Establishment of national diagnostic reference levels in dental cone beam computed tomography in Switzerland.

Marine Deleu1,2, Dorothea Dagassan3, Isabelle Berg4, Julie Bize1, Karl Dula5, Vincent Lenoir2, Barbara Ott6, Julien G Ott7, Francis R Verdun1, Robert Weber8, Panagiotis Ziglinas9, Cesare Zingariello2, Marta Sans Merce2.   

Abstract

OBJECTIVES: The aim of this study was to establish diagnostic reference levels (DRLs) in the field of dental maxillofacial and ear-nose-throat (ENT) practices using cone beam CT (CBCT) in Switzerland.
METHODS: A questionnaire was sent to owners of CBCTs in Switzerland; to a total of 612 institutions. The answers were analyzed for each indication, provided that enough data were available. The DRLs were defined as the 75th percentile of air kerma product distribution (PKA).
RESULTS: 227 answers were collected (38% of all centers). Third quartile of PKA values were obtained for five dental indications: 662 mGy cm² for wisdom tooth, 683 mGy cm² for single tooth implant treatment, 542 mGy cm² for tooth position anomalies, 569 mGy cm² for pathological dentoalveolar modifications, and 639 mGy cm² for endodontics. The standard field of view (FOV) size of 5 cm in diameter x 5 cm in height was proposed.
CONCLUSIONS: Large ranges of FOV and PKA were found for a given indication, demonstrating the importance of establishing DRLs as well as FOV recommendations in view of optimizing the present practice. For now, only DRLs for dental and maxillofacial could be defined; because of a lack of ENT data, no DRL values for ENT practices could be derived from this survey.

Entities:  

Keywords:  Clinical DRL; Cone beam computed tomography; Dental Radiography; Diagnostic reference level; Dose area product

Mesh:

Year:  2020        PMID: 32267774      PMCID: PMC7461735          DOI: 10.1259/dmfr.20190468

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


Introduction

Over the last 10 years, the number of cone beam CT (CBCT) devices used in dental practices or head and neck examinations has noticeably increased[1] (Figure 1). Concretely, there were about 700 installed devices in Switzerland at the beginning of 2019 (612 at the time of the survey, which translates to about 1.2 units per 10,000 inhabitants) and new devices are continually being installed.
Figure 1.

Evolution of the number of CBCTs in dental, maxilla, and ENT in Switzerland between 2010 and 2019

Evolution of the number of CBCTs in dental, maxilla, and ENT in Switzerland between 2010 and 2019 Technical progress in dentistry and in CBCT-technologies, as well as the availability of CBCT scans, has led to an increase of the latter’s use and in which large variations of patient exposure have been reported.[2] This means that even when the added value of using CBCT is clearly stated, the variety of image quality levels available (due in particular to the wide range of the available voxel sizes and detector doses) and the variability of the chosen field of view (FOV) sizes has led to heterogeneous patient exposure levels. This situation has prompted some countries to organize surveys meant to help introduce a few diagnostic reference levels (DRL) as defined by the International Commission on Radiological Protection (ICRP)[3] for the most commonly performed indications. At the moment, only two countries have proposed a set of DRL for dental CBCT. Both countries published using the global dose indicator air kerma product (PKA), with only one considering the FOV used.[4-6] It is worth mentioning that the introduction of DRL is also supported by the ESR EuroSafe imaging initiative to promote optimal patient exposures. In addition, although some CBCT indications do present certain advantages for the patient, there is still no medical consensus among practitioners for some other indications. In this context, an optimization strategy has to be established involving all interested parties and considering all the parameters that influence image quality (in particular voxel size and detector dose) and patient exposure (that is the previous parameters and the size of the FOV). The aim of this study was to propose a set of DRLs intended to help practitioners optimize their practices with data that not only focus on the dose indicator itself but also give some hint about the volume of data acquisition. In this way, practitioners should get a comparable level of image quality for a given clinical indication. To reach this objective, a working group involving dentists, medical physicists, and radiology technicians was established to define the most frequent indications and prepare a questionnaire that was then sent to all centers within Switzerland using a CBCT for the head and neck regions.

