| Literature DB >> 32238662 |
Yukitoshi Ikeya1, Toshiko Nakai1, Nobuhiro Murata1, Masaki Monden1, Akihito Ogaku1, Koichiro Hori1, Ryuta Watanabe1, Masaru Arai1, Yasuo Okumura1.
Abstract
To avoid the negative effects associated with pacing, pacemakers are designed to achieve a pacing cadence as close to physiological pacing as possible. In closed-loop stimulation (CLS; a type of rate-responsive functionality used in pacemakers), the changes in impedance (which correlates with the contractility of the myocardium around the lead tip electrode) are tracked, and the paced heart rate is adjusted accordingly. We herein report a case in which we implanted a pacemaker in a post-tricuspid valve replacement patient. A ventricular lead positioned in the coronary vein exhibited good CLS functionality, and the patient's dizziness and heart failure improved.Entities:
Keywords: CLS; closed-loop stimulation; coronary vein lead; left ventricular pacing
Mesh:
Year: 2020 PMID: 32238662 PMCID: PMC7184078 DOI: 10.2169/internalmedicine.4018-19
Source DB: PubMed Journal: Intern Med ISSN: 0918-2918 Impact factor: 1.271
Figure 1.The electrocardiography at the initial visit. The patient’s heart rate was 20 beat per minute, and complete right bundle branch block (QRS: 144 ms) was observed.
Figure 2.The post-implantation electrocardiography at rest under pacing. The pacemaker settings were VVI-closed-loop stimulation (CLS) mode, basic rate of 60, and sensor rate of 120 ppm. The heart rate is 77 beat per minute, indicating that the CLS function is working.
Results of 24-hour Holter ECG Monitoring after Pacemaker Implantation.
| Total heart beats (/day) | 100,108 | |
| Maximum heart rate (bpm) | 121 | |
| Minimum heart rate (bpm) | 59 | |
| Average heart rate (bpm) | 70 | |
| Longest RR interval (sec) | 1.22 |
ECG: electrocardiography, bpm: beats per minute
Figure 3.Post-implantation X-ray images taken at the outpatient clinic after discharge. Postero-anterior (left) and lateral (right) views. A pacing lead (arrow) placed in the coronary vein (lateral branch) is shown. No findings indicating pulmonary congestion or pleural effusion are seen.