Closed loop stimulation is effective in improving heart rate and blood pressure response to mental stress: report of a single-chamber pacemaker study in patients with chronotropic incompetent atrial fibrillation.

Riccardo Proietti1, Gianmauro Manzoni, Luigi Di Biase, Gianluca Castelnuovo, Leonida Lombardi, Camilla Fundarò, Nadia Vegliante, Giada Pietrabissa, Pasquale Santangeli, Robert A Canby, Antonio Sagone, Maurizio Viecca, Andrea Natale.   

Abstract

INTRODUCTION: Closed-loop stimulation (CLS) is a form of rate-adaptive pacing capable of providing an effective pacing rate profile not only during physical exercise but also during mental stress. To test its effectiveness, CLS and accelerometer sensor (AS) rate response were compared intraindividually during a mental stress test (MST).
METHODS: Thirty-six patients (mean age 78.9 ± 6.4 years) implanted with a pacemaker with the CLS algorithm (Cylos, Biotronik, Berlin, Germany) underwent MSTs in different pacing configurations: nonrate-adaptive mode (VVI), AS mode (VVIR), and CLS mode, respectively. A modified Stroop test was used in order to induce mental stress. Heart rate (HR), systolic blood pressure (SBP), diastolic blood pressure, and pacing percentage burden were collected for 5 minutes before, during, and 5 minutes after the test.
RESULTS: Mean peak-HR during MST was significantly higher in CLS configuration than in VVIR and VVI modes (92.8 ± 12.6 vs 78.9 ± 6.5 vs 77.8 ± 7.5; P ≤ 0.001). The average HR increase during MST was also higher in CLS configuration than in VVIR and VVI modes (22.7 ± 16.7 vs 8.2 ± 8.6 vs 6.6 ± 6.3; P ≤ 0.001). The percentage of pacing beats during MST was higher in CLS configuration than with the other two algorithms (48.4 ± 17.9 vs 27.4 ± 17.5 vs 25.8 ± 17.6; P ≤ 0.001). The average peak-SBP was significantly higher during MST in CLS mode than in VVIR and VVI configurations (172.6 ± 15.5 vs 156.7 ± 12.2 vs 145.5 ± 13.7; P ≤ 0.001). The mean SBP increase showed a similar behavior (51.8 ± 24.7 vs 18.4 ± 13.7 vs 16.4 ± 10.3; P ≤ 0.001).
CONCLUSION: CLS algorithm in a single-chamber device is more effective than AS in detecting an hemodynamic demand due to an emotional stress and supplying a proper HR increase. These results are even more surprising compared to previous data in dual-chamber pacemakers, because they imply that CLS algorithm can provide an appropriate rate-modulation in patients with AF and chronotropic incompetence. ©2012, The Authors. Journal compilation ©2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22680238     DOI: 10.1111/j.1540-8159.2012.03445.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  12 in total

Review 1.  Physical Activity Capture Technology With Potential for Incorporation Into Closed-Loop Control for Type 1 Diabetes.

Authors:  Vikash Dadlani; James A Levine; Shelly K McCrady-Spitzer; Eyal Dassau; Yogish C Kudva
Journal:  J Diabetes Sci Technol       Date:  2015-10-18

2.  Closed loop stimulation with His bundle lead placement.

Authors:  Leonardo Marinaccio; Daniele Giacopelli; Giuliana Ginocchio; Francesco Vetta; Domenico Marchese
Journal:  J Cardiol Cases       Date:  2022-04-16

3.  A transient, closed-loop network of wireless, body-integrated devices for autonomous electrotherapy.

Authors:  Yeon Sik Choi; Hyoyoung Jeong; Rose T Yin; Raudel Avila; Anna Pfenniger; Jaeyoung Yoo; Jong Yoon Lee; Andreas Tzavelis; Young Joong Lee; Sheena W Chen; Helen S Knight; Seungyeob Kim; Hak-Young Ahn; Grace Wickerson; Abraham Vázquez-Guardado; Elizabeth Higbee-Dempsey; Bender A Russo; Michael A Napolitano; Timothy J Holleran; Leen Abdul Razzak; Alana N Miniovich; Geumbee Lee; Beth Geist; Brandon Kim; Shuling Han; Jaclyn A Brennan; Kedar Aras; Sung Soo Kwak; Joohee Kim; Emily Alexandria Waters; Xiangxing Yang; Amy Burrell; Keum San Chun; Claire Liu; Changsheng Wu; Alina Y Rwei; Alisha N Spann; Anthony Banks; David Johnson; Zheng Jenny Zhang; Chad R Haney; Sung Hun Jin; Alan Varteres Sahakian; Yonggang Huang; Gregory D Trachiotis; Bradley P Knight; Rishi K Arora; Igor R Efimov; John A Rogers
Journal:  Science       Date:  2022-05-26       Impact factor: 63.714

4.  Incorrect Holter-ECG analysis caused by the pacemaker delivering small high-frequency currents for thoracic impedance measurement.

Authors:  Masaomi Chinushi; Hitoshi Tachikawa; Yuko Chinushi; Toshio Yamaguchi; Osamu Saitoh; Takashi Tsuda
Journal:  J Cardiol Cases       Date:  2017-09-12

Review 5.  Relationships among norepinephrine levels, exercise capacity, and chronotropic responses in heart failure patients.

Authors:  Liza Grosman-Rimon; Evan Wright; Solomon Sabovich; Jordan Rimon; Sagi Gleitman; Doron Sudarsky; Alla Lubovich; Itzhak Gabizon; Spencer D Lalonde; Sharon Tsuk; Michael A McDonald; Vivek Rao; David Gutterman; Ulrich P Jorde; Shemy Carasso; Erez Kachel
Journal:  Heart Fail Rev       Date:  2022-03-24       Impact factor: 4.214

6.  Enhancing behavioral change with motivational interviewing: a case study in a Cardiac Rehabilitation Unit.

Authors:  Giada Pietrabissa; Martina Ceccarini; Maria Borrello; Gian Mauro Manzoni; Annamaria Titon; Ferruccio Nibbio; Mariella Montano; Gianandrea Bertone; Luca Gondoni; Gianluca Castelnuovo
Journal:  Front Psychol       Date:  2015-03-18

7.  Use of Biotronik closed loop pacemaker to treat recurrent syncope in pediatric patient with dysautonomia.

Authors:  Jennifer Shortland; Orhan Uzun; Deirdre Wilson; Graham A Stuart; Mark A Walsh
Journal:  HeartRhythm Case Rep       Date:  2016-09-15

8.  Atrial auto-short phenomenon as a rare cause of ventricular lead failure in a pediatric dual chamber pacemaker patient.

Authors:  Nathalie Noessler; Martin Koestenberger; Stefan Kurath-Koller
Journal:  Pacing Clin Electrophysiol       Date:  2020-02-26       Impact factor: 1.976

9.  Treating the mind to improve the heart: the summon to cardiac psychology.

Authors:  J P Ginsberg; Giada Pietrabissa; Gian Mauro Manzoni; Gianluca Castelnuovo
Journal:  Front Psychol       Date:  2015-08-04

10.  Effective Pacing Intervention by Closed-loop Stimulation Using a Coronary Vein Lead in a Post-tricuspid Valve Replacement Patient.

Authors:  Yukitoshi Ikeya; Toshiko Nakai; Nobuhiro Murata; Masaki Monden; Akihito Ogaku; Koichiro Hori; Ryuta Watanabe; Masaru Arai; Yasuo Okumura
Journal:  Intern Med       Date:  2020-04-01       Impact factor: 1.271

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.