How to choose a point-of-care testing for troponin.

Ya-Hui Lin1, Yang Zhang1, Yu-Tao Liu1, Kai Cui1, Jin-Suo Kang1, Zhou Zhou1.   

Abstract

BACKGROUND: Point-of-care (POC) cTn assays are needed when the central laboratory is unable to provide timely results to the emergency department. Many POC devices are available. The prospect of choosing them is daunting. In order to provide a quick decision-making reference for POC cTn device selection comparing them to the central laboratory, seven POC devices commonly employed by emergency department were evaluated.
METHODS: Firstly, we reviewed all devices package inserts. Secondly, we evaluated several POC cTn assays for imprecision, linearity, and correlation with central laboratory assays according to CLSI EP protocols. The linear regression analyses were performed only for the detectable concentrations. Five cTnI devices (Alere Triage, BioMerieux Vidas, Mitsubishi Pathfast, ReLIA TZ-301, and Radiometer AQT90) were evaluated against a contemporary cTnI assay (Beckman Access II Accu TnI). Two cTnT assays (Radiometer AQT90 and Roche Cobas h232) were compared to a high-sensitivity (hs) cTnT method (Roche Cobas e601).
RESULTS: For cTn levels around the 99th percentile upper reference limits (URLs) of the comparator assays, imprecision could not be assessed for the Alere, BioMerieux, and Cobas h232 as they gave undetectable readings due to a lack of assay sensitivity. Imprecision (CV) was unacceptably high for the ReLIA (33.3%). On account of this precision metric, these four assays were deemed unsuitable. Regression analyses showed acceptable linearity for all the POC devices. The correlation coefficients for ReLIA, BioMerieux, Cobas h232, and Radiometer cTnT were >0.95. Unlike the cTnT devices, the cTnI assays employ different capture and detection antibodies leading to non-commutable results. The POC cTn results were concordant with their comparator-Radiometer cTnT 90%, Pathfast cTnI 85%, and Radiometer cTnI 75%.
CONCLUSION: Our study provides the procedure and essential data to guide selection of a POC cTn device. Of the point-of-care devices, methods evaluated Radiometer AQT90 (cTnI and cTnT) and Pathfast might be considered.
© 2020 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals, Inc.

Entities:  

Keywords:  analytical evaluation; cardiac troponin; point-of-care testing

Mesh:

Substances:

Year:  2020        PMID: 32222055      PMCID: PMC7370753          DOI: 10.1002/jcla.23263

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


INTRODUCTION

Measurements of cardiac troponin (cTn) are vital in the management and diagnosis of myocardial injury and are the preferred cardiac biomarker.1, 2 This is especially critical in the emergency department (ED) where rapid decision‐making in chest pain patients improves outcome and decreases mortality.3, 4, 5 The main focus in the ED is to discharge patients home as soon as possible or to admit them to the hospital ward. Laboratory tests are vital for making such decisions in the ED.6 In fact, laboratory tests are requested in up to half of the patients presenting at the ED.7 Clinical guidelines stipulate a maximum turnaround time of 60 minutes (preferably 30 minutes) from blood draw to results availability.8 The central laboratory is often challenged to deliver such rapid test results.9 This has given rise to the introduction of many point‐of‐care (POC) devices that can provide rapid troponin results.10, 11, 12, 13 However, different POC cTn assays exhibit variable analytical sensitivity and accuracy14 and give results discordant with those of the central laboratory.15 Laboratorians and clinicians need a handy information resource to decide on which POC cTn to adopt. While a sensitive POC cTn is preferred for rapid diagnosis of acute myocardial infarction,16 expert committees of Cardiology societies concede that assays with imprecision of up to 20% may be used for diagnosis and risk stratification.17 Thus, we undertook to evaluate the analytical performance of seven POC cTn assays (5 cTnI and 2 cTnT) against our central laboratory cTnI assay and a high‐sensitivity cTnT (hs‐cTnT) method according to Clinical and Laboratory Standards Institute (CLSI) protocols.