Methods and materials

Questionnaire

A peer-group of experts in CBCT of the head and neck area in Switzerland—including dentists, medical physicists and radiology technicians—was created with the aim of proposing a set of clinical indications in order to then establish DRLs related to the practice. After several meetings, the group of experts accepted a final version of the questionnaire with discipline-specific indications of dentistry, maxillofacial surgery, and otorhinolaryngology. The questionnaire contained the following information: general information about the practice, number of CBCT and panoramic X-ray examinations performed per year, if examinations were performed for other referring physicians, and if examinations on patients under age 18 were performed. A list of 19 dental and maxillofacial indications (Table 1), and 20 ENT indications (Table 2), was proposed to enable physicians to then choose the 5 most frequent indications used in their practices (dental/ENT practice or hospital). For each of these five indications, the following questions had to be answered: the FOV (expressed as: diameter x height, in cm x cm) used; the frequency of examinations (per week, month, or year); the indicated exposure parameters (air kerma product (PKA) and/or the volume CT dose index (CTDIvol) expressed in mGy cm², mGy respectively) and if possible the tube high voltage (kVp) and the total tube current exposure time product (mAs) used for the acquisition.
Table 1.

Dental and maxilla indications associated with the number of response (percentage of centre it represents) and the number of examinations per year (percentage of the practice it represents)

IndicationsNumber of responsesAnnual frequencies
(percentage)(percentage)
Wisdom teeth176 (18)5869 (24)
Implant on maxilla or mandible170 (18)5140 (21)
Implant on maxilla with sinuslift (unilateral/bilateral)152 (16)3097 (13)
Form and position anomalies of tooth and their relations with surrounding structures106 (11)1980 (8)
Modification of dentoalveolar pathologies (e.g. cyst, periodontal and periapical lesions)90 (9)1748 (7)
Virtual planning of prosthetic implant's restoration65 (7)1867 (8)
Endodontic53 (5)1002 (4)
Odontogenic tumor, bone pathologies, lesions: cysts, maxilla bone tumor30 (3)327 (1)
Orthodontics24 (2)832 (3)
Sinus pathologies24 (2)752 (3)
Dental and maxillofacial traumatology (zygomatic bone, mandible, frontal sinus, orbits, nose, maxilla sinus (pre- or post-op))19 (2)436 (2)
Periodontology16 (2)94 (0)
Bone damage of temporomandibular joint (arthrosis)14 (1)870 (4)
Foreign body localization at face and jaw level10 (1)274 (1)
Diagnostic and operating planning of complex cranial malformations (cleft palate: alveoplasty planning, dental germs localization, osteotomy)9 (1)354 (1)
Osteitis, osteomyelitis, necrosis, osteoporosis7 (1)91 (0)
Diagnostic of superior airway1 (0)10 (0)
Salivary lithiasisNo response
Representation of the lachrymal duct with contrast agent
Table 2.

Ear, nose & throat indications associated with the number of response (percentage of center it represents) and the number of examinations per year (percentage of the practice it represents)

IndicationsNumber of responsesAnnual frequency
(percentage)(percentage)
Anterior skull base
Refractory or recurrent rhinosinusitis (with or without nasal polyps)6 (18)700 (30)
Sinonasal disease of dental origin5 (15)154 (7)
Pre-operative work-up5 (15)145 (6)
Nasal obstruction, rhinorrhea, hyposmia or facial pain without clinical correlation4 (12)700 (30)
Sinonasal and skull base masses/tumors3 (9)27 (1)
Nasal fractures2 (6)3 (0)
Radiopaque foreign bodies1 (3)1 (0)
CSF leakNo response
Congenital anomalies of the nose
Nasolacrimal duct disease (local application of contrast medium)
Lateral skull base
Post-cochlear implant control3 (9)60 (3)
Pre-operative work-up2 (6)408 (17)
Cholesteatoma2 (6)168 (7)
Chronic suppurative otitis mediaNo response
Otosclerosis
Petrous apex disease
Temporal bone tumors
Temporal bone fractures
CSF leak
Congenital anomalies of the temporal bone
Radiopaque foreign bodies
Conductive hearing loss of unknown origin

CSF, cerebrospinal fluid.

Dental and maxilla indications associated with the number of response (percentage of centre it represents) and the number of examinations per year (percentage of the practice it represents) Ear, nose & throat indications associated with the number of response (percentage of center it represents) and the number of examinations per year (percentage of the practice it represents) CSF, cerebrospinal fluid.

