In-Vitro MPI-guided IVOCT catheter tracking in real time for motion artifact compensation.

Florian Griese1,2, Sarah Latus3, Matthias Schlüter3, Matthias Graeser1,2, Matthias Lutz4, Alexander Schlaefer3, Tobias Knopp1,2.   

Abstract

PURPOSE: Using 4D magnetic particle imaging (MPI), intravascular optical coherence tomography (IVOCT) catheters are tracked in real time in order to compensate for image artifacts related to relative motion. Our approach demonstrates the feasibility for bimodal IVOCT and MPI in-vitro experiments.
MATERIAL AND METHODS: During IVOCT imaging of a stenosis phantom the catheter is tracked using MPI. A 4D trajectory of the catheter tip is determined from the MPI data using center of mass sub-voxel strategies. A custom built IVOCT imaging adapter is used to perform different catheter motion profiles: no motion artifacts, motion artifacts due to catheter bending, and heart beat motion artifacts. Two IVOCT volume reconstruction methods are compared qualitatively and quantitatively using the DICE metric and the known stenosis length.
RESULTS: The MPI-tracked trajectory of the IVOCT catheter is validated in multiple repeated measurements calculating the absolute mean error and standard deviation. Both volume reconstruction methods are compared and analyzed whether they are capable of compensating the motion artifacts. The novel approach of MPI-guided catheter tracking corrects motion artifacts leading to a DICE coefficient with a minimum of 86% in comparison to 58% for a standard reconstruction approach.
CONCLUSIONS: IVOCT catheter tracking with MPI in real time is an auspicious method for radiation free MPI-guided IVOCT interventions. The combination of MPI and IVOCT can help to reduce motion artifacts due to catheter bending and heart beat for optimized IVOCT volume reconstructions.

Entities:  

Year:  2020        PMID: 32231378      PMCID: PMC7108728          DOI: 10.1371/journal.pone.0230821

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


1 Introduction

Optical coherence tomography (OCT) enables a high-resolution imaging of tissue structures [1-3]. In the field of cardiovascular diseases intravascular OCT (IVOCT) imaging is applied to assess the vascular wall structures and observe plaque formations and related stenosis lengths [4, 5]. IVOCT highly benefits from a second imaging modality in order to align its catheter tip position within the global coordinate system of the patient. Using digital subtraction angiography (DSA), ionizing radiation is introduced and only 2D projections of the catheter tip positions are observed. Different methods have been presented to determine the 3D vascular shape using a combination of IVOCT and angiographic images. For example, a co-registration of both imaging modalities is applied to align the images to each other [6-9]. An improved 3D volume reconstruction method uses the information of both the vessel center line as well as the 3D catheter trajectory determined in bi-plane angiographic frames [10]. Most of the recent volume reconstruction methods assume a static imaging scenario neglecting heart beat motion, arterial vasomotion, and catheter bending leading to motion artifacts. Nevertheless, several publications depict a relevant influence of motion artifacts on the IVOCT volume reconstructions. For example, an irregular formation of stent struts are related to heart beat motion [11, 12]. In a pre-clinical scenario a setup for ECG triggered IVOCT imaging with a duration of less than one second is proposed [13], hence heart beat motion artifacts can be minimized. Micro-motor catheters are proposed in order to deal with the problem of imaging artifacts due to bending of proximally rotated catheters [13, 14]. However, the miniaturization of high-speed motors is a challenging and expensive task. Thus, a medically approved IVOCT catheter with micro motor has not been presented yet. Consequently, motion artifacts due to catheter bending and arterial vasomotion still arise in clinical scenarios and have an influence on the quantification of plaque formations. In addition, a contrast agent (iodine) is necessary for DSA imaging, which can be problematic in some patients with kidney diseases [15-17]. As an alternative, magnetic particle imaging (MPI) spatially resolves the distribution of superparamagnetic iron oxide nanoparticles (SPION) in 4D at high temporal resolution by using the particle’s non-linear magnetization characteristics [18, 19]. MPI applies static and oscillating magnetic fields to visualize the SPIONs. Thus in contrast to DSA, no ionizing radiation is induced to the patient. The changing magnetic fields are operated within the safety constraints of the peripheral nerve stimulation [20] and specific absorption rate (SAR) [21-23]. The SPIONs are biodegradable and decomposed within the liver [24, 25]. Furthermore, MPI provides 3D information over time, while DSA only provides 2D projections over time. An advanced biplane DSA measurements can be post-processed to gain a 3D information over time, which however leads to a doubled radiation exposure. Further, MPI has demonstrated its beneficial usage in several interventional applications such as catheter tracking, stenosis identification and stenosis clearing [26-28]. Catheters and guide wires are coated with magnetic markers to track their position with MPI in real time [29-33]. The first bimodal experiments combining IVOCT and MPI are presented in [34, 35]. The 3D vessel center line can be estimated from static MPI images. With the help of the estimated vessel center line the IVOCT images are oriented in 3D space to reconstruct the vessel volume. In this work, we track the IVOCT catheter by labeling its tip with an MPI visible marker. Using real time MPI imaging we get the catheter position over time allowing for motion compensation. Both imaging modalities are registered to each other using a time synchronization. An experimental setup, including a custom built IVOCT adapter, enables the generation of different catheter motion profiles. In all experiments the MPI-tracked catheter trajectory is used to reconstruct a 4D IVOCT volume. A straight 3D printed vessel phantom with integrated stenosis is imaged. In a first experiment the plausibility and statistical error of the MPI catheter tracking is analyzed using a constant catheter velocity. In the following experiments motion artifacts due to catheter bending and heart beat are simulated. The reduction of motion artifacts and their statistical deviation are analyzed. The reduced artifacts in the reconstructed volumes are shown by quantitative measurements using the Dice similarity coefficient (DICE) factor and the estimated stenosis lengths in comparison to the ground-truth shape of the phantom.

