Literature DB >> 25315142

Analytical evaluation of point of care cTnT and clinical performances in an unselected population as compared with central laboratory highly sensitive cTnT.

Anne Marie Dupuy1, Mustapha Sebbane2, François Roubille3, Thibault Coste1, Anne Sophie Bargnoux1, Stéphanie Badiou1, Nils Kuster1, Jean Paul Cristol4.   

Abstract

OBJECTIVES: To report the analytical performances of the Radiometer AQT90 FLEX® cTnT assay (Neuilly-Plaisance, France) and to evaluate the concordance with hs-cTnT results from central laboratory for the diagnosis of acute myocardial infarction (AMI) at baseline and during a short follow-up among unselected patients admitted in emergency room or cardiology department. DESIGN AND METHODS: Analytical performances of AQT90 FLEX® cTnT immunoassay included imprecision study with determination of a coefficient of variation at 10% and 20%, linearity, and limit of detection. The concordance study was based on samples obtained from 170 consecutive patients with chest pain suggestive of acute coronary syndrome (ACS) admitted in the emergency room or cardiology department. The kinetic study (within 62 additional samples 3h later) was based on absolute delta criterion and the combination of relative change of 30% with absolute change of 7ng/L.
RESULTS: The cTnT assay from Radiometer was evaluated as clinically usable, although less sensitive than the Roche hs-cTnT assay as demonstrated by the concordance and the kinetic studies.
CONCLUSIONS: In non-selected population, the cTnT AQT Flex© assay on AQT90© with kinetic change at 3h, provides similar clinical classification of patients, particularly for AMI group as compared to central laboratory hs-cTnT assay and could be suitable for clinical use.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Diagnostic performance; Highly sensitive cardiac troponin; Kinetic change; Point of care

Mesh:

Substances:

Year:  2014        PMID: 25315142     DOI: 10.1016/j.clinbiochem.2014.09.027

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  3 in total

1.  Clinical value of cardiac color ultrasound and cardiac troponin T combined with dynamic electrocardiogram in treatment of acute pulmonary embolism.

Authors:  Hongjun Li; Lixin Kang; Yongqiang Sun
Journal:  Exp Ther Med       Date:  2017-12-18       Impact factor: 2.447

2.  The Usefulness and Limitations of Point-of-care Cardiac Troponin Measurement in the Emergency Department.

Authors:  Kenichiro Suzuki; Kimiaki Komukai; Kotaro Nakata; Ryeonshi Kang; Yuhei Oi; Eri Muto; Yusuke Kashiwagi; Mitsutoshi Tominaga; Satoru Miyanaga; Tetsuya Ishikawa; Kenji Okuno; Masahiko Uzura; Michihiro Yoshimura
Journal:  Intern Med       Date:  2018-02-09       Impact factor: 1.271

3.  How to choose a point-of-care testing for troponin.

Authors:  Ya-Hui Lin; Yang Zhang; Yu-Tao Liu; Kai Cui; Jin-Suo Kang; Zhou Zhou
Journal:  J Clin Lab Anal       Date:  2020-03-28       Impact factor: 2.352

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.