MATERIALS AND METHODS

Analyzers

Five POC cTnI assays—Alere Triage Cardio3TnI (Alere International), BioMerieux Vidas Troponin I Ultra (Biomerieux), Mitsubishi Pathfast cTnI (LSI Medience Corporation), ReLIA TZ‐301 cTnI (Ruilai Biological Engineering), and Radiometer AQT90 TnI (Radiometer Medical), were evaluated against the central laboratory cTnI assay (Access II Accu TnI, Beckman Coulter). Two POC cTnT assays—Radiometer AQT90 cTnT (Radiometer Medical) and Roche Cobas h232 (Roche Diagnostics), were compared to a central laboratory hs‐cTnT assay—Roche Cobas e601 (Roche Diagnostics).

Sample preparation

Samples were collected into Greiner serum tubes (VACUETTE® 454204, Greiner Bio‐One), centrifuged at 3000 g for 10 minutes, and immediately made into multiple aliquots. One aliquot was tested on the AccuTnI (Beckman Access II) and the results released to the ED, while the other aliquots were stored at −80°C until used. Before analysis, the samples were thawed at room temperature, centrifuged at 3000 g for 10 minutes, and the supernatant tested on the various cTn platforms.

Analytical performances

Imprecision studies were based on the CLSI EP15‐A protocol (duplicate measurements twice per day on 3 levels for 5 consecutive days) using pooled human serum.18 The cTn concentration of the low serum pool was within the respective assay's 99th percentile upper reference limits (URL), while the medium and high level pools were around fivefold and 10‐fold of their respective URLs. Troponin values at the 99th percentile with good imprecision (CV ≤ 10%) are “guideline acceptable,” intermediate imprecision (CV 10%‐20%) is “clinically usable,” and poor imprecision (CV > 20%) is not acceptable.19 To demonstrate the linearity of low cTn concentrations, we employed the CLSI EP06‐A protocol by proportionally mixing high and low human serum pools.20 Six concentrations were measured twice at each level. Correlation studies between central laboratory assays and POC assays were performed on the same forty samples from consecutive patients admitted to Fuwai hospital according to the CLSI EP09‐A3.21 These samples covered the most frequent cTn concentration ranges (cTnI: 0.01‐37 µg/L; cTnT: 0.005‐10 µg/L) encountered in our chest pain patients. The same reagent lots of each assay were used throughout the evaluation.

Statistical analyses

All statistical analyses were performed on MedCalc Statistical Software version 15.2.2 (MedCalc Software). Least squares linear regression was employed for linearity assessment, and Passing‐Bablok regression analysis was used for correlation studies. Values below the limit of detection were replaced with one half the limit of detection. A P value < .05 was considered statistically significant.

Others

This study was conducted according to the Declaration of Helsinki and approved by the Institutional Review Board of Fuwai Hospital (IRB approval number 2016‐809).

RESULTS

Review package inserts

Firstly, the analytical characteristics of the seven POC cTn assays and two central laboratory assays according to the manufacturer's claims were summarized in Table 1.
Table 1

Analytical characteristics from nine quantitative devices

Company · platform · assayLoD, (µg/L)99th percentile upper reference limit, µg/LTime, (min)Volume, (µL)Sample typeAmino acid residues of epitopes recognized by Antibodies
Capture antibodiesDetection antibodies
cTnI
Beckman Access II Accu TnI0.010.041255Plasma/Serum41‐4924‐40
Alere Triage cTnI0.050.0520250Plasma/Whole blood27‐3983‐93, 190‐196
bioMerieux Vidas cTnI Ultra0.010.0120200Plasma/Whole blood41‐49, 22‐2987‐91, 7B9
Mitsubishi Pathfast cTnI0.020.02917100Plasma/Serum41‐4971‐116, 163‐209
ReLIA TZ‐301 cTnI0.020.039870Plasma/Serum/Whole blood41‐49, 83‐9324‐40
Radiometer AQT90 cTnI0.010.023181500Plasma/Serum/Whole blood41‐49, 190‐196137‐149
cTnT
Roche Cobas e601 hs‐cTnT0.0050.014950Plasma/Serum125‐131136‐147
Radiometer AQT90 cTnT0.010.017121500Plasma/Serum/Whole blood125‐131136‐147
Roche Cobas h232 Cardiac cTnT0.05NA12150Whole blood125‐131136‐147

Abbreviations: cTnI, cardiac troponin I; cTnT, cardiac troponin T; LoD, limit of detection.