2 Materials and methods

2.1 Experimental setup

The experimental setup is composed of a pre-clinical MPI scanner [36], a custom built IVOCT imaging adapter, a spectral domain OCT system (Telesto I, Thorlabs), and a control unit as shown in Fig 1. A straight vessel phantom with an inner diameter of 2.5 mm and total length of 20 mm is positioned within the MPI field of view (FoV). A stenosis with a length of 1.5 mm and an inner diameter of 1.5 mm is integrated in the phantom (see Fig 1, CAD sketch). A 3D printer (Form 2, Formlabs) based on stereolithography is used to build the phantom out of gray resin. An IVOCT catheter (Dragonfly Duo Kit, Abbott) with an outer diameter of 0.9 mm is used. The catheter consists of an optical fiber covered by a tight and flexible protection, which can rotate freely within a hollow plastic catheter. Within the catheter tip a prism directs the infrared light to the surface. To enable a MPI-based catheter tracking, the catheter tip is coated with a thin layer of magnetic lacquer (1 μL Magneto Magnetic Lacquer, Hand & Nail Harmony), as seen in Fig 1. The lacquer dries quickly on the catheter tip. The OCT images acquired with the marked catheter provide suitable image quality, whereas the phantom structures are still apparent for later segmentation algorithms.
Fig 1

Experimental setup.

A vessel phantom with a stenosis (a) is positioned within the MPI FoV. In the CAD sketch of the phantom (bottom right), the phantom and entire stenosis dimensions are depicted. The stenosis has a diameter and length of 1.5 mm. Triggered by MPI, an IVOCT catheter (b) is rotated and pulled backwards through the phantom using a custom built adapter. The catheter tip (blue dot) is coated with magnetic lacquer (c) without covering the OCT prism. The cropped MPI FoV is highlighted with a red box.

Experimental setup.

A vessel phantom with a stenosis (a) is positioned within the MPI FoV. In the CAD sketch of the phantom (bottom right), the phantom and entire stenosis dimensions are depicted. The stenosis has a diameter and length of 1.5 mm. Triggered by MPI, an IVOCT catheter (b) is rotated and pulled backwards through the phantom using a custom built adapter. The catheter tip (blue dot) is coated with magnetic lacquer (c) without covering the OCT prism. The cropped MPI FoV is highlighted with a red box.

2.1.1 MPI acquisition parameters

For the MPI measurements a pre-clinical MPI scanner is used together with a custom-built receive coil [37]. The scanner excites the particles with three orthogonal sinusoidal excitation fields with frequencies f = 2.5/102 MHz, f = 2.5/96 MHz, and f = 2.5/99 MHz. The magnetic field strength is set to 12 mT in all three directions while the gradient strength is set to 2.0 T m-1 in z-direction and 1.0 T m-1 in the x- and y-directions. The imaging period is 21.54 ms which equals a frame rate of 46.43 Hz. The FoV has a size of 24 mm × 24 mm × 12 mm and the MPI data acquisition is conducted with the system software Paravision (Bruker). In order to reconstruct an MPI image using the frequency space approach [38], a calibration scan is required. This scan moves a small delta sample filled with SPIONs through the FoV while the system response at all attended positions is measured. The acquired data is used to set up the MPI system matrix, which characterizes the relation between the induced voltage signal and the particle distribution. In this work, the system matrix is acquired at 35 × 25 × 13 positions which cover a total volume of 35 mm × 25 mm × 13 mm. To prevent artifacts at the FoV boundaries, the calibration volume is chosen to be larger than the system FoV in all directions [39]. The delta sample has a size of 1 mm × 1 mm × 1 mm and is filled with 1 μL undiluted magnetic lacquer.

2.1.2 IVOCT acqusition parameters

The OCT system with an A-scan rate of fOCT = 91 kHz uses a central wavelength of 1315 nm. The axial OCT FoV is about 2.66 mm in air, whereas each A-scan consists of 512 pixels. The phantom is filled with distilled water yielding a pixel spacing of 4.5 μm between catheter and inner phantom wall, assuming a refractive index of 1.33. The custom-built catheter adapter enables a simultaneous rotation and translation of the IVOCT catheter. A rotational frequency of frot = 6.25 Hz is used for all experiments. The center pullback velocity v0 = −1.25 mm s−1 is varied during the experiments to simulate different motion artifacts. In all experiments, the catheter is pulled back over a total distance of s = 25 mm.

2.2 MPI image reconstruction and image processing

In frequency space MPI, the inverse problem to reconstruct an MPI image is treated with a first-order Tikhonov-regularized least-squares approach where is the MPI system matrix, is the measurement vector and is the particle-concentration vector. This least-squares problem is iteratively solved by using the Kaczmarz method. The Kaczmarz method converges quickly for nearly orthogonal matrices, which is the case for MPI [40, 41]. For the MPI reconstruction and data processing the Julia packages MPIFiles.jl [42] and MPIReco.jl [43] are used. The number of Kaczmarz iterations is set to 3 whereas the regularization parameter λ is set to λ = λ0 ⋅ 10−3, where λ0 = trace( )N−1. These reconstruction parameters have been optimized regarding the visual impression of the reconstructed MPI images.

