Christian Velten1,2, Lee Goddard1,2, Kyoungkeun Jeong1,2, Madhur K Garg1,2, Wolfgang A Tomé1,2.
Abstract
PURPOSE: To assess clinically relevant image quality metrics (IQMs) of helical fan beam kilovoltage (kV) fan beam computed tomography (CT). METHODS AND MATERIALS: kVCT IQMs were evaluated on an Accuray Radixact unit equipped with helical fan beam kVCT to assess the capabilities of this newly available modality. kVCT IQMs were evaluated and compared to a kVCT simulator and linear accelerator-based cone beam CTs (CBCT) using a commercial CBCT image quality phantom. kVCTs were acquired on the Accuray Radixact for all combinations of kVp and mAs in fine mode using a 440-mm field of view (FOV). Evaluated IQMs were spatial resolution, overall uniformity, subject contrast, contrast-to-noise ratio (CNR), and effective slice thickness. Imaging dose was assessed for planar kV imaging.Entities:
Year: 2021 PMID: 35036634 PMCID: PMC8749200 DOI: 10.1016/j.adro.2021.100862
Source DB: PubMed Journal: Adv Radiat Oncol ISSN: 2452-1094
kV system operating characteristics before and after the hard- and software upgrade to ClearRT
| Protocol | Voltage (kV) | Tube current (mA) | Air kerma per exposure (mGy) | MeankVp | HVL (mmAl) | Total filtration (mmAl) | Spatial resolution (lp/mm) | Minimum uniformity | Contrast | CNR |
|---|---|---|---|---|---|---|---|---|---|---|
| Before upgrade | ||||||||||
| Thorax (XXS) | 100 | 0.8 | 0.051 | 102.3 | 8.9 | 21.4 | 1.17 | 98.3% | 38.2% | 44.7 |
| Thorax (XS) | 100 | 1.0 | 0.063 | 102.4 | 8.9 | 21.6 | 1.17 | 98.5% | 38.1% | 49.1 |
| Thorax (S) | 120 | 0.8 | 0.090 | 118.0 | 9.9 | 21.8 | 1.17 | 98.7% | 36.2% | 51.2 |
| Thorax (M) | 120 | 1.0 | 0.112 | 124.3 | 9.8 | 19.7 | 1.17 | 98.7% | 36.5% | 53.4 |
| Thorax (L) | 120 | 1.6 | 0.114 | 122.3 | 9.9 | 20.6 | 1.16 | 99.0% | 36.9% | 64.1 |
| Pelvis (S) | 120 | 1.0 | 0.140 | 121.7 | 9.8 | 19.8 | 1.17 | 98.6% | 36.8% | 51.4 |
| Pelvis (M) | 120 | 1.25 | 0.179 | 123.6 | 9.9 | 20.3 | 0.94 | 96.9% | 60.7% | 62.4 |
| Pelvis (L) | 120 | 2.0 | 0.222 | 120.2 | 9.8 | 20.6 | 1.17 | 99.0% | 37.7% | 64.8 |
| Pelvis (XL) | 140 | 4.0 | 0.705 | 142.1 | 10.9 | 21.0 | - | - | - | - |
| After upgrade | ||||||||||
| Thorax (XXS) | 100 | 0.8 | 0.018 | 103.7 | 8.9 | 21.4 | 0.98 | 97.5% | 47.9% | 48.5 |
| Thorax (XS) | 100 | 1.0 | 0.022 | 103.4 | 8.9 | 21.4 | 0.97 | 97.7% | 48.4% | 54.8 |
| Thorax (S) | 120 | 0.8 | 0.032 | 119.3 | 9.9 | 21.6 | 0.97 | 97.9% | 47.1% | 53.3 |
| Thorax (M) | 120 | 1.0 | 0.038 | 124.1 | 9.9 | 21.7 | 0.97 | 98.0% | 48.0% | 60.8 |
| Thorax (L) | 120 | 1.6 | 0.039 | 122.4 | 9.8 | 19.8 | - | - | - | - |
| Pelvis (S) | 120 | 1.0 | 0.047 | 120.0 | 9.9 | 21.7 | 0.97 | 98.1% | 47.1% | 57.4 |
| Pelvis (M) | 120 | 1.25 | 0.063 | 123.5 | 10.0 | 21.9 | 0.97 | 98.3% | 47.9% | 66.4 |
| Pelvis (L) | 120 | 2.0 | 0.075 | 122.6 | 9.8 | 19.5 | - | - | - | - |
| Pelvis (XL) | 140 | 2.5 | 0.151 | 144.6 | 10.87 | 20.9 | - | - | - | - |
| Pelvis (XXL) | 140 | 4.0 | 0.238 | 142.0 | 10.9 | 21.4 | - | - | - | - |
Abbreviations: CNR = contrast-to-noise ratio; HVL = half value layer; kV = kilovoltage.