Analytical characteristics from nine quantitative devices Abbreviations: cTnI, cardiac troponin I; cTnT, cardiac troponin T; LoD, limit of detection.

Evaluate the possible candidates

The imprecision of all devices assessed is shown in Table 2. For the low‐level serum pool, the central laboratory instruments showed a CV of 5.90% for the Roche hs‐cTnT and 8.59% for the Beckman Accu TnI, respectively. At this cTn concentration, the imprecision of the Radiometer AQT90 cTnI and AQT90 cTnT (7.96% and 9.53%, respectively) was quite respectable while that for the Mitsubishi Pathfast cTnI (13.37%) was clinically acceptable. For the low‐level serum pool, the Alere cTnI, BioMerieux cTnI, and Roche Cobas h232 cTnT gave undetectable readings, while the imprecision of the ReLIA cTnI was guideline unacceptable (>20%).19 Imprecision for the medium and high serum pools was less than 10% for most of the methods except the Alere cTnI and ReLIA cTnI (CV between 15% and 20%).
Table 2

Imprecision summary

Assay platformMean concentration (µg/L)

%CV

Within run Total

Judgment
cTnI
Beckman Access II AccuTnILow0.0334.488.59Guideline Acceptable
Medium0.3822.695.26
High1.5875.039.59
Alere Triage cTnILow<0.05NANAClinically Acceptable
Medium0.07416.5516.65
High0.4727.5511.08
BioMerieux Vidas cTnI UltraLow<0.01NANAGuideline Acceptable
Medium0.4302.774.15
High1.0171.172.18
Mitsubishi Pathfast cTnILow0.00612.5513.37Clinically Acceptable
Medium0.1047.768.58
High0.36114.7715.02
ReLIA TZ‐301 cTnILow0.04433.2432.32Unacceptable
Medium0.40717.0815.74
High1.4006.588.89
Radiometer AQT90 cTnILow0.0117.697.96Guideline Acceptable
Medium0.1443.713.35
High0.3043.834.40
cTnT
Roche Cobas e601 hs‐cTnTLow0.0133.445.90Guideline Acceptable
Medium0.1481.392.73
High0.5931.211.54
Radiometer AQT90 cTnTLow0.0146.769.53Guideline Acceptable
Medium0.1043.934.99
High0.5842.782.99
Roche Cobas h232 Cardiac cTnTLow<0.05NANAUnacceptable
Medium0.05‐0.10NANA
High0.4155.75010.587

Abbreviations: CV, coefficient of variation; cTnI, cardiac troponin I; cTnT, cardiac troponin T; NA: not applicable.

Imprecision summary %CV Within run Total Abbreviations: CV, coefficient of variation; cTnI, cardiac troponin I; cTnT, cardiac troponin T; NA: not applicable. Regression analyses showed that all devices have acceptable linearity. The average recovery at lower cTn concentrations (close to the 99th percentile URL) was uneven among different devices ranging from 71.2% to 128.7%. The recovery at higher cTn concentrations was satisfactory ranging from 80.9% to 102.3% (See Table 3). The recovery of Beckman · Access II · Accu TnI and Roche · Cobas e601 · hs‐cTnT was 98.8% and 99.2% at lower cTn concentrations, and 97.6% and 99.4% at higher cTn concentrations, respectively.
Table 3

Linearity of seven quantitative POCT cTn devices with “in‐house” prepared samples

Company · platform · assaySlopeIntercept R 2 Recovery (%)
At lower concentrationsAt higher concentrations
cTnI
Beckman · Access II · AccuTnI0.999−0.4220.99898.897.6
Alere · Triage · cTnI0.987−0.4920.98071.291.9
BioMerieux · Vidas · cTnI Ultra0.9940.0670.999128.7102.3
Mitsubishi · Pathfast · cTnI0.987−0.2230.98190.980.9
ReLIA · TZ‐301 · cTnI1.004−0.2400.99998.396.3
Radiometer · AQT90 · cTnI1.002−0.0220.99996.498.1
cTnT
Roche · Cobas e601 · hs‐cTnT0.9950.0210.99999.299.4
Radiometer · AQT90 · cTnT0.998−0.0260.999116.398.4
Roche · Cobas h232 · Cardiac cTnT0.9870.0280.999112.4101.9

Abbreviations: cTnI, cardiac troponin I; cTnT, cardiac troponin T.