Data analysis

Data were analyzed using Matlab software v. R2017b. First, general questions about the practice were analyzed, such as the number of centers conducting examinations for other physicians, the number of centers conducting examinations on pediatric patients, as well as a comparison between the number of CBCTs vs panoramic X-rays per year. The dosimetric quantities displayed by the different devices were then analyzed in order to choose which should be considered for establishing the DRLs. The answers concerning the five most frequent examinations reported by the users were analyzed in terms of number of responses for each indication and frequency of examination. This allowed us to define the indications where enough statistics could be obtained (Tables 1 and 2). DRL values are defined by the International Commission on Radiological Protection (ICRP)[3] as the 75th percentile of the dose indicator distribution. Thus, after deciding which was the most common dose indicator used, the first, median, and third quartiles were evaluated using all the data available without considering the number of acquisitions made in the centers. In order to orient any users willing to optimize his/her practice, it was also decided to analyze the FOV distribution for each selected indication. Thus, not only the global dose indicator would be available but also some hint about the median FOV used. These data would enable the user to determine the reason behind his/her noncompliance with the DRL. For example, if the FOV is adequate then the noncompliance might be a result of an inadequate level of image quality or a problem with image detector performance. This is similar to the philosophy used when dealing with CT imaging since the analysis of the CTDIvol/DLP (dose–length product) makes it possible to verify the adequacy of the scanned length. Among the many small, medium, or large FOV available there are variations among the various units used in data acquisition and voxel sizes. In order to correct for the differences of FOV available, the third quartile PKA values were normalized to the most used FOV according to the following relationship: Where the is the PKA given by the device, the is the FOV used by the physician for the examination and the is the FOV chosen to normalize the PKA to the most used FOV.

Results

Despite several contact attempts, only 227 (38%) institutes out of the 612 contacted answered the questionnaire. Almost half of the responding centers perform examinations for other referring physicians [107/110 (yes/no)] and most institutes do pediatric examinations [121/96 (yes/no)]. The total number of annual panoramic X-rays (approximately 51,000) is higher than CBCT examinations (23,000) in those centers functioning with both devices. The analysis of the data showed that 7 indications of the 39 proposed in the list represented 76.4% of the total examinations number (20,703/27,109), dental, maxillofacial and ENT indications confounded. Each of these 7 indications represented more than 1000 examinations per year. The following seven proposed indications appeared to be particularly relevant: Wisdom teeth, Implant on maxilla or mandible, Tooth form and position anomalies and their relationship to surrounding structures, Modification of dentoalveolar pathologies (e.g. cyst, periodontal, and periapical lesions), Endodontics, Implant on maxilla with sinuslift (unilateral/bilateral), Virtual planning of restoration of prosthetic implant. From our results and as found in the literature,[7,8] PKA is displayed by most of the devices. The choice of PKA as a dosimetric quantity was consistent with actual tendencies and necessary to keep the maximum amount of data. 39 centers had to be excluded because PKA values were not provided. For each of the indications studied, we established a frequency distribution of the PKA values. See Table 3, leading to the following third quartile values: wisdom teeth (662 mGy cm2), implant on maxilla or mandible (683 mGy cm2), tooth form and position anomalies and their relationship to surrounding structures (542 mGy cm2), modification of dentoalveolar pathologies (e.g. cyst, periodontal, and periapical lesions) (569 mGy cm2), endodontics (639 mGy cm2).
Table 3.

PKA (mGy.cm) values obtained for the five indications chosen

IndicationsFirst quartileSecond quartileThird quartile
Wisdom tooth215421662
Implant on maxilla or mandible218442683
Form and position anomalies of tooth and their relations with surrounding structures203342542
Modification of dentoalveolar pathologies (e.g. cyst, periodontal and periapical lesions)259421569
Endodontics297421639

PKA, air kerma product.

