR; Rapid testing; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 32388471      PMCID: PMC7187839          DOI: 10.1016/j.jcv.2020.104390

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Introduction

In early January 2020, SARS-CoV-2 was first identified as the likely causative agent of a cluster of cases of viral pneumonia in the city of Wuhan, China [1]. The novel virus is situated in the ‘sarbecovirus’ subgenus along with its genetically distinct relative, the original SARS-coronavirus [2]. SARS-CoV-2 saw rapid spread worldwide eventually prompting the WHO to declare a ‘global health emergency’ by the end of January [3]. Outbreak scenarios present a unique challenge for diagnostic laboratories. Particularly in the case of respiratory viruses such as SARS-CoV-2, clinical symptoms can be largely indistinguishable from other common respiratory pathogens such as e.g. Influenza [4] and polymerase chain reaction (PCR) assays are necessary to confirm or rule out the novel virus [5]. A variety of suitable assays were made available early on during the outbreak, notably by Corman et al. [6] and the CDC, which were swiftly adopted by many labs in Europe and around the world. However, their overall testing capacity remained limited [7]. We and others have previously demonstrated how automation in molecular diagnostics enables easy scaling of testing capacity by substantially cutting back hands-on time for PCR-assays [8,9]. For the assay presented in this study, we used a fully automated random-access platform for molecular diagnostics, handling everything from extraction, amplification, signal detection to reporting of results [10]. For RNA targets, the time-to-result is approximately 80 min, given optimal conditions. The availability of an open mode allows for the rapid implementation of lab developed tests (LDT). The aim of this study was to adapt and evaluate a previously published SARS-CoV-2 PCR assay (by Corman et al. (6)) for the NeuModx 96 system.

Materials and methods

Envelope (E) Gene-LDT assay setup (NeuMoDx 96 system)

Primers (fwd: 5´-ACAGGTACGTTAATAGTTAATAGCmGT-3´, rev 5´-ATATTGCAGCAGTACGCACAmCA-3´) and probe (5´-Fam-ACACTAGCC/ZEN/ATCCTTACTGCGCTTCG-Iowa Black FQ-3′) used for the Envelope (E) Gene-LDT were custom made and purchased from IDT DNA Technologies (Coralville, USA). Both primers were modified with 2′-O-methyl bases in their penultimate base to prevent formation of primer dimers (mG or mC). A double-quenched probe was used in order to reduce background fluorescence. In accordance with instructions issued by the manufacturer, a 6x Primer/Probe mix was prepared and 5 μL of the mix were loaded into the LDT-Strip well by well for each reaction (e.g. 400 nM primers, 75 nM probe per reaction). For a complete run protocol see the test-summary displayed in Table 1 . Reagents and materials for extraction and PCR were from NeuMoDx inc. (Ann Arbor, USA; distributed by QIAGEN) and used according to instructions by the manufacturer.
Table 1

NeuMoDx-Software run-protocol summary displaying settings and PCR protocol.

NeuMoDx software setup:
Ct Calling Algorithm: Second Derivative
Result Type: Qualitative
Speciment Type: TransportMedium
 Specimen Aspirate Volume (μL): 400
 Lysis: 600 s. (Lysis Buffer 4)
Target: SPC2, Speciment Type: TransportMedium
 (Internal Control)
 Reporter: Yellow (530/555)
 Peak Minimum Cycle: 25
 Peak Maximum Cycle: 40
 Minimum End Point Fluorescence: 1000
 Minimum Peak Height: 10
Target: FAM (Sarbeco-E), Speciment Type: TransportMedium
 Reporter: Green (470/510)
 Peak Minimum Cycle: 25
 Peak Maximum Cycle: 40
 Minimum End Point Fluorescence: 1000
 Minimum Peak Height: 10
PCR Stage: RT (Hold, 900 s, 50 °C)
PCR Stage: InActivation (Hold, 240 s, 95 °C)
PCR Stage: Cycle (Cycle, 50 Cycles)
 Step Denature: 6 s, at 95 °C, No Detect
 Step Anneal: 19 s, at 60 °C, Detect
NeuMoDx-Software run-protocol summary displaying settings and PCR protocol.

