| Literature DB >> 32375358 |
Marco Salvatore1, Annalisa Amato1, Giovanna Floridia2, Federica Censi1, Gianluca Ferrari1, Fabrizio Tosto1, Rita Padoan3, Valeria Raia4, Natalia Cirilli5, Giuseppe Castaldo6, Ettore Capoluongo6, Ubaldo Caruso7, Carlo Corbetta8, Domenica Taruscio1.
Abstract
(1) Background: Diagnostic testing for cystic fibrosis (CF) is based on a sweat chloride test (SCT) considering the appropriate signs and symptoms of the disease and results of a gene mutation analysis. In 2014, the Istituto Superiore di Sanità (ISS) established a pilot Italian external quality assessment program for CF SCT (Italian EQA-SCT), which is now a third party service carried out by the ISS. (2)Entities:
Keywords: active participation; choloride; cystic fibrosis; diagnosis; external quality assessment; quality; sweat test
Year: 2020 PMID: 32375358 PMCID: PMC7246827 DOI: 10.3390/ijerph17093196
Source DB: PubMed Journal: Int J Environ Res Public Health ISSN: 1660-4601 Impact factor: 3.390
Figure 1Italian EQA-SCT scheme calendar.
Figure 2Number and regional distribution of laboratories participating within the period 2015–2019.
Characteristics of the 12 different samples dispatched between 2015 and 2019.
| 2015−16 | 2016−17 | 2017−18 | 2018−19 | ||
|---|---|---|---|---|---|
|
| Expected Chloride concentration (mEq/L) | 20 | 35,66 | 31,6 | 48 |
| Mock chloride quantity | 120 | 86 | 180 | 100 | |
| Mock age | 3 years | 30 days | 2 months | 60 years | |
| Expected interpretation | normal | borderline | borderline | borderline | |
|
| Expected Chloride concentration (mEq/L) | 40 | 54,96 | 61,6 | 69,25 |
| Mock chloride quantity | 80 | 120 | 80 | 60 | |
| Mock age | 2 months | 62 years | 42 years | 1 month | |
| Expected interpretation | borderline | borderline | pathologic | pathologic BUT | |
|
| Expected Chloride concentration (mEq/L) | 200 | 110,72 | 104,0 | 80,75 |
| Mock chloride quantity | 130 | 92 | 280 | 130 | |
| Mock age | 16 years | 20 years | 20 days | 8 years | |
| Expected interpretation | pathologic BUT | pathologic | pathologic | pathologic | |
Legend: the Italian EQA-SCT experts and coordinators dispatched samples with mock clinical information and technical data (namely “Expected Chloride concentration”, “Mock chloride quantity” and “Mock age”) to test the participant’s consistency of a sweat chloride result with the correct range and the clinical interpretation of the result (“Expected interpretation”).All the laboratories that participated in our survey, routinely adopt intra-laboratory quality control systems to monitor the accuracy and precision of their procedure. The results of patients (including those related to our survey) are considered valid and reported when the accuracy and precision parameters obtained from the control cards are < 2 SDs.
Figure 3Participation in the Italian EQA-STC.
Figure 4Total score of Laboratories not participating in all Italian EQA-SCT rounds (a) and of those participating in the 2015–2016 and 2018–2019 rounds (b).
Figure 5Average score of the eleven laboratories participating in all Italian EQA-SCT rounds.
Figure 6Average score details of the eleven laboratories participating in the Italian EQA-SCT from 2015 to 2019.