2.2.1 MPI-Guided catheter tracking

The 4D MPI images are block averaged with a factor of two over time prior to reconstruction, which leads to a temporal resolution of fMPI = 23.2 Hz. The set of MPI images is denoted by () with I(, t) where is the position and t is the time. The catheter localization is performed in three steps as generally described for more than one marker in [44]. At first, a threshold filter is applied to each image in order to separate the marker from the background. This results in the data set with where Θ ∈ [0, 1] denotes the relative threshold. In our case the relative threshold is chosen to be Θ = 0.35. In a second step, the connected region , with the highest maximal intensity value max is identified by connected-component labeling of Iseg(Ω, t), . Finally, the position of the catheter marker is obtained by calculating the center of mass of the voxel intensities of the corresponding connected region in the MPI image I. The accuracy for this sub-voxel approach is within the sub-millimeter range and the catheter position is determined only within cropped MPI FoV robustly. The positions M1 to M2 denote the positions when the catheter enters and leaves this cropped MPI FoV. Hence, we crop the MPI FoV for later 4D reconstruction methods. In x-direction the cropped MPI FoV has a length of approximately 10 mm. Outside this cropped MPI FoV the catheter position could not be determined as robust since more image artifacts are introduced by the rotation of the catheter. These outer positions are not considered for the later reconstruction methods. Additionally, outliers are removed with a Ransac algorithm and extreme outliers are excluded via thresholding. In these extreme cases, the images are affected by noise and the localization algorithm falsely detects a high intensity noise voxel as a marker position. Further, the trajectories are smoothed to ensure a continuous trajectory.

2.3 Volume reconstruction methods

We refer to two different methods as IVOCT catheter marker tracking (MT) and input parameter (IP) based volume reconstruction, respectively. For both reconstruction methods, the inner phantom wall is segmented in the IVOCT data using a semi-automatic algorithm [34, 35, 45]. Especially in the narrowed phantom parts, some manual corrections are applied. As a result the distance r between catheter and phantom wall is given for each A-scan. 3D point clouds are generated based on the MT and IP method, whereas their envelopes are used to quantify the volume reconstructions.

2.3.1 Input parameter (IP) method

On the basis of the known input parameters of the custom built adapter (pullback and rotational speed) we take the distances r for each OCT A-scans and place a respective point in a 3D coordinate system. We assume a constant pullback and rotational velocity and align the 3D phantom boundary points on a helix with constant pitch p0 = v0/fOCT and angle θ0 = 360° ⋅ frot/fOCT.

2.3.2 Marker tracking (MT) method

Using the MPI-guided IVOCT catheter tracking we can arrange the OCT A-scans along the actual catheter trajectory. A temporal synchronization of both imaging systems allows for image registration (Fig 2). The 4D volume reconstruction method is separated in two parts. First, the OCT and MPI data sets are registered via temporal correlation. The measurements are synchronized via a trigger signal sent from MPI to the IVOCT system. The related time events can be seen on the time line in Fig 3. One second after the MPI trigger arises (ttrigger), the catheter motion profile and OCT A-scan acquisition starts (tOCT,0). The time stamps t and t are related to the MPI volumes, wherein the catheter tip enters and leaves the cropped MPI FoV. Once the catheter motion profile is finished (tOCT,end) the MPI measurement is stopped subsequently (tMPI,end).
Fig 2

Exemplary IVOCT and MPI data.

The OCT A-scans are arranged over time (top). The segmented phantom boundary is highlighted in red. For three time stamps t, the related MPI signals from catheter tip are shown within the CAD sketch (bottom).

Fig 3

Time axis for synchronizing OCT device and pullback device with the help of the MPI trigger signal.

Exemplary IVOCT and MPI data.

The OCT A-scans are arranged over time (top). The segmented phantom boundary is highlighted in red. For three time stamps t, the related MPI signals from catheter tip are shown within the CAD sketch (bottom). Then, we place points at distance r in 3D space considering both the spatial and temporal dependencies of MPI and OCT data. Due to substantial noise of the y- and z-component of the estimated 4D catheter trajectory, we only consider the x-coordinate (in pullback direction) as catheter position over time. For two successive catheter positions we determine the distance in space Δx and time ΔtMT and distribute the meanwhile acquired A-scans equidistantly. Based on the given catheter rotation frot, OCT frequency fOCT, and MPI volume rate fMPI up to four catheter positions are observed per catheter rotation. Assuming a constant catheter rotation frot, the A-scans are oriented with a fixed angle difference θ0 around the actual catheter trajectory.