Assessment of Limit of detection and intra-/inter-run variability

For analytical evaluation, in-vitro transcribed RNA (IVT-RNA) of the viral E-gene was generated as described previously [6] using the following primers: 5´- TACTAATACGACTCACTATAGATACAGGTACGTTAATAGTTAATAGCGT-3´ and 5´-ttttttttgtatacATATTGCAGCAGTACGCACACA-3´. IVT-RNA was adjusted for copy-numbers to a predefined RNA standard obtained from “European virus archive” (EVA), (https://www.european-virus-archive.com). A total of 4 different concentrations (400, 100, 40 and 10 copies/mL) and negative control, 8 replicates each, were used to determine Limit of detection (LoD) by probit-analysis (MedCalc, MedCalc Software Ltd). Inter-run and intra-run variability were evaluated using spiked swab samples containing IVT-RNA at approximately 5× and 10× LoD, running 5 repeats each on two different days.

Evaluation of cross-reactivity

In order to rule out potential cross-reactivity with other organisms present in respiratory swabs, a set of predetermined clinical samples containing a variety of respiratory pathogens and external quality assessment panel samples (INSTAND e.V., Germany) were selected and subjected to the E-Gene-LDT.

Comparing of clinical performance

Clinical specimens used for this study were oropharyngeal and nasopharyngeal swabs (E-Swab collection kits, Copan, Italy). Prior to analysis, 1 mL Roche cobas PCR medium (≤ 40 % guanidine hydrochloride in Tris-HCL buffer) was added to the sample for pre-lysis and inactivation. Samples were then briefly vortexed before being loaded into the instrument. A total of 176 clinical were prepared, split into aliquots and tested in parallel on both systems. Samples that did not yield valid results on the NeuMoDx system are reported as “Invalid”.

Results

Assessment of analytical performance

Limit of detection was determined as 95.55 cp/mL at 95 % probability of detection (CI 63.56 cp/mL – 241.46 cp/mL). Intra-/inter-run variability yielded median Ct values of 27.045 (+/- 0.695 ct) and 27.640 (+/- 1.14 ct) for 10× LoD and 5× LoD respectively. This is in line with previously published data for comparable assays [11]. There were no false positive results in cross-reactivity experiments indicating solid specificity, see Table 2 .
Table 2

Clinical samples and external quality control samples (provided by INSTAND e.V., Düsseldorf, Germany) were tested for potential cross-reactivity with the E-Gene-LDT.

Positive clinical samplesNumberResult
hCoV 229E2Negative
hCoV HKU12Negative
Influenza A3Negative
Influenza A H1N12Negative
Influenza B2Negative
RSV3Negative
Rhino-/Enterovirus2Negative
Human Metapneumovirus2Negative
Parainfluenzavirus 31Negative
Adenovirus1Negative
Boca-virus2Negative
Mycoplasma pneumoniae1Negative
Chlamydophila pneumoniae1Negative
Pneumocystis jirovecii1Negative
External quality assessment panels (INSTAND)
MERS Coronavirus2Negative
hCoV NL631Negative
hCoV 229E1Negative
hCoV OC431Negative
Parainfluenzavirus 21Negative
Parainfluenzavirus 31Negative
Total number tested:32
Clinical samples and external quality control samples (provided by INSTAND e.V., Düsseldorf, Germany) were tested for potential cross-reactivity with the E-Gene-LDT.

Comparing clinical performance

Clinical performance of the assay was analyzed by comparing the E-Gene-LDT to the reference method within the lab, the cobas6800-based “SARS-CoV-2 UCT” assay [11]. Inhibition rate was 6.3 % (11/176 samples, all of which were tested negative in the reference assay). Positive agreement was 100 % (35/35) and negative agreement was 99.2 % (129/130), see Table 3 . Median ct of clinical samples was 25.15 (IQR 20.87–27.04) for the NeuMoDx E-Gene-LDT and 28.16 (IQR 24.36–31.66) for the cobas6800 SARS-CoV-2 UCT. A single discrepant sample occurred, returning positive on the NeuMoDx system (ct 28.73, close to LoD) and negative on the cobas6800. Root cause investigation revealed that this patient had previously been diagnosed with COVID-19 elsewhere.
Table 3

Results of clinical samples were compared between the novel E-Gene-LDT and the routine assay (SARS-CoV UCT on the cobas6800 system).