Entries with “-” could not be evaluated.
Image quality metrics for ClearRT (fine and coarse modes) and technique-matched CT simulator.
| ClearRT kVCT (fine, 440 mm FOV) | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Head (100 kV) | Thorax (120 kV) | Pelvis (140 kV) | ||||||||||
| Small | Medium | Large | Small | Medium | Large | X-large | Small | Medium | Large | X-large | ||
| Spatial resolution | (lp/mm) | 0.28 | 0.28 | 0.28 | 0.28 | 0.28 | 0.30 | 0.28 | 0.27 | 0.29 | 0.28 | 0.28 |
| Manual spat. res. | (lp/mm) | 0.45 | 0.45 | 0.45 | 0.45 | 0.45 | 0.45 | 0.45 | 0.40 | 0.40 | 0.40 | 0.40 |
| Overall uniformity | (%) | 98.8 | 99.0 | 99.2 | 99.0 | 99.0 | 99.3 | 99.1 | 99.0 | 99.1 | 99.0 | 99.2 |
| Contrast | (%) | 10.1 | 10.3 | 10.2 | 10.2 | 9.9 | 9.9 | 10.0 | 9.8 | 9.9 | 10.0 | 9.9 |
| CNR | 18.2 | 23.7 | 29.1 | 24.8 | 30.7 | 32.7 | 30.4 | 22.8 | 29.9 | 31.7 | 34.9 | |
| Slice thickness | (mm) | 4.0 | 3.9 | 3.9 | 4.5 | 4.6 | 4.5 | 4.8 | 4.2 | 4.5 | 4.4 | 4.7 |
| ClearRT kVCT (coarse, 440 mm FOV) | ||||||||||||
| Head (100 kV) | Thorax (120 kV) | Pelvis (140 kV) | ||||||||||
| Small | Medium | Large | Small | Medium | Large | X-large | Small | Medium | Large | X-large | ||
| Spatial resolution | (lp/mm) | 0.26 | 0.28 | 0.27 | 0.26 | 0.27 | 0.27 | 0.26 | 0.27 | 0.26 | 0.25 | 0.26 |
| Manual spat. res. | (lp/mm) | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 | 0.40 |
| Overall uniformity | (%) | 97.6 | 97.3 | 97.9 | 97.3 | 97.8 | 97.8 | 97.4 | 97.2 | 98.0 | 98.0 | 98.2 |
| Contrast | (%) | 9.4 | 9.5 | 9.7 | 9.4 | 9.7 | 9.3 | 9.6 | 9.7 | 9.6 | 9.6 | 9.5 |
| CNR | 7.9 | 11.3 | 13.1 | 11.8 | 11.0 | 14.7 | 11.4 | 11.9 | 12.7 | 14.6 | 14.3 | |
| Slice thickness | (mm) | 4.1 | 3.7 | 4.3 | 3.9 | 5.4 | 4.5 | 4.4 | 5.8 | 5.2 | 5.0 | 4.6 |
| Matched CT-simulator | ||||||||||||
| Head (100 kV) | Thorax (120 kV) | Pelvis (140 kV) | ||||||||||
| Small | Medium | Large | Small | Medium | Large | X-large | Small | Medium | Large | X-large | ||
| Spatial resolution | (lp/mm) | 0.45 | 0.45 | 0.40 | 0.39 | 0.43 | 0.46 | 0.41 | 0.39 | 0.42 | 0.42 | 0.40 |
| Manual spat. res. | (lp/mm) | 0.65 | 0.65 | 0.65 | 0.60 | 0.65 | 0.65 | 0.60 | 0.60 | 0.65 | 0.60 | 0.60 |
| Overall uniformity | (%) | 98.3 | 98.5 | 98.8 | 99.2 | 99.4 | 99.4 | 99.5 | 99.3 | 99.5 | 99.6 | 99.6 |
| Contrast | (%) | 10.3 | 10.4 | 10.5 | 10.2 | 10.2 | 10.2 | 10.2 | 10.0 | 10.0 | 10.0 | 10.0 |
| CNR | 10.1 | 12.6 | 14.4 | 20.4 | 23.3 | 24.0 | 24.8 | 19.7 | 21.8 | 30.5 | 31.8 | |
| Slice thickness | (mm) | 2.3 | 2.7 | 2.7 | 4.0 | 4.0 | 4.0 | 3.6 | 3.9 | 3.8 | 4.1 | 4.0 |
Abbreviations: CNR = contrast-to-noise ratio; CT = computed tomography; FOV = field of view; kV = kilovoltage.