2.4 Experiments

We perform three experiments with the stenosis phantom repeating each experiment three times. As a first experiment, we conduct a standard pullback profile (SP) with a constant pullback velocity v0 and a pullback distance s. As a second experiment, a bending artifact profile (BA) is used to simulate the non-linear pullback of the catheter when the catheter is decelerated due to a bending and is then suddenly accelerated due to its elastic material. At first, the catheter is pulled back with velocity v0 for the first 10 mm. Afterwards the catheter is simulated to be stuck and its velocity is set to v1 = −0.625 mm s−1 for the next 5 mm. Finally, the velocity is set back to the initial velocity v0 for the last 10 mm to simulate the elastic contraction of the catheter. The distances over time of the BA profile are shown in Fig 4a).
Fig 4

a) In case of the BA profile, the catheter is pulled backwards with a velocity v0 = −1.25 mm s−1 over the first 10 mm, then the velocity is reduced to v1 = −0.625 mm s−1 for the next 5 mm, afterwards the velocity is increased back to v0 for the last 10 mm. b) In case of HBA profile, the IVOCT catheter is pulled backwards over the first 15 mm, then the catheter is moved forward for 5 mm. Last, the catheter is pulled backwards in the original direction again for 15 mm. The catheter moves with the initial velocity v0 for all motion directions.

a) In case of the BA profile, the catheter is pulled backwards with a velocity v0 = −1.25 mm s−1 over the first 10 mm, then the velocity is reduced to v1 = −0.625 mm s−1 for the next 5 mm, afterwards the velocity is increased back to v0 for the last 10 mm. b) In case of HBA profile, the IVOCT catheter is pulled backwards over the first 15 mm, then the catheter is moved forward for 5 mm. Last, the catheter is pulled backwards in the original direction again for 15 mm. The catheter moves with the initial velocity v0 for all motion directions. As a third experiment, we perform a measurement with a heart beat motion artifact (HBA) profile. A heart beat artifact is related to the heart contraction and the relative vessel motion w.r.t. the IVOCT catheter. This artifact results in multiple acquisitions of the same blood vessel part due to a back and forth movement of the vessel (Fig 5). We use a catheter motion profile (HBA) that simulates this relative motion. The velocity is set to v0 for the first 15 mm. Subsequently, the velocity is inverted to −v0 for the following 5 mm to imitate the heart beat movement. Afterwards the velocity is adjusted back to v0 for the last 15 mm. The distance over time of the HBA profile is shown in Fig 4b).
Fig 5

Exemplary sketch of heart beat motion artifact.

Due to heart contraction the imaged artery is deformed for time stamp t2. Meanwhile, the catheter tip (blue dot) moves continuously backwards. After heart contraction (t3) the artery gets back to its original shape (t1). Again, the catheter motion is continued in between. This relative motion between catheter and artery leads to multiple IVOCT imaging of the sketched stenosis (black).

Exemplary sketch of heart beat motion artifact.

Due to heart contraction the imaged artery is deformed for time stamp t2. Meanwhile, the catheter tip (blue dot) moves continuously backwards. After heart contraction (t3) the artery gets back to its original shape (t1). Again, the catheter motion is continued in between. This relative motion between catheter and artery leads to multiple IVOCT imaging of the sketched stenosis (black). The described phantom experiments are performed in-vitro and do not involve human subjects. The datasets are acquired with the described imaging devices (pre-clinical MPI scanner, spectral-domain OCT system). The MPI data processing is implemented in Julia, while the OCT data processing is written in Matlab.

3 Results

The results are divided into three parts. First, the positions and the resulting velocities determined by 4D MPI catheter tracking are validated for three motion profiles SP, BA, and HBA. The mean absolute error (MAE) is calculated for the distance traveled only in x and for the distance traveled in x, y, z. The same is done for the velocities of the profiles. Second, we compare the IP and the MT volume reconstruction using the IVOCT and MPI data from the standard profile. The influence on both reconstruction methods in terms of bending artifacts is analyzed for the BA profile. Additionally, the HBA profile is used to investigate heart beat artifacts on both reconstruction methods. Third, the DICE factor is calculated for both reconstruction methods. In addition, the stenosis length is quantified for all reconstruction methods/profiles and compared to its ground-truth value.

3.1 Statistical validation of 4D MPI catheter tracking

For the standard profile the distance in x over time between M1 = 18 mm and M2 = 6 mm is shown in Fig 6a). From 18 mm to 11 mm the tracked x positions (black) are in good agreement with the expected x positions (red).
Fig 6

a) MPI measurements for standard profile: The measured distance in x over time between 18 mm and 11 mm is in good agreement with the expected values. Expect in the last part the positions in x marginally deviate. b) MPI measurements for BA profile: The first change of velocity is not captured by the measurement. After the first change the measured distances in x over time are in agreement with the different velocity v0 and v1 with only slight deviations.

a) MPI measurements for standard profile: The measured distance in x over time between 18 mm and 11 mm is in good agreement with the expected values. Expect in the last part the positions in x marginally deviate. b) MPI measurements for BA profile: The first change of velocity is not captured by the measurement. After the first change the measured distances in x over time are in agreement with the different velocity v0 and v1 with only slight deviations. Between 11 mm to 6 mm the tracked x positions (black) seem to diverge slightly from the expected values (red). The mean values are used to fit a regression line (blue). The MAE for the distance in x is 0.44 mm ± 0.44 mm. The absolute error (AE) of the velocity using the regression line is 0.21 mm s-1 with a relative error (RE) of 16.8% as given in Table 1. For the BA profile the distance in x over time between 18 mm and 6 mm is presented in Fig 6b). Overall the tracked x positions (black) are in good agreement with the expected x positions (red). Only in the first segment a small deviation is visible. Again, all three measurements are shown as a box plot and illustrate the distribution of the tracked positions. The mean values are used to determine a regression line (blue).
Table 1

The mean absolute error (MAE) is given for the distance in x-direction with its standard deviation (SD).

