T Kammer1, S Beck, A Thielscher, U Laubis-Herrmann, H Topka. 1. Department of Neurobiology, Max-Planck-Institut for Biological Cybernetics, Spemannstrasse 38, D-72076, Tübingen, Germany. thomas.kammer@tuebingen.mpg.de
Abstract
OBJECTIVES: To evaluate the stimulation effectiveness of different magnetic stimulator devices with respect to pulse waveform and current direction in the motor cortex. METHODS: In 8 normal subjects we determined motor thresholds of transcranial magnetic stimulation in a small hand muscle. We used focal figure-of-eight coils of 3 common stimulators (Dantec Magpro, Magstim 200 and Magstim Rapid) and systematically varied current direction (postero-anterior versus antero-posterior, perpendicular to the central sulcus) as well as pulse waveform (monophasic versus biphasic). The coil position was kept constant with a stereotactic positioning device. RESULTS: Motor thresholds varied consistently with changing stimulus parameters, despite substantial interindividual variability. By normalizing the values with respect to the square root of the energy of the capacitors in the different stimulators, we found a homogeneous pattern of threshold variations. The normalized Magstim threshold values were consistently higher than the normalized Dantec thresholds by a factor of 1.3. For both stimulator types the monophasic pulse was more effective if the current passed the motor cortex in a postero-anterior direction rather than antero-posterior. In contrast, the biphasic pulse was weaker with the first upstroke in the postero-anterior direction. We calculated mean factors for transforming the intensity values of a particular configuration into that of another configuration by normalizing the different threshold values of each individual subject to his lowest threshold value. CONCLUSIONS: Our transformation factors allow us to compare stimulation intensities from studies using different devices and pulse forms. The effectiveness of stimulation as a function of waveform and current direction follows the same pattern as in a peripheral nerve preparation (J Physiol (Lond) 513 (1998) 571).
OBJECTIVES: To evaluate the stimulation effectiveness of different magnetic stimulator devices with respect to pulse waveform and current direction in the motor cortex. METHODS: In 8 normal subjects we determined motor thresholds of transcranial magnetic stimulation in a small hand muscle. We used focal figure-of-eight coils of 3 common stimulators (Dantec Magpro, Magstim 200 and Magstim Rapid) and systematically varied current direction (postero-anterior versus antero-posterior, perpendicular to the central sulcus) as well as pulse waveform (monophasic versus biphasic). The coil position was kept constant with a stereotactic positioning device. RESULTS: Motor thresholds varied consistently with changing stimulus parameters, despite substantial interindividual variability. By normalizing the values with respect to the square root of the energy of the capacitors in the different stimulators, we found a homogeneous pattern of threshold variations. The normalized Magstim threshold values were consistently higher than the normalized Dantec thresholds by a factor of 1.3. For both stimulator types the monophasic pulse was more effective if the current passed the motor cortex in a postero-anterior direction rather than antero-posterior. In contrast, the biphasic pulse was weaker with the first upstroke in the postero-anterior direction. We calculated mean factors for transforming the intensity values of a particular configuration into that of another configuration by normalizing the different threshold values of each individual subject to his lowest threshold value. CONCLUSIONS: Our transformation factors allow us to compare stimulation intensities from studies using different devices and pulse forms. The effectiveness of stimulation as a function of waveform and current direction follows the same pattern as in a peripheral nerve preparation (J Physiol (Lond) 513 (1998) 571).
Authors: Xiaoming Du; Fow-Sen Choa; Ann Summerfelt; Malle A Tagamets; Laura M Rowland; Peter Kochunov; Paul Shepard; L Elliot Hong Journal: Exp Brain Res Date: 2014-11-16 Impact factor: 1.972
Authors: Jack L Lancaster; Shalini Narayana; Dennis Wenzel; James Luckemeyer; John Roby; Peter Fox Journal: Hum Brain Mapp Date: 2004-08 Impact factor: 5.038
Authors: Xiaoming Du; Laura M Rowland; Ann Summerfelt; Andrea Wijtenburg; Joshua Chiappelli; Krista Wisner; Peter Kochunov; Fow-Sen Choa; L Elliot Hong Journal: Brain Stimul Date: 2018-05-04 Impact factor: 8.955