Anne Marie Dupuy1, Mustapha Sebbane2, François Roubille3, Thibault Coste1, Anne Sophie Bargnoux1, Stéphanie Badiou1, Nils Kuster1, Jean Paul Cristol4.
Abstract
OBJECTIVES: To report the analytical performances of the Radiometer AQT90 FLEX® cTnT assay (Neuilly-Plaisance, France) and to evaluate the concordance with hs-cTnT results from central laboratory for the diagnosis of acute myocardial infarction (AMI) at baseline and during a short follow-up among unselected patients admitted in emergency room or cardiology department. DESIGN AND METHODS: Analytical performances of AQT90 FLEX® cTnT immunoassay included imprecision study with determination of a coefficient of variation at 10% and 20%, linearity, and limit of detection. The concordance study was based on samples obtained from 170 consecutive patients with chest pain suggestive of acute coronary syndrome (ACS) admitted in the emergency room or cardiology department. The kinetic study (within 62 additional samples 3h later) was based on absolute delta criterion and the combination of relative change of 30% with absolute change of 7ng/L.
RESULTS: The cTnT assay from Radiometer was evaluated as clinically usable, although less sensitive than the Roche hs-cTnT assay as demonstrated by the concordance and the kinetic studies.
CONCLUSIONS: In non-selected population, the cTnT AQT Flex© assay on AQT90© with kinetic change at 3h, provides similar clinical classification of patients, particularly for AMI group as compared to central laboratory hs-cTnT assay and could be suitable for clinical use.
OBJECTIVES: To report the analytical performances of the Radiometer AQT90 FLEX® cTnT assay (Neuilly-Plaisance, France) and to evaluate the concordance with hs-cTnT results from central laboratory for the diagnosis of acute myocardial infarction (AMI) at baseline and during a short follow-up among unselected patients admitted in emergency room or cardiology department. DESIGN AND METHODS: Analytical performances of AQT90 FLEX® cTnT immunoassay included imprecision study with determination of a coefficient of variation at 10% and 20%, linearity, and limit of detection. The concordance study was based on samples obtained from 170 consecutive patients with chest pain suggestive of acute coronary syndrome (ACS) admitted in the emergency room or cardiology department. The kinetic study (within 62 additional samples 3h later) was based on absolute delta criterion and the combination of relative change of 30% with absolute change of 7ng/L.
RESULTS: The cTnT assay from Radiometer was evaluated as clinically usable, although less sensitive than the Roche hs-cTnT assay as demonstrated by the concordance and the kinetic studies.
CONCLUSIONS: In non-selected population, the cTnT AQT Flex© assay on AQT90© with kinetic change at 3h, provides similar clinical classification of patients, particularly for AMI group as compared to central laboratory hs-cTnT assay and could be suitable for clinical use.
Copyright © 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
Entities:
Keywords:
Diagnostic performance; Highly sensitive cardiac troponin; Kinetic change; Point of care
Mesh:
Substances:
Year: 2014
PMID: 25315142 DOI: 10.1016/j.clinbiochem.2014.09.027
Source DB: PubMed Journal: Clin Biochem ISSN: 0009-9120 Impact factor: 3.281