Additionally, the absolute error (AE) along with the relative error (RE) of the velocity using the 3D regression line between t and t is also reported.

ErrorsStandard profile
MAE Trajectory 1D-x [mm]0.44 ± 0.44
AE (RE) Velocity [mm/s]0.21 (16.8%)

The mean absolute error (MAE) is given for the distance in x-direction with its standard deviation (SD).

Additionally, the absolute error (AE) along with the relative error (RE) of the velocity using the 3D regression line between t and t is also reported. The MAE for the distance in x is 0.26 mm ± 0.16 mm for the first segment, 0.35 mm ± 0.11 mm for the second segment and 0.20 mm ± 0.22 mm for the third segment. The AE and RE regarding the velocity of the regression in all three segments for the BA profile are given in Table 2.
Table 2

The mean absolute error (MAE) for the distance in x-direction for the BA profile is given with its standard deviation (SD).

The AE along with the RE of the velocity using the 3D regression line between t and t is also reported.

ErrorsBA profile segment 1BA profile segment 2BA profile segment 3
MAE Trajectory 1D-x [mm]0.26 ± 0.160.35 ± 0.110.20 ± 0.22
AE (RE) Velocity [mm/s]0.44 (35.4%)0.07 (10.8%)0.22 (17.9%)

The mean absolute error (MAE) for the distance in x-direction for the BA profile is given with its standard deviation (SD).

The AE along with the RE of the velocity using the 3D regression line between t and t is also reported. For the HBA profile the distances in x over time between 18 mm and 6 mm are shown in Fig 7a. Overall, the tracked x positions resemble the movement of the catheter being pulled back and forth. The turning points between velocities −v0, v0 and again −v0 can be clearly identified. However, in the first two segments the velocity is underestimated as the tracked x positions are not in full accordance with the expected x positions (red). In the third segments the tracked x positions agree with the expected values. The distances in y and z over time are presented in Fig 7b and 7c) and shows that the stenosis phantom has been inserted slightly diagonal as the y-values increase and the z-values decrease depending on the x-position. For a straight insertion we would expect a straight line in both dimensions. Only at the time points when the velocities change the tracked positions in x differ from the expected positions in x. The three measurements are depicted as box plots to show the distribution of the measurements. The regression lines for each segment are plotted in blue. The mean absolute error for the distance in x is 0.64 mm ± 36 mm for the first segment, 0.51 mm ± 55 mm for the second segment and 0.38 mm ± 45 mm for the third segment. In Fig 7d) the velocity in x over time is shown and the inversion of the velocity is visible. The absolute and relative error of the velocity using the regression line in x are 0.38 mm/s (30.0%) for the first segment, 0.49 mm/s (39.4%) for the second segment and 0.04 mm/s (3.2%) for the third segment. The errors regrading the HBA profile are given in Table 3.
Fig 7

a) The measured distances in x agree with the set catheter movement. The turning points are clearly visible. The velocity, however, is underestimated in the first two segments. In the third segment the measured distances in x agree with the expected positions in x. b) The measure distance in y and c) z shows that the stenosis phantom is inserted slightly diagonal and the back and forth movement is also noticeable in the y and z dimension. d) The inversion of the velocity is visible and the mean velocity value are within the range of the expected velocities. However, the spread of the velocity is quite high.

Table 3

The mean absolute error for the distance in x-direction with its standard deviation (SD) is presented for the HBA profile.

The absolute and relative error of the velocity using the 3D regression line between t and t is also reported.

ErrorsHBA profile segment 1HBA profile segment 2HBA profile segment 3
MAE Trajectory 1D-x [mm]0.64 ± 0.360.51 ± 0.550.38 ± 0.45
AE (RE) Velocity [mm/s]0.38 (30.0%)0.49 (39.4%)0.04 (3.2%)
a) The measured distances in x agree with the set catheter movement. The turning points are clearly visible. The velocity, however, is underestimated in the first two segments. In the third segment the measured distances in x agree with the expected positions in x. b) The measure distance in y and c) z shows that the stenosis phantom is inserted slightly diagonal and the back and forth movement is also noticeable in the y and z dimension. d) The inversion of the velocity is visible and the mean velocity value are within the range of the expected velocities. However, the spread of the velocity is quite high.

The mean absolute error for the distance in x-direction with its standard deviation (SD) is presented for the HBA profile.

The absolute and relative error of the velocity using the 3D regression line between t and t is also reported.

3.2 Volume reconstructions

In Fig 8 the 4D volume reconstructions are compared for all motion profiles and both reconstruction methods. The volumes are shown with x cropped to the MPI FoV. A ground-truth volume with boundary information created by the parameters from the CAD sketch is depicted as a reference. The 4D boundary points are colored related to the underlying time, whereas the color map is shifted with respect to the time values of the positions . The envelopes of all volume reconstructions show deviations compared to the ground-truth volume. The stenosis lengths are highlighted with red arrows. The MT reconstruction method leads to stenosis lengths and relative positions that are almost equal to the ground-truth volume for all motion profiles. The IP reconstruction method shows a larger deviation of the stenosis relative position. Furthermore, an obvious deviation of the depicted stenosis length using the IP volume reconstruction method are depicted for the BA and HBA profiles. Especially, for the BA profile with underlying deceleration of the catheter, the length is obviously increased.
Fig 8

Reconstructed volumes for all motion profiles w.r.t. the ground-truth volume (top) for the cropped MPI FoV.

