| Literature DB >> 32272943 |
Emilia Persson1,2, Christian Jamtheim Gustafsson3,4, Petra Ambolt3, Silke Engelholm3, Sofie Ceberg5, Sven Bäck3, Lars E Olsson3,4, Adalsteinn Gunnlaugsson3.
Abstract
BACKGROUND: Retrospective studies on MRI-only radiotherapy have been presented. Widespread clinical implementations of MRI-only workflows are however limited by the absence of guidelines. The MR-PROTECT trial presents an MRI-only radiotherapy workflow for prostate cancer using a new single sequence strategy. The workflow incorporated the commercial synthetic CT (sCT) generation software MriPlanner™ (Spectronic Medical, Helsingborg, Sweden). Feasibility of the workflow and limits for acceptance criteria were investigated for the suggested workflow with the aim to facilitate future clinical implementations.Entities:
Keywords: Acceptance criteria; Clinical implementation; MRI only; Prospective; Prostate; Synthetic CT
Mesh:
Year: 2020 PMID: 32272943 PMCID: PMC7147064 DOI: 10.1186/s13014-020-01513-7
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1The MR-PROTECT MRI-only study workflow divided into the following five categories: 1. Imaging, 2. Post imaging, 3. Treatment plan creation, 4. Quality assurance and 5. Treatment delivery. Included patients underwent the workflow from category 1 to 5, following the tasks within each category as indicated by the white arrows. White boxes within each category indicate tasks inherited from the conventional dual-modality workflow. Grey boxes are new tasks specific to MRI-only. Black boxes are tasks incorporated in the study workflow for QA-purposes during the implementation procedure, but will probably not be needed in future clinical routine of MRI-only. No image registration between CT and MR was needed in this MRI-only workflow to facilitate treatment planning. In a future clinical routine MRI-only workflow, the black boxes can be excluded and replaced by appropriate QA-routines not including CT-imaging. Along some tasks in category 1–3, there are green check marks, which indicate the use of an electronic questionnaire. The electronic questionnaire items are shown below each of the corresponding check marks. Definitions: magnetic resonance (MR), computed tomography (CT), large field of view (L-FoV), treatment planning system (TPS), gold fiducial marker (GFM), clinical target volume (CTV), digitally reconstructed radiograph (DRR), Hounsfield Unit (HU), quality assurance (QA)
Worst and mean observed distortion measured from March 2017 to May 2018 on the MR-scanner
| Geometric distortion over 15 months (mm) | ||||
|---|---|---|---|---|
| Radial distance from isocenter | < 100 | 100–150 | 150–200 | 200–250 |
| Mean distortion (1 SD) [range] | 0.2 (0.0) [0.1–0.2] | 0.3 (0.0) [0.2–0.4] | 0.5 (0.1) [0.4–0.7] | 1.9 (0.1) [1.7–2.0] |
| Mean of max distortion (1 SD) [range] | 0.6 (0.1) [0.4–0.7] | 0.8 (0.2) [0.6–1.1] | 1.6 (0.3) [1.4–2.6] | 7.9 (0.3) [7.4–8.6] |
MR-parameter check deviations during the study
| Deviation in MR-image parameter | Number of patients | Comment |
|---|---|---|
| Extended FoV | 7 | Used FoV intendedly defined > 44.8 cm. Accepted solution for larger patients |
| Number of image slices not as specified | 3 | Specific absorption rate (SAR) limitation transcended, due to low patient body weight |
| MR-sequence protocol order | 2 | Order of MRI-sequences in protocol were changed by operator mistakes |
| Script error | 1 | MR-parameter template non-compliance due to deviations in sequence-prescanning options |