PKA (mGy.cm) values obtained for the five indications chosen PKA, air kerma product. For each of the indications studied, we also established a frequency distribution of the FOV size, as shown in Figure 2. The most used FOV in practice was 25 cm2 (comprising values of FOV between 20 and 25 cm2) (n = 159, 40.6%), except for the indications of implant planning on maxilla or mandible with sinuslift (unilateral/bilateral) which had an FOV larger than 25 cm² in 64.2% (n = 43/67) of the cases and virtual planning of prosthetic implant restoration which had an FOV larger than 25 cm² in 86.2% (n = 25/29) of the cases.
Figure 2.

Field of view area (in cm²) occurrence for each of the 7 indications studied.

Field of view area (in cm²) occurrence for each of the 7 indications studied. Unfortunately, the survey did not gather enough information for subindications regarding implants on maxilla with sinuslift (unilateral/bilateral) or virtual planning of prosthetic implant restoration. Indeed, each subindication would imply the use of a different FOV. We therefore could not establish a reliable DRL value for these 2 indications, which represented 217 of the 999 responses. Figure 3 (a) and (b) show the histograms and the box plots of the distribution of the PKA values obtained for the wisdom tooth indication. Figure 3 (c) and (d) show the histograms and the box plots of the PKA values normalized to the most used FOV. Similar histograms were established for each of the indications studied.
Figure 3.

(a, b) show the histograms and the box plots of the distribution of the PKA values obtained for the wisdom tooth indication. Figure 3 (c,d) show the histograms and the box plots of the PKA values normalized to the most used FOV. Similar histograms were established for each of the indications studied. FOV, field of view; PKA, air kerma product.

(a, b) show the histograms and the box plots of the distribution of the PKA values obtained for the wisdom tooth indication. Figure 3 (c,d) show the histograms and the box plots of the PKA values normalized to the most used FOV. Similar histograms were established for each of the indications studied. FOV, field of view; PKA, air kerma product.

Discussion

Our survey results show that panoramic X-rays are still the reference examination in dental practice, indicating that Swiss dentists and maxillofacial surgeons are still favoring this imaging technique in which patient exposure is lower than in CBCT.[9,10] Due to the limited number of CBCT units compared to standard X-ray units (11,517 panoramic devices installed in Switzerland) used for dental care, it seems evident that there are more panoramic X-rays performed in Switzerland than CBCT examinations. As mentioned previously, more than 40% of all CBCT acquisitions performed used a FOV of about 25 cm². We found this result quite encouraging since it showed that an optimization process concerning the choice of the acquisition volume was already being performed. Most indications contain the visualization of anatomical landmarks like wisdom teeth, especially their root tips, the mandibular canal, or the outer margin of the alveolar bone in the maxilla or mandible. As CBCT systems are also used for pediatric examinations, it is interesting to note that some constructors add pediatric FOV such as 4 x 4 cm, as well as propose adapted mA values as recommended in the literature.[11-15] In our study, because the indication of CBCTs on wisdom teeth may also be used for pediatrics, this might be the reason why so many CBCT users responded that they acquire child CBCT scans. Nevertheless, pediatrics protocols still need to be standardized since smaller FOV and adapted protocol are not available on several devices. The indications studied were all dental indications. For ENT imaging, the number of responding centers was very low (only six centers answered, and with incomplete data). The lack of data might come from the fact that the survey asked for the five indications most frequently used in dental and ENT confounded; the responses showed that dental applications were more frequent than ENT applications in the institutes, which were mostly centers specialized in dental indications. Nevertheless, since the number of annual ENT examinations reached 700 for those few institutes performing ENT examinations, it could be interesting to organize a similar survey focused only on specific ENT indications. The obtained FOV were classified into different groups based on the literature[14,16,17]: small for FOV ≤40 cm2, medium for FOV between 40 and 100 cm2, and large for FOV ≥ 00 cm2; examples of FOVs for each category are available in Table 4. In our survey, since the indications were all dental, the small FOV category and the 5 x 5 cm FOV were the most used corresponding to a single tooth acquisition. The large FOV is not recommended in dental practice as the FOV must be limited to the region of interest.[14] This category is reserved for ENT practices where imaging the whole skull may be needed.
Table 4.

FOVs examples for the three different categories

CategoriesCommon FOVs in cm Ø x cm height (size in cm2)
S: Small4 × 4 (16 cm2)5 × 5 (25 cm2)6 × 6 (36 cm2)8 × 5 (40 cm2)
M: Medium5 × 10 (50 cm2)8 × 8 (64 cm2)8 × 9 (72 cm2)
L: Large10 × 10 (100 cm2)15 × 15 (225 cm2)17 × 23 (391 cm2)

FOV, field of view.

