BACKGROUND: Flow diverting stent (FDS) devices have revolutionized the treatment of large and complex brain aneurysms, but there is still room for improvement, particularly on the flow diversion properties and technical challenges associated with stent deployment. In this study we compared flow diversion properties between the new generation Surpass Evolve (Stryker) and the Pipeline Flex (Medtronic) devices by quantitatively evaluating intra-aneurysmal flow modification.Entities:
Keywords: aneurysm; angiography; blood flow; flow diverter
Mesh:
Year: 2020 PMID: 32317371 PMCID: PMC7402444 DOI: 10.1136/neurintsurg-2019-015696
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Comparison between device specifications for Surpass Evolve and Pipeline Flex flow diverting stents
| Evolve | Pipeline | |
| Design |
|
|
| No. of wires | 64 | 48 |
| Type | Braided | Braided |
| Braid angle | Higher | Lower |
| Pore density | Higher | Lower |
| Wire thickness | Thinner | Thicker |
| Metal coverage | Similar | Similar |
| Composition | 80% cobalt chromium / | 75% cobalt chromium / |
Figure 1Representative views of 3D Vaso-CT imaging demonstrating wall opposition after angioplasty of both a Pipeline Flex (4.25×25 mm) and a Surpass Evolve (4.25×25 mm) stent, sequentially placed in the same patient-specific silicone aneurysm model. MAFA-R, mean aneurysm flow amplitude ratio.
Figure 2Representative mean aneurysm flow amplitude analysis displaying averaged detector velocity fields (white dots with lines representing the direction of travel) superimposed on a velocity magnitude color map (m2/s; blue=slower, red=faster) for each flow diverting stent, before and after stent placement. Aneurysm region of interest highlighted in blue. MAFA-R, mean aneurysm flow amplitude ratio.
Summary of aneurysm flow ratios between Evolve and Pipeline devices. Mean aneurysmal flow amplitude ratio values within the aneurysm sac, calculated from pre- and post Evolve and Pipeline deployment, were compared in four patient specific silicone models
| Model | Evolve | Pipeline | ||||
| MAFApre | MAFApost | Ratio | MAFApre | MAFApost | Ratio | |
| A | 2.4 | 1.6 | 0.67 | 1.9 | 1.4 | 0.74 |
| B | 4.1 | 2.4 | 0.59 | 3.9 | 2.8 | 0.72 |
| C | 3.4 | 2.4 | 0.71 | 2.3 | 1.7 | 0.74 |
| D | 4.8 | 2.4 | 0.50 | 6.3 | 3.9 | 0.62 |
| Average | 0.62 | 0.71 | ||||
| SD | ±0.09 | ±0.06 | ||||
MAFA, mean aneurysmal flow amplitude.