The distances x = 18 and x = 8 mm correspond to the time points tM1 and tM2. The IP and MT volume reconstructions are labeled (left). The phantom boundary points are colored w.r.t. the time color map (right). The stenosis lengths are depicted with red arrows.

Reconstructed volumes for all motion profiles w.r.t. the ground-truth volume (top) for the cropped MPI FoV.

The distances x = 18 and x = 8 mm correspond to the time points tM1 and tM2. The IP and MT volume reconstructions are labeled (left). The phantom boundary points are colored w.r.t. the time color map (right). The stenosis lengths are depicted with red arrows. In order to consider the complete pullback time for the BA and HBA profile, the related 4D volumes without cropping the x-axis are shown in Figs 9 and 10, respectively. Considering the BA profile (Fig 9), the overall IP volume results in an increased length with constant helical pitch p0. In contrast, the MT volume does not overestimate the total volume and especially the stenosis length. The varying catheter velocity, as depicted in Fig 6b), is apparent for the MT method by different densities of boundary points between and x2 = 10.2 mm compared to the points between x2 and x3 = 8 mm. In case of the HBA profile (Fig 10), the IP volume also shows a relevant overestimation of the total volume. Furthermore, in the volume reconstruction beyond x = 8 mm a second stenosis appears. The MT volume again presents an improved reconstruction method. Considering the tracked catheter motion, the 3D boundary points are arranged over time such that several boundary points overlay each other between x1 and x4 = 14 mm. Hence, the colored 4D volume (bottom) represents a catheter trajectory with a turning point within the stenosis.
Fig 9

Complete IP volume reconstruction compared to MT volume reconstruction for the bending profile.

Fig 10

Complete IP volume reconstruction compared to MT volume reconstruction for the heart beat profile.

3.3 Quantitative volume results

We determine the envelopes of the 3D boundary points of the IP and MT methods and quantify the volume reconstruction results using the DICE metric whereas U are the 2D projected shapes of the reconstructed envelopes for method IP and MT, respectively, compared to the ground-truth 2D projected shapes V for all angles from 1 to N = 180°. The mean DICE for all repetitions are listed in Table 4. The stenosis length is determined in x-direction as full at half width of the envelope decay of the volume shapes for all reconstructions, profiles and experimental repetitions. In case of the HBA profile, the stenosis length is determined as summation of the two stenosis lengths.
Table 4

DICE quantification and related stenosis lengths in mm for 3D and 4D reconstruction methods for motion profiles SP, BA, and HBA, respectively.

SP profileBA profileHBA profile
DICE IP0.860.660.58
DICE MT0.880.890.86
Stenosis IP (RE)1.52 (1.3%)3.91 (160%)3.82 (154%)
Stenosis MT (RE)1.15 (23%)1.49 (0.6%)1.19 (21%)

Discussion

The 4D catheter trajectory is tracked by the MPI for three different catheter motion profiles. The statistical validation of all motion profiles and repetitions reveal small MAEs in x-direction of around 0.5mm, which is in good accordance with the estimated determination accuracy [44]. The ground truth for the trajectory in terms of the position in x, y and z directions is not known, since it is hardly possible to track the catheter’s position within the MPI scanner with a second instance, e.g., an optical system. The ground truth for the trajectory is only known in terms of the pullback velocity and distance in 3D over a defined period. Therefore, the calculated MAEs in x-direction contain a small uncertainty because the y and z ground truth positions are assumed to be constant zero. The absolute and relative errors of the velocities determined by a regression line in 3D are comparable to the ground truth velocity of the IVOCT adapter. They show varying relative errors in the range of 3.2%–39% for all motion profiles and their segments. For the BA profile the first change of velocity could not be captured by the measured values. One reason could be that the velocity change is close to the border of the FoV. The voxel intensities representing the marked catheter tip have circle shaped form in the MPI images. If this circle shape has not entered the FoV completely, the center of mass localization algorithm might misinterpret the position. In addition, in case of the HBA profile deviations occur around the turning points seen in Fig 7a), which lead to an underestimation of the velocity. These deviations can be linked to the catheter setup with a proximal actuator such that the pullback is increased by the shrinkage and stretching of the flexible catheter. It is also worth noting that the back and forth motion of the catheter is also visible in the y- and z-positions seen in Fig 7b and 7c) as the vessel phantom is not placed perfectly in accordance with the x-axis. The novel MT volume reconstruction method based on MPI catheter tracking demonstrates a qualitative improvement in comparison to the IP method (Fig 8). Even in case of the SP profile without additional motion artifacts, the IP method shows worse results by means of the DICE metric. The illustrated results of the motion artifact profiles underline the need of a catheter tracking over time. In addition to the catheter tracking in 3D [7, 10, 34, 35], the time synchronization of the IVOCT and MPI data leads to an optimized arrangement of OCT A-scans in 3D space. The DICE metrics and stenosis lengths in Table 4 emphasize the relevant errors in case of the IP method. Nevertheless, inaccuracies in volume shapes occur for all methods and profiles (DICE< 0.9), as other imaging artifacts have an influence on the IVOCT and MPI data. For example, non-uniform rotational distortions (NURD) of the catheter might appear due to the catheter setup. Furthermore, the boundary segmentation in the IVOCT data as well as the catheter tip segmentation in the MPI data contain inaccuracies. In future work, the results can be further improved by a correction of additional artifacts and image enhancements. On the one hand, a rotation tracking with MPI may be possible with an asymmetric marking of the catheter tip. On the other hand, the phantom centerline and catheter trajectory can be tracked using a multi-contrast MPI imaging approach [29–32, 46, 47] visualizing the marker and the blood pool tracer inside the phantom. Both approaches can be used to minimize the effect of NURD artifacts.