FOVs examples for the three different categories FOV, field of view. In spite of a certain FOV homogeneity, a large dose indicator distribution was obtained for all the indications studied. This mostly depends on the device used and on the parameters chosen.[11,18] It is important to mention that devices propose a limited choice of parameters and all users seem to use similar default values. For other devices, a wide range of exposure can be delivered (up to a factor of three) even when the FOV is kept constant by varying the kV, mA, and voxel size parameters.[16,19] The outcome of this study is a set of DRL values for Switzerland. Moreover, third quartile PKA values related to the most frequently used FOV have been derived. Indeed, in order to enable dentists to optimize their imaging techniques when dealing with CBCT examinations, third quartile PKA values normalized to the recommended FOV following equation 1 as presented in Table 5 have been provided.
Table 5.

Proposed DRLs and optimization values (third quartile normalized to the FOV) for Switzerland

IndicationsThird quartile values (mGy.cm2) normalized to a FOV of 5 cm Ø x 5 cm heightDRL proposal (mGy.cm2)
Wisdom tooth439662
Single implant on maxilla and mandible447683
Form and position anomalies of tooth* and their relations with surrounding structures451542
Modification of dentoalveolar pathologies (e.g. cyst, periodontal and periapical lesions)459569
Endodontics639639

DRL, diagnostic reference level; FOV, field of view.

Proposed DRLs and optimization values (third quartile normalized to the FOV) for Switzerland DRL, diagnostic reference level; FOV, field of view. At the moment, national surveys on CBCT have only been done in the UK and Finland.[2,4-6] For the indication of a single implant, UK obtained a value of 265 mGy cm² for adults with an average FOV area of 67 cm². Both countries decided to use the data without adapting the values to the most used FOV. Historically, the UK provided a value of 250 mGy cm²[6] for a recommended FOV of 4 x 4 cm. This value was deduced from the 75th percentile of the data distribution from 41 devices normalized to a FOV of 4 x 4 cm. They named this value an achievable dose since they used the normalized value. Finnish DRL[4] values are based on the Turnbull-Smith master thesis.[2] They published DRLs for four indications; from these four indications, only two are common to our study. The proposed Swiss DRLs are slightly higher than the Finnish ones: Wisdom tooth—662 mGy cm² for Switzerland (439 mGy cm² normalized to the most common FOV) compared to 380 mGy cm² for Finland; Single implant on maxilla or mandible—683 mGy cm² for Switzerland (447 mGy cm² normalized to the most common FOV) compared to 360 mGy cm² for Finland. When comparing the Swiss value for a single implant 683 mGy cm² with the UK value of 265 mGy cm² for 75th percentile and 458 for the 95th percentile, we can conclude that protocols in Switzerland need to be optimized. One of the reasons why the Swiss values are higher than the UK or Finnish values could be the use of high-quality images when it is sometimes not necessary. The values used for the comparison among different countries have been summarized in Table 6.
Table 6.

Comparison of DRLs proposal and optimization values (third quartile normalized to the FOV) with other countries DRLs (mGy cm²)

IndicationsSwiss third quartile values normalized to the recommended FOV recommended FOV)Swiss DRLFinnish DRL (4)UK achievable dose (recommended FOV) (2)UK DRL (5)
Wisdom tooth439 (5 cm Ø x 5 cm)662380
Single implant on maxilla or mandible447 (5 cm Ø x 5 cm)683360250 (4 cm Ø x 4 cm)265

DRL, diagnostic reference level; FOV, field of view.

Comparison of DRLs proposal and optimization values (third quartile normalized to the FOV) with other countries DRLs (mGy cm²) DRL, diagnostic reference level; FOV, field of view. The obtained values for PKA should be used with caution if conversion factors between PKA and effective dose are used to obtain an estimation of the effective dose. Indeed, kVp values range from 50 kV till almost 100 kV and different filtrations might be used, making it almost impossible to convert PKA to effective dose in a straightforward, linear manner. Unlike other devices, dental CBCT provide PKA values predetermined by the manufacturers instead of measured ones by a built in PKA meter. Therefore, the PKA values provided in this study might differ from the real measured values; this is a source of uncertainty that has not been considered in our analysis.