Linearity of seven quantitative POCT cTn devices with “in‐house” prepared samples Abbreviations: cTnI, cardiac troponin I; cTnT, cardiac troponin T. As per CLSI document EP09‐A3,21 the correlation between devices and the comparator assays using Passing‐Bablok regression is summarized in Table 4. The correlation coefficients for ReLIA TZ‐301 and BioMerieux Vidas versus Beckman Accu TnI were surprisingly good (>0.95) while that between Radiometer AQT90 cTnI and Beckman Accu cTnI were slightly lower at 0.86.
Table 4

Regression analyses between POC cTn devices against Beckman Access II or Roche Cobas e601

 Passing‐Bablok RegressionCorrelation% of undetectable values
Linear modelBias at 99th URL, µg/L R P value
cTnI
ReLIA TZ‐301a y = −0.021 + 0.931x40%0.969<.00010
Mitsubishi Pathfasta y = −0.001 + 4.033x290%0.934<.000117.5
Alere Triagea y = 0.001 + 1.734x220%0.875<.000125
BioMerieux Vidasa y = −0.009 + 2.033x28%0.959<.000125
Radiometer AQT90a y = −0.040 + 9.805x465%0.864<.000130
cTnT
Roche Cobas h232b y = −0.021 + 0.456x14%0.969<.000135
Radiometer AQT90b y = 0.011 + 0.940x193%0.982<.000125

Abbreviations: cTnT, cardiac troponin T; hs‐cTnT; cTnI, cardiac troponin I.

Linear regression analysis against Beckman Access II cTnI.

Linear regression analysis against Roche Cobas e601.

Regression analyses between POC cTn devices against Beckman Access II or Roche Cobas e601 Abbreviations: cTnT, cardiac troponin T; hs‐cTnT; cTnI, cardiac troponin I. Linear regression analysis against Beckman Access II cTnI. Linear regression analysis against Roche Cobas e601. The Mitsubishi Pathfast, BioMerieux Vidas, Alere Triage, and Radiometer AQT90 cTnI values were consistently lower than the Beckman Accu TnI (see Figure 1). For cTnT, the correlation coefficients for Roche Cobas h232 and Radiometer AQT90 results versus Roche Cobas e601 were 0.969 and 0.982, respectively.
Figure 1

Passing‐Bablok regression analysis of POC cTn assays. A, ReLIA TZ‐301, (B) BioMerieux Vidas, (C) Alere Triage, (D) Mitsubishi Pathfast, (E)Radiometer AQT90 cTnI, (F) Radiometer AQT90 cTnT, and (G) Roche Cobas h232. Legend: Solid line is Passing and Bablok regression; dot line is identity; dash line is 95% CI bands

Passing‐Bablok regression analysis of POC cTn assays. A, ReLIA TZ‐301, (B) BioMerieux Vidas, (C) Alere Triage, (D) Mitsubishi Pathfast, (E)Radiometer AQT90 cTnI, (F) Radiometer AQT90 cTnT, and (G) Roche Cobas h232. Legend: Solid line is Passing and Bablok regression; dot line is identity; dash line is 95% CI bands

Analyze candidate POC cTn assays

For the three suitable candidate POC cTn assays, we performed a concordance analysis according to their cTn result categories—99P (Table 5). Concordance between Roche Cobas hs‐cTnT and Radiometer cTnT was 90% (36/40). For the Beckman Accu cTnI, the concordance was 85% (34/40) and 75% (30/40) with Mitsubishi Pathfast and Radiometer AQT90, respectively.
Table 5

Concordance between result categories of POC devices against Beckman Access II or Roche Cobas e601

 Radiometer cTnI (µg/L)Mitsubishi Pathfast (µg/L)
Beckman · Access II · AccuTnI<LoDLoD–99P>99 P<LoDLoD–99P>99 P
<LoD (4)0.004  0.0030.001 
LoD – 99P (1)0.001  0.001  
>LoD (35)0.0050.0040.0260.004 0.031
Sub‐total (40)0.010.0040.0260.0080.0010.031

Abbreviations; 99P, 99th percentile URL; cTnI, cardiac troponin I; cTnT, cardiac troponin T; LoD, limit of detection.