Conclusion

A novel approach for MPI-guided IVOCT catheter tracking is presented considering both the 3D catheter trajectory and the time synchronization of IVOCT and MPI data in order to reconstruct volumes of a known vessel phantom shape in 4D. A DICE coefficient of up to 89% is achieved for different IVOCT motion artifact studies. The presented approach estimates the stenosis length for simulated artifacts more precisely with a relative error of up to 0.6% in comparison to 160% of the standard method. 3 Jan 2020 PONE-D-19-33061 In-Vitro MPI-Guided IVOCT Catheter Tracking in Real Time for Motion Artifact Compensation PLOS ONE Dear Mr. Griese, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. We would appreciate receiving your revised manuscript by Feb 17 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols Please include the following items when submitting your revised manuscript: A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'. A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'. An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'. Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out. We look forward to receiving your revised manuscript. Kind regards, Wolfgang Rudolf Bauer, M.D., Ph.D. Academic Editor PLOS ONE Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf 2. In your methods section, please provide more information about the provenance of the dataset, and clarify whether the study involved human subjects. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes ********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: The manuscript "In-Vitro MPI-Guided IVOCT Catheter Tracking in Real Time for Motion Artifact Compensation" describes a combination of IntraVascular Optical Coherence Tomography with Magnetic Particle Imaging. The authors use MPI for determining the position as well as the velocity of the catheter during OCT. The use of a magnetic lacquer enables MPI to detect the marked areas of the catheter as has been previously shown in literature. This approach allows a radiation-free bimodal application of significant (pre-)clinical relevance. Today's standard uses as reference DSA (X-ray) imaging methods. The manuscript is well written and understandable, the figures are comprehensive. Nevertheless, there are a few issues that need to be resolved prior to a possible publication: l22: Please choose: "proximally rotated catheters" or "rotated proximal catherters" depending on what your want to express. l74: You are claiming the lacquer does not affect the OCT beam profile. Please either give a suitable reference or show your measurements supporting this claim. l92: then -> than l97: Either "contains" or "consists of" but not "contains of". l136: What is your criterium for declaring a value an "outlier"? l137: What kind of smoothing did you apply? l184: While mathematically correct, please avoid speaking of "decreased" in this context, rather use "set". A decrease leading to a higher absolute velocity if counter-intuitive for the reader. fig4+6: I recommend to combine these two figures only displaying the contents of 4b and 6a. A constant velocity (4a) needs no depiction and I do not see the benefit of having the velocities graphically presented in 6b. l210: You are stating that the tracked positions are in good agreement with expected values from 18mm to 10mm. Looking at the error bars of your measured values I would rather cut this argument short at 11mm. fig7 (caption): Here you state the good agreement was from 18mm down to 6mm. This is (a) in contradiction to your own statement in the text (18mm-10mm) and also please consider my previous comment. l215: The values of AE and RE given here do not agree with the ones given in tab1. Please verify! l218: You are addressing "a small deviation" of the measured values from the expected ones. Looking at fig7b the measurements seems not to capture the first change in velocity at all since it starts at a wrong slope. Please discuss this result in more detail! Are there issues with higher velocities? l227 and fig8: Similarly to the above case the velocity seems to be underestimated while you are constating a good accordance. Please comment on the too small slopes for higher velocities. l236/7: Three times the units (mm) are missing. l239/40 and tab3: The values given for AE/RE do not agree between table and text. Good job! Your Reviewer ********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: No [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step. 20 Jan 2020 The responds to the reviewer and editor comments is given in die document "Answers to the reviewers". Submitted filename: Rebuttal__Tracking_IVOCT_catheter.pdf Click here for additional data file. 10 Mar 2020 In-Vitro MPI-Guided IVOCT Catheter Tracking in Real Time for Motion Artifact Compensation PONE-D-19-33061R1 Dear Dr. Griese, We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements. Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication. Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org. If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. With kind regards, Wolfgang Rudolf Bauer, M.D., Ph.D. Academic Editor PLOS ONE Additional Editor Comments (optional): Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressed ********** 2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: Yes ********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: Yes ********** 6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: All my remarks to the original manuscript have been fully addressed. From my side there are no more issues to be resolved prior to a possible publication. Your reviewer. ********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Volker C. Behr 13 Mar 2020 PONE-D-19-33061R1 In-Vitro MPI-Guided IVOCT Catheter Tracking in Real Time for Motion Artifact Compensation Dear Dr. Griese: I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department. If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org. For any other questions or concerns, please email plosone@plos.org. Thank you for submitting your work to PLOS ONE. With kind regards, PLOS ONE Editorial Office Staff on behalf of Prof. Wolfgang Rudolf Bauer Academic Editor PLOS ONE
  36 in total

1.  Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation.