Fig. 2Comparison of target and OAR-doses for sCT and CT dose-distributions. Deviation showed in % of the prescribed dose of 78 Gy for the DVH-parameters in the clinical protocol used for treatment planning. Outliers are represented as black circles. Three patients in the study cohort had dose deviations between sCT and CT that exceeded 2% in CTVmin, PTVD98, Rectum D10 or Rectum D15 seen as red diamonds. The remaining population had deviations below 2% for all DVH-parameters. The explanation to the deviations above 2% was rectal gas in close connection to the CTV in the CT-images. This was concluded by replacing the rectal gas for these three patients with HU = 0 in the TPS and recalculation of the plan. All deviations were after recalculation below 2% in comparison to the sCT dose-distribution and the outliers were eliminated
Global gamma pass-rates for comparison between sCT and CT dose-distributions using a 15% dose cut-off
| Gamma criteria | Gamma pass rate ± 1 SD [range] (%) |
|---|---|
| 3%/3 mm | 99.8 ± 0.2 [99.2–100] |
| 3%/2 mm | 99.7 ± 0.3 [98.9–100] |
| 2%/2 mm | 99.7 ± 0.3 [98.7–100] |
| 2%/1 mm | 99.1 ± 0.4 [98.0–99.7] |
Fig. 3Difference between sCT-CBCT and CT-CBCTregistrations (left column) and CT-kV and sCT-kV registrations (right column) are shown for the three translational directions, x (first row), y (second row) and z (third row). Used bin size were 0.25 mm
MRI protocol
| T2 Propeller (Tra, Cor and Sag) | Large FOV T2 | MEGRE | |
|---|---|---|---|
| Sequence order in protocol | Tra 1, Cor 4 and Sag 5 | 2 | 3 |
| Primary field of application | Target definition support | sCT-generation GFM-definition Target definition Risk organ definition | GFM-identification |
| Scan sequence | 2D - Periodically Rotated Overlapping Parallel Lines with Enhanced Reconstruction | 2D Fast Recovery Fast Spin Echo | 2D Multi-Echo Fast Gradient Echo |
| Slice orientation | Tra, Cor, Sag | Tra | Tra |
| Frequence encoding direction | – | Right-Left | Anterior-Posterior |
| Field of view (frequency encoding direction) (cm) | 22 | 44.8 | 24 |
| Field of view (phase encoding direction) (cm) | 22 | 31.4 | 24 |
| Matrix size (frequency encoding direction) | 352 | 640 | 164 |
| Matrix size (phase encoding direction) | 352 | 512 | 164 |
| Scan pixel size (frequency encoding direction) (mm) | 0.63 | 0.7 | 1.46 |
| Scan pixel size (phase encoding direction) (mm) | 0.63 | 0.61 | 1.46 |
| Reconstructed pixel size (frequency encoding direction) (mm) | 0.43 | 0.44 | 0.47 |
| Reconstructed pixel size (phase encoding direction) (mm) | 0.43 | 0.44 | 0.47 |
| Bandwidth per pixel (Hz) | 473 | 390 | 508 |
| Acqusition time | 04:43, 04:17, 03:19 | 07:00 | 05:36 |
| Slice thickness (mm) | 2.8, 3, 3 | 2.5 | 2.8 |
| Slice gap (mm) | 0 | 0 | 0 |
| Repetition time (ms) | 9151, 8293, 7381 | 15,000 | 1000 |
| Number of slices | 32, 29, 23 | 88 | 34 |
| Echo time (ms) | 96, 96, 109 | 96 | 2.38–23.6 |
| Inter-echo time | – | – | 3.03 ms |
| Refocusing flip angle (deg) | 120 | 130 | – |
| Echo train length | 28, 28, 32 | 15 | – |
| Number of averages | 2.10, 2, 2 | 1 | 2 |
| Number of echoes | 1 | 1 | 8 |
| Intensity correction | Yes (SCIC) | Yes (SCIC) | Yes (SCIC) |
| Intensity filter | – | None | None |
| 3D geometry correction | Not available | Yes | Yes |
| Shimming | Yes (Auto) | Yes (Auto) | Yes (Auto) |
| Flow compensation direction | – | Slice direction | – |
| RF transmit mode | Multi transmit | Multi transmit | Multi transmit |