| Literature DB >> 32216082 |
Christian Velten1, Robert Boyd1, Kyoungkeun Jeong1, Madhur K Garg1,2, Wolfgang A Tomé1,2.
Abstract
Megavoltage computed tomography (MVCT) image quality metrics were evaluated on an Accuray Radixact unit to recommend scan settings for the implementation of a consistent adaptive radiotherapy program. Megavoltage computed tomography image quality was evaluated and compared to a kilovoltage CT (kVCT) simulator using a commercial cone beam computed tomography image quality phantom. Megavoltage computed tomographies were acquired on the Accuray Radixact using fine, normal, and coarse pitches, with all available reconstruction slice thicknesses, each of which were reconstructed using standard and iterative reconstruction (IR). Image quality metrics (IQM) were evaluated using DoseLab: automatically and manually calculated spatial resolution, subject contrast, and contrast-to-noise ratio (CNR). Scanning time was 15.6 s/cm for fine, 8.1 s/cm for normal, and 5.6 s/cm for coarse pitch. Automatically evaluated spatial resolutions ranged from 0.39, 0.41, to 0.42 lp/mm for standard reconstruction and from 0.24, 0.21, to 0.18 lp/mm for soft-tissue IR, respectively, with general IR yielding values in between these. Spatial resolution for kVCT was measured to be at least 0.42 lp/mm. Contrast was consistent across MVCT settings with 8.1 ± 0.2%, while kVCT contrast was 10.27 ± 0.05%. CNR was calculated to be 3.3 ± 0.4 for standard reconstruction, 7.4 ± 0.4 for general IR, and 12.0 ± 1.9 for soft-tissue IR. It was found that increasing reconstruction slice thickness for a given pitch does not improve IQMs. Based on the consistency of contrast metrics across pitch values and the only slightly reduced spatial resolution using normal compared to fine pitch, we recommend the use of normal pitch with 2 mm slice thickness to maximize image quality for ART while limiting scanning time. Only for sites for which improved CNR is required and reduced spatial resolution is acceptable, soft-tissue IR is recommended.Entities:
Keywords: MVCT; adaptive radiotherapy; image quality; tomotherapy
Mesh:
Year: 2020 PMID: 32216082 PMCID: PMC7286013 DOI: 10.1002/acm2.12859
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Computed tomography slices for standard protocol kilovoltage CT with 2.5 mm slice thickness (row 1) and fine, normal, and coarse pitch (rows 2–4) with standard reconstruction megavoltage CT with 1, 2, and 3 mm slice thicknesses.
Fig. 2Computed tomography slices for normal pitch megavoltage computed tomography (2 mm slice thickness) using standard (row 1), general iterative (row 2), and soft‐tissue iterative (row 3) image reconstruction techniques. Spatial resolution worsens and noise decreases from standard to soft‐tissue image reconstruction, while contrast (right column) appears comparable.
Summary of axial in‐plane image quality metrics acquired using the kV‐cone beam computed tomography (CBCT) phantom.
| Pitch | kVCT | MVCT‐fine | MVCT‐normal | MVCT‐coarse |
|---|---|---|---|---|
| Slice thickness (mm) | 2.5 | 1 | 2 | 3 |
| Scan time (s/cm) | 15.6 | 8.1 | 5.6 |
IR‐G: general iterative image reconstruction; IR‐ST soft‐tissue image reconstruction; kVCT, kilovoltage computed tomography; MVCT, megavoltage CT; SR: Standard image reconstruction.
Summary of sagittal plane image quality metrics acquired using the kV‐CBCT phantom.
| Pitch | MVCT‐fine | MVCT‐normal | MVCT‐coarse |
|---|---|---|---|
| Slice thickness (mm) | 1 | 2 | 3 |
IR‐G: general iterative image reconstruction; IR‐ST soft‐tissue image reconstruction; kVCT, kilovoltage computed tomography; MVCT, megavoltage CT; SR: Standard image reconstruction.