Conclusion

The use of CBCT in dental, maxillae, and ENT practices has increased over recent years in line with the increasing exposure of the Swiss population to medical examinations, which represented 1.4 mSv/year in 2013 compared to 1.0 mSv/year in 1998.[20] DRLs are a valuable tool for physicians to optimize their practice. Indeed, the choices of parameters like FOV, voxel size, and mAs have a strong influence on the dose delivered to the patient and on image quality. Yet, until now, no values were available in the field of CBCT for dental, maxillae, and ENT examinations. This first survey of dose in dental, maxilla, and ENT CBCT gave us a good initial view on how CBCT is being used in dental practice in Switzerland and will be helpful in an approach of future optimization. The DRLs in this field are quite new and a second iteration will most probably be needed. Nevertheless, Swiss DRLs are in good agreement to other existing values. For ENT, a new survey was launched in mid-2019.
  14 in total

Review 1.  Dental cone beam CT: A review.

Authors:  Timo Kiljunen; Touko Kaasalainen; Anni Suomalainen; Mika Kortesniemi
Journal:  Phys Med       Date:  2015-10-23       Impact factor: 2.685

2.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

3.  Clinical recommendations regarding use of cone beam computed tomography in orthodontics. [corrected]. Position statement by the American Academy of Oral and Maxillofacial Radiology.

Authors: 
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2013-08

4.  Determination of size-specific exposure settings in dental cone-beam CT.

Authors:  Ruben Pauwels; Reinhilde Jacobs; Ria Bogaerts; Hilde Bosmans; Soontra Panmekiate
Journal:  Eur Radiol       Date:  2016-04-23       Impact factor: 5.315

5.  A preliminary study to determine the diagnostic reference level using dose-area product for limited-area cone beam CT.

Authors:  A Endo; T Katoh; S B Vasudeva; I Kobayashi; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013-02-18       Impact factor: 2.419

6.  Effective doses from panoramic radiography and CBCT (cone beam CT) using dose area product (DAP) in dentistry.

Authors:  H S Shin; K C Nam; H Park; H U Choi; H Y Kim; C S Park
Journal:  Dentomaxillofac Radiol       Date:  2014-03-10       Impact factor: 2.419

Review 7.  Cone-beam CT in paediatric dentistry: DIMITRA project position statement.

Authors:  Anne Caroline Oenning; Reinhilde Jacobs; Ruben Pauwels; Andreas Stratis; Mihaela Hedesiu; Benjamin Salmon
Journal:  Pediatr Radiol       Date:  2017-11-15

8.  Estimation of the radiation dose for pediatric CBCT indications: a prospective study on ProMax3D.

Authors:  Maria Marcu; Mihaela Hedesiu; Benjamin Salmon; Ruben Pauwels; Andreas Stratis; Anne Caroline Costa Oenning; Mike E Cohen; Reinhilde Jacobs; Mihaela Baciut; Raluca Roman; Cristian Dinu; Horatiu Rotaru; Ioan Barbur
Journal:  Int J Paediatr Dent       Date:  2018-01-22       Impact factor: 3.455

9.  A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation dose.

Authors:  R Pauwels; O Silkosessak; R Jacobs; R Bogaerts; H Bosmans; S Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2014-04-08       Impact factor: 2.419

10.  Dosimetry analysis of panoramic-imaging devices in different-sized phantoms.

Authors:  Muizz A Wahid; Ella Choi; David S MacDonald; Nancy L Ford
Journal:  J Appl Clin Med Phys       Date:  2017-02-25       Impact factor: 2.102

View more
  2 in total

1.  Assessment of Regional Diagnostic Reference Levels in Dental Radiography in Tamil Nadu.

Authors:  Amal Jose; A Saravana Kumar; K N Govindarajan; Sunil Dutt Sharma
Journal:  J Med Phys       Date:  2022-03-31

2.  Radiographic imaging in relation to the mandibular third molar: a survey among oral surgeons in Sweden.

Authors:  Josefine Cederhag; Anna Truedsson; Per Alstergren; Xie-Qi Shi; Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2021-10-01       Impact factor: 3.573

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.