Concordance between result categories of POC devices against Beckman Access II or Roche Cobas e601 Abbreviations; 99P, 99th percentile URL; cTnI, cardiac troponin I; cTnT, cardiac troponin T; LoD, limit of detection.

DISCUSSION

The primary goal of cTn assays is to aid in the early assessment of acute myocardial infarction especially in hospital emergency departments. Therefore, the analytical imprecision at the decision limit (at the 99th percentile URL) of different cTn assays is vital for clinical decision‐making. For this imprecision metric, the optimal imprecision (CV ≤ 10%) is considered “guideline acceptable.” Intermediate imprecision (10%‐20%) is “clinically usable”,17 but imprecision >20% is not acceptable.19 Information on POC cTn assay imprecision is frequently obtained from manufacturer's package inserts22 or published literature under ideal, non‐routine laboratory conditions. In the current study, the total imprecision of POC cTn assays is compared to central laboratory assays by simultaneously measuring common pooled human sera containing cTn concentrations covering the clinical decision range—99th percentile URL, fivefold URL, and 10‐fold URL. Although POCT devices are optimized for whole blood samples, we used serum in our investigations since no stable whole blood materials are available. Besides, manufacturers employ serum‐based liquid controls and report precision data in their package inserts based on serum or plasma. It is noteworthy that two of the seven POC cTn assays (Radiometer AQT90 cTnI and cTnT) achieved precision that is guideline acceptable, while one other assay (Mitsubishi Pathfast cTnI) met the clinically usable designation. The Cobas h232 cTnT assay is semi‐quantitative; cTnT values < 0.050 µg/L are reported as negative, while those between 0.050 and 0.100 µg/L are reported as positive; actual cTnT results are only reported for values above 0.100 µg/L. Roche has since improved the Cobas h232 to render it fully quantitative with a measuring range of 0.040‐2.000 µg/L.23 At the time, our evaluation was performed, the improved Cobas h232 was not available to us as it had not yet obtained regulatory approval for clinical use. Our results confirm the discrepant cTn concentrations obtained between different POC devices and the central laboratory analyzers. Even on an aggregate level by categories (below LoD, LoD to 99P, and above 99P), the discordance is not insignificant—10% with the Radiometer cTnT, but 15%‐25% with the suitable cTnI devices (Mitsubishi Pathfast and Radiometer). This discrepancy may be caused by a variety of reasons including the use of various capture and detection antibodies/labels, reacting conditions, detection principles, and calculation methods.24 In addition, the available and unique cTn standard reference material, SRM 2921, is a trimeric I‐T‐C complex which is difficult to simulate accurate cTn levels in vivo. Besides, SRM 2921 appears to have poor stability due to factors such as time and temperature.25 Besides, ALP‐dependent and chemiluminescent immunoassays have been reported to show better total imprecision at the URL than cTn assays employing fluorophores or gold particles.26 The difference in antibody specificities has been a key factor hindering the harmonization of POCT measurements, especially for the cTnI assays. It is noteworthy that the epitopes recognized by the capture and detection antibodies employed by the Beckman Access II as well as the ReLIA TZ‐301 are located at the stable mid‐region of cTnI.27 However, the epitopes recognized by the detection antibodies of the Mitsubishi Pathfast, Alere Triage, and Radiometer AQT90 are close to the less stable C‐terminus of cTnI at amino acids 163‐209, 190‐196, and 137‐149, respectively.28 Our results show that the measured values of these POC assays were significantly lower than Beckman Access II when the same samples were tested (see Figure 1). With regard to cTnT, the results of Radiometer AQT90 and Roche Cobas e601 are similar and with good correlation since identical capture and detection antibodies are employed. However, the correlation coefficients were only marginally dissimilar for Cobas h232 and AQT90‐TnT versus Cobas e601. Thus even when antibodies employed recognize the same antigenic epitopes, cTnT concentrations from POCT devices are unlikely to equate to those of the central laboratory.29 Harmonization of POC cTn testing is sorely needed because cTn may be tested on both POC and central laboratory immunochemistry analyzers in the same patient with chest pain presenting to the emergency department. It is important that results from the POC cTn performed at the emergency department are comparable to those from the central laboratory. When choosing POC cTn assays, the analytical characteristics of the various POC devices especially their precision at around the 99th percentile URL should be carefully examined first. Instruments with high imprecision should be excluded. The package insert data (Table 1) reveal that the Cobas h232 has no statement on the 99th percentile URL and is unlikely to meet acceptability criteria. Moreover, identical limits of detection and the 99th percentile URLs are stated for the Alere cTnI and the Biomerieux cTnI; thus, a high imprecision at the URL can be expected. This view is confirmed by the poor CV at the low serum pool (Table 2) for Cobas h232, Alere cTnI, Biomerieux cTnI, and ReLia cTnI rendering them “not acceptable.” Our study provides laboratories and emergency department personnel some objective data for choosing a suitable POC cTn assay as well as not to pursue unlikely candidates. POC cTn has decreased time in the emergency department (ED) for patients with chest pain,5 and the degree of benefit also depends on the setting in which it is used especially acceptance by ED personnel. In the pre‐hospital phase, POC cTn has helped in the triage of chest pain patients to the appropriate hospital facility.30 Beyond hospitals, POC cTn has reduced costs from unnecessary referrals in primary care settings but at the expense of some missed cases of small or early myocardial infarction due to the limitation of assay sensitivity.31 However, higher sensitivity POC cTn platforms are looming on the horizon and promises to change the ED diagnostic landscape.32, 33, 34 Our study provided the step‐by‐step guide to choose POC for cTn and a head‐head comparison of 7 POC troponin devices. It is aimed at providing a quick decision‐making reference for POC cTn device selection comparing them to the central laboratory cTnI and hs‐cTnT. This is the first POC cTn study conducted in the real‐world conditions prevailing in the Asia‐Pacific. Besides, there is paucity of peer‐reviewed literature verifying manufacturers’ analytical claims. The study has some limitations in that we did not conduct reference interval studies or clinical performance studies of the different POC cTn platforms. This information is already available in part in the published literature.14, 29, 35, 36, 37, 38 Further studies on larger population samples are needed.39 In another hand, POC cTnI assays were compared against a contemporary assay, whereas cTnT assays were compared against a high‐sensitivity assay. As we known, the imprecision or sensitivity of high‐sensitivity troponin is better than contemporary troponin. It means that the standards of POC for troponin T were higher troponin I.