Authors:  Guillermo J Tearney; Evelyn Regar; Takashi Akasaka; Tom Adriaenssens; Peter Barlis; Hiram G Bezerra; Brett Bouma; Nico Bruining; Jin-man Cho; Saqib Chowdhary; Marco A Costa; Ranil de Silva; Jouke Dijkstra; Carlo Di Mario; Darius Dudek; Darius Dudeck; Erling Falk; Erlin Falk; Marc D Feldman; Peter Fitzgerald; Hector M Garcia-Garcia; Hector Garcia; Nieves Gonzalo; Juan F Granada; Giulio Guagliumi; Niels R Holm; Yasuhiro Honda; Fumiaki Ikeno; Masanori Kawasaki; Janusz Kochman; Lukasz Koltowski; Takashi Kubo; Teruyoshi Kume; Hiroyuki Kyono; Cheung Chi Simon Lam; Guy Lamouche; David P Lee; Martin B Leon; Akiko Maehara; Olivia Manfrini; Gary S Mintz; Kyiouchi Mizuno; Marie-angéle Morel; Seemantini Nadkarni; Hiroyuki Okura; Hiromasa Otake; Arkadiusz Pietrasik; Francesco Prati; Lorenz Räber; Maria D Radu; Johannes Rieber; Maria Riga; Andrew Rollins; Mireille Rosenberg; Vasile Sirbu; Patrick W J C Serruys; Kenei Shimada; Toshiro Shinke; Junya Shite; Eliot Siegel; Shinjo Sonoda; Shinjo Sonada; Melissa Suter; Shigeho Takarada; Atsushi Tanaka; Mitsuyasu Terashima; Troels Thim; Thim Troels; Shiro Uemura; Giovanni J Ughi; Heleen M M van Beusekom; Antonius F W van der Steen; Gerrit-Anne van Es; Gerrit-Ann van Es; Gijs van Soest; Renu Virmani; Sergio Waxman; Neil J Weissman; Giora Weisz
Journal:  J Am Coll Cardiol       Date:  2012-03-20       Impact factor: 24.094

2.  Heartbeat OCT: in vivo intravascular megahertz-optical coherence tomography.

Authors:  Tianshi Wang; Tom Pfeiffer; Evelyn Regar; Wolfgang Wieser; Heleen van Beusekom; Charles T Lancee; Geert Springeling; Ilona Krabbendam; Antonius F W van der Steen; Robert Huber; Gijs van Soest
Journal:  Biomed Opt Express       Date:  2015-11-23       Impact factor: 3.732

3.  Optical coherence tomography.

Authors:  D Huang; E A Swanson; C P Lin; J S Schuman; W G Stinson; W Chang; M R Hee; T Flotte; K Gregory; C A Puliafito
Journal:  Science       Date:  1991-11-22       Impact factor: 47.728

4.  The impact of Fourier-Domain optical coherence tomography catheter induced motion artefacts on quantitative measurements of a PLLA-based bioresorbable scaffold.

Authors:  N S van Ditzhuijzen; A Karanasos; N Bruining; M van den Heuvel; O Sorop; J Ligthart; K Witberg; H M Garcia-Garcia; F Zijlstra; D J Duncker; H M M van Beusekom; E Regar
Journal:  Int J Cardiovasc Imaging       Date:  2014-05-16       Impact factor: 2.357

5.  Online reconstruction of 3D magnetic particle imaging data.

Authors:  T Knopp; M Hofmann
Journal:  Phys Med Biol       Date:  2016-05-16       Impact factor: 3.609

Review 6.  Contrast-induced nephrotoxicity: clinical landscape.

Authors:  R W Katzberg; C Haller
Journal:  Kidney Int Suppl       Date:  2006-04       Impact factor: 10.545

Review 7.  Contrast-induced acute kidney injury.

Authors:  Peter A McCullough
Journal:  J Am Coll Cardiol       Date:  2008-04-15       Impact factor: 24.094

Review 8.  Ultrasmall superparamagnetic iron oxides (USPIOs): a future alternative magnetic resonance (MR) contrast agent for patients at risk for nephrogenic systemic fibrosis (NSF)?

Authors:  Edward A Neuwelt; Bronwyn E Hamilton; Csanad G Varallyay; William R Rooney; Robert D Edelman; Paula M Jacobs; Suzanne G Watnick
Journal:  Kidney Int       Date:  2008-10-08       Impact factor: 10.612

9.  Anatomically correct three-dimensional coronary artery reconstruction using frequency domain optical coherence tomographic and angiographic data: head-to-head comparison with intravascular ultrasound for endothelial shear stress assessment in humans.

Authors:  Michail I Papafaklis; Christos V Bourantas; Taishi Yonetsu; Rocco Vergallo; Anna Kotsia; Shimpei Nakatani; Lampros S Lakkas; Lambros S Athanasiou; Katerina K Naka; Dimitrios I Fotiadis; Charles L Feldman; Peter H Stone; Patrick W Serruys; Ik-Kyung Jang; Lampros K Michalis
Journal:  EuroIntervention       Date:  2015-08       Impact factor: 6.534

10.  Magnetic Particle / Magnetic Resonance Imaging: In-Vitro MPI-Guided Real Time Catheter Tracking and 4D Angioplasty Using a Road Map and Blood Pool Tracer Approach.

Authors:  Johannes Salamon; Martin Hofmann; Caroline Jung; Michael Gerhard Kaul; Franziska Werner; Kolja Them; Rudolph Reimer; Peter Nielsen; Annika Vom Scheidt; Gerhard Adam; Tobias Knopp; Harald Ittrich
Journal:  PLoS One       Date:  2016-06-01       Impact factor: 3.240