Literature DB >> 28407148

Effects of closed-loop stimulation vs. DDD pacing on haemodynamic variations and occurrence of syncope induced by head-up tilt test in older patients with refractory cardioinhibitory vasovagal syncope: the Tilt test-Induced REsponse in Closed-loop Stimulation multicentre, prospective, single blind, randomized study.

Pietro Palmisano1, Gabriele Dell'Era2, Vincenzo Russo3, Maria Zaccaria1, Rolando Mangia1, Miriam Bortnik2, Federica De Vecchi2, Ailia Giubertoni2, Fabiana Patti2, Andrea Magnani2, Gerardo Nigro3, Anna Rago3, Eraldo Occhetta2, Michele Accogli1.   

Abstract

Aims: Closed-loop stimulation (CLS) seemed promising in preventing the recurrence of vasovagal syncope (VVS) in patients with a cardioinhibitory response to head-up tilt test (HUTT) compared with conventional pacing. We hypothesized that the better results of this algorithm are due to its quick reaction in high-rate pacing delivered in the early phase of vasovagal reflex, which increase the cardiac output and the blood pressure preventing loss of consciousness. Methods and results: This prospective, randomized, single-blind, multicentre study was designed as an intra-patient comparison and enrolled 30 patients (age 62.2 ± 13.5 years, males 60.0%) with cardioinhibitory VVS, carrying a dual-chamber pacemaker incorporating CLS algorithm. Two HUTTs were performed one week apart: one during DDD-CLS 60-130/min pacing and the other during DDD 60/min pacing; patients were randomly and blindly assigned to two groups: in one the first HUTT was performed in DDD-CLS (n = 15), in the other in DDD (n = 15). Occurrence of syncope and haemodynamic variations induced by HUTT was recorded during the tests. Compared with DDD, DDD-CLS significantly reduced the occurrence of syncope induced by HUTT (30.0% vs. 76.7%; P < 0.001). In the patients who had syncope in both DDD and DDD-CLS mode, DDD-CLS significantly delayed the onset of syncope during HUTT (from 20.8 ± 3.9 to 24.8 ± 0.9 min; P = 0.032). The maximum fall in systolic blood pressure recorded during HUTT was significantly lower in DDD-CLS compared with DDD (43.2 ± 30.3 vs. 65.1 ± 25.8 mmHg; P = 0.004).
Conclusion: In patients with cardioinhibitory VVS, CLS reduces the occurrence of syncope induced by HUTT, compared with DDD pacing. When CLS is not able to abort the vasovagal reflex, it seems to delay the onset of syncope.

Entities:  

Mesh:

Year:  2018        PMID: 28407148     DOI: 10.1093/europace/eux015

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  8 in total

Review 1.  Pacing in neurocardiogenic/vasovagal syncope.

Authors:  Richard Sutton
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2018-05-25

Review 2.  [Autonomic nervous system and reflex syncope].

Authors:  Ralph Bosch
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2021-06-29

Review 3.  Pacing for Vasovagal Syncope.

Authors:  Rakesh Gopinathannair; Benjamin C Salgado; Brian Olshansky
Journal:  Arrhythm Electrophysiol Rev       Date:  2018-06

Review 4.  Pacing in vasovagal syncope: Physiology, pacemaker sensors, and recent clinical trials-Precise patient selection and measurable benefit.

Authors:  Richard Sutton; Jelle S Y de Jong; Julian M Stewart; Artur Fedorowski; Frederik J de Lange
Journal:  Heart Rhythm       Date:  2020-02-06       Impact factor: 6.343

5.  Closed-looped stimulation cardiac pacing for recurrent vasovagal syncope: A systematic review and meta-analysis.

Authors:  Pattara Rattanawong; Tanawan Riangwiwat; Pakawat Chongsathidkiet; Wasawat Vutthikraivit; Nath Limpruttidham; Narut Prasitlumkum; Napatt Kanjanahattakij; Chanavuth Kanitsoraphan
Journal:  J Arrhythm       Date:  2018-08-03

6.  Tilt testing remains a valuable asset.

Authors:  Richard Sutton; Artur Fedorowski; Brian Olshansky; J Gert van Dijk; Haruhiko Abe; Michele Brignole; Frederik de Lange; Rose Anne Kenny; Phang Boon Lim; Angel Moya; Stuart D Rosen; Vincenzo Russo; Julian M Stewart; Roland D Thijs; David G Benditt
Journal:  Eur Heart J       Date:  2021-05-01       Impact factor: 35.855

7.  Effective Pacing Intervention by Closed-loop Stimulation Using a Coronary Vein Lead in a Post-tricuspid Valve Replacement Patient.

Authors:  Yukitoshi Ikeya; Toshiko Nakai; Nobuhiro Murata; Masaki Monden; Akihito Ogaku; Koichiro Hori; Ryuta Watanabe; Masaru Arai; Yasuo Okumura
Journal:  Intern Med       Date:  2020-04-01       Impact factor: 1.271

Review 8.  Cardiac Pacing in Cardioinhibitory Reflex Syncope: Clinical Use of Closed-loop Stimulation.

Authors:  Marco Tomaino; Vincenzo Russo; Daniele Giacopelli; Alessio Gargaro; Michele Brignole
Journal:  Arrhythm Electrophysiol Rev       Date:  2021-12
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.