CONFLICT OF INTEREST

We have no conflicts of interest to declare.

AUTHOR CONTRIBUTIONS

All the authors have accepted responsibility for the entire content of this submitted the study and approved submission.
  35 in total

1.  Degradation of troponin I in serum or plasma: mechanisms, and analytical and clinical implications.

Authors:  Giuseppe Lippi; Gianfranco Cervellin
Journal:  Semin Thromb Hemost       Date:  2012-02-16       Impact factor: 4.180

2.  Being rational about (im)precision: a statement from the Biochemistry Subcommittee of the Joint European Society of Cardiology/American College of Cardiology Foundation/American Heart Association/World Heart Federation Task Force for the definition of myocardial infarction.

Authors:  Allan S Jaffe; Fred S Apple; David A Morrow; Bertil Lindahl; Hugo A Katus
Journal:  Clin Chem       Date:  2010-04-01       Impact factor: 8.327

Review 3.  Cardiac biomarkers and the case for point-of-care testing.

Authors:  Barry McDonnell; Stephen Hearty; Paul Leonard; Richard O'Kennedy
Journal:  Clin Biochem       Date:  2009-02-07       Impact factor: 3.281

4.  Analytical evaluation of point of care cTnT and clinical performances in an unselected population as compared with central laboratory highly sensitive cTnT.