SARS-CoV UCT cobas6800
PositiveNegative
E-Gene-LDT NeuMoDx 96Positive351
Negative0129
Invalid011
Total number:176
Results of clinical samples were compared between the novel E-Gene-LDT and the routine assay (SARS-CoV UCT on the cobas6800 system).

Discussion

Reliance on manual PCR setups is one of the fundamental limitations in molecular diagnostics when it comes to scalability and speed during outbreak scenarios such as the current SARS-CoV-2 situation. A study by Reusken et al. reported readiness to test for the novel Coronavirus by the end of January 2020 in almost all countries of the European union, but with a capacity of 250 tests per week or less for the vast majority of them [7]. Similar issues were reported early on in China, where testing could not be performed for all suspected cases due to limitations in capacity [12]. In a recent study we demonstrated that a previously published TaqMan based SARS-CoV-2 RT-PCR assay, endorsed by ECDC and WHO, can be adapted to run on an automated batch-based high-throughput system, the cobas6800 [11]. Incidentally, Roche recently released their own SARS-CoV-2 assay for this system under “Emergency Use Authorization” by the FDA [13]. However, taking into consideration sample registration, pretreatment, preparation of batches, and generating reports, it usually takes more than 5 h before results can be made available to clinicians [14]. Consequently, alternative workflows are required to enable fast-tracking of high-priority samples. The NeuMoDx 96 system is a fully automated RT-PCR platform, performing extraction, amplification and signal detection without requiring any human interaction. it provides random-access capabilities, turn-around times of 80 min for RNA targets and a throughput of 144 samples/8 h [15]. In this study we have adapted the SARS-CoV-2 RT-PCR assay by Corman et al. [6] for use on the NeuMoDx 96 automated system. Analytical and clinical performance was comparable to the cobas6800-based reference assay [11], showing an LoD of approximately 100 copies/mL and positive and negative agreement of 100 % and 99.2 % respectively. The relatively high inhibition rate of 6.3 % suggests that sample preparation procedures can be further optimized. During the preparation of this manuscript, several commercial assays were released offering rapid random-access testing (< 80 min), including Xpert Xpress (Cepheid), QIAstat-Dx (QIAGEN) and ID NOW COVID-19 (Abbott) [16,17], however clinical evaluation of these assays is not yet available in peer-reviewed literature.

Conclusion

In this study we have adapted a publicly available SARS-CoV-2 screening assay for use on the open mode of the NeuMoDx 96 system. The assay demonstrates comparable analytical and clinical performance to established LDTs currently in use for SARS-CoV-2 diagnostics. Due to its random-access capabilities and short turn-around times (80 min), the system is well suited for automating medium-throughput routine SARS-CoV-2 testing, or as an addition to high-throughput systems to allow fast-tracking for highly urgent clinical samples.

Authors contribution

ML, SP, MA, NF, AS, SK, and UM conceptualized and supervised the study. DN performed the experiments. DN, ML and SP wrote the initial draft of the manuscript. All authors agreed to the publication of the final manuscript.

Declaration of Competing Interest

All authors declare no conflict of interest.
  12 in total

1.  Clinical evaluation of multiplex RT-PCR assays for the detection of influenza A/B and respiratory syncytial virus using a high throughput system.

Authors:  Ulrich Eigner; Svenja Reucher; Nadine Hefner; Sandrine Staffa-Peichl; Melissa Kolb; Ulrike Betz; Martin Holfelder; Gene Spier; Susanne Pfefferle; Marc Lütgehetmann
Journal:  J Virol Methods       Date:  2019-04-01       Impact factor: 2.014

Review 2.  Ten years of R&D and full automation in molecular diagnosis.

