| Literature DB >> 32289720 |
Lari M Koponen1, Stefan M Goetz2, Debara L Tucci3, Angel V Peterchev4.
Abstract
BACKGROUND: Accurate data on the sound emitted by transcranial magnetic stimulation (TMS) coils is lacking.Entities:
Keywords: Coil click; Hearing safety; Sound; Stimulation strength; TMS; Transcranial magnetic stimulation
Mesh:
Year: 2020 PMID: 32289720 PMCID: PMC7263763 DOI: 10.1016/j.brs.2020.03.004
Source DB: PubMed Journal: Brain Stimul ISSN: 1876-4754 Impact factor: 8.955
Tested TMS stimulators and coils from Magstim (UK), MagVenture (Denmark), and Neuronetics (USA).
| Stimulator | Coil |
|---|---|
| Magstim Rapid 2 (P/N 3013-US) | 70mm Double Coil (P/N 9925–00)[ |
| MagVenture MagPro X100 incl. MagOption (P/N 9016E0731)[ | CeB60 (P/N 9016E0482) |
| Neuronetics NeuroStar TMS Therapy System (ver. 1.0, P/N 81–00315-000) | NeuroStar (ver. 1.0, P/N 81–00900-000) |
Connected to stimulator with adapter (Magstim, P/N 3110e00).
“Standard biphasic” pulse mode was used.
Scaling laws for reported peak SPL and rTMS sound level.
| Effect | Change in peak SPL | Change in rTMS sound level |
|---|---|---|
| Stimulation strength | ||
| Repetition rate | 0 | |
| Distance from coil[ |
Valid down to distance of 5 cm [14].
Fig. 1.E-field and sound characteristics of various measured TMS coils. (A) Peak E-field value has approximately linear relationship with stimulator output. (B) Pulse durations of air-core coils are approximately constant, whereas pulse duration of NeuroStar iron-core coil starts to drop after about 30% MSO. (C) Stimulation strength estimated with neural model, where 50.3% MSO of Magstim 70mm Double Coil is set to correspond to 100% RMT. (D) Peak sound pressure level (SPL) with Z-weighting, i.e., flat response curve. (E) Duration for which sound is within 20 dB of peak value. (F) Simulated sound level of a 20 Hz rTMS pulse train with A-weighting (slow time weighting).
Fig. 2.1/3-octave spectra of tested coils at estimated 100% RMT (solid line) and at 100% MSO (dash-dotted line). Solid black line shows ambient sound spectrum. The spectra below 300 Hz are affected by the acoustics of the soundproof chamber (for more detail, see Supplementary material). For all coils, spectra power peak is between 1000 and 10,000 Hz, i.e.,within hearing range (20e20,000 Hz). Spectra at 100% RMT were produced from closest measured stimulator output by assuming that sound waveform is proportional to pulse energy. Due to the 0.2 s time window for each sample, the sound level corresponds to that of a 5 Hz rTMS pulse train.
Fig. 3.For each tested coil, the rTMS pulse repetition rate corresponding to an estimated airborne sound level of 85 dB(A) for the subject as a function of stimulation strength (in % of average RMT).