Authors:  Anne Marie Dupuy; Mustapha Sebbane; François Roubille; Thibault Coste; Anne Sophie Bargnoux; Stéphanie Badiou; Nils Kuster; Jean Paul Cristol
Journal:  Clin Biochem       Date:  2014-10-12       Impact factor: 3.281

5.  Third universal definition of myocardial infarction.

Authors:  Kristian Thygesen; Joseph S Alpert; Allan S Jaffe; Maarten L Simoons; Bernard R Chaitman; Harvey D White; Kristian Thygesen; Joseph S Alpert; Harvey D White; Allan S Jaffe; Hugo A Katus; Fred S Apple; Bertil Lindahl; David A Morrow; Bernard A Chaitman; Peter M Clemmensen; Per Johanson; Hanoch Hod; Richard Underwood; Jeroen J Bax; Robert O Bonow; Fausto Pinto; Raymond J Gibbons; Keith A Fox; Dan Atar; L Kristin Newby; Marcello Galvani; Christian W Hamm; Barry F Uretsky; Ph Gabriel Steg; William Wijns; Jean-Pierre Bassand; Phillippe Menasché; Jan Ravkilde; E Magnus Ohman; Elliott M Antman; Lars C Wallentin; Paul W Armstrong; Maarten L Simoons; James L Januzzi; Markku S Nieminen; Mihai Gheorghiade; Gerasimos Filippatos; Russell V Luepker; Stephen P Fortmann; Wayne D Rosamond; Dan Levy; David Wood; Sidney C Smith; Dayi Hu; José-Luis Lopez-Sendon; Rose Marie Robertson; Douglas Weaver; Michael Tendera; Alfred A Bove; Alexander N Parkhomenko; Elena J Vasilieva; Shanti Mendis
Journal:  Eur Heart J       Date:  2012-08-24       Impact factor: 29.983

6.  Equal clinical performance of a novel point-of-care cardiac troponin I (cTnI) assay with a commonly used high-sensitivity cTnI assay.

Authors:  Per Venge; Lian van Lippen; Sabine Blaschke; Michael Christ; Felicitas Geier; Evangelos Giannitsis; Emil Hagström; Pierre Hausfater; Mehdi Khellaf; Johannes Mair; David Pariente; Volkher Scharnhorst; Veronique Semjonow
Journal:  Clin Chim Acta       Date:  2017-03-25       Impact factor: 3.786

Review 7.  Cardiac markers and their point-of-care testing for diagnosis of acute myocardial infarction.

Authors:  Zhen Yang; Dao Min Zhou
Journal:  Clin Biochem       Date:  2006-06-07       Impact factor: 3.281

8.  Analytical evaluation of a new point of care system for measuring cardiac Troponin I.

Authors:  Danielle Wm Kemper; Veronique Semjonow; Femke de Theije; Diederick Keizer; Lian van Lippen; Johannes Mair; Bernadette Wille; Michael Christ; Felicitas Geier; Pierre Hausfater; David Pariente; Volkher Scharnhorst; Joyce Curvers; Jeroen Nieuwenhuis
Journal:  Clin Biochem       Date:  2016-11-12       Impact factor: 3.281

Review 9.  Analytical characteristics of high-sensitivity cardiac troponin assays.

Authors:  Fred S Apple; Paul O Collinson
Journal:  Clin Chem       Date:  2011-09-30       Impact factor: 8.327

10.  2015 ESC Guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation: Task Force for the Management of Acute Coronary Syndromes in Patients Presenting without Persistent ST-Segment Elevation of the European Society of Cardiology (ESC).

Authors:  Marco Roffi; Carlo Patrono; Jean-Philippe Collet; Christian Mueller; Marco Valgimigli; Felicita Andreotti; Jeroen J Bax; Michael A Borger; Carlos Brotons; Derek P Chew; Baris Gencer; Gerd Hasenfuss; Keld Kjeldsen; Patrizio Lancellotti; Ulf Landmesser; Julinda Mehilli; Debabrata Mukherjee; Robert F Storey; Stephan Windecker
Journal:  Eur Heart J       Date:  2015-08-29       Impact factor: 29.983

View more
  2 in total

1.  How to choose a point-of-care testing for troponin.

Authors:  Ya-Hui Lin; Yang Zhang; Yu-Tao Liu; Kai Cui; Jin-Suo Kang; Zhou Zhou
Journal:  J Clin Lab Anal       Date:  2020-03-28       Impact factor: 2.352

Review 2.  The role of antibody-based troponin detection in cardiovascular disease: A critical assessment.

Authors:  Hui Ma; Arabelle Cassedy; Richard O'Kennedy
Journal:  J Immunol Methods       Date:  2021-07-28       Impact factor: 2.303

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.