Authors:  Gilbert Greub; Roland Sahli; René Brouillet; Katia Jaton
Journal:  Future Microbiol       Date:  2016       Impact factor: 3.165

3.  Generating timely molecular diagnostic test results: workflow comparison of the cobas® 6800/8800 to Panther.

Authors:  Gudrun Aretzweiler; Susanne Leuchter; Christian O Simon; Ed Marins; Andre Frontzek
Journal:  Expert Rev Mol Diagn       Date:  2019-09-17       Impact factor: 5.225

4.  Multicenter Evaluation of NeuMoDx Group B Streptococcus Assay on the NeuMoDx 288 Molecular System.

Authors:  C L Emery; R F Relich; T H Davis; S A Young; M D Sims; B L Boyanton
Journal:  J Clin Microbiol       Date:  2019-01-30       Impact factor: 5.948

5.  Clinical Evaluation of the cobas SARS-CoV-2 Test and a Diagnostic Platform Switch during 48 Hours in the Midst of the COVID-19 Pandemic.

Authors:  Mario Poljak; Miša Korva; Nataša Knap Gašper; Kristina Fujs Komloš; Martin Sagadin; Tina Uršič; Tatjana Avšič Županc; Miroslav Petrovec
Journal:  J Clin Microbiol       Date:  2020-05-26       Impact factor: 5.948

6.  Early Transmission Dynamics in Wuhan, China, of Novel Coronavirus-Infected Pneumonia.

Authors:  Qun Li; Xuhua Guan; Peng Wu; Xiaoye Wang; Lei Zhou; Yeqing Tong; Ruiqi Ren; Kathy S M Leung; Eric H Y Lau; Jessica Y Wong; Xuesen Xing; Nijuan Xiang; Yang Wu; Chao Li; Qi Chen; Dan Li; Tian Liu; Jing Zhao; Man Liu; Wenxiao Tu; Chuding Chen; Lianmei Jin; Rui Yang; Qi Wang; Suhua Zhou; Rui Wang; Hui Liu; Yinbo Luo; Yuan Liu; Ge Shao; Huan Li; Zhongfa Tao; Yang Yang; Zhiqiang Deng; Boxi Liu; Zhitao Ma; Yanping Zhang; Guoqing Shi; Tommy T Y Lam; Joseph T Wu; George F Gao; Benjamin J Cowling; Bo Yang; Gabriel M Leung; Zijian Feng
Journal:  N Engl J Med       Date:  2020-01-29       Impact factor: 176.079

7.  Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.

Authors:  Victor M Corman; Olfert Landt; Marco Kaiser; Richard Molenkamp; Adam Meijer; Daniel Kw Chu; Tobias Bleicker; Sebastian Brünink; Julia Schneider; Marie Luisa Schmidt; Daphne Gjc Mulders; Bart L Haagmans; Bas van der Veer; Sharon van den Brink; Lisa Wijsman; Gabriel Goderski; Jean-Louis Romette; Joanna Ellis; Maria Zambon; Malik Peiris; Herman Goossens; Chantal Reusken; Marion Pg Koopmans; Christian Drosten
Journal:  Euro Surveill       Date:  2020-01

8.  In Vitro Diagnostic Assays for COVID-19: Recent Advances and Emerging Trends.

Authors:  Sandeep Kumar Vashist
Journal:  Diagnostics (Basel)       Date:  2020-04-05

Review 9.  Diagnostic Testing for Severe Acute Respiratory Syndrome-Related Coronavirus 2: A Narrative Review.

Authors:  Matthew P Cheng; Jesse Papenburg; Michaël Desjardins; Sanjat Kanjilal; Caroline Quach; Michael Libman; Sabine Dittrich; Cedric P Yansouni
Journal:  Ann Intern Med       Date:  2020-04-13       Impact factor: 25.391

10.  Evaluation of a quantitative RT-PCR assay for the detection of the emerging coronavirus SARS-CoV-2 using a high throughput system.

Authors:  Susanne Pfefferle; Svenja Reucher; Dominic Nörz; Marc Lütgehetmann
Journal:  Euro Surveill       Date:  2020-03
View more
  30 in total

Review 1.  COVID-19 Pandemic: from Molecular Biology, Pathogenesis, Detection, and Treatment to Global Societal Impact.