Image quality metrics for clinically used patient setup imaging
| Head | Thorax | Pelvis | Pelvis large | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ClearRT (coarse-S) | CBCT | iCBCT | ClearRT (coarse-S) | CBCT | iCBCT | ClearRT (normal-L) | CBCT | iCBCT | ClearRT (fine-XL) | CBCT | iCBCT | ||
| Spatial resolution | (lp/mm) | 0.26 | 0.43 | 0.45 | 0.26 | 0.38 | 0.37 | 0.27 | 0.37 | 0.37 | 0.28 | 0.37 | 0.36 |
| Manual spat. res. | (lp/mm) | 0.40 | 0.70 | 0.60 | 0.40 | 0.50 | 0.45 | 0.45 | 0.70 | 0.40 | 0.40 | 0.50 | 0.40 |
| Overall uniformity | (%) | 97.6 | 95.2 | 96.5 | 97.3 | 98.0 | 97.6 | 98.7 | 94.6 | 99.0 | 99.2 | 98.5 | 99.0 |
| Contrast | (%) | 9.4 | 10.7 | 10.3 | 9.4 | 10.5 | 10.5 | 9.7 | 10.3 | 10.5 | 9.9 | 10.4 | 10.4 |
| CNR | 7.9 | 4.1 | 4.4 | 11.8 | 11.2 | 7.90 | 18.4 | 14.2 | 53.8 | 34.9 | 20.4 | 61.7 | |
| Slice thickness | (mm) | 4.1 | 2.4 | 2.3 | 3.9 | 2.6 | 2.4 | 4.3 | 2.5 | 3.0 | 4.7 | 2.5 | 2.5 |
Abbreviations: CBCT = cone beam computed tomography; CNR = contrast-to-noise ratio.
ClearRT modes were chosen based on the closest imaging dose (CTDIvol).
Fig. 1Computed tomography (CT) simulation (a), kilovoltage cone beam CT (kV-CBCT) (b), ClearRT kVCT (c), and megavoltage (MV) CT (d) of a patient treated for prostate cancer. The ClearRT scan was acquired with pelvis medium settings in fine mode showing comparable contrast and spatial resolution to CBCT.
Fig. 2Computed tomography (CT) simulation (a), kilovoltage cone beam CT (kV-CBCT) (b), ClearRT kVCT (c), and megavoltage (MV) CT (d) of a patient treated for head and neck cancer of the larynx. The ClearRT scan was acquired with head medium settings in normal mode showing considerable loss of signal between the highly attenuating bony anatomy. The CBCT was acquired in full-fan head mode with limited field of view, exhibiting minor CBCT and photon starvation artifacts.
Fig. 3Computed tomography (CT) simulation (a), kilovoltage cone beam CT (kV-CBCT) (b), ClearRT kVCT (c), and megavoltage (MV) CT (d) of a patient treated for grade 4 glioma. The ClearRT scan was acquired with head small settings in normal mode exhibiting windmill and beam hardening artifacts in brain tissue. Notably, the craniotomy defect in the left posterior cranium is not visible only on the ClearRT scan. The CBCT shows qualitatively similar noise levels in soft tissue to those of MVCT.