Literature DB >> 23686973

New generation of Flow Diverter (surpass) for unruptured intracranial aneurysms: a prospective single-center study in 37 patients.

Joost De Vries1, Jeroen Boogaarts, Anouk Van Norden, Ajay K Wakhloo.   

Abstract

BACKGROUND AND
PURPOSE: In patients harboring intracranial aneurysms, the major goal in treatment is to prevent bleeding. A new generation of an endoluminal device (Surpass Flow Diverter [Surpass]) was developed to reconstruct parent artery and occlude the aneurysm. We present our clinical and angiographic single-center experience.
METHODS: Patients with a wide range of complex unruptured aneurysms were treated with the Surpass placed in the parent artery and bridging the aneurysm. Clinical and angiographic follow-up were performed at 6 months. Data were prospectively collected.
RESULTS: Thirty seven patients (mean age, 56 years; range, 32-79), harboring 49 unruptured aneurysms were treated at our center. All except 1 patient were treated with a single device. Successful delivery of the device was achieved in all patients. All 35 nonbifurcation aneurysm necks were covered completely, whereas 14 bifurcation aneurysms were only partially covered. There was no major periprocedural morbidity or mortality. During follow-up, 4 patients (10.4%) experienced transient neurological deficit. One patient (3%) developed a minor stroke at 4-month follow-up with persistent neurological deficit. Twelve patients had neurological symptoms related to their aneurysm and 7 showed improvement of these symptoms during follow-up. At 6-month follow-up, 29 of 31 aneurysms studied that had complete neck coverage showed a complete occlusion (94%) including 1 case with a 95% to 100% occlusion, whereas 5 of the 10 bifurcation aneurysms were occluded.
CONCLUSIONS: Our study shows high safety and efficacy profile of a new generation endoluminal device in treatment of complex intracranial aneurysms. Long-term studies of treated bifurcation aneurysms are needed.

Entities:  

Keywords:  flow diverter; hemorrhage; intervention; intracranial aneurysm; stroke; treatment

Mesh:

Year:  2013        PMID: 23686973     DOI: 10.1161/STROKEAHA.111.000434

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  46 in total

Review 1.  Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review.

Authors:  Xianli Lv; Hongchao Yang; Peng Liu; Youxiang Li
Journal:  Neuroradiol J       Date:  2016-02-02

Review 2.  Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review.

Authors:  Francesco Briganti; Giuseppe Leone; Mariano Marseglia; Giuseppe Mariniello; Ferdinando Caranci; Arturo Brunetti; Francesco Maiuri
Journal:  Neuroradiol J       Date:  2015-08-27

3.  The FRED flow-diverter stent for intracranial aneurysms: clinical study to assess safety and efficacy.

Authors:  M A Möhlenbruch; C Herweh; L Jestaedt; S Stampfl; S Schönenberger; P A Ringleb; M Bendszus; M Pham
Journal:  AJNR Am J Neuroradiol       Date:  2015-02-26       Impact factor: 3.825

4.  Initial Experience with p64: A Novel Mechanically Detachable Flow Diverter for the Treatment of Intracranial Saccular Sidewall Aneurysms.

Authors:  S Fischer; M Aguilar-Pérez; E Henkes; W Kurre; O Ganslandt; H Bäzner; H Henkes
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

Review 5.  Emerging Technologies in Flow Diverters and Stents for Cerebrovascular Diseases.

Authors:  Michael Karsy; Jian Guan; Andrea A Brock; Anubhav Amin; Min S Park
Journal:  Curr Neurol Neurosci Rep       Date:  2017-10-28       Impact factor: 5.081

6.  Flow Diversion for Intracranial Aneurysm Management: A New Standard of Care.

Authors:  Fawaz Al-Mufti; Krishna Amuluru; Chirag D Gandhi; Charles J Prestigiacomo
Journal:  Neurotherapeutics       Date:  2016-07       Impact factor: 7.620

7.  Mid-term follow-up of staged bilateral internal carotid artery aneurysm treatment with Pipeline embolization.

Authors:  Juan G Tejada; Gloria Vv Lopez; Jerry Me Koovor; Kalen Riley; Mesha Martinez
Journal:  Interv Neuroradiol       Date:  2019-06-03       Impact factor: 1.610

8.  Surpass flow diverter in the treatment of intracranial aneurysms: a prospective multicenter study.

Authors:  A K Wakhloo; P Lylyk; J de Vries; C Taschner; J Lundquist; A Biondi; M Hartmann; I Szikora; L Pierot; N Sakai; H Imamura; N Sourour; I Rennie; M Skalej; O Beuing; A Bonafé; F Mery; F Turjman; P Brouwer; E Boccardi; L Valvassori; S Derakhshani; M W Litzenberg; M J Gounis
Journal:  AJNR Am J Neuroradiol       Date:  2014-08-14       Impact factor: 3.825

Review 9.  On Flow Diversion: The Changing Landscape of Intracerebral Aneurysm Management.

Authors:  A A Dmytriw; K Phan; J M Moore; V M Pereira; T Krings; A J Thomas
Journal:  AJNR Am J Neuroradiol       Date:  2019-03-20       Impact factor: 3.825

Review 10.  [Management of paraophthalmic aneurysms : Review of endovascular treatment strategies].

Authors:  P Bhogal; M Aguilar Pérez; G Sauder; H Bäzner; O Ganslandt; H Henkes
Journal:  Ophthalmologe       Date:  2018-02       Impact factor: 1.059

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.