Authors:  Shivani Sood; Vaishali Aggarwal; Diwakar Aggarwal; Sushil K Upadhyay; Katrin Sak; Hardeep Singh Tuli; Manoj Kumar; Jayant Kumar; Shivangi Talwar
Journal:  Curr Pharmacol Rep       Date:  2020-07-27

Review 2.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

3.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Sarah Berhane; Melissa Taylor; Ada Adriano; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Julie Domen; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; Matthew Df McInnes; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2021-03-24

4.  SARS-CoV-2 sample-to-answer nucleic acid testing in a tertiary care emergency department: evaluation and utility.

Authors:  Pia Jokela; Anu E Jääskeläinen; Hanna Jarva; Tanja Holma; Maarit J Ahava; Laura Mannonen; Maija Lappalainen; Satu Kurkela; Raisa Loginov
Journal:  J Clin Virol       Date:  2020-08-27       Impact factor: 3.168

Review 5.  The estimation of diagnostic accuracy of tests for COVID-19: A scoping review.

Authors:  Dierdre B Axell-House; Richa Lavingia; Megan Rafferty; Eva Clark; E Susan Amirian; Elizabeth Y Chiao
Journal:  J Infect       Date:  2020-08-31       Impact factor: 6.072

6.  Patient Characteristics and Clinical Course of COVID-19 Patients Treated at a German Tertiary Center during the First and Second Waves in the Year 2020.

Authors:  Thomas Theo Brehm; Andreas Heyer; Kevin Roedl; Dominik Jarczak; Axel Nierhaus; Michael F Nentwich; Marc van der Meirschen; Alexander Schultze; Martin Christner; Walter Fiedler; Nicolaus Kröger; Tobias B Huber; Hans Klose; Martina Sterneck; Sabine Jordan; Benno Kreuels; Stefan Schmiedel; Marylyn M Addo; Samuel Huber; Ansgar W Lohse; Stefan Kluge; Julian Schulze Zur Wiesch
Journal:  J Clin Med       Date:  2021-05-24       Impact factor: 4.241

7.  The handling of SARS-CoV-2 associated deaths - infectivity of the body.

Authors:  Carolin Edler; Antonia Fitzek; Jan-Peter Sperhake; Ann Sophie Schröder; Benjamin Ondruschka; Klaus Püschel; Julia Schädler; Axel Heinemann; Fabian Heinrich; Marc Lütgehetmann; Susanne Pfefferle; Martin Aepfelbacher
Journal:  Forensic Sci Med Pathol       Date:  2021-06-02       Impact factor: 2.007

8.  Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection.

Authors:  Jacqueline Dinnes; Jonathan J Deeks; Ada Adriano; Sarah Berhane; Clare Davenport; Sabine Dittrich; Devy Emperador; Yemisi Takwoingi; Jane Cunningham; Sophie Beese; Janine Dretzke; Lavinia Ferrante di Ruffano; Isobel M Harris; Malcolm J Price; Sian Taylor-Phillips; Lotty Hooft; Mariska Mg Leeflang; René Spijker; Ann Van den Bruel
Journal:  Cochrane Database Syst Rev       Date:  2020-08-26

9.  Challenges of oral medicine specialists at the time of COVID-19 pandemic.

Authors:  Arkadiusz Dziedzic; Elena M Varoni
Journal:  Oral Dis       Date:  2020-07-29       Impact factor: 4.068

10.  Analysis of Co-inhibitory Receptor Expression in COVID-19 Infection Compared to Acute Plasmodium falciparum Malaria: LAG-3 and TIM-3 Correlate With T Cell Activation and Course of Disease.

Authors:  Marissa Herrmann; Sophia Schulte; Nils H Wildner; Melanie Wittner; Thomas Theo Brehm; Michael Ramharter; Robin Woost; Ansgar W Lohse; Thomas Jacobs; Julian Schulze Zur Wiesch
Journal:  Front Immunol       Date:  2020-08-26       Impact